  <front>
    <journal-meta><journal-id journal-id-type="publisher">AMT</journal-id><journal-title-group>
    <journal-title>Atmospheric Measurement Techniques</journal-title>
    <abbrev-journal-title abbrev-type="publisher">AMT</abbrev-journal-title><abbrev-journal-title abbrev-type="nlm-ta">Atmos. Meas. Tech.</abbrev-journal-title>
  </journal-title-group><issn pub-type="epub">1867-8548</issn><publisher>
    <publisher-name>Copernicus Publications</publisher-name>
    <publisher-loc>Göttingen, Germany</publisher-loc>
  </publisher></journal-meta>
    <article-meta>
      <article-id pub-id-type="doi">10.5194/amt-13-941-2020</article-id><title-group><article-title>Increase of the particle hit rate in a laser single-particle mass spectrometer
by pulse delayed extraction technology</article-title><alt-title>Increase of the particle hit rate</alt-title>
      </title-group><?xmltex \runningtitle{Increase of the particle hit rate}?><?xmltex \runningauthor{Y.~Chen et al.}?>
      <contrib-group>
        <contrib contrib-type="author" corresp="no" rid="aff1 aff2">
          <name><surname>Chen</surname><given-names>Ying</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff3">
          <name><surname>Kozlovskiy</surname><given-names>Viacheslav</given-names></name>
          
        <ext-link>https://orcid.org/0000-0002-0921-5521</ext-link></contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1 aff2">
          <name><surname>Du</surname><given-names>Xubing</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff5">
          <name><surname>Lv</surname><given-names>Jinnuo</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff4">
          <name><surname>Nikiforov</surname><given-names>Sergei</given-names></name>
          
        <ext-link>https://orcid.org/0000-0001-7510-4355</ext-link></contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1 aff2">
          <name><surname>Yu</surname><given-names>Jiajun</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff6">
          <name><surname>Kolosov</surname><given-names>Alexander</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1 aff2">
          <name><surname>Gao</surname><given-names>Wei</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1 aff2">
          <name><surname>Zhou</surname><given-names>Zhen</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1 aff2">
          <name><surname>Huang</surname><given-names>Zhengxu</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="yes" rid="aff1 aff2">
          <name><surname>Li</surname><given-names>Lei</given-names></name>
          <email>lileishdx@163.com</email>
        <ext-link>https://orcid.org/0000-0002-6211-1668</ext-link></contrib>
        <aff id="aff1"><label>1</label><institution>Institute of Mass Spectrometer and Atmospheric Environment, Jinan
University, Guangzhou, 510632, China</institution>
        </aff>
        <aff id="aff2"><label>2</label><institution>Guangdong Provincial Engineering Research Center for On-Line Source
Apportionment System of Air Pollution, <?xmltex \hack{\break}?>Guangzhou, 510632, China</institution>
        </aff>
        <aff id="aff3"><label>3</label><institution>Chernogolovka Branch of the N. N. Semenov Federal Research Center
for Chemical Physics, Russian Academy <?xmltex \hack{\break}?>of Sciences, Chernogolovka, 142432,
Russia</institution>
        </aff>
        <aff id="aff4"><label>4</label><institution>Prokhorov General Physics Institute, Russian Academy of Science,
Moscow, 119991, Russia</institution>
        </aff>
        <aff id="aff5"><label>5</label><institution>Guangzhou Hexin Analytical Instrument Company, Ltd., Guangzhou,
510530, China</institution>
        </aff>
        <aff id="aff6"><label>6</label><institution>N. N. Semenov Federal Research Center for Chemical Physics, Russian
Academy of Sciences, <?xmltex \hack{\break}?>Moscow, 119991, Russia</institution>
        </aff>
      </contrib-group>
      <author-notes><corresp id="corr1">Lei Li (lileishdx@163.com)</corresp></author-notes><pub-date><day>28</day><month>February</month><year>2020</year></pub-date>
      
      <volume>13</volume>
      <issue>2</issue>
      <fpage>941</fpage><lpage>949</lpage>
      <history>
        <date date-type="received"><day>17</day><month>April</month><year>2019</year></date>
           <date date-type="rev-request"><day>3</day><month>June</month><year>2019</year></date>
           <date date-type="rev-recd"><day>14</day><month>December</month><year>2019</year></date>
           <date date-type="accepted"><day>6</day><month>January</month><year>2020</year></date>
      </history>
      <permissions>
        <copyright-statement>Copyright: © 2020 Ying Chen et al.</copyright-statement>
        <copyright-year>2020</copyright-year>
      <license license-type="open-access"><license-p>This work is licensed under the Creative Commons Attribution 4.0 International License. To view a copy of this licence, visit <ext-link ext-link-type="uri" xlink:href="https://creativecommons.org/licenses/by/4.0/">https://creativecommons.org/licenses/by/4.0/</ext-link></license-p></license></permissions><self-uri xlink:href="https://amt.copernicus.org/articles/amt-13-941-2020.html">This article is available from https://amt.copernicus.org/articles/amt-13-941-2020.html</self-uri><self-uri xlink:href="https://amt.copernicus.org/articles/amt-13-941-2020.pdf">The full text article is available as a PDF file from https://amt.copernicus.org/articles/amt-13-941-2020.pdf</self-uri>
      <abstract><title>Abstract</title>
    <p id="d1e212">A single-particle mass spectrometer (SPMS) can provide a
wealth of valuable information on chemical and physical parameters of
individual particles in real time. One of the main performance criteria of
the instrument is efficiency of particle detection (hit rate). Most SPMS
instruments use constant electrical field (DC) extraction, where stationary
high voltage is applied to the extraction electrodes. As the aerosol
particles initially carry a certain charge, those with a high amount to
charge can be deflected by this electric field and lost, thus decreasing the
hit rate. We realized that the delayed extraction technique can eliminate the
stochastic dispersion of the particle beam caused by their deflection in the
stationary electric field. As the result, the hit rate of the instrument can
be significantly improved. Also, as the effect of the deflection in the
electric field is mass dependent, it can cause distortion of the measured
size distribution of the particles. Hence, the delayed extraction technique
can bring the recorded distribution closer to the actual one. We found that
the delayed extraction technique provides a mass resolution improvement as well
as increases the hit rate. The gain in the hit rate depends on the type of
particles. It can be 2 orders of magnitude for model particles and up to
2–4 times for ambient particles. In the present work we report experiments
and results showing the effect of the delayed extraction on the beam
divergence caused by particle charge, the hit rate improvement, and the
effect of the delayed extraction on the measured particle size distribution.</p>
  </abstract>
    </article-meta>
  </front>
<body>
      

<sec id="Ch1.S1" sec-type="intro">
  <label>1</label><title>Introduction</title>
      <p id="d1e224">Aerosol particles have a strong impact on the climate, environment and human
health. These effects are closely related to physical and chemical
properties of the particles. An in-depth understanding of the physical and
chemical properties of individual particles is important for studying
various effects of aerosols. A single-particle mass spectrometer (SPMS) is
an analytical tool that can provide the particle size and chemical
composition of individual particles in real time. There are many
publications that have reviewed the principle, structure, and applications of a
SPMS in detail (Murphy, 2007; Nash et al., 2006; Noble and Prather,
2000; and Pratt and Prather, 2012). SPMSs usually use an aerodynamic lens to focus
and form an aerosol beam and to transfer the beam from the atmosphere into
the vacuum system. As a common arrangement, a two-beam sizing laser measures
the particle speed and size right before the “sized” particle passes into
a high-power pulsed laser beam. When the particles are<?pagebreak page942?> evaporated and
ionized by the laser radiation, positive and negative ions are formed. These
ions are accelerated by the strong electric field in the time-of-flight mass
spectrometer (TOF MS) and then detected and recorded by a data acquisition
system based on a fast analog-to-digital converter card (ADC). Compared to
the traditional offline particle methods, the SPMS measurements can be
conducted without a preliminary collection and preparation of the particles,
providing a rapid real-time analysis. At present, a SPMS is mainly used for the
characterization of physical and chemical properties of the particles,
atmospheric chemical process analysis, atmospheric aerosol analysis, and
other environmental and material science applications (Bi et al., 2011; Kim
et al., 2012; Li et al., 2014; Middlebrook et al., 2003; and Roth et al., 2016).
A SPMS is fast, and detection time for a single particle is typically on the
order of a few milliseconds. However, it is common to accumulate single-particle data over a period of time to obtain statistically significant
information about particle diversity. The application of a SPMS to
atmospheric aerosol analysis and investigation of chemical processes with
atmospheric aerosols requires instruments with a high efficiency of particle
detection (hit rate). Hit rate is an important parameter of a SPMS related to
its performance. It is generally defined as the ratio of the number of
spectra generated by laser ionization to the number of particles detected by
sizing lasers. There are many aspects that affect the hit rate of a SPMS,
including the particle size, physical and chemical properties (Dall'Osto
et al., 2006; Zelenyuk et al., 2009) of the particle, and the instrument
design (Gemayel et al., 2016; Zhao et al., 2005).</p>
      <p id="d1e227">It is well known (Dodd, 1953; Gunn and Woessner, 1956; and Woessner and Gunn,
1956) that aerosol particles usually carry a certain amount of surface
charge. If such particles enter the stationary electrical field in the
extraction region of a SPMS, they can be deflected as they move through the
ion source region. Once the particle trajectory has shifted from the center
of the laser beam, the ionization efficiency can drop down noticeably
(Su et al., 2004). Hence the particle deflection in the ion
source results in a decrease in the hit rate. In order to reduce the
influence of the electrical field on the particle beam trajectory, an
electrostatic neutralizer can be installed between the particle introduction
and particle ionization regions. The electrostatic neutralizer generally
uses a radioactive source based on <inline-formula><mml:math id="M1" display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">210</mml:mn></mml:msup></mml:math></inline-formula>Po, <inline-formula><mml:math id="M2" display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">241</mml:mn></mml:msup></mml:math></inline-formula>Am, or <inline-formula><mml:math id="M3" display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">85</mml:mn></mml:msup></mml:math></inline-formula>Kr, or it uses an X-ray source
(Demokritou et al., 2004; Kousaka et al., 1981; Nicosia et al., 2018;
Pratt et al., 2009; and Tsai et al., 2005). However, the management of
radioactive sources is a difficult task; the equipment is relatively
expensive, and it is very inconvenient to use it in field measurements
because of strict regulations restricting the handling, transport, and
storage of radioactive sources (Nicosia et al., 2018).</p>
      <p id="d1e257">Another problem for a SPMS is particle beam divergence. There are many
instruments that use Nd:YAG laser with fourth harmonic generation (266 nm
wavelength) as the ionization source. Since the Nd:YAG laser with lamp
pumping requires a more than 100 <inline-formula><mml:math id="M4" display="inline"><mml:mrow class="unit"><mml:mi mathvariant="normal">µ</mml:mi></mml:mrow></mml:math></inline-formula>s delay between the lamp ignition and
the laser pulse, it is necessary to place the laser ionization source at a
considerable distance (about 10 cm) downstream of the second particle-sizing
laser. The particle beam is dispersed on the way from the sizing laser to
the ionizing laser, and the increase in the beam width leads to a decrease
in the hit rate. This problem can be solved using a laser with a shorter delay
time, such as excimer or nitrogen pulsed gas lasers (Alaime et al., 1983;
Trimborn et al., 2002). It is well known that the delay time for the pulsed
gas laser is less than 100 ns. The small delay time makes it possible to
reduce the distance between the sizing laser and the ion source, which will
be limited by the mass spectrometer design in this case. The effect can be especially
bigger for irregular nonspherical particles (Zelenyuk et
al., 2006). Shortening the distance between the sizing laser and the
ionizing laser can compensate for the influence of the particle shape to some
extent (Cziczo et al., 2006; Thomson et al., 1997).</p>
      <p id="d1e268">Zelenyuk et al. (2009) also studied the shift of the trigger
signals generated by particles with different sizes at the sizing laser,
which resulted in the hit rate being affected by the particle size. As a
result, the dynamic laser trigger system was developed to implement the
trigger compensation for the particles of different sizes, thereby
achieving an improvement in the hit rate. However, this method is directed
to solve the positional spread of the aerosol particles along the axis; it
does not contribute to solving the problem of the lowered hit rate caused by
deflection of the aerosol in the direction perpendicular to the particle
beam axis. This deflection is caused by the static electrical field and depends
on the ion source dimensions and the electrical field strength.
Unfortunately, the ion source cannot be built short enough to neglect
particle deflection. But it is possible to switch off the electrical field
by using a delayed extraction (DE) technique (Li et al., 2018;
Chudinov et al., 2019) for ion sampling.</p>
      <p id="d1e272">In our previous works we have used the double exponential pulse delayed
extraction technique to improve the mass resolution of a SPMS (Chudinov et
al., 2019; Li et al., 2018). This solution provides a mass resolution
enhancement of about 2–3 times over the broad mass range in comparison to
the usual constant electrical field (DC) extraction. It is important that in
the delayed extraction, ion source particles pass to the middle of the ion
source with no DC electrical field. Therefore, the influence of the particle
charge on its trajectory will be neglected in comparison with the usual ion
source with a stationary electrical field. In this paper we present our
investigation of the effect of stationary and pulse extraction on the
particle trajectory and the hit rate of the SPMS instrument.</p>
</sec>
<sec id="Ch1.S2">
  <label>2</label><title>Instruments and methods</title>
      <p id="d1e283">The SPMS used in this study is a commercial instrument from Hexin Analytical
Instrument Co., Ltd. (Guangzhou), which is normally called a single-particle
aerosol mass<?pagebreak page943?> spectrometer (SPAMS). The principle and design of the
SPAMS has been described elsewhere in detail (Li et al., 2011).
The commercial SPAMS was modified in two ways. First, the delayed extraction
technique is used instead of the original DC field extraction technology.
The superposition of the rectangular and the exponentially shaped extraction
pulses simultaneously improves the resolution over the wide mass range for
positive and negative ions. The detailed information on the instrument
changes required to implement the delayed extraction is provided elsewhere
(Li et al., 2018). After the improvement, a positive ion mass resolution
of <inline-formula><mml:math id="M5" display="inline"><mml:mrow><mml:mo>&gt;</mml:mo><mml:mn mathvariant="normal">1000</mml:mn></mml:mrow></mml:math></inline-formula> full width at half maximum (FWHM) and a negative ion mass resolution of
<inline-formula><mml:math id="M6" display="inline"><mml:mrow><mml:mo>&gt;</mml:mo><mml:mn mathvariant="normal">2000</mml:mn></mml:mrow></mml:math></inline-formula> FWHM were achieved. The second improvement is the use of
multichannel superimposed signal acquisition technology. The native signal
from the microchannel plate (MCP) detector is equally split in two channels with different
amplification ratios and acquired by two equal 8 bit ADC. Thus, the high
dynamic range data acquisition system is capable of acquiring signals
ranging from 5 to 20 000 mV, and the dynamic range is nearly 40 times higher
than that of the original SPAMS. This solution enables one, for instance, to
detect high-intensity alkali metal ion peaks together with very weak ion
signals from other elements or molecules at the same time (Shen et
al., 2018).</p>
      <p id="d1e306">The test particles used in the experiment were standard polystyrene latex
microspheres (PSL) with particle sizes of 320, 510, 720, 960, and 1400 nm
purchased from Duke Scientific. The commercial aerosol generator TSI 9302
was used to atomize the aqueous solution of the PSL and to produce a
controlled beam of particles. The airflow with PSL particles was dried by a
diffusion drying tube and then passed to the SPAMS for analysis.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F1"><?xmltex \currentcnt{1}?><label>Figure 1</label><caption><p id="d1e311">Experimental setup. The glass plate position and
extraction region dimensions are shown for two cases: <bold>(a)</bold> stationary
electrical field extraction and <bold>(b)</bold> delayed extraction.</p></caption>
        <?xmltex \igopts{width=241.848425pt}?><graphic xlink:href="https://amt.copernicus.org/articles/13/941/2020/amt-13-941-2020-f01.png"/>

      </fig>

      <p id="d1e327">For an independent particle size distribution measurement, the glass plate
was installed under the ion source as shown in Fig. 1. The distance between
the center of the ion source and the plate surface was 31.5 mm. After the
glass plate was exposed to the particle flow, the photo of the particles
collected on the glass plate surface was acquired by an Olympus CX31RTSF
microscope. The density profile of the collected particles was extracted by
free ImageJ software. Trajectories of particles shown in
Fig. 1 illustrate the deflection in the electrical field. The deflection
depends on the charge and mass of the particle. The deflection and
trajectory shape is discussed below in the Sect. 3.1</p>
<sec id="Ch1.S2.SS1">
  <label>2.1</label><title>Key factors affecting the efficiency</title>
      <p id="d1e337">Figure 1 shows a schematic representation of the SPAMS ion source. The Nd:YAG laser beam is focused in the center of the ion source where particles
are ionized by laser radiation. The 266 nm laser operates in TEM<inline-formula><mml:math id="M7" display="inline"><mml:msub><mml:mi/><mml:mn mathvariant="normal">00</mml:mn></mml:msub></mml:math></inline-formula> mode
with a Gaussian distribution. The <inline-formula><mml:math id="M8" display="inline"><mml:mrow><mml:msup><mml:mi>e</mml:mi><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:mrow></mml:math></inline-formula> width of the focused Gaussian laser
beam is about 300 <inline-formula><mml:math id="M9" display="inline"><mml:mrow class="unit"><mml:mi mathvariant="normal">µ</mml:mi></mml:mrow></mml:math></inline-formula>m. Factors that actually affect the ionization
efficiency are the total laser pulse energy and the location of the particle
in the focused laser spot. As the particle size is more than 300 times
smaller than that of the focused laser beam width, each particle is ionized
by a virtually uniform laser power density. But when the beam of the
particles is wider than the diameter of the laser beam in the ionization
region, the particles in the beam can be exposed to very different laser
energy fluencies, resulting in a great difference in ionization efficiency.
Therefore, when the particles travel near the edge of the laser spot, it
becomes impossible to generate enough ions to detect such particles.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F2" specific-use="star"><?xmltex \currentcnt{2}?><label>Figure 2</label><caption><p id="d1e373">Comparison of hit rate dependencies for different sizes of PSL
particles in two cases: <bold>(a)</bold> constant electric field (DC extraction) and <bold>(b)</bold> pulsed electric field (delayed extraction; DE).</p></caption>
          <?xmltex \igopts{width=341.433071pt}?><graphic xlink:href="https://amt.copernicus.org/articles/13/941/2020/amt-13-941-2020-f02.png"/>

        </fig>

      <p id="d1e388">In our experiments, the Ultra compact Q-switched Nd:YAG laser (Quantel Laser) was
used with a pulse width of 7.2 ns. The pulse energy of the laser was set to
0.6 mJ, which corresponds to a laser power density in the focal point of
about <inline-formula><mml:math id="M10" display="inline"><mml:mrow><mml:mn mathvariant="normal">2</mml:mn><mml:mo>×</mml:mo><mml:msup><mml:mn mathvariant="normal">10</mml:mn><mml:mn mathvariant="normal">8</mml:mn></mml:msup></mml:mrow></mml:math></inline-formula> W cm<inline-formula><mml:math id="M11" display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">2</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>.</p>
      <p id="d1e419">When the delayed extraction technique is not used, the two DC high-voltage
potentials are constantly applied to the extraction electrodes of the
original bipolar SPAMS ion source. When the charged particles enter the
space of the ion source, they start to be deflected under the action of the
DC electrical field. The amount of deflection is related to the speed and
the mass of the particles, the amount of the surface charge, and the
electrical field strength. Thus, during the flight through the ion source,
the particles gradually deviate from the laser center point, resulting in
the decrease in the ionization efficiency and a significant drop in the hit
rate.</p>
      <p id="d1e422">When the delayed extraction technique is used, the particles enter the
extraction region while the electrical field is kept at zero. Then the laser
ionizes the particle, and the high voltage extraction pulse is applied with 100 ns
delay after the laser pulse. So there is no electric field between the ion
source plates before the particles are ionized by the laser. Thus, the
charged aerosol particles are not deflected by the<?pagebreak page944?> electric field force
during their flight, thereby achieving a higher hit rate.</p>
</sec>
</sec>
<sec id="Ch1.S3">
  <label>3</label><title>Results and discussions</title>
<sec id="Ch1.S3.SS1">
  <label>3.1</label><title>The experiment with PSL particles</title>
      <p id="d1e441">The influence of the electric field on the hit rate was studied by using PSL
beads of five different particle sizes. In Fig. 2 the dependencies of the hit
rate on the electrical field strength for DC and DE modes are shown. It can
be seen that under the same DC electric field strength (Fig. 2a), the hit
rate of the particles from 320 to 1400 nm increases, indicating that the
effect of the electric field on the small particles is more pronounced. As
the DC electric field strength increases, the hit rate of the particles of
each size gradually decreases, indicating that the particles are deflected
by the electric field force.</p>
      <p id="d1e444">Under the DE mode, the hit rate for the particles of all sizes is obviously
increased in comparison with the DC case (Fig. 2b). When the pulsed
electric field strength exceeds 60 kV m<inline-formula><mml:math id="M12" display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>, the hit rate of the
particles above 520 nm is close to 100 %, indicating that the flight path
without an electrical field is essential for the increase in the hit rate of
the particles. The hit rate of the 320 nm PSL particles is 40 %, which is
much lower than 100 % of other particles. This can be due to the focusing
effect of the aerodynamic lens. The divergence of the particle beam itself
is higher than that for the particles above 500 nm, which results in a
decrease in the hit rate. In addition, when the electric field strength
increases from 0 to 60 kV m<inline-formula><mml:math id="M13" display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>, the hit rate gradually increases from
5 % to 100 %. We speculate that the increase in the hit rate with the
electrical field strength in this range may be caused by the increase in the
resolving power of the TOF MS (Cotter, 1994). The lower the extraction
electrical field strength is, the bigger the turnaround time in the TOF MS is,
which in turn decreases the resolving power of the instrument. In the case
of a large turnaround time, the peak width increases and the peak height
decreases. This, in turn, can result in a loss of measured ion current
because of an ADC threshold adjusted to reject the low-intensity peaks
accounted as noise. The loss of the recorded ion current can in turn result
in a miscount of some particles, thus decreasing the hit rate. Also, for the
low extraction electric field, ion divergence can be enhanced, resulting
in a loss of ion current during the flight of the ion to the detector.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F3"><?xmltex \currentcnt{3}?><label>Figure 3</label><caption><p id="d1e473">The gain in the hit rate achieved for different sizes of
PSL particles plotted over the extraction electric field strength.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://amt.copernicus.org/articles/13/941/2020/amt-13-941-2020-f03.png"/>

        </fig>

      <p id="d1e483">In Fig. 3 dependencies of the hit rate gain achieved for the DE mode compared
to the DC mode case are shown. It can be seen that for different particle sizes,
the hit rate gain increases sharply with the increase in the electric field
strength, especially for small particles. For 320 nm particles, the hit
rate can be increased by more than 100 times at an electrical field strength
over 180 kV m<inline-formula><mml:math id="M14" display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>. This result shows that the DE technique can
effectively eliminate the influence of the electric field intensity on the
flight path of the particles, avoid the use of additional peripheral devices
such as electrostatic neutralizers, and result in an improvement in the hit
rate.</p>
      <p id="d1e498">It is worth noting that the mass spectra obtained with DE have a higher
resolution, but they are approximately the same set of peaks as those obtained
with the injection of ions in a constant field (CF). Peaks in the mass
spectrum obtained<?pagebreak page945?> with DE are narrower and, on average, have a larger
amplitude. The influence of a <inline-formula><mml:math id="M15" display="inline"><mml:mrow><mml:mo>∼</mml:mo><mml:mn mathvariant="normal">100</mml:mn></mml:mrow></mml:math></inline-formula> ns delay is not reflected in
the composition of peaks in acquired mass spectra. Mass spectra of PSL
particles acquired in DE and DC modes are presented in the Supplement for comparison. The mechanisms of ion formation as part of a
complex of processes during laser exposure to a particle are of great
interest and are the subject of our further research. In this work we did
not focus on these issues, since they are not directly related to the
observed increase in the hit rate.</p>
      <p id="d1e511">In order to further confirm that the particles are deflected by the action
of the electric field, we conducted several further experiments. Taking the
720 nm PSL microsphere as an example, we compared the hit rate of 720 nm PSL
beads under the constant DC electrical field only to the hit rate of those under the DC electric
field with an electrostatic neutralizer and to the hit rate under the delayed extraction combined with or without a neutralizer. Figure 4 shows the hit rate value for a measurement
period of <inline-formula><mml:math id="M16" display="inline"><mml:mrow><mml:mo>∼</mml:mo><mml:mn mathvariant="normal">30</mml:mn></mml:mrow></mml:math></inline-formula> min. It can be seen that the particle hit
rate after the use of the electrostatic neutralizer has increased from an
average level of 20 % to a level exceeding 85 %, indicating that the
surface charge of the particles has a great influence on the hit rate.
However, the average hit rate of the particles obtained by the DE technique
is close to 96 %, which is slightly higher than that of the hit rate
measured in the case when the electrostatic neutralizer was used. This
difference can be caused by a residual charge remaining on the particles
even after the usage of an X-ray neutralizer (TSI 3088). The higher hit rate
achieved by delayed extraction is caused by the absence of electric field
influence on the particle motion.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F4"><?xmltex \currentcnt{4}?><label>Figure 4</label><caption><p id="d1e526">Hit rate count for four cases: DC extraction with no
neutralizer, DC extraction with the neutralizer on, delayed extraction with no
neutralizer, and delayed extraction with the neutralizer on.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://amt.copernicus.org/articles/13/941/2020/amt-13-941-2020-f04.png"/>

        </fig>

      <p id="d1e535">A further two experiments were performed to measure particle beam divergence
under the influence of the electrical field formed between two electrodes of the
extraction region of the SPMS. The particles were collected on the surface of
the glass plate, as it is shown in Fig. 1. In the first experiment (Fig. a, b) no voltage was applied to the electrodes, and the resulting spot just
represents a 720 nm particle beam profile as a result of the aerodynamic dynamic
lens action. The profile was fitted by a Gauss envelope, and the average
beam width extracted from the Gaussian curve is a little more than 0.1 mm.
The second experiment (Fig. 6a, b) was performed with DC potentials
(<inline-formula><mml:math id="M17" display="inline"><mml:mrow><mml:mo>+</mml:mo><mml:mo>/</mml:mo><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1000</mml:mn></mml:mrow></mml:math></inline-formula> V) applied to the extraction region electrodes, in accordance with
the usual working conditions of the DC extraction mode. The trace of
the same 720 nm PSL beads on the glass plate is shown in Fig. 6a.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F5" specific-use="star"><?xmltex \currentcnt{5}?><label>Figure 5</label><caption><p id="d1e555">The distribution of 720 nm particles collected on the
glass plate without electrical field. <bold>(a)</bold> Photo of the spot from particles
collected on the glass plate. <bold>(b)</bold> The density profile extracted from <bold>(a)</bold> by
ImageJ software and the Gaussian approximation of the profile.</p></caption>
          <?xmltex \igopts{width=341.433071pt}?><graphic xlink:href="https://amt.copernicus.org/articles/13/941/2020/amt-13-941-2020-f05.png"/>

        </fig>

      <?pagebreak page946?><p id="d1e573">The particle divergence is obviously increased in this case. Using the
ImageJ program, the trace profile was extracted, and then
the Gaussian fitting was uploaded on the data. The space distribution of the
particles can be easily transformed into a particle charge distribution as we
deal with the calibrated PSL beads. The ion beam displacement at the exit of
the extraction region (Fig. 6a, b) was analyzed by Liu et
al. (1995), and it can be described by the simple equation:
            <disp-formula id="Ch1.E1" content-type="numbered"><label>1</label><mml:math id="M18" display="block"><mml:mrow><mml:mi mathvariant="normal">Δ</mml:mi><mml:msub><mml:mi>x</mml:mi><mml:mn mathvariant="normal">1</mml:mn></mml:msub><mml:mo>=</mml:mo><mml:mstyle displaystyle="true"><mml:mfrac style="display"><mml:mn mathvariant="normal">1</mml:mn><mml:mn mathvariant="normal">2</mml:mn></mml:mfrac></mml:mstyle><mml:mi>a</mml:mi><mml:msup><mml:mi>t</mml:mi><mml:mn mathvariant="normal">2</mml:mn></mml:msup><mml:mo>=</mml:mo><mml:mstyle displaystyle="true"><mml:mfrac style="display"><mml:mn mathvariant="normal">1</mml:mn><mml:mn mathvariant="normal">2</mml:mn></mml:mfrac></mml:mstyle><mml:mstyle displaystyle="true"><mml:mfrac style="display"><mml:mi>q</mml:mi><mml:mi>m</mml:mi></mml:mfrac></mml:mstyle><mml:mstyle displaystyle="true"><mml:mfrac style="display"><mml:mrow><mml:msub><mml:mi>V</mml:mi><mml:mi>H</mml:mi></mml:msub></mml:mrow><mml:mrow><mml:msub><mml:mi>m</mml:mi><mml:mi mathvariant="normal">p</mml:mi></mml:msub><mml:mi>d</mml:mi></mml:mrow></mml:mfrac></mml:mstyle><mml:msup><mml:mfenced open="(" close=")"><mml:mstyle displaystyle="true"><mml:mfrac style="display"><mml:mrow><mml:msub><mml:mi>L</mml:mi><mml:mn mathvariant="normal">1</mml:mn></mml:msub></mml:mrow><mml:mrow><mml:msub><mml:mi>U</mml:mi><mml:mi mathvariant="normal">p</mml:mi></mml:msub></mml:mrow></mml:mfrac></mml:mstyle></mml:mfenced><mml:mn mathvariant="normal">2</mml:mn></mml:msup><mml:mo>.</mml:mo></mml:mrow></mml:math></disp-formula>
          Here <inline-formula><mml:math id="M19" display="inline"><mml:mi>q</mml:mi></mml:math></inline-formula> and <inline-formula><mml:math id="M20" display="inline"><mml:mrow><mml:msub><mml:mi>m</mml:mi><mml:mi mathvariant="normal">p</mml:mi></mml:msub></mml:mrow></mml:math></inline-formula> are particle charge and mass, <inline-formula><mml:math id="M21" display="inline"><mml:mi>d</mml:mi></mml:math></inline-formula> and <inline-formula><mml:math id="M22" display="inline"><mml:mrow><mml:msub><mml:mi>V</mml:mi><mml:mi>H</mml:mi></mml:msub></mml:mrow></mml:math></inline-formula> are the
distance and voltage difference between two flat extraction electrodes,
<inline-formula><mml:math id="M23" display="inline"><mml:mrow><mml:msub><mml:mi>L</mml:mi><mml:mn mathvariant="normal">1</mml:mn></mml:msub></mml:mrow></mml:math></inline-formula> is the electrode height, <inline-formula><mml:math id="M24" display="inline"><mml:mrow><mml:msub><mml:mi>U</mml:mi><mml:mi mathvariant="normal">p</mml:mi></mml:msub></mml:mrow></mml:math></inline-formula> is the particle longitudinal
velocity, <inline-formula><mml:math id="M25" display="inline"><mml:mi>a</mml:mi></mml:math></inline-formula> is the acceleration of the particle caused by the extraction
electrical field, and <inline-formula><mml:math id="M26" display="inline"><mml:mi>t</mml:mi></mml:math></inline-formula> is the particle dwell time in the extraction region.
Thus, the particle trajectories are parabolic, scaled by their mass and
charge. The equation was derived from a simplified assumption that the
particle moving between the plates is accelerated by the constant horizontal
electrostatic force. Then the particle with the horizontal speed component moves until
it hits the glass gathering plate. The resulting particle displacement in
the position of the gathering glass plate is as follows:
            <disp-formula id="Ch1.E2" content-type="numbered"><label>2</label><mml:math id="M27" display="block"><mml:mrow><mml:mi mathvariant="normal">Δ</mml:mi><mml:mi>x</mml:mi><mml:mo>=</mml:mo><mml:mi mathvariant="normal">Δ</mml:mi><mml:msub><mml:mi>x</mml:mi><mml:mn mathvariant="normal">1</mml:mn></mml:msub><mml:mo>+</mml:mo><mml:mi mathvariant="normal">Δ</mml:mi><mml:msub><mml:mi>x</mml:mi><mml:mn mathvariant="normal">2</mml:mn></mml:msub><mml:mo>=</mml:mo><mml:mstyle displaystyle="true"><mml:mfrac style="display"><mml:mn mathvariant="normal">1</mml:mn><mml:mn mathvariant="normal">2</mml:mn></mml:mfrac></mml:mstyle><mml:mstyle displaystyle="true"><mml:mfrac style="display"><mml:mrow><mml:mi>q</mml:mi><mml:msub><mml:mi>V</mml:mi><mml:mi>H</mml:mi></mml:msub></mml:mrow><mml:mrow><mml:msub><mml:mi>m</mml:mi><mml:mi mathvariant="normal">p</mml:mi></mml:msub><mml:mi>d</mml:mi></mml:mrow></mml:mfrac></mml:mstyle><mml:mstyle displaystyle="true"><mml:mfrac style="display"><mml:mrow><mml:msub><mml:mi>L</mml:mi><mml:mn mathvariant="normal">1</mml:mn></mml:msub><mml:mfenced open="(" close=")"><mml:mrow><mml:msub><mml:mi>L</mml:mi><mml:mn mathvariant="normal">1</mml:mn></mml:msub><mml:mo>+</mml:mo><mml:mn mathvariant="normal">2</mml:mn><mml:msub><mml:mi>L</mml:mi><mml:mn mathvariant="normal">2</mml:mn></mml:msub></mml:mrow></mml:mfenced></mml:mrow><mml:mrow><mml:msubsup><mml:mi>U</mml:mi><mml:mi mathvariant="normal">p</mml:mi><mml:mn mathvariant="normal">2</mml:mn></mml:msubsup></mml:mrow></mml:mfrac></mml:mstyle><mml:mo>.</mml:mo></mml:mrow></mml:math></disp-formula>
          Here <inline-formula><mml:math id="M28" display="inline"><mml:mrow><mml:msub><mml:mi>L</mml:mi><mml:mn mathvariant="normal">2</mml:mn></mml:msub></mml:mrow></mml:math></inline-formula> is the distance from the lower edge of the extraction electrode
to the gathering glass plate surface. The resulting particle displacement is
proportional to the particle charge; so from the trace in Fig. 6a we can
extract the particle charge distribution shown in Fig. 6b. The average
particle charge (in units of e) extracted from the Gaussian envelope is 60 elementary
charges. There are multiple publications in which this electrification charge
was measured for different types of particles (Dodd, 1953). The
results reported in the literature (40 charges per 700 nm road dust
particle; Forsyth et al., 1998; 40 charges per 400 nm PSL particle; Kleefsman et al., 2008) are in reasonable agreement with our estimation made for 720 nm PSL beads.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F6" specific-use="star"><?xmltex \currentcnt{6}?><label>Figure 6</label><caption><p id="d1e830">The distribution of 720 nm particles moved through the
electrical field formed with a 2000 V potential difference between the
electrodes in the extraction region collected on the glass plate. <bold>(a)</bold> Photo
of the trace from particles on the glass plate. <bold>(b)</bold> The extracted
particle charge distribution from the density profile in <bold>(a)</bold> by ImageJ
software and the Gaussian approximation of the profile.</p></caption>
          <?xmltex \igopts{width=398.338583pt}?><graphic xlink:href="https://amt.copernicus.org/articles/13/941/2020/amt-13-941-2020-f06.png"/>

        </fig>

      <p id="d1e848">Using known particle mass, velocity, extraction voltage, and electrodes
dimensions, we have extracted the particle displacement per unit charge in
the middle point on which the ionizing laser is focused. For the parameters
of our instrument this displacement is 12.5 <inline-formula><mml:math id="M29" display="inline"><mml:mrow class="unit"><mml:mi mathvariant="normal">µ</mml:mi></mml:mrow></mml:math></inline-formula>m z<inline-formula><mml:math id="M30" display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>, where <inline-formula><mml:math id="M31" display="inline"><mml:mi>z</mml:mi></mml:math></inline-formula> is the
number of elementary charges on the particle. Hence, a particle with an
average of 20 elementary charges will be deflected <inline-formula><mml:math id="M32" display="inline"><mml:mrow><mml:mo>∼</mml:mo><mml:mn mathvariant="normal">250</mml:mn></mml:mrow></mml:math></inline-formula> <inline-formula><mml:math id="M33" display="inline"><mml:mrow class="unit"><mml:mi mathvariant="normal">µ</mml:mi></mml:mrow></mml:math></inline-formula>m from the ionizing laser spot center. Taking into account the Gaussian
profile of the focused laser beam of 300 <inline-formula><mml:math id="M34" display="inline"><mml:mrow class="unit"><mml:mi mathvariant="normal">µ</mml:mi></mml:mrow></mml:math></inline-formula>m, the hit efficiency will be
greatly affected by such a displacement.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F7" specific-use="star"><?xmltex \currentcnt{7}?><label>Figure 7</label><caption><p id="d1e907"><bold>(a)</bold> Particles size distribution of real aerosol
particles. Blue circles – DC extraction; red asterisks – delayed extraction.
<bold>(b)</bold> The hit rate gain achieved by delayed extraction (DE) plotted over the
particles size.</p></caption>
          <?xmltex \igopts{width=341.433071pt}?><graphic xlink:href="https://amt.copernicus.org/articles/13/941/2020/amt-13-941-2020-f07.png"/>

        </fig>

</sec>
<sec id="Ch1.S3.SS2">
  <label>3.2</label><title>Comparison of environmental testing</title>
      <p id="d1e929">The comparison of the detection efficiency for the instrument with DC
extraction and with DE used for the actual atmospheric particulate matter
detection is shown in Fig. 7a and b. Aerosol particles from laboratory room air
were analyzed by the SPMS used in this work. In order to ensure that aerosol
composition is stable during the experiment, the DC and DE modes of
operation were switched alternately after 10 min. Four sets of both DC
and DE extraction data were acquired in 80 min. The hit rate was
calculated as the ratio of the number of ionized particles to the number of
sized particles. The hit rate data were extracted for the groups of
particles selected by size; bin value was 20 nm while the whole range was
200–1000 nm. The data of four sets of 10 min each, both for DC and DE
extraction modes, were summed. In the DC mode, in 40 min, a total of
12 174 particles was recorded, of which 4569 were ionized, while in the DE
mode, 12 030 particles were recorded, and 8228 were ionized for the same
time. It is seen from Fig. 7a that the center of the distribution obtained
with DC extraction (blue circles) is shifted to the higher particles size.
This behavior corresponds to our results obtained for the model PSL
particles. Figure 7b shows the hit rate gain from the DE technique to the DC
extraction, obtained from the data plotted in Fig. 7a. It should be noted
that usage of delayed extraction eliminates the distortion of particle
distribution, which results from the dispersion of particle deflection by
their size in the case of DC extraction. The integral hit rate for the
experiment with the ambient particles is <inline-formula><mml:math id="M35" display="inline"><mml:mrow><mml:mo>∼</mml:mo><mml:mn mathvariant="normal">25</mml:mn></mml:mrow></mml:math></inline-formula> % higher in the
case of delayed extraction (Fig. 7a), while the gain in hit rate achieved for
small particles is <inline-formula><mml:math id="M36" display="inline"><mml:mrow><mml:mo>∼</mml:mo><mml:mn mathvariant="normal">4</mml:mn></mml:mrow></mml:math></inline-formula> times bigger than that measured for DC
extraction (Fig. 7b). The increased effect for small particles can be
explained by Eq. (1). The smaller the <inline-formula><mml:math id="M37" display="inline"><mml:mrow><mml:mi>m</mml:mi><mml:mo>/</mml:mo><mml:mi>q</mml:mi></mml:mrow></mml:math></inline-formula> of the particle, the bigger its
deflection in the ion source (extraction region). Hence, the hit rate of
smaller particles is decreased more, and the gain caused by DE should be
higher.</p>
      <p id="d1e964">The hit rate of the atmospheric aerosol is lower than that of the standard
PSL beads mainly due to the shape of the particles. In the actual
atmosphere, the shape of the particles is much more complicated than that of the PSL
sphere. For example, the aged carbon particles tend to be spherical, and the
hit rate is relatively high, while the fresh black carbon<?pagebreak page947?> particles exhibit
a chain structure, so the divergence during their flight is probably bigger,
resulting in lower hit rates (Ghan et al., 2012). In
general, using the delayed extraction technique, the hit rate of environmental
particles with a size of <inline-formula><mml:math id="M38" display="inline"><mml:mrow><mml:mo>&lt;</mml:mo><mml:mn mathvariant="normal">500</mml:mn></mml:mrow></mml:math></inline-formula> nm can be noticeably improved, and
the obtained particle size distribution becomes closer to the real one.</p>
</sec>
</sec>
<sec id="Ch1.S4" sec-type="conclusions">
  <label>4</label><title>Conclusions</title>
      <p id="d1e986">Aerosols often carry surface charges. When charged particles fly through the
SPMS ion source, they can be deflected by the action of constant electric
field, resulting in a substantial drop in the hit rate. When delayed
extraction is used in a SPMS, the instrument efficiency is improved because
the particles are not deflected by the action of the stationary electric
field. The absence of the DC electrical field in this method allows one to
avoid the use of the auxiliary external devices such as electrostatic
neutralizers, maximize the hit rate of the aerosol particles, and decrease
the distortion of the measured size distribution of particles. Experiments
have shown that for PSL microspheres, the smaller the particle size, the
greater the gain in the hit rate. There is no dependence of the hit rate on
the measured particle size for particles with a diameter of more than 500 nm
when DE is used. For the real application it means that this method
minimizes losses in the hit rate, which in turn affects the measured size
distribution of the particles.</p>
      <p id="d1e989">The experiments with environmental aerosol have also shown that SPAMS with
delayed extraction technology can improve the average detection efficiency of
the actual atmospheric aerosol by more than 25 % and improve the hit rate
for small (<inline-formula><mml:math id="M39" display="inline"><mml:mrow><mml:mo>&lt;</mml:mo><mml:mn mathvariant="normal">500</mml:mn></mml:mrow></mml:math></inline-formula> nm) particles by more than 4 times. It should be
noted that the charge of the particles is connected with their parameters.
It means that a preselection of particles based on their charge can be used
to extract extra information from the data acquired by a SPMS. It could be
done, for example, by using two pairs of deflecting plates before<?pagebreak page948?> the ion
source. By choosing the deflection voltage, only particles with a specific
mass<inline-formula><mml:math id="M40" display="inline"><mml:mo>/</mml:mo></mml:math></inline-formula>charge ratio will pass this double deflector and can be ionized in the
ion source.</p>
</sec>

      
      </body>
    <back><notes notes-type="dataavailability"><title>Data availability</title>

      <p id="d1e1014">These data can be publicly accessible in free.</p>
  </notes><app-group>
        <supplementary-material position="anchor"><p id="d1e1017">The supplement related to this article is available online at: <inline-supplementary-material xlink:href="https://doi.org/10.5194/amt-13-941-2020-supplement" xlink:title="pdf">https://doi.org/10.5194/amt-13-941-2020-supplement</inline-supplementary-material>.</p></supplementary-material>
        </app-group><notes notes-type="authorcontribution"><title>Author contributions</title>

      <p id="d1e1026">LL designed the study; YC, XD, and LL performed the experiments;
LL, VK, SN, XD, and ZXH conducted the data analysis; LL, VK, and SN wrote
the paper with the input from all authors.</p>
  </notes><notes notes-type="competinginterests"><title>Competing interests</title>

      <p id="d1e1032">The authors declare that they have no conflict of interest.</p>
  </notes><ack><title>Acknowledgements</title><p id="d1e1038">We would like to thank engineer Wang Jingjing of the Guangzhou Hexin Mass Spectrometer Co., Ltd., for his technical support and Alexandre Loboda and Adam Strock for the language proof.</p></ack><notes notes-type="financialsupport"><title>Financial support</title>

      <p id="d1e1043">This research has been supported by the Guangdong Provincial Natural Science Foundation-Ph.D. Launched Vertical Collaborative Project (grant no. 2017A030310384), the Guangdong International Cooperation Project (grant no. 2018A05056020), 2013–2020 Program of Basic Research of the Russian State Academies of Sciences (nos. AAAA-A18-118112690060-9,
AAAA-A19-119121790091-3),
and the National Key Research Project (grant no. 2017YFC0209506).</p>
  </notes><notes notes-type="reviewstatement"><title>Review statement</title>

      <p id="d1e1049">This paper was edited by Mingjin Tang and reviewed by four anonymous referees.</p>
  </notes><ref-list>
    <title>References</title>

      <ref id="bib1.bib1"><label>1</label><?label 1?><mixed-citation>Alaime, C., Bertrand, J., Bonnet, L., El Masri, Y., Palffy, L., Pellegrin,
P., and Prieels, R.: Stabilization of the light output delay of a pulsed
nitrogen laser, Nucl. Instrum. Methods, 207,
423–427, <ext-link xlink:href="https://doi.org/10.1016/0167-5087(83)90653-1" ext-link-type="DOI">10.1016/0167-5087(83)90653-1</ext-link>, 1983.</mixed-citation></ref>
      <ref id="bib1.bib2"><label>2</label><?label 1?><mixed-citation>Bi, X., Zhang, G., Li, L., Wang, X., Li, M., Sheng, G., Fu, J., and Zhou,
Z.: Mixing state of biomass burning particles by single particle aerosol
mass spectrometer in the urban area of PRD, China, Atmos. Environ.,
45, 3447–3453, <ext-link xlink:href="https://doi.org/10.1016/j.atmosenv.2011.03.034" ext-link-type="DOI">10.1016/j.atmosenv.2011.03.034</ext-link>,
2011.</mixed-citation></ref>
      <ref id="bib1.bib3"><label>3</label><?label 1?><mixed-citation>Chudinov, A., Li, L., Zhou, Z., Huang, Z., Gao, W., Yu, J., Nikiforov, S.,
Pikhtelev, A., Bukharina, A., and Kozlovskiy, V.: Improvement of peaks
identification and dynamic range for bi-polar Single Particle Mass
Spectrometer, Int. J. Mass Spectrom., 436, 7–17,
<ext-link xlink:href="https://doi.org/10.1016/j.ijms.2018.11.013" ext-link-type="DOI">10.1016/j.ijms.2018.11.013</ext-link>, 2019.</mixed-citation></ref>
      <ref id="bib1.bib4"><label>4</label><?label 1?><mixed-citation>Cziczo, D. J., Thomson, D. S., Thompson, T. L., DeMott, P. J., and Murphy,
D. M.: Particle analysis by laser mass spectrometry (PALMS) studies of ice
nuclei and other low number density particles, Int. J. Mass Spectrom., 258, 21–29, <ext-link xlink:href="https://doi.org/10.1016/j.ijms.2006.05.013" ext-link-type="DOI">10.1016/j.ijms.2006.05.013</ext-link>, 2006.</mixed-citation></ref>
      <ref id="bib1.bib5"><label>5</label><?label 1?><mixed-citation>Dall'Osto, M., Harrison, R. M., Beddows, D. C. S., Freney, E. J., Heal, M.
R., and Donovan, R. J.: Single-particle detection efficiencies of aerosol
time-of-flight mass spectrometry during the North Atlantic marine boundary
layer experiment, Environ. Sci. Technol., 40, 5029–5035,
<ext-link xlink:href="https://doi.org/10.1021/es050951i" ext-link-type="DOI">10.1021/es050951i</ext-link>, 2006.</mixed-citation></ref>
      <ref id="bib1.bib6"><label>6</label><?label 1?><mixed-citation>Demokritou, P., Lee, S. J., Ferguson, S. T., and Koutrakis, P.: A compact
multistage (cascade) impactor for the characterization of atmospheric
aerosols, J. Aerosol Sci., 35, 281–299, <ext-link xlink:href="https://doi.org/10.1016/j.jaerosci.2003.09.003" ext-link-type="DOI">10.1016/j.jaerosci.2003.09.003</ext-link>, 2004.</mixed-citation></ref>
      <ref id="bib1.bib7"><label>7</label><?label 1?><mixed-citation>Dodd, E. E.: The Statistics of Liquid Spray and Dust Electrification by the
Hopper and Laby Method, J. Appl. Phys., 24, 73–80, <ext-link xlink:href="https://doi.org/10.1063/1.1721137" ext-link-type="DOI">10.1063/1.1721137</ext-link>, 1953.</mixed-citation></ref>
      <ref id="bib1.bib8"><label>8</label><?label 1?><mixed-citation>Gemayel, R., Hellebust, S., Temime-Roussel, B., Hayeck, N., Van Elteren, J. T., Wortham, H., and Gligorovski, S.: The performance and the characterization of laser ablation aerosol particle time-of-flight mass spectrometry (LAAP-ToF-MS), Atmos. Meas. Tech., 9, 1947–1959, <ext-link xlink:href="https://doi.org/10.5194/amt-9-1947-2016" ext-link-type="DOI">10.5194/amt-9-1947-2016</ext-link>, 2016.</mixed-citation></ref>
      <ref id="bib1.bib9"><label>9</label><?label 1?><mixed-citation>Ghan, S. J., Liu, X., East
er, R. C., Zaveri, R., Rasch, P. J., Yoon, J.-H.,
and Eaton, B.: Toward a Minimal Representation of Aerosols in Climate
Models: Comparative Decomposition of Aerosol Direct, Semidirect, and
Indirect Radiative Forcing, J. Climate, 25, 6461–6476, <ext-link xlink:href="https://doi.org/10.1175/jcli-d-11-00650.1" ext-link-type="DOI">10.1175/jcli-d-11-00650.1</ext-link>, 2012.</mixed-citation></ref>
      <ref id="bib1.bib10"><label>10</label><?label 1?><mixed-citation>Gunn, R. and Woessner, R. H.: Measurements of the systematic
electrification of aerosols, J. Coll. Sci. Imp. U. Tok., 11, 254–259,
<ext-link xlink:href="https://doi.org/10.1016/0095-8522(56)90050-2" ext-link-type="DOI">10.1016/0095-8522(56)90050-2</ext-link>, 1956.</mixed-citation></ref>
      <ref id="bib1.bib11"><label>11</label><?label 1?><mixed-citation>Kim, M., Cahill, J. F., Su, Y., Prather, K. A., and Cohen, S. M.:
Postsynthetic ligand exchange as a route to functionalization of “inert”
metal–organic frameworks, Chem. Sci., 3, 126–130, <ext-link xlink:href="https://doi.org/10.1039/C1SC00394A" ext-link-type="DOI">10.1039/C1SC00394A</ext-link>, 2012.</mixed-citation></ref>
      <ref id="bib1.bib12"><label>12</label><?label 1?><mixed-citation>Kousaka, Y., Okuyama, K., and Endo, Y.: Calibration of differential mobility
analyser by visual method, J. Aerosol Sci., 12, 339–348,
<ext-link xlink:href="https://doi.org/10.1016/0021-8502(81)90023-9" ext-link-type="DOI">10.1016/0021-8502(81)90023-9</ext-link>, 1981.</mixed-citation></ref>
      <ref id="bib1.bib13"><label>13</label><?label 1?><mixed-citation>Li, L., Huang, Z., Dong, J., Li, M., Gao, W., Nian, H., Fu, Z., Zhang, G.,
Bi, X., Cheng, P., and Zhou, Z.: Real time bipolar time-of-flight mass
spectrometer for analyzing single aerosol particles, Int. J. Mass Spectrom., 303, 118–124, <ext-link xlink:href="https://doi.org/10.1016/j.ijms.2011.01.017" ext-link-type="DOI">10.1016/j.ijms.2011.01.017</ext-link>, 2011.</mixed-citation></ref>
      <ref id="bib1.bib14"><label>14</label><?label 1?><mixed-citation>Li, L., Li, M., Huang, Z., Gao, W., Nian, H., Fu, Z., Gao, J., Chai, F., and
Zhou, Z.: Ambient particle characterization by single particle aerosol mass
spectrometry in an urban area of Beijing, Atmos. Environ., 94,
323–331, <ext-link xlink:href="https://doi.org/10.1016/j.atmosenv.2014.03.048" ext-link-type="DOI">10.1016/j.atmosenv.2014.03.048</ext-link>, 2014.</mixed-citation></ref>
      <ref id="bib1.bib15"><label>15</label><?label 1?><mixed-citation>Li, L., Liu, L., Xu, L., Li, M., Li, X., Gao, W., Huang, Z., and Cheng, P.:
Improvement in the Mass Resolution of Single Particle Mass Spectrometry
Using Delayed Ion Extraction, Int. J. Mass Spectrom., 29, 2105–2109, <ext-link xlink:href="https://doi.org/10.1007/s13361-018-2037-4" ext-link-type="DOI">10.1007/s13361-018-2037-4</ext-link>, 2018.</mixed-citation></ref>
      <ref id="bib1.bib16"><label>16</label><?label 1?><mixed-citation>Liu, P., Ziemann, P. J., Kittelson, D. B., and McMurry, P. H.: Generating
Particle Beams of Controlled Dimensions and Divergence: II. Experimental
Evaluation of Particle Motion in Aerodynamic Lenses and Nozzle Expansions,
Aerosol Sci. Technol., 22, 314–324, <ext-link xlink:href="https://doi.org/10.1080/02786829408959749" ext-link-type="DOI">10.1080/02786829408959749</ext-link>, 1995.</mixed-citation></ref>
      <?pagebreak page949?><ref id="bib1.bib17"><label>17</label><?label 1?><mixed-citation>Middlebrook, A. M., Murphy, D. M., Lee, S.-H., Thomson, D. S., Prather, K.
A., Wenzel, R. J., Liu, D.-Y., Phares, D. J., Rhoads, K. P., Wexler, A. S.,
Johnston, M. V., Jimenez, J. L., Jayne, J. T., Worsnop, D. R., Yourshaw, I.,
Seinfeld, J. H., and Flagan, R. C.: A comparison of particle mass
spectrometers during the 1999 Atlanta Supersite Project, J. Geophys. Res.-Atmos., 108,  8424, <ext-link xlink:href="https://doi.org/10.1029/2001jd000660" ext-link-type="DOI">10.1029/2001jd000660</ext-link>, 2003.</mixed-citation></ref>
      <ref id="bib1.bib18"><label>18</label><?label 1?><mixed-citation>Murphy, D. M.: The design of single particle laser mass spectrometers, Mass
Spectrom. Rev., 26, 150–165, <ext-link xlink:href="https://doi.org/10.1002/mas.20113" ext-link-type="DOI">10.1002/mas.20113</ext-link>,
2007.</mixed-citation></ref>
      <ref id="bib1.bib19"><label>19</label><?label 1?><mixed-citation>Nash, D. G., Baer, T., and Johnston, M. V.: Aerosol mass spectrometry: An
introductory review, Int. J. Mass Spectrom., 258, 2–12,
<ext-link xlink:href="https://doi.org/10.1016/j.ijms.2006.09.017" ext-link-type="DOI">10.1016/j.ijms.2006.09.017</ext-link>, 2006.</mixed-citation></ref>
      <ref id="bib1.bib20"><label>20</label><?label 1?><mixed-citation>Nicosia, A., Manodori, L., Trentini, A., Ricciardelli, I., Bacco, D.,
Poluzzi, V., Di Matteo, L., and Belosi, F.: Field study of a soft X-ray
aerosol neutralizer combined with electrostatic classifiers for nanoparticle
size distribution measurements, Particuology, 37, 99–106, <ext-link xlink:href="https://doi.org/10.1016/j.partic.2017.08.001" ext-link-type="DOI">10.1016/j.partic.2017.08.001</ext-link>, 2018.</mixed-citation></ref>
      <ref id="bib1.bib21"><label>21</label><?label 1?><mixed-citation>Noble, C. A. and Prather, K. A.: Real-time single particle mass
spectrometry: a historical review of a quarter century of the chemical
analysis of aerosols, Mass Spectrom. Rev., 19, 248–274, <ext-link xlink:href="https://doi.org/10.1002/1098-2787(200007)19:4&lt;248::AID-MAS3&gt;3.0.CO;2-I" ext-link-type="DOI">10.1002/1098-2787(200007)19:4&lt;248::AID-MAS3&gt;3.0.CO;2-I</ext-link>,
2000.</mixed-citation></ref>
      <ref id="bib1.bib22"><label>22</label><?label 1?><mixed-citation>Pratt, K. A., Mayer, J. E., Holecek, J. C., Moffet, R. C., Sanchez, R. O.,
Rebotier, T. P., Furutani, H., Gonin, M., Fuhrer, K., Su, Y., Guazzotti, S.,
and Prather, K. A.: Development and Characterization of an Aircraft Aerosol
Time-of-Flight Mass Spectrometer, Anal. Chem., 81, 1792–1800,
<ext-link xlink:href="https://doi.org/10.1021/ac801942r" ext-link-type="DOI">10.1021/ac801942r</ext-link>, 2009.</mixed-citation></ref>
      <ref id="bib1.bib23"><label>23</label><?label 1?><mixed-citation>Pratt, K. A. and Prather, K. A.: Mass spectrometry of atmospheric
aerosols – Recent developments and applications, Part II: On-line mass
spectrometry techniques, 31, 17–48, <ext-link xlink:href="https://doi.org/10.1002/mas.20330" ext-link-type="DOI">10.1002/mas.20330</ext-link>, 2012.</mixed-citation></ref>
      <ref id="bib1.bib24"><label>24</label><?label 1?><mixed-citation>Roth, A., Schneider, J., Klimach, T., Mertes, S., van Pinxteren, D., Herrmann, H., and Borrmann, S.: Aerosol properties, source identification, and cloud processing in orographic clouds measured by single particle mass spectrometry on a central European mountain site during HCCT-2010, Atmos. Chem. Phys., 16, 505–524, <ext-link xlink:href="https://doi.org/10.5194/acp-16-505-2016" ext-link-type="DOI">10.5194/acp-16-505-2016</ext-link>, 2016.
</mixed-citation></ref><?xmltex \hack{\newpage}?>
      <ref id="bib1.bib25"><label>25</label><?label 1?><mixed-citation>Shen, W., Dai, X., Huang, Z. X., Hou, Z. H., Cai, W. G., Du, X. B., Zhou,
Z., Li, M., and Li, L.: Improvement of the Dynamic Range of Data Acquisition
System in Single Particle Mass Spectrometry, J. Chinese Cinemas., 39, 331–336, <ext-link xlink:href="https://doi.org/10.7538/zpxb.2017.0119" ext-link-type="DOI">10.7538/zpxb.2017.0119</ext-link>, 2018.</mixed-citation></ref>
      <ref id="bib1.bib26"><label>26</label><?label 1?><mixed-citation>Su, Y., Sipin, M. F., Furutani, H., and Prather, K. A.: Development and
Characterization of an Aerosol Time-of-Flight Mass Spectrometer with
Increased Detection Efficiency, Anal. Chem., 76, 712–719,
<ext-link xlink:href="https://doi.org/10.1021/ac034797z" ext-link-type="DOI">10.1021/ac034797z</ext-link>, 2004.</mixed-citation></ref>
      <ref id="bib1.bib27"><label>27</label><?label 1?><mixed-citation>Thomson, D. S., Middlebrook, A. M., and Murphy, D. M.: Thresholds for
Laser-Induced Ion Formation from Aerosols in a Vacuum Using Ultraviolet and
Vacuum-Ultraviolet Laser Wavelengths, Aerosol Sci. Technol., 26,
544–559, <ext-link xlink:href="https://doi.org/10.1080/02786829708965452" ext-link-type="DOI">10.1080/02786829708965452</ext-link>, 1997.</mixed-citation></ref>
      <ref id="bib1.bib28"><label>28</label><?label 1?><mixed-citation>Trimborn, A., Hinz, K. P., and Spengler, B.: Online analysis of atmospheric
particles with a transportable laser mass spectrometer during LACE 98,
J. Geophys. Res.-Atmos., 107, LAC 13-11-LAC 13-10,
<ext-link xlink:href="https://doi.org/10.1029/2001JD000590" ext-link-type="DOI">10.1029/2001JD000590</ext-link>, 2002.</mixed-citation></ref>
      <ref id="bib1.bib29"><label>29</label><?label 1?><mixed-citation>
Tsai, C.-J., Lin, J.-S., Deshpande, C., and Liu, L.-C.: Electrostatic
Charge Measurement and Charge Neutralization of Fine Aerosol
Particles during the Generation Process, Part. Part. Syst. Charact., 22 293–298, https://doi.org/10.1002/ppsc.200500961, 2005.</mixed-citation></ref>
      <ref id="bib1.bib30"><label>30</label><?label 1?><mixed-citation>Woessner, R. H. and Gunn, R.: Measurements related to the fundamental
processes of aerosol electrification, J. Coll. Sci. Imp. U. Tok., 11, 69–76,
<ext-link xlink:href="https://doi.org/10.1016/0095-8522(56)90021-6" ext-link-type="DOI">10.1016/0095-8522(56)90021-6</ext-link>, 1956.</mixed-citation></ref>
      <ref id="bib1.bib31"><label>31</label><?label 1?><mixed-citation>Zelenyuk, A., Cai, Y., and Imre, D.: From Agglomerates of Spheres to
Irregularly Shaped Particles: Determination of Dynamic Shape Factors from
Measurements of Mobility and Vacuum Aerodynamic Diameters, Aerosol Sci. Technol., 40, 197–217, <ext-link xlink:href="https://doi.org/10.1080/02786820500529406" ext-link-type="DOI">10.1080/02786820500529406</ext-link>, 2006.</mixed-citation></ref>
      <ref id="bib1.bib32"><label>32</label><?label 1?><mixed-citation>Zelenyuk, A., Yang, J., Imre, D., and Choi, E.: Achieving Size Independent
Hit-Rate in Single Particle Mass Spectrometry, Aerosol Sci. Technol., 43, 305–310, <ext-link xlink:href="https://doi.org/10.1080/02786820802637915" ext-link-type="DOI">10.1080/02786820802637915</ext-link>, 2009.</mixed-citation></ref>
      <ref id="bib1.bib33"><label>33</label><?label 1?><mixed-citation>Zhao, Y., Bein, K. J., Wexler, A. S., Misra, C., Fine, P. M., and Sioutas,
C.: Field evaluation of the versatile aerosol concentration enrichment
system (VACES) particle concentrator coupled to the rapid single-particle
mass spectrometer (RSMS-3), J. Geophys. Res.-Atmos.,
110, D07S02, <ext-link xlink:href="https://doi.org/10.1029/2004jd004644" ext-link-type="DOI">10.1029/2004jd004644</ext-link>, 2005.</mixed-citation></ref>

  </ref-list></back>
    <!--<article-title-html>Increase of the particle hit rate in a laser single-particle mass spectrometer by pulse delayed extraction technology</article-title-html>
<abstract-html><p>A single-particle mass spectrometer (SPMS) can provide a
wealth of valuable information on chemical and physical parameters of
individual particles in real time. One of the main performance criteria of
the instrument is efficiency of particle detection (hit rate). Most SPMS
instruments use constant electrical field (DC) extraction, where stationary
high voltage is applied to the extraction electrodes. As the aerosol
particles initially carry a certain charge, those with a high amount to
charge can be deflected by this electric field and lost, thus decreasing the
hit rate. We realized that the delayed extraction technique can eliminate the
stochastic dispersion of the particle beam caused by their deflection in the
stationary electric field. As the result, the hit rate of the instrument can
be significantly improved. Also, as the effect of the deflection in the
electric field is mass dependent, it can cause distortion of the measured
size distribution of the particles. Hence, the delayed extraction technique
can bring the recorded distribution closer to the actual one. We found that
the delayed extraction technique provides a mass resolution improvement as well
as increases the hit rate. The gain in the hit rate depends on the type of
particles. It can be 2 orders of magnitude for model particles and up to
2–4 times for ambient particles. In the present work we report experiments
and results showing the effect of the delayed extraction on the beam
divergence caused by particle charge, the hit rate improvement, and the
effect of the delayed extraction on the measured particle size distribution.</p></abstract-html>
<ref-html id="bib1.bib1"><label>1</label><mixed-citation>
Alaime, C., Bertrand, J., Bonnet, L., El Masri, Y., Palffy, L., Pellegrin,
P., and Prieels, R.: Stabilization of the light output delay of a pulsed
nitrogen laser, Nucl. Instrum. Methods, 207,
423–427, <a href="https://doi.org/10.1016/0167-5087(83)90653-1" target="_blank">https://doi.org/10.1016/0167-5087(83)90653-1</a>, 1983.
</mixed-citation></ref-html>
<ref-html id="bib1.bib2"><label>2</label><mixed-citation>
Bi, X., Zhang, G., Li, L., Wang, X., Li, M., Sheng, G., Fu, J., and Zhou,
Z.: Mixing state of biomass burning particles by single particle aerosol
mass spectrometer in the urban area of PRD, China, Atmos. Environ.,
45, 3447–3453, <a href="https://doi.org/10.1016/j.atmosenv.2011.03.034" target="_blank">https://doi.org/10.1016/j.atmosenv.2011.03.034</a>,
2011.
</mixed-citation></ref-html>
<ref-html id="bib1.bib3"><label>3</label><mixed-citation>
Chudinov, A., Li, L., Zhou, Z., Huang, Z., Gao, W., Yu, J., Nikiforov, S.,
Pikhtelev, A., Bukharina, A., and Kozlovskiy, V.: Improvement of peaks
identification and dynamic range for bi-polar Single Particle Mass
Spectrometer, Int. J. Mass Spectrom., 436, 7–17,
<a href="https://doi.org/10.1016/j.ijms.2018.11.013" target="_blank">https://doi.org/10.1016/j.ijms.2018.11.013</a>, 2019.
</mixed-citation></ref-html>
<ref-html id="bib1.bib4"><label>4</label><mixed-citation>
Cziczo, D. J., Thomson, D. S., Thompson, T. L., DeMott, P. J., and Murphy,
D. M.: Particle analysis by laser mass spectrometry (PALMS) studies of ice
nuclei and other low number density particles, Int. J. Mass Spectrom., 258, 21–29, <a href="https://doi.org/10.1016/j.ijms.2006.05.013" target="_blank">https://doi.org/10.1016/j.ijms.2006.05.013</a>, 2006.
</mixed-citation></ref-html>
<ref-html id="bib1.bib5"><label>5</label><mixed-citation>
Dall'Osto, M., Harrison, R. M., Beddows, D. C. S., Freney, E. J., Heal, M.
R., and Donovan, R. J.: Single-particle detection efficiencies of aerosol
time-of-flight mass spectrometry during the North Atlantic marine boundary
layer experiment, Environ. Sci. Technol., 40, 5029–5035,
<a href="https://doi.org/10.1021/es050951i" target="_blank">https://doi.org/10.1021/es050951i</a>, 2006.
</mixed-citation></ref-html>
<ref-html id="bib1.bib6"><label>6</label><mixed-citation>
Demokritou, P., Lee, S. J., Ferguson, S. T., and Koutrakis, P.: A compact
multistage (cascade) impactor for the characterization of atmospheric
aerosols, J. Aerosol Sci., 35, 281–299, <a href="https://doi.org/10.1016/j.jaerosci.2003.09.003" target="_blank">https://doi.org/10.1016/j.jaerosci.2003.09.003</a>, 2004.
</mixed-citation></ref-html>
<ref-html id="bib1.bib7"><label>7</label><mixed-citation>
Dodd, E. E.: The Statistics of Liquid Spray and Dust Electrification by the
Hopper and Laby Method, J. Appl. Phys., 24, 73–80, <a href="https://doi.org/10.1063/1.1721137" target="_blank">https://doi.org/10.1063/1.1721137</a>, 1953.
</mixed-citation></ref-html>
<ref-html id="bib1.bib8"><label>8</label><mixed-citation>
Gemayel, R., Hellebust, S., Temime-Roussel, B., Hayeck, N., Van Elteren, J. T., Wortham, H., and Gligorovski, S.: The performance and the characterization of laser ablation aerosol particle time-of-flight mass spectrometry (LAAP-ToF-MS), Atmos. Meas. Tech., 9, 1947–1959, <a href="https://doi.org/10.5194/amt-9-1947-2016" target="_blank">https://doi.org/10.5194/amt-9-1947-2016</a>, 2016.
</mixed-citation></ref-html>
<ref-html id="bib1.bib9"><label>9</label><mixed-citation>
Ghan, S. J., Liu, X., East
er, R. C., Zaveri, R., Rasch, P. J., Yoon, J.-H.,
and Eaton, B.: Toward a Minimal Representation of Aerosols in Climate
Models: Comparative Decomposition of Aerosol Direct, Semidirect, and
Indirect Radiative Forcing, J. Climate, 25, 6461–6476, <a href="https://doi.org/10.1175/jcli-d-11-00650.1" target="_blank">https://doi.org/10.1175/jcli-d-11-00650.1</a>, 2012.
</mixed-citation></ref-html>
<ref-html id="bib1.bib10"><label>10</label><mixed-citation>
Gunn, R. and Woessner, R. H.: Measurements of the systematic
electrification of aerosols, J. Coll. Sci. Imp. U. Tok., 11, 254–259,
<a href="https://doi.org/10.1016/0095-8522(56)90050-2" target="_blank">https://doi.org/10.1016/0095-8522(56)90050-2</a>, 1956.
</mixed-citation></ref-html>
<ref-html id="bib1.bib11"><label>11</label><mixed-citation>
Kim, M., Cahill, J. F., Su, Y., Prather, K. A., and Cohen, S. M.:
Postsynthetic ligand exchange as a route to functionalization of “inert”
metal–organic frameworks, Chem. Sci., 3, 126–130, <a href="https://doi.org/10.1039/C1SC00394A" target="_blank">https://doi.org/10.1039/C1SC00394A</a>, 2012.
</mixed-citation></ref-html>
<ref-html id="bib1.bib12"><label>12</label><mixed-citation>
Kousaka, Y., Okuyama, K., and Endo, Y.: Calibration of differential mobility
analyser by visual method, J. Aerosol Sci., 12, 339–348,
<a href="https://doi.org/10.1016/0021-8502(81)90023-9" target="_blank">https://doi.org/10.1016/0021-8502(81)90023-9</a>, 1981.
</mixed-citation></ref-html>
<ref-html id="bib1.bib13"><label>13</label><mixed-citation>
Li, L., Huang, Z., Dong, J., Li, M., Gao, W., Nian, H., Fu, Z., Zhang, G.,
Bi, X., Cheng, P., and Zhou, Z.: Real time bipolar time-of-flight mass
spectrometer for analyzing single aerosol particles, Int. J. Mass Spectrom., 303, 118–124, <a href="https://doi.org/10.1016/j.ijms.2011.01.017" target="_blank">https://doi.org/10.1016/j.ijms.2011.01.017</a>, 2011.
</mixed-citation></ref-html>
<ref-html id="bib1.bib14"><label>14</label><mixed-citation>
Li, L., Li, M., Huang, Z., Gao, W., Nian, H., Fu, Z., Gao, J., Chai, F., and
Zhou, Z.: Ambient particle characterization by single particle aerosol mass
spectrometry in an urban area of Beijing, Atmos. Environ., 94,
323–331, <a href="https://doi.org/10.1016/j.atmosenv.2014.03.048" target="_blank">https://doi.org/10.1016/j.atmosenv.2014.03.048</a>, 2014.
</mixed-citation></ref-html>
<ref-html id="bib1.bib15"><label>15</label><mixed-citation>
Li, L., Liu, L., Xu, L., Li, M., Li, X., Gao, W., Huang, Z., and Cheng, P.:
Improvement in the Mass Resolution of Single Particle Mass Spectrometry
Using Delayed Ion Extraction, Int. J. Mass Spectrom., 29, 2105–2109, <a href="https://doi.org/10.1007/s13361-018-2037-4" target="_blank">https://doi.org/10.1007/s13361-018-2037-4</a>, 2018.
</mixed-citation></ref-html>
<ref-html id="bib1.bib16"><label>16</label><mixed-citation>
Liu, P., Ziemann, P. J., Kittelson, D. B., and McMurry, P. H.: Generating
Particle Beams of Controlled Dimensions and Divergence: II. Experimental
Evaluation of Particle Motion in Aerodynamic Lenses and Nozzle Expansions,
Aerosol Sci. Technol., 22, 314–324, <a href="https://doi.org/10.1080/02786829408959749" target="_blank">https://doi.org/10.1080/02786829408959749</a>, 1995.
</mixed-citation></ref-html>
<ref-html id="bib1.bib17"><label>17</label><mixed-citation>
Middlebrook, A. M., Murphy, D. M., Lee, S.-H., Thomson, D. S., Prather, K.
A., Wenzel, R. J., Liu, D.-Y., Phares, D. J., Rhoads, K. P., Wexler, A. S.,
Johnston, M. V., Jimenez, J. L., Jayne, J. T., Worsnop, D. R., Yourshaw, I.,
Seinfeld, J. H., and Flagan, R. C.: A comparison of particle mass
spectrometers during the 1999 Atlanta Supersite Project, J. Geophys. Res.-Atmos., 108,  8424, <a href="https://doi.org/10.1029/2001jd000660" target="_blank">https://doi.org/10.1029/2001jd000660</a>, 2003.
</mixed-citation></ref-html>
<ref-html id="bib1.bib18"><label>18</label><mixed-citation>
Murphy, D. M.: The design of single particle laser mass spectrometers, Mass
Spectrom. Rev., 26, 150–165, <a href="https://doi.org/10.1002/mas.20113" target="_blank">https://doi.org/10.1002/mas.20113</a>,
2007.
</mixed-citation></ref-html>
<ref-html id="bib1.bib19"><label>19</label><mixed-citation>
Nash, D. G., Baer, T., and Johnston, M. V.: Aerosol mass spectrometry: An
introductory review, Int. J. Mass Spectrom., 258, 2–12,
<a href="https://doi.org/10.1016/j.ijms.2006.09.017" target="_blank">https://doi.org/10.1016/j.ijms.2006.09.017</a>, 2006.
</mixed-citation></ref-html>
<ref-html id="bib1.bib20"><label>20</label><mixed-citation>
Nicosia, A., Manodori, L., Trentini, A., Ricciardelli, I., Bacco, D.,
Poluzzi, V., Di Matteo, L., and Belosi, F.: Field study of a soft X-ray
aerosol neutralizer combined with electrostatic classifiers for nanoparticle
size distribution measurements, Particuology, 37, 99–106, <a href="https://doi.org/10.1016/j.partic.2017.08.001" target="_blank">https://doi.org/10.1016/j.partic.2017.08.001</a>, 2018.
</mixed-citation></ref-html>
<ref-html id="bib1.bib21"><label>21</label><mixed-citation>
Noble, C. A. and Prather, K. A.: Real-time single particle mass
spectrometry: a historical review of a quarter century of the chemical
analysis of aerosols, Mass Spectrom. Rev., 19, 248–274, <a href="https://doi.org/10.1002/1098-2787(200007)19:4&lt;248::AID-MAS3&gt;3.0.CO;2-I" target="_blank">https://doi.org/10.1002/1098-2787(200007)19:4&lt;248::AID-MAS3&gt;3.0.CO;2-I</a>,
2000.
</mixed-citation></ref-html>
<ref-html id="bib1.bib22"><label>22</label><mixed-citation>
Pratt, K. A., Mayer, J. E., Holecek, J. C., Moffet, R. C., Sanchez, R. O.,
Rebotier, T. P., Furutani, H., Gonin, M., Fuhrer, K., Su, Y., Guazzotti, S.,
and Prather, K. A.: Development and Characterization of an Aircraft Aerosol
Time-of-Flight Mass Spectrometer, Anal. Chem., 81, 1792–1800,
<a href="https://doi.org/10.1021/ac801942r" target="_blank">https://doi.org/10.1021/ac801942r</a>, 2009.
</mixed-citation></ref-html>
<ref-html id="bib1.bib23"><label>23</label><mixed-citation>
Pratt, K. A. and Prather, K. A.: Mass spectrometry of atmospheric
aerosols – Recent developments and applications, Part II: On-line mass
spectrometry techniques, 31, 17–48, <a href="https://doi.org/10.1002/mas.20330" target="_blank">https://doi.org/10.1002/mas.20330</a>, 2012.
</mixed-citation></ref-html>
<ref-html id="bib1.bib24"><label>24</label><mixed-citation>
Roth, A., Schneider, J., Klimach, T., Mertes, S., van Pinxteren, D., Herrmann, H., and Borrmann, S.: Aerosol properties, source identification, and cloud processing in orographic clouds measured by single particle mass spectrometry on a central European mountain site during HCCT-2010, Atmos. Chem. Phys., 16, 505–524, <a href="https://doi.org/10.5194/acp-16-505-2016" target="_blank">https://doi.org/10.5194/acp-16-505-2016</a>, 2016.

</mixed-citation></ref-html>
<ref-html id="bib1.bib25"><label>25</label><mixed-citation>
Shen, W., Dai, X., Huang, Z. X., Hou, Z. H., Cai, W. G., Du, X. B., Zhou,
Z., Li, M., and Li, L.: Improvement of the Dynamic Range of Data Acquisition
System in Single Particle Mass Spectrometry, J. Chinese Cinemas., 39, 331–336, <a href="https://doi.org/10.7538/zpxb.2017.0119" target="_blank">https://doi.org/10.7538/zpxb.2017.0119</a>, 2018.
</mixed-citation></ref-html>
<ref-html id="bib1.bib26"><label>26</label><mixed-citation>
Su, Y., Sipin, M. F., Furutani, H., and Prather, K. A.: Development and
Characterization of an Aerosol Time-of-Flight Mass Spectrometer with
Increased Detection Efficiency, Anal. Chem., 76, 712–719,
<a href="https://doi.org/10.1021/ac034797z" target="_blank">https://doi.org/10.1021/ac034797z</a>, 2004.
</mixed-citation></ref-html>
<ref-html id="bib1.bib27"><label>27</label><mixed-citation>
Thomson, D. S., Middlebrook, A. M., and Murphy, D. M.: Thresholds for
Laser-Induced Ion Formation from Aerosols in a Vacuum Using Ultraviolet and
Vacuum-Ultraviolet Laser Wavelengths, Aerosol Sci. Technol., 26,
544–559, <a href="https://doi.org/10.1080/02786829708965452" target="_blank">https://doi.org/10.1080/02786829708965452</a>, 1997.
</mixed-citation></ref-html>
<ref-html id="bib1.bib28"><label>28</label><mixed-citation>
Trimborn, A., Hinz, K. P., and Spengler, B.: Online analysis of atmospheric
particles with a transportable laser mass spectrometer during LACE 98,
J. Geophys. Res.-Atmos., 107, LAC 13-11-LAC 13-10,
<a href="https://doi.org/10.1029/2001JD000590" target="_blank">https://doi.org/10.1029/2001JD000590</a>, 2002.
</mixed-citation></ref-html>
<ref-html id="bib1.bib29"><label>29</label><mixed-citation>
Tsai, C.-J., Lin, J.-S., Deshpande, C., and Liu, L.-C.: Electrostatic
Charge Measurement and Charge Neutralization of Fine Aerosol
Particles during the Generation Process, Part. Part. Syst. Charact., 22 293–298, https://doi.org/10.1002/ppsc.200500961, 2005.
</mixed-citation></ref-html>
<ref-html id="bib1.bib30"><label>30</label><mixed-citation>
Woessner, R. H. and Gunn, R.: Measurements related to the fundamental
processes of aerosol electrification, J. Coll. Sci. Imp. U. Tok., 11, 69–76,
<a href="https://doi.org/10.1016/0095-8522(56)90021-6" target="_blank">https://doi.org/10.1016/0095-8522(56)90021-6</a>, 1956.
</mixed-citation></ref-html>
<ref-html id="bib1.bib31"><label>31</label><mixed-citation>
Zelenyuk, A., Cai, Y., and Imre, D.: From Agglomerates of Spheres to
Irregularly Shaped Particles: Determination of Dynamic Shape Factors from
Measurements of Mobility and Vacuum Aerodynamic Diameters, Aerosol Sci. Technol., 40, 197–217, <a href="https://doi.org/10.1080/02786820500529406" target="_blank">https://doi.org/10.1080/02786820500529406</a>, 2006.
</mixed-citation></ref-html>
<ref-html id="bib1.bib32"><label>32</label><mixed-citation>
Zelenyuk, A., Yang, J., Imre, D., and Choi, E.: Achieving Size Independent
Hit-Rate in Single Particle Mass Spectrometry, Aerosol Sci. Technol., 43, 305–310, <a href="https://doi.org/10.1080/02786820802637915" target="_blank">https://doi.org/10.1080/02786820802637915</a>, 2009.
</mixed-citation></ref-html>
<ref-html id="bib1.bib33"><label>33</label><mixed-citation>
Zhao, Y., Bein, K. J., Wexler, A. S., Misra, C., Fine, P. M., and Sioutas,
C.: Field evaluation of the versatile aerosol concentration enrichment
system (VACES) particle concentrator coupled to the rapid single-particle
mass spectrometer (RSMS-3), J. Geophys. Res.-Atmos.,
110, D07S02, <a href="https://doi.org/10.1029/2004jd004644" target="_blank">https://doi.org/10.1029/2004jd004644</a>, 2005.
</mixed-citation></ref-html>--></article>