<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing with OASIS Tables v3.0 20080202//EN" "journalpub-oasis3.dtd">
<article xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:oasis="http://docs.oasis-open.org/ns/oasis-exchange/table" dtd-version="3.0"><?xmltex \makeatother\@nolinetrue\makeatletter?>
  <front>
    <journal-meta>
<journal-id journal-id-type="publisher">AMT</journal-id>
<journal-title-group>
<journal-title>Atmospheric Measurement Techniques</journal-title>
<abbrev-journal-title abbrev-type="publisher">AMT</abbrev-journal-title>
<abbrev-journal-title abbrev-type="nlm-ta">Atmos. Meas. Tech.</abbrev-journal-title>
</journal-title-group>
<issn pub-type="epub">1867-8548</issn>
<publisher><publisher-name>Copernicus GmbH</publisher-name>
<publisher-loc>Göttingen, Germany</publisher-loc>
</publisher>
</journal-meta>

    <article-meta>
      <article-id pub-id-type="doi">10.5194/amt-8-541-2015</article-id><title-group><article-title>A new airborne laser-induced fluorescence instrument for in situ
detection of formaldehyde throughout the troposphere and<?xmltex \hack{\break}?> lower stratosphere</article-title>
      </title-group><?xmltex \runningtitle{A new airborne laser-induced fluorescence instrument}?><?xmltex \runningauthor{M. Cazorla et al.}?>
      <contrib-group>
        <contrib contrib-type="author" corresp="no" rid="aff1 aff5">
          <name><surname>Cazorla</surname><given-names>M.</given-names></name>
          
        <ext-link>https://orcid.org/0000-0001-5295-2968</ext-link></contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1 aff2">
          <name><surname>Wolfe</surname><given-names>G. M.</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1">
          <name><surname>Bailey</surname><given-names>S. A.</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff1 aff3">
          <name><surname>Swanson</surname><given-names>A. K.</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="no" rid="aff4">
          <name><surname>Arkinson</surname><given-names>H. L.</given-names></name>
          
        </contrib>
        <contrib contrib-type="author" corresp="yes" rid="aff1">
          <name><surname>Hanisco</surname><given-names>T. F.</given-names></name>
          <email>thomas.hanisco@nasa.gov</email>
        <ext-link>https://orcid.org/0000-0001-9434-8507</ext-link></contrib>
        <aff id="aff1"><label>1</label><institution>Atmospheric Chemistry and Dynamics Lab, NASA Goddard Space
Flight Center, Greenbelt, MD, USA</institution>
        </aff>
        <aff id="aff2"><label>2</label><institution>Joint Center for Earth Systems Technology, University of
Maryland Baltimore County, Baltimore, MD, USA</institution>
        </aff>
        <aff id="aff3"><label>3</label><institution>Goddard Earth Sciences Technology and Research, University
of Maryland Baltimore County, Baltimore, MD, USA</institution>
        </aff>
        <aff id="aff4"><label>4</label><institution>Department of Atmospheric and Oceanic Science, University
of Maryland, College Park, MD, USA</institution>
        </aff>
        <aff id="aff5"><label>*</label><institution>now at: Universidad San Francisco de Quito, Instituto de
Investigaciones Atmosféricas, Colegio de Ciencias e Ingeniería,
Diego de Robles y Vía Interoceánica, Quito, Ecuador</institution>
        </aff>
      </contrib-group>
      <author-notes><corresp id="corr1">T. F. Hanisco (thomas.hanisco@nasa.gov)</corresp></author-notes><pub-date><day>3</day><month>February</month><year>2015</year></pub-date>
      
      <volume>8</volume>
      <issue>2</issue>
      <fpage>541</fpage><lpage>552</lpage>
      <history>
        <date date-type="received"><day>21</day><month>July</month><year>2014</year></date>
           <date date-type="rev-request"><day>19</day><month>August</month><year>2014</year></date>
           <date date-type="rev-recd"><day>13</day><month>November</month><year>2014</year></date>
           <date date-type="accepted"><day>9</day><month>December</month><year>2014</year></date>
           
      </history>
      <permissions>
<license license-type="open-access">
<license-p>This work is licensed under a Creative Commons Attribution 3.0 Unported License. To view a copy of this license, visit <ext-link ext-link-type="uri" xlink:href="http://creativecommons.org/licenses/by/3.0/">http://creativecommons.org/licenses/by/3.0/</ext-link></license-p>
</license>
</permissions><self-uri xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015.html">This article is available from https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015.html</self-uri>
<self-uri xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015.pdf">The full text article is available as a PDF file from https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015.pdf</self-uri>


      <abstract>
    <p>The NASA In Situ Airborne Formaldehyde (ISAF) instrument is a
high-performance laser-based detector for gas-phase formaldehyde (HCHO). ISAF
uses rotational-state specific laser excitation at 353 nm for laser-induced
fluorescence (LIF) detection of HCHO. A number of features make ISAF ideal
for airborne deployment, including (1) a compact, low-maintenance fiber
laser, (2) a single-pass design for stable signal response, (3) a
straightforward inlet design, and (4) a stand-alone data acquisition system. A
full description of the instrument design is given, along with detailed
performance characteristics. The accuracy of reported mixing ratios is
<inline-formula><mml:math display="inline"><mml:mo>±</mml:mo></mml:math></inline-formula>10 % based on calibration against IR and UV absorption of a primary
HCHO standard. Precision at 1 Hz is typically better than 20 % above
100 pptv, with uncertainty in the signal background contributing most to
variability at low mixing ratios. The 1 Hz detection limit for a
signal <inline-formula><mml:math display="inline"><mml:mo>/</mml:mo></mml:math></inline-formula> noise ratio of 2 is 36 pptv for 10 mW of laser power, and the
<inline-formula><mml:math display="inline"><mml:mi>e</mml:mi></mml:math></inline-formula> fold time response at typical sample flow rates is 0.19 s. ISAF has
already flown on several field missions and platforms with excellent results.</p>
  </abstract>
    </article-meta>
  </front>
<body>
      

      <?xmltex \hack{\newpage}?>
<sec id="Ch1.S1" sec-type="intro">
  <title>Introduction</title>
      <p>Formaldehyde (HCHO) is a ubiquitous constituent of the Earth's atmosphere. It
is primarily produced during hydrocarbon oxidation, with an estimated source
strength of 250 <inline-formula><mml:math display="inline"><mml:mo>±</mml:mo></mml:math></inline-formula> 54 Tg yr<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> (Fortems-Cheiney et al., 2012). In
the background troposphere, the oxidation of methane (CH<inline-formula><mml:math display="inline"><mml:mrow><mml:msub><mml:mi/><mml:mn mathvariant="normal">4</mml:mn></mml:msub><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> by hydroxyl
radical (OH) is the main source. Photochemical degradation of anthropogenic
and biogenic hydrocarbons greatly enhances HCHO production in the lower
troposphere, with the most significant precursor being biogenic
hydrocarbon isoprene. HCHO is also emitted through fuel combustion (Olaguer
et al., 2009; Luecken et al., 2012), biomass burning (Yokelson et al., 2013)
and vegetation (DiGangi et al., 2011). These sources are generally minor
globally compared to secondary production, but they may be significant
locally. HCHO is lost via photolysis and reaction with OH, with a typical
daytime lifetime of 2–3 h. Wet and dry deposition are minor sinks globally,
but can be significant locally, such as during precipitation. Typical mixing
ratios range from tens of parts per trillion by volume (pptv) in pristine air
to several hundred parts per billion by volume (ppbv) near combustion
sources.</p>
      <p>HCHO is a tracer for, and active participant in, multiple atmospheric
processes. As a byproduct of hydrocarbon oxidation, HCHO can provide
quantitative constraints on the photochemical link between primary emissions
and secondary pollutants (Chatfield et al., 2010; Duncan et al., 2010). A
number of studies have employed satellite-based HCHO observations to better
constrain isoprene emission inventories (Palmer et al., 2003, 2006; Millet et
al., 2008; Marais et al., 2012). Because of its short lifetime, the mixing
ratio of HCHO in the upper troposphere is typically 10 to 100 times less than
that near continental surfaces. Convection can pump surface air to the upper
troposphere very rapidly, and HCHO is a valuable marker for the efficiency of
this process (Barth et al., 2007; Fried et al., 2008) and a participant in
ensuing chemistry (Apel et al., 2012). Furthermore, photolysis of HCHO is a
potent source of radicals and thus helps to propagate chemical cycles
(Edwards et al., 2011, 2013).</p>
      <p>A handful of techniques are currently available to measure HCHO in the
atmosphere; a review of these methods and inter-comparisons can be found
elsewhere (Gilpin, et al., 1997; Cardenas et al., 2000; Hak et al., 2005;
Wisthaler et al., 2008; Kaiser et al., 2014). Laser-based and spectroscopic
methods are generally employed when high sensitivity and time response are
required. The most common among these include tunable diode laser-based
absorption spectroscopy (Weibring et al., 2007; Fried et al., 2008),
differential optical absorption spectroscopy (DOAS) (Baidar et al., 2013) and
laser-induced fluorescence (LIF) (Hottle et al., 2009). Diode laser
spectroscopy and DOAS rely on molecular absorption in the IR and UV regions,
respectively, while LIF utilizes UV-stimulated emission. DOAS can be applied
for in situ or remote (column-averaged) observations, and a similar principle
underlies satellite-based HCHO observations (Chance et al., 2000). Indeed,
HCHO is one of only a few hydrocarbons observable from space, and
high-quality in situ observations are critical for retrieval validation.</p>
      <p>Instrumentation for airborne observations is held to stringent engineering
standards. An ideal instrument is small, lightweight, low power, and able to
withstand vibrations and large temperature variations. Low maintenance,
calibration stability, and fast time response (<inline-formula><mml:math display="inline"><mml:mo>≥</mml:mo></mml:math></inline-formula> 1 Hz) are also
desirable. For HCHO, a substantial dynamic range – roughly 4 orders of
magnitude – is also necessary to adequately sample mixing ratios throughout
the troposphere and lower stratosphere. Here, we present the NASA In Situ Airborne Formaldehyde (ISAF) instrument, a new LIF-based HCHO detector that
meets or exceeds all of the above requirements. After a brief synopsis of the
LIF method in Sect. 2, we provide a detailed description of instrument
construction (Sect. 3) and performance characteristics (Sect. 4). Section 5
gives an example of observations taken during a recent field campaign.</p>
</sec>
<sec id="Ch1.S2">
  <title>Principle of operation</title>
      <p>LIF is a well-established spectroscopic method for selective measurement of
trace gases. LIF detection of HCHO was first reported by Becker et al. (1975)
and refined by Möhlmann (1985). Hottle et al. (2009) applied the
technique to in situ atmospheric measurements. In these approaches a tunable
laser is used to ensure specificity to HCHO and a pulsed operation is used to
minimize background noise. Figure 1 illustrates the key processes involved. A
narrow-band laser, tuned to a specific rovibronic absorption line in the
353–355 nm region of the <inline-formula><mml:math display="inline"><mml:mi>A</mml:mi></mml:math></inline-formula>–<inline-formula><mml:math display="inline"><mml:mi>X</mml:mi></mml:math></inline-formula> electronic transition, promotes a
fraction of the HCHO molecules to the first excited electronic state. The
majority of these excited HCHO molecules are quenched back to the ground
electronic state by collisions with air. A small fraction (<inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 1 %)
undergo vibrational relaxation and subsequently relax to the ground
electronic state through fluorescence, emitting a structured band of photons
in the range of 400–470 nm (Becker et al., 1975; Shibuya et al., 1979;
Möhlmann, 1985). Fluorescence occurs with an exponential decay over
several hundred nanoseconds (Fig. 1, right). The number of fluorescence
photons per unit of laser power is proportional to the mixing ratio of HCHO in
the sample gas. This relationship is readily quantified with a known
standard.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F1"><caption><p>The spectral <bold>(a)</bold> and temporal <bold>(b)</bold> features of
laser induced fluorescence detection. In <bold>(a)</bold> the laser excites a
single rotational transition at 353.16 nm. The fluorescence occurs after
collisional relaxation and appears at longer wavelengths. In <bold>(b)</bold>
time gating of fluorescence detection provides additional discrimination
between prompt and longer timescale processes. Laser excitation and elastic
scattering processes that include Raman, Rayleigh, and chamber reflections
have timescales that match the laser pulse (25 ns full-width). Fluorescence
has a lifetime of tens of nanoseconds and the signal extends for more than 100 ns
beyond the end of the laser pulse and elastic scattering. The delayed gate
captures HCHO fluorescence while the full gate captures both fluorescence and
scattered light.</p></caption>
        <?xmltex \igopts{width=184.942913pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f01.pdf"/>

      </fig>

      <p>For ISAF, the chosen absorption feature is centered at 353.163 nm, which has
a cross-section of 3.77 <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 10<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn>19</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup></mml:math></inline-formula> at 150 Torr (Co et
al., 2005). While this cross-section is <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 7 % lower than the peak
value in this band, it is chosen because of the presence of several unique
adjacent peaks, notably a triplet feature (see Fig. 6). This area of the
spectrum serves as a “fingerprint” for HCHO, simplifying the
wavelength-tracking algorithm. The offline position is only 0.005 nm away
from the peak. This wavelength difference provides a high differential
cross-section for HCHO, but the differential is negligible for species that
have non-structured absorptions in this region. This allows for precise and
selective measurements of HCHO. To date, we have not identified any
interfering species that has a measurable differential absorption and
fluorescence between the online and offline positions.</p>
</sec>
<sec id="Ch1.S3">
  <title>Instrument description</title>
      <p>ISAF is comprised of four main subsystems: a UV fiber laser, an optical bench
comprising the detection and reference cells, gas handling components
(pressure controller, inlet system, tubing, valves and pump), and the data
acquisition computer. Each of these subsystems is described in greater detail
below. Aside from the vacuum pump, all components are contained in a single
custom-built chassis with dimensions of 60 cm length <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 43 cm
width <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 35 cm height. The layout of the chassis is shown in Fig. 2.
This chassis is designed to run on 28 V DC, which is supplied directly from
aircraft power or through an external AC/DC interface unit. The instrument
itself weighs 25 kg and draws 200 W. An external pump, computer, and
control box are used for operation on larger aircraft with on-board
operators. Including all peripherals, the total instrument package weighs
60 kg and draws a peak power of 600 W.</p>
<sec id="Ch1.S3.SS1">
  <title>Laser</title>
      <p>Pulsed, tunable UV laser radiation is provided by a NovaWave (now Thermo
Scientific) TFL 3000 fiber laser. The laser uses a master oscillator power
amplifier architecture to provide narrow wavelength, tunable, pulsed light at
353 nm. A single distributed feedback laser operating at 1060 nm is chopped
with an acousto-optic modulator and amplified in two stages to 2 W with a
series of diode-pumped, semiconductor-doped, polarization-maintaining optical
fibers. The 1060 nm output is fiber-coupled into a harmonic generation
module containing two periodically poled nonlinear crystals and collimating
optics. The harmonic generation housing is mounted on the optical bench
(Fig. 2) and coupled with free-space optics. The repetition rate is 300 kHz,
with a pulse width of 25 ns and a typical output power of 10–40 mW. The
output wavelength is tunable to <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 0.0001 nm, permitting resolution of
individual rovibronic absorption features. Computer control is provided via
RS232. The entire unit weighs <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 12 kg and draws <inline-formula><mml:math display="inline"><mml:mo>&lt;</mml:mo></mml:math></inline-formula> 60 W, making it
an attractive alternative to larger laser systems typically used for LIF.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F2"><caption><p>A top view of the ISAF instrument layout. The instrument chassis is
60 <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 43 <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 35 cm and 25 kg. The major subsystems are
labeled: (A) pressure controller, (B) optical bench, (C) tunable fiber laser,
(D) harmonic generation housing, and (E) data acquisition system.</p></caption>
          <?xmltex \igopts{width=184.942913pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f02.pdf"/>

        </fig>

</sec>
<sec id="Ch1.S3.SS2">
  <title>Optical bench</title>
      <p>The optical bench is shown in greater detail in Figs. 3 and 4. The bench is
an aluminum sandwich panel with 1.25 cm thick honeycomb core and with
0.1 cm cladding. The bench is mounted to the instrument chassis with
vibration damping mounts (Sorbothane) to minimize vibrations. Figure 3 shows
the optical bench and the path of the laser and gas flow. Figure 4 shows a
cross-sectional view of the detection cell. The simplicity and single-laser
pass design of this system is a major advantage for field applications;
alignment rarely degrades throughout flight campaigns or during shipping.
Likewise, the instrument shows no effects in high vibration environments or
in aircraft maneuvers. Both cells are heated to a constant temperature of
35 <inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mo>∘</mml:mo></mml:msup></mml:math></inline-formula>C, and capability exists to heat other parts of the instrument if
needed (i.e., for high-altitude applications).</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F3"><caption><p>This diagram shows a top view of the optical bench with the harmonic
generation housing and platform removed. The laser propagation is indicated
by the solid (open to air) and dashed (inside the cells) blue lines. The sample flow is indicated by the black arrows. The
components are (A) piezo actuated mirror (3); (B) pressure sensor;
(C) thermistor; (D) sample detection cell; (E) fixed mirror; (F) sample
heater; (G) reference cell (2); (H) beam splitter; (I) reference cell;
(J) formaldehyde source.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f03.pdf"/>

        </fig>

      <p>UV laser light is collimated at 2 mm diameter at the output of the harmonic
stage with a lens (Thor LA4600-UV, <inline-formula><mml:math display="inline"><mml:mi>f</mml:mi></mml:math></inline-formula> <inline-formula><mml:math display="inline"><mml:mo>=</mml:mo></mml:math></inline-formula> 100 mm) and directed to the
detection cell via four dielectric-coated mirrors (Edmund 47-323). Two of
these mirrors, along with the two between the detection and reference cells,
utilize piezo-driven mounts (Newport Agilis) for remote alignment. The laser
beam enters and exits the reduced-pressure detection cell (Fig. 4) through
V-type antireflection-coated fused silica windows (Edmund 48-207) tilted at
5<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mo>∘</mml:mo></mml:msup></mml:math></inline-formula> to redirect reflections from the window surfaces out of the beam
propagation path. The arms of the detection cell contain a series of circular
baffles with 2.5–3.5 mm apertures to eliminate stray light. Baffles and
cell walls are coated with a combination of Avian black paint (Avian
Technologies), black anodize, and molybdenum oxide treated aluminum
(Insta-Black 380, EPI). The innermost baffles are coated with an ultra-black
custom carbon nanotube treatment (Hagopian, 2011). The laser beam, gas flow,
and focus of the collection lens intersect orthogonally in the center of the
cell. Along this detection axis, photons are imaged with an aspheric lens
(Edmund 6-321) through a 370 nm long pass absorption filter (Edmund 66-049)
and a 400 nm long pass interference filter (Omega 3RD400LP). The image is
partially focused with another lens (Thor LA1608, <inline-formula><mml:math display="inline"><mml:mi>f</mml:mi></mml:math></inline-formula> <inline-formula><mml:math display="inline"><mml:mo>=</mml:mo></mml:math></inline-formula> 75 mm) onto a
photomultiplier tube (PMT) (Hamamatsu 7360). A spherical mirror (Thor CM254,
<inline-formula><mml:math display="inline"><mml:mi>f</mml:mi></mml:math></inline-formula> <inline-formula><mml:math display="inline"><mml:mo>=</mml:mo></mml:math></inline-formula> 25 mm) is located opposite the aspheric lens to improve collection
efficiency. Since this is a fluorescence measurement, only the sample inside
the imaging volume of the collection lens (the center of the cell) is
detected.</p>
      <p>After the detection cell, a beam sampler (Thor BSF05-UV) diverts
<inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 4 % of the laser to a power monitor comprised of a diffuser (Thor
DGUV10-600), filter (Thor FGUV11), and amplified photodiode (OSI 555-UV). The
photodiode has a large active area (1 cm<inline-formula><mml:math display="inline"><mml:mrow><mml:msup><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msup><mml:mo>)</mml:mo></mml:mrow></mml:math></inline-formula> to ensure that the entire
laser beam is detected. The diffuser is needed to spread the laser over
enough of the photodiode to avoid inhomogeneities in the photodiode
responsivity. The main laser beam is directed into a reference cell (Fig. 3,
lower right), which is used to monitor the laser wavelength. This cell is
nearly identical to the sample cell shown in Fig. 4; the main differences are
shorter arms and a neutral density filter (Thor NE30A) to limit fluorescence
signal intensity. A heated cartridge filled with powdered urea formaldehyde
glue (DAP Weldwood) provides a large concentration of HCHO (<inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 1 ppmv).
This cell is static (no gas flow) but is maintained at the same pressure as
the sample cell by coupling the arms to the sample cell exhaust. This is
critical for maintaining the same absorption line width in both cells, as the
reference cell signal is used to correct sample cell data (see Sect. 3.5). A
normally closed solenoid valve and an activated charcoal scrubber are placed
inline to limit sample cell contamination. The laser beam is ultimately
dumped into a second power monitor, which is primarily used to gauge laser
alignment.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F4"><caption><p>This diagram shows a side-view of the cross-section of the detection
cell. The blue line indicates the laser path and propagation direction.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f04.pdf"/>

        </fig>

</sec>
<sec id="Ch1.S3.SS3">
  <title>Gas handling</title>
      <p>The primary components of the air sampling system include the aircraft
inlet, pressure controller, and vacuum pump. Some care must be taken with
the inlet system to minimize adsorption and production/loss of HCHO on
surfaces (Wert et al., 2002). Outgassing of HCHO from tubing walls,
physisorption and desorption of HCHO, and the chemical or catalytic
conversion of oxygenated hydrocarbons on metal surfaces to HCHO are primary
concerns in the design of the sampling system. Fast flows, low surface area
fluorocarbon or fluorocarbon-coated surfaces, and minimal dead space are the
design features. For example, there are no valves or sensors between the tip
of the inlet and the instrument; the sampling path is continuous tubing with
VCO (Swagelok) fittings. Furthermore, all metal surfaces are coated with
fluorocarbon (FluoroPel, Cytonix), all tubing is PTFE or FEP, and pressure,
temperature, and flow are measured after the detection cell.</p>
      <p>Figure 5 depicts the inlet design used on the NASA DC-8 aircraft. The inlet
tube is a 45 cm length of 0.48 cm ID Silcosteel (Restek) coated with
FluoroPel mounted at 90<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mo>∘</mml:mo></mml:msup></mml:math></inline-formula> to the airflow to reject particles. The tip
of the inlet is chamfered at 15<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mo>∘</mml:mo></mml:msup></mml:math></inline-formula> to enhance the particle rejection and
is heated to 30–60 <inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mo>∘</mml:mo></mml:msup></mml:math></inline-formula>C inside the winglet to prevent condensation and
sticking. A tapered aluminum nacelle oriented parallel to the direction of
flight serves as a flow straightener and generates a ram pressure equal to
<inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 1.15 times that of ambient at the inlet tip, driving an air flow of
10–25 sLm through the inlet. The corresponding volumetric flow at typical
pressures and temperatures is 25–65 L min<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>. Inside the aircraft, air
is carried via 1 m of 0.7 cm ID PTFE tubing to the instrument chassis and
is sub-sampled at 2–4 sLm. The excess flow passes through a flow meter
(Honeywell) and exits behind the inlet. The residence time in the heated
Silcosteel portion of the tube is 8–20 ms, with an additional 35–100 ms
in the transfer tubing.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F5"><caption><p>This diagram shows a cross-section view of the particle rejection
inlet mounted on a window plate. A ram air nacelle mounted on an airfoil
30 cm from the plate generates ram pressure that directs the ambient air
through the heated tube mounted at 90<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mo>∘</mml:mo></mml:msup></mml:math></inline-formula> to the flow. A small fraction
(2 sLm) of the heated air enters the instrument and the remainder
(<inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 20 sLm) returns to ambient. The plumbing on the cabin side of the
window plate is not drawn to scale. There is 1 m of tubing between the inlet
and detection pickoff.</p></caption>
          <?xmltex \igopts{width=184.942913pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f05.pdf"/>

        </fig>

      <p>Inside the chassis, sample air passes through a pressure controller (MKS
640A) heated to 35 <inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mo>∘</mml:mo></mml:msup></mml:math></inline-formula>C, followed by a <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 20 cm length of 0.4 cm
ID PFA tubing connected to the detection cell. The residence time in the
detection cell is <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 115 ms. Sample pressure, temperature, and flow rate
are monitored immediately downstream of the cell to minimize surface area. A
small amount of sample air (<inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 100 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">3</mml:mn></mml:msup></mml:math></inline-formula> s<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>) is drawn through the laser
arms to reduce contamination from painted surfaces or dead space. Sample air
is exhausted through the vacuum pump (Agilent IDP-3 or Vacuubrand MD-1).
During instrument shutdown, contamination is minimized by backfilling the
cell with ambient air scrubbed by a Drierite/molecular sieve cartridge, which
reduces HCHO to <inline-formula><mml:math display="inline"><mml:mo>&lt;</mml:mo></mml:math></inline-formula> 100 pptv.</p>
</sec>
<sec id="Ch1.S3.SS4">
  <title>Data acquisition</title>
      <p>All hardware interfacing is handled with an all-solid-state CompactRIO
(National Instruments) with realtime operating system and a field
programmable gate array (FPGA). The FPGA and acquisition software are
configured for fast photon counting and autonomous operation. Six 200 MHz
counters are used to integrate PMT signals at a nominal rate of 10 Hz.
Pulses from the sample cell PMT are sampled by three of these counters, one
non-gated (continuous) and two each with a separate gate (Fig. 1, right).
The “full” gate samples both scattered laser light and HCHO fluorescence,
while the “delayed” gate excludes the laser. The reference cell gate is the
same width as the full gate. Laser power is also monitored at 10 Hz. Other
diagnostics (laser parameters, temperature, pressure, flow, GPS, and
aircraft data) are recorded at 1 Hz.</p>
      <p>Figure 6 illustrates the standard data acquisition routine. Online counting
is done for 2.5 s (25 points), followed by 0.5 s (5 points) offline. Laser
drift (mainly due to temperature of the seed laser diode) requires an active
line-locking algorithm. The online position is alternately shifted to a
slightly (<inline-formula><mml:math display="inline"><mml:mo>±</mml:mo></mml:math></inline-formula>0.0001 nm) lower or higher wavelength on adjacent online
cycles, and the reference cell is monitored for maximum signal. When the
laser is tuned precisely to the online position the reference cell is at the
maximum signal. When the laser is tuned 0.0001 nm (or 0.0002 nm) off of the
peak position the signal drops by about <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 5 % (or
<inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 10 %). The resulting small changes in the reference cell signal
are used to maintain the laser on the peak of the line. The signal is also
used to correct the sample signal by the <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 5 % (or
<inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 10 %) needed to normalize the signal to the signal at the peak
position.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F6"><caption><p>These plots show the wavelength (top) and fluorescence signal
(bottom) during a scan of the laser, followed by the standard data
acquisition mode. The computer algorithm uses the reference cell signal to
maintain the laser on the large rotational feature at 353.163 nm. The
contributions of species other than HCHO to the fluorescence signal are
removed by subtracting the signal at the offline position. The difference
between the signal at the online position and the offline position is
proportional to [HCHO].</p></caption>
          <?xmltex \igopts{width=184.942913pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f06.pdf"/>

        </fig>

      <p>Once per hour, a full scan over the HCHO fingerprint is recorded. This is
primarily used to optimize the harmonic generation crystal temperatures and
as a performance diagnostic. For example, the presence of interfering species
can be detected in the spectrum. To date, we have not detected any features
other than those shown in Fig. 6. This is not surprising. To be detected, the
potential species must absorb at 353.163 nm, fluoresce in the 400–470 nm
band, and have a significant fluorescence lifetime.</p><?xmltex \hack{\newpage}?>
</sec>
<sec id="Ch1.S3.SS5">
  <title>Data reduction</title>
      <p>Both the full and delay gate signals can be used to derive HCHO mixing
ratios. The delay gate excludes <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 53 % of the fluorescence signal,
but it is also relatively insensitive to laser scatter (from Raman, Rayleigh,
or particulates) and gives superior signal <inline-formula><mml:math display="inline"><mml:mo>/</mml:mo></mml:math></inline-formula> noise (S <inline-formula><mml:math display="inline"><mml:mo>/</mml:mo></mml:math></inline-formula> N) ratios below
<inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 550 pptv HCHO. Thus, the delay gate signal is typically used, and
all results discussed below stem from this signal. The full gate data
provides a useful metric for scatter and other potential laser-related
issues.</p>
      <p>The detected signal is proportional to the abundance of HCHO and the laser
power. To the first order, we can determine HCHO from the ratio of the laser
power-normalized difference between the online and offline signals and the
calibration factor. To minimize systematic noise, however, we use additional
steps to convert raw counts to mixing ratios. These analysis steps are
(1) normalization by laser power; (2) background subtraction, determined by a
smoothed interpolation of offline signals; (3) normalization by the reference
cell signal to account for deviations from the maximum absorption wavelength
due to the line-locking algorithm; and (4) application of a
pressure-dependent calibration factor (see Sect. 4.1).</p>
</sec>
</sec>
<sec id="Ch1.S4">
  <title>Performance</title>
<sec id="Ch1.S4.SS1">
  <title>Sensitivity</title>
      <p>Detection sensitivity (signal response per unit of HCHO) is a function of laser
power, collection efficiency, PMT response, and counter gating. As mentioned
above, signals are normalized by laser power. Because of the single-pass
design, laser alignment does not affect sensitivity except in extreme cases
when the laser is partially blocked. In practice, the laser maintains
alignment for entire field campaigns and usually during shipments to and from
deployment locations. During flight, the ratio of the two power monitors (at
the exits of the detection and reference cells) is constant to <inline-formula><mml:math display="inline"><mml:mo>±</mml:mo></mml:math></inline-formula>3 %.</p>
      <p>Collection efficiency depends on the physical characteristics of the sample
cell, including the types of collection optics (lenses and filters) and their
relative spacing. We do not expect the collection efficiency to change under
normal operation. However, catastrophic events, such as the degradation of
optics from contamination could reduce the sensitivity. The responsivity of
the PMT should be stable in the short term (months to years), but it can
degrade over the lifetime of the PMT. The gated counter timing is a fixed
parameter and is optimized for signal <inline-formula><mml:math display="inline"><mml:mo>/</mml:mo></mml:math></inline-formula> noise. Changes in the gate timing
relative to the laser pulse could affect the sensitivity. These changes are
easily diagnosed in data reduction and can be corrected by using the
non-delayed gates.</p>
      <p>In practice, instrument sensitivity is determined via calibration against a
primary standard, consisting of a 500–700 ppbv mixture of HCHO in N<inline-formula><mml:math display="inline"><mml:msub><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msub></mml:math></inline-formula>
(Scott Gas/Air Liquide). This mixture is calibrated every few months
(typically before and after a field deployment) with an MKS Multigas 2031
Fourier transform infrared spectrometer using manufacturer-provided IR cross-sections tied to gravimetric permeation tube calibrations. The 1<inline-formula><mml:math display="inline"><mml:mi mathvariant="italic">σ</mml:mi></mml:math></inline-formula>
uncertainty in mixing ratio from any single determination is typically
2–3 %. HCHO mixtures are also stable over time; for example, five
calibrations performed on a single tank over the course of a year give a
mixing ratio of 611 <inline-formula><mml:math display="inline"><mml:mo>±</mml:mo></mml:math></inline-formula> 8 ppbv and show no trend.</p>
      <p>To better constrain the calibration to literature standards, a series of
experiments were performed to compare concentrations derived from Fourier transform infrared spectroscopy (FTIR) and
long-path UV absorption. The UV absorption cell was a custom-built, 200 cm
long, 1.1 cm diameter stainless-steel tube fitted with windows, gas ports,
and a pressure gauge. Visible light from a Cathodeon deuterium lamp was
filtered (Edmund U-330) and focused through the cell and measured with an
Ocean Optics Maya 2000 spectrometer. The cell pressure was held above atmosphere
(1110–1150 Torr) with a gas flow rate of 6 sLm. Absorption measurements
were carried out simultaneously in both the IR and UV systems using a high
concentration (nominally 11 ppmv) HCHO standard. Literature UV cross-sections (Meller and Moortgat, 2000) in the 285–370 nm range were fit to
the measured absorption spectra (nominal resolution <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 1 nm) to obtain
HCHO concentrations and mixing ratios. Based on 11 such experiments, the
mixing ratios from UV and IR are 10.49 <inline-formula><mml:math display="inline"><mml:mo>±</mml:mo></mml:math></inline-formula> 0.20 ppmv and
10.96 <inline-formula><mml:math display="inline"><mml:mo>±</mml:mo></mml:math></inline-formula> 0.02 ppmv, respectively. The ratio of UV <inline-formula><mml:math display="inline"><mml:mo>/</mml:mo></mml:math></inline-formula> IR <inline-formula><mml:math display="inline"><mml:mo>=</mml:mo></mml:math></inline-formula> 0.9571
is applied to all IR-derived values; for example, the IR-derived average
mixing ratio of 611 ppbv is corrected to 585 ppbv. By making this
adjustment, our calibration is tied to the Meller and Moortgat (2000) cross-section rather than the MKS spectrometer.</p>
      <p>Calibrations of ISAF are achieved via standard additions to zero air. Zero
air is delivered either from an ultra zero air cylinder or from a zero air
generator (Sabio 2020) with an additional Drierite/molecular sieve scrubber
to reduce HCHO mixing ratios to 30 pptv or less. Small flows of the primary
HCHO standard (1–50 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">3</mml:mn></mml:msup></mml:math></inline-formula> s<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>) are diluted into a 5 sLm flow of zero air to
yield HCHO concentrations in the range of 0–5 ppbv. Signal is plotted
against the diluted HCHO mixing ratio, and the slope of this relationship
gives the instrument sensitivity (see inset of Fig. 7). This sensitivity
varies with sample cell pressure due to the combined effects of number
density, absorption line broadening, and quenching. Figure 7 shows this
pressure dependence as determined from a number of standard addition
experiments. Peak sensitivity occurs at a cell pressure of 200 mbar. In
flight, the system is configured to maintain a constant pressure that is
dictated by the ceiling of the aircraft. For example, on the NASA DC-8, the
cell pressure is typically held at 100 mbar, and the sensitivity is
75 counts s<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> mW<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> ppbv<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>. Propagation of uncertainties in
the HCHO concentration, determined by both the absolute UV absorption
measurement and dilution flow rates, gives an estimated accuracy of
<inline-formula><mml:math display="inline"><mml:mo>±</mml:mo></mml:math></inline-formula>10 %.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F7"><caption><p>This plot shows the sensitivity of the instrument to HCHO versus
pressure. The inset shows an example of a calibration using an HCHO standard
addition to a flow of air. Several calibrations at one pressure are used to
determine the pressure dependence.</p></caption>
          <?xmltex \igopts{width=184.942913pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f07.pdf"/>

        </fig>

      <p>Calibrations are generally carried out in the laboratory both before and
after field missions. Because HCHO adheres to surfaces in tubing, fittings,
regulators and flow meters, several days are needed to ensure consistent
results, especially with new PTFE or FEP tubing. Calibrations can be
performed in the field during campaigns, if needed. In practice, these
calibrations are used to monitor instrument performance. We have never
changed the instrument sensitivity calibration numbers during a campaign.
Based on four field campaigns spanning 2 years, sensitivity varies by less than
10 % over a single mission and by less than 30 % between missions.
Given the difficulty of adding a known amount of HCHO, it is likely that the
sensitivity of the instrument was constant over those 2 years and that the
variability in the calibrations is due to differences in calibration
conditions and personnel.</p>
</sec>
<sec id="Ch1.S4.SS2">
  <title>Precision</title>
      <p>The instrument precision is primarily limited by processes that generate
noise or affect the background (offline) signal. The largest potential
sources of noise, Rayleigh and Raman scattering of laser light, are reduced
with a combination of optical filters, absorbing surfaces (e.g., baffles), and
gated photon counting. Laboratory tests indicate that scattering contributes
less than 7 counts s<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> mW<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> to the background signal at
100 mbar.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F8"><caption><p>This plot shows the measured and theoretical 1<inline-formula><mml:math display="inline"><mml:mi mathvariant="italic">σ</mml:mi></mml:math></inline-formula> precision as
a percentage of HCHO concentration. The inset shows an example of the mean
and standard deviation for constant HCHO. The average value (24 pptv) and
the standard deviation (17 pptv) are used to determine the measured
precision (70 % in this case) shown by the blue diamonds. The solid line
is the theoretical maximum precision based on the counting statistics of the
signal level.</p></caption>
          <?xmltex \igopts{width=184.942913pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f08.pdf"/>

        </fig>

      <p>Figure 8 shows the short-term precision estimated from the normalized
standard deviation (<inline-formula><mml:math display="inline"><mml:mrow><mml:msub><mml:mi mathvariant="italic">σ</mml:mi><mml:mi>x</mml:mi></mml:msub><mml:mo>/</mml:mo><mml:mo>&lt;</mml:mo><mml:mi>x</mml:mi><mml:mo>&gt;</mml:mo></mml:mrow></mml:math></inline-formula>) for data segments of 10 to 15 min at
a range of constant mixing ratios. Also displayed is the inverse square root
dependence expected from Poisson counting statistics. Above mixing ratios of
<inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 100 pptv, the precision is better than 20 % and exhibits the
expected slope but is somewhat higher than the Poisson limit. Some fraction
of this offset is due to uncertainty in the background. This would also
explain the reduced precision observed at the lowest mixing ratios. It is
also likely that some “extra” variability is contributed by the calibration
gas addition system, which would be directly proportional to the HCHO mixing
ratio. We note that small changes in room temperature can cause small
fluctuations in the amount of HCHO added to the calibration gas flow,
presumably from increased or decreased desorption of HCHO from surfaces
(i.e.,
regulator, valves, flow meters, and fittings). The true instrument precision
thus lies between the measured and predicted values in Fig. 8.</p>
      <p>In theory, time-averaging will improve instrument precision. Figure 9 shows
the mean-normalized Allen deviation for 60 min of sampling at a constant
mixing ratio of 5.1 ppbv. At short averaging times, precision improves as
the inverse square root of integration time, consistent with white noise as
the dominant source of variability. Averaging from the native data rate of
0.1 s to the typical reporting interval of 1 s improves precision by a
factor of 3. Averaging to the “optimal” interval of 130 s improves
precision by a factor of 18, although natural variability would likely
dominate on this timescale in the real atmosphere.</p>
</sec>
<sec id="Ch1.S4.SS3">
  <title>Detection limit</title>
      <p>For counting statistics, the signal <inline-formula><mml:math display="inline"><mml:mo>/</mml:mo></mml:math></inline-formula> noise ratio is defined as

                <disp-formula id="Ch1.E1" content-type="numbered"><mml:math display="block"><mml:mrow><mml:mfrac><mml:mi>S</mml:mi><mml:mi>N</mml:mi></mml:mfrac><mml:mo>=</mml:mo><mml:mfrac><mml:mrow><mml:mi>p</mml:mi><mml:msub><mml:mi>C</mml:mi><mml:mi mathvariant="normal">f</mml:mi></mml:msub><mml:mo>[</mml:mo><mml:mi mathvariant="normal">HCHO</mml:mi><mml:msub><mml:mo>]</mml:mo><mml:mi>t</mml:mi></mml:msub></mml:mrow><mml:msqrt><mml:mrow><mml:mi>p</mml:mi><mml:msub><mml:mi>C</mml:mi><mml:mi mathvariant="normal">f</mml:mi></mml:msub><mml:mo>[</mml:mo><mml:mi mathvariant="normal">HCHO</mml:mi><mml:mo>]</mml:mo><mml:mi>t</mml:mi><mml:mo>+</mml:mo><mml:mn mathvariant="normal">2</mml:mn><mml:mi>p</mml:mi><mml:mi>B</mml:mi><mml:mi>t</mml:mi></mml:mrow></mml:msqrt></mml:mfrac><mml:mo>,</mml:mo></mml:mrow></mml:math></disp-formula>

          where <inline-formula><mml:math display="inline"><mml:mi>p</mml:mi></mml:math></inline-formula> is laser power, <inline-formula><mml:math display="inline"><mml:mrow><mml:msub><mml:mi>C</mml:mi><mml:mi mathvariant="normal">f</mml:mi></mml:msub></mml:mrow></mml:math></inline-formula> is the calibration factor
(sensitivity), [HCHO] is the mixing ratio, <inline-formula><mml:math display="inline"><mml:mi>B</mml:mi></mml:math></inline-formula> is the background count rate,
and <inline-formula><mml:math display="inline"><mml:mi>t</mml:mi></mml:math></inline-formula> is the integration time. For a typical laser power of 10 mW, a
sensitivity of 75 counts s<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> mW<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> ppbv<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>, a background
count rate of 8 counts s<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> mW<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>, and an integration time of
1 s, the estimated detection limit is 18 pptv for S <inline-formula><mml:math display="inline"><mml:mo>/</mml:mo></mml:math></inline-formula> N <inline-formula><mml:math display="inline"><mml:mo>=</mml:mo></mml:math></inline-formula> 1 or
36 pptv for S <inline-formula><mml:math display="inline"><mml:mo>/</mml:mo></mml:math></inline-formula> N <inline-formula><mml:math display="inline"><mml:mo>=</mml:mo></mml:math></inline-formula> 2. This value agrees well with the 1<inline-formula><mml:math display="inline"><mml:mi mathvariant="italic">σ</mml:mi></mml:math></inline-formula>
variability of 17 pptv observed at low mixing ratios (inset of Fig. 8). Time
averaging will improve the detection limit in proportion to the precision
(Fig. 9). The detection limit scales as the inverse square root of laser
power. The laser is capable of producing as much as 40 mW of power; thus,
ISAF can potentially achieve a factor of two reduction in the detection limit
by operating at higher laser power. For example, the instrument operated at
20 mW during its first deployment and achieved higher S <inline-formula><mml:math display="inline"><mml:mo>/</mml:mo></mml:math></inline-formula> N. More
recently, however we choose to operate at reduced power (10 mW) to extend
the life of the laser diodes and nonlinear crystals.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F9"><caption><p>The mean normalized Allen deviation is shown for 60 min of sampling
at a constant HCHO concentration of 5.1 ppbv. The short integration times
show a <inline-formula><mml:math display="inline"><mml:mrow><mml:msup><mml:mi mathvariant="italic">τ</mml:mi><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn><mml:mo>/</mml:mo><mml:mn mathvariant="normal">2</mml:mn></mml:mrow></mml:msup></mml:mrow></mml:math></inline-formula> increase in precision, consistent with white noise.</p></caption>
          <?xmltex \igopts{width=184.942913pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f09.pdf"/>

        </fig>

      <p>The inset of Fig. 8 is a typical example of the signal measured with zero air
added to the calibration system. Zero air and ultra zero air have
20–50 pptv HCHO, but we can obtain lower values by passing the air through
a molecular sieve. With clean FEP or PFA tubing, measurements of zero or near-zero are easily obtained, even after exposure to 10 or 20 ppbv of HCHO. The
ISAF instrument has no measurable chemical offset or bias of HCHO and shows
little if any “memory” of exposure to high concentrations of HCHO (see
Sects. 4.4 and 4.5 for more detail). However, the uncertainty shown in the
inset of Fig. 8 is typical at these low concentrations. At very low
fluorescence photon count rates, small changes in the laser scatter between
the online and offline position can lead to an offset in the measurement.
This is because we assume that the laser scatter is equal at the online and
offline wavelengths. Though the wavelength difference is small (0.005 nm)
there is some laser pointing change due to walk-off in the harmonic
generation stage between these wavelengths, and potentially a small
difference in laser scatter. This difference can result in a negative or
positive offset. We assign an uncertainty of <inline-formula><mml:math display="inline"><mml:mo>±</mml:mo></mml:math></inline-formula>10 pptv to account for
this potential offset.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F10"><caption><p>The time response of the instrument is demonstrated with a 10 ms
pulse of HCHO added to the sample air flow in the lab. The measured HCHO
shows the rapid rise expected with the flush time of <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 115 ms. The
decay of 190 ms indicates second-order effects such as wall interactions and
the presence of small volumes with longer flush times.</p></caption>
          <?xmltex \igopts{width=184.942913pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f10.pdf"/>

        </fig>

</sec>
<sec id="Ch1.S4.SS4">
  <title>Time response</title>
      <p>Though the nominal sampling frequency is 10 Hz, the flush time of sampling
volumes can limit the effective instrument time response. Minimizing the
response time is critical for applications that require measurements of fast
changes in concentration, such as eddy covariance and sampling of discrete
plumes and for eliminating contamination from exposure to high concentrations
of HCHO. For typical inlet pressures and flow rates, the plug-flow transit
time from the tip of the inlet to the instrument chassis is 43–120 ms. This
creates a delay between sampling and detection but should not cause
significant “smearing” of concentration fluctuations, as the flow is
typically turbulent (<italic>Re</italic> <inline-formula><mml:math display="inline"><mml:mo>∼</mml:mo></mml:math></inline-formula> 3000–6000). Such smearing can,
however, occur in the pressure controller and sample cell. Based on the
internal cell volume of 60 cm<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mn mathvariant="normal">3</mml:mn></mml:msup></mml:math></inline-formula> and a typical volumetric flow rate of
31 L min<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula> (2.75 sLm at 100 mbar and 35 <inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mo>∘</mml:mo></mml:msup></mml:math></inline-formula>C), the expected
<inline-formula><mml:math display="inline"><mml:mrow><mml:mn mathvariant="normal">1</mml:mn><mml:mo>/</mml:mo><mml:mi>e</mml:mi></mml:mrow></mml:math></inline-formula> cell flush time is 115 ms. Figure 10 shows the instantaneous response
to a 10 ms pulse of HCHO, delivered via a fast switching valve (IEP series,
The Lee Company). The observed decay time is 190 ms, somewhat slower than
that predicted from flow characteristics. The extra delay is likely due to
the pressure controller, which acts as a critical orifice and adds to the
internal volume of the sampling system. A faster time response can be
achieved by increasing the flow rate through the sample cell, which depends
on the conductance of the pressure controller and the pumping speed.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F11"><caption><p>In situ measurements of HCHO are shown for the flight of 2 June on
the NASA DC-8 during the Deep Convective Clouds and Chemistry (DC3) campaign in 2012. The top shows 1 s HCHO and GPS
altitude. The lower portions are expanded views of the boundary layer (BL)
leg and the upper tropospheric (UT) leg. The fast response to large changes
in ambient HCHO are evident in the BL leg. The sensitivity (better than
20 pptv s<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula>) is evident in the UT leg.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f11.pdf"/>

        </fig>

      <p>The fast time response is an important feature of
contamination-free detection at low concentrations of HCHO. HCHO continuously adsorbs and desorbs from
surfaces inside the instrument sampling system. This reversible process can
lead to biases, especially when low concentrations are sampled after exposure
of the instrument to very high concentrations. The effects of reversible
adsorption and desorption of HCHO from surfaces is minimized by having a
large mass flow, small volumes, and low surface area, the same features that
lead to the fast time response. The measured time response includes both the
effects of volumetric displacement and adsorption/desorption. The impact on
in situ measurements is demonstrated in Sect. 4.5.</p>
</sec>
<sec id="Ch1.S4.SS5">
  <title>Field demonstration</title>
      <p>As of March 2014, ISAF had flown successfully on four field missions spanning
three platforms and a wide range of chemical conditions. For many of these
flights, the instrument has flown autonomously, without an operator.
Figure 11 presents observations from a single flight on the NASA DC-8 during
the Deep Convective Clouds and Chemistry (DC3) experiment, which took place
over the central US in May and June 2012 (Barth et al., 2015). On this
flight, the DC-8 sampled at altitudes of 0–12 km, with HCHO mixing ratios
ranging from 50–5500 pptv. The expanded views in Fig. 11 show a boundary
layer leg and an upper tropospheric leg to illustrate the time response,
dynamic range and sensitivity of ISAF. During DC3, ISAF was flown
simultaneously with a well-established instrument, the University of Colorado DFGAS
(Difference Frequency Generation Absorption Spectrometer) (Weibring et al.,
2007). Preliminary analysis indicates excellent agreement between these two
instruments, and a rigorous inter-comparison will be the subject of a future
publication.</p>

      <?xmltex \floatpos{t}?><fig id="Ch1.F12"><caption><p>The time series obtained during the SEAC4RS campaign on the NASA DC-8
on 26 August 2013. The panel <bold>(a)</bold> shows the HCHO measured at a
constant pressure altitude of 5 km as the DC-8 sampled the plume from the Rim
Fire in Stanislaus National Forest as the plume traveled eastward over
California and Nevada. Note the linear and logarithmic scaling.
Panels <bold>(b)</bold> and <bold>(c)</bold> show a window of 30 s starting at
23:07 UTC expanded with data averaged at 10 and 1 Hz for linear and
logarithmic scaling. Panels <bold>(c)</bold> and <bold>(d)</bold> show the same for
the 30 s window starting at 23:13 UTC.</p></caption>
          <?xmltex \igopts{width=236.157874pt}?><graphic xlink:href="https://www.atmos-meas-tech.net/8/541/2015/amt-8-541-2015-f12.pdf"/>

        </fig>

      <p>The time response of the instrument in flight is illustrated in Fig. 12.
These data were obtained during the Studies of Emissions and Atmospheric
Composition, Clouds and Climate Coupling Regional Surveys (SEAC4RS) campaign
in 2013. During this flight on 26 August, the DC-8 sampled the smoke plume
from the Rim Fire in Stanislaus National Forest as it passed from California
to Nevada. In this plume, measured at a constant pressure altitude of 5 km,
the HCHO abundance reached over 200 ppbv. The data is shown in linear and
log scale to illustrate the several orders of magnitude change in HCHO during
this segment. The observations demonstrate the ability of ISAF to measure low
concentrations of HCHO even after exposure to very high concentrations for
several minutes.</p>
      <p>The expanded views provide 30 s snapshots of two regions where the abundance
of HCHO is highly variable. The 10 Hz data is shown along with the 1 Hz
data to demonstrate the fast response of the instrument. These 10 Hz data
show much more structure than the 1 Hz and at the edges of plume filaments
the transitions are much steeper. For example, the edge of the plume at
time <inline-formula><mml:math display="inline"><mml:mo>=</mml:mo></mml:math></inline-formula> 22 s in Fig. 12b and c drops from an abundance of 65 to 25 ppb
(<inline-formula><mml:math display="inline"><mml:mrow><mml:mn mathvariant="normal">1</mml:mn><mml:mo>/</mml:mo><mml:mi>e</mml:mi></mml:mrow></mml:math></inline-formula>) in 0.3 s. If this were limited by instrument response alone, that is
if the plume were to have an abrupt boundary, the corresponding instrument
time constant would be 0.3 s. Several filaments have steeper edges, near
0.2 s, while most show broader edges that indicate more diffuse boundaries.
These steep edges are consistent with the ISAF time response of around
0.2 s shown in Fig. 10. These examples – fast time response and the ability
to measure highly variable concentrations – are important demonstrations of
the small effect of the reversible adsorption and desorption of HCHO in the
ISAF inlet and instrument. Reversible adsorption and desorption are clearly factors because the time constant is larger than the volumetric flush
time, but the effects are small, on the order of 100 ms.</p>

<?xmltex \floatpos{t}?><table-wrap id="Ch1.T1"><caption><p>Summary of the ISAF instrument capability.</p></caption><oasis:table frame="topbot"><oasis:tgroup cols="2">
     <oasis:colspec colnum="1" colname="col1" align="left"/>
     <oasis:colspec colnum="2" colname="col2" align="left"/>
     <oasis:thead>
       <oasis:row>  
         <oasis:entry colname="col1">Size</oasis:entry>  
         <oasis:entry colname="col2">38 <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 43 <inline-formula><mml:math display="inline"><mml:mo>×</mml:mo></mml:math></inline-formula> 60 cm</oasis:entry>
       </oasis:row>
       <oasis:row>  
         <oasis:entry colname="col1">Weight</oasis:entry>  
         <oasis:entry colname="col2">25 kg</oasis:entry>
       </oasis:row>
       <oasis:row>  
         <oasis:entry colname="col1">Power</oasis:entry>  
         <oasis:entry colname="col2">200 W</oasis:entry>
       </oasis:row>
       <oasis:row>  
         <oasis:entry colname="col1">Precision</oasis:entry>  
         <oasis:entry colname="col2">20 pptv s<inline-formula><mml:math display="inline"><mml:msup><mml:mi/><mml:mrow><mml:mo>-</mml:mo><mml:mn mathvariant="normal">1</mml:mn></mml:mrow></mml:msup></mml:math></inline-formula></oasis:entry>
       </oasis:row>
       <oasis:row>  
         <oasis:entry colname="col1">Time response</oasis:entry>  
         <oasis:entry colname="col2"><inline-formula><mml:math display="inline"><mml:mo>&lt;</mml:mo></mml:math></inline-formula> 0.2 s</oasis:entry>
       </oasis:row>
       <oasis:row>  
         <oasis:entry colname="col1">Accuracy</oasis:entry>  
         <oasis:entry colname="col2"><inline-formula><mml:math display="inline"><mml:mo>±</mml:mo></mml:math></inline-formula>10 %</oasis:entry>
       </oasis:row>
       <oasis:row rowsep="1">  
         <oasis:entry colname="col1">Data rate</oasis:entry>  
         <oasis:entry colname="col2">1 s (100 ms on request)</oasis:entry>
       </oasis:row>
     </oasis:thead>
     <oasis:tbody>
       <oasis:row>  
         <oasis:entry namest="col1" nameend="col2">Operation on the DC-8 uses pump (10 kg <inline-formula><mml:math display="inline"><mml:mo>+</mml:mo></mml:math></inline-formula> 300 W) </oasis:entry>
       </oasis:row>
       <oasis:row>  
         <oasis:entry namest="col1" nameend="col2"> and a control interface box (10 kg) </oasis:entry>
       </oasis:row>
       <oasis:row>  
         <oasis:entry namest="col1" nameend="col2">Operation on the P3 and GV uses pump (5 kg <inline-formula><mml:math display="inline"><mml:mo>+</mml:mo></mml:math></inline-formula> 200 W)  </oasis:entry>
       </oasis:row>
       <oasis:row>  
         <oasis:entry namest="col1" nameend="col2"> and a control interface box (10 kg) </oasis:entry>
       </oasis:row>
     </oasis:tbody>
   </oasis:tgroup></oasis:table></table-wrap>

</sec>
</sec>
<sec id="Ch1.S5" sec-type="conclusions">
  <title>Summary</title>
      <p>The NASA ISAF instrument is a compact, high-sensitivity, field-proven
instrument for airborne observations of HCHO throughout the troposphere and
lower stratosphere. The instrument capabilities are summarized in Table 1.
It has already flown on the NASA DC-8, the NOAA WP-3D and the NCAR G-V
aircraft on four aircraft field campaigns: DC3, Southeast Nexus (SENEX),
SEAC4RS, and CONvective Transport of Active Species in the Tropics
(CONTRAST). Future work will include modifications for deployment on
high-altitude aircraft, such as the NASA ER-2 or WB-57, to achieve an even
more complete observational data set.</p>
</sec>

      
      </body>
    <back><ack><title>Acknowledgements</title><p>This research was funded by the Goddard Internal Research and Development
(IRAD) program and the NASA ROSES grant NNH10ZDA001N-SEAC4RS. The NASA
Postdoctoral Program provided funding for M. Cazorla.<?xmltex \hack{\newline}?><?xmltex \hack{\newline}?> Edited by: E. C. Apel</p></ack><?xmltex \hack{\newpage}?><?xmltex \hack{\newpage}?><ref-list>
    <title>References</title>

      <ref id="bib1.bib1"><label>1</label><mixed-citation>Apel, E. C., Olson, J. R., Crawford, J. H., Hornbrook, R. S., Hills, A. J.,
Cantrell, C. A., Emmons, L. K., Knapp, D. J., Hall, S., Mauldin III, R. L.,
Weinheimer, A. J., Fried, A., Blake, D. R., Crounse, J. D., Clair, J. M. St.,
Wennberg, P. O., Diskin, G. S., Fuelberg, H. E., Wisthaler, A., Mikoviny, T.,
Brune, W., and Riemer, D. D.: Impact of the deep convection of isoprene and
other reactive trace species on radicals and ozone in the upper troposphere,
Atmos. Chem. Phys., 12, 1135–1150, <ext-link xlink:href="http://dx.doi.org/10.5194/acp-12-1135-2012" ext-link-type="DOI">10.5194/acp-12-1135-2012</ext-link>, 2012.</mixed-citation></ref>
      <ref id="bib1.bib2"><label>2</label><mixed-citation>Baidar, S., Oetjen, H., Coburn, S., Dix, B., Ortega, I., Sinreich, R., and
Volkamer, R.: The CU Airborne MAX-DOAS instrument: vertical profiling of
aerosol extinction and trace gases, Atmos. Meas. Tech., 6, 719–739,
<ext-link xlink:href="http://dx.doi.org/10.5194/amt-6-719-2013" ext-link-type="DOI">10.5194/amt-6-719-2013</ext-link>, 2013.</mixed-citation></ref>
      <ref id="bib1.bib3"><label>3</label><mixed-citation>Barth, M. C., Kim, S. W., Skamarock, W. C., Stuart, A. L., Pickering, K. E.,
and Ott, L. E.: Simulations of the redistribution of formaldehyde, formic
acid, and peroxides in the 10 July 1996 Stratospheric-Tropospheric
Experiment: Radiation, Aerosols, and Ozone deep convection storm, J. Geophys.
Res.-Atmos., 112, D13310, <ext-link xlink:href="http://dx.doi.org/10.1029/2006jd008046" ext-link-type="DOI">10.1029/2006jd008046</ext-link>, 2007.</mixed-citation></ref>
      <ref id="bib1.bib4"><label>4</label><mixed-citation>Barth, M. C., Cantrell, C. A., Brune, W. H., Rutledge, S. A., Crawford, J.
H., Huntrieser, H., Carey, L. D., MacGorman, D., Weisman, M., Pickering, K.
E., Bruning, E., Anderson, B., Apel, E., Biggerstaff, M., Campos, T.,
Campuzano-Jost, P., Cohen, R. C., Crounse, J. D.,  Day, D. A., Diskin, G.,
Flocke, F., Fried, A., Garland, C., Heikes, B., Honomichi, S., Hornbrook, R.,
Huey, L. G., Jimenez, J., Lang, T., Lichtenstern, M., Mikoviny, T.,
O'Sullivan, D., Pan, L., Peischl, J., Pollack, I., Riemer, D., Ryerson, T.,
Schlager, H., St. Clair, J. M., Walega, J., Weibring, P., Weinheimer, A.,
Wennberg, P. O., Wisthaler, A., Wooldridge, P. J., and Ziegler, C.: The deep
convective clouds and chemistry (DC3) field campaign, J. Geophys. Res.,
submitted, 2015.</mixed-citation></ref>
      <ref id="bib1.bib5"><label>5</label><mixed-citation>Becker, K. H., Schurath, U., and Tatarczyk, T.: Fluorescence determination
of low formaldehyde concentrations in air by dye laser excitation, Appl.
Optics, 14, 310–313, <ext-link xlink:href="http://dx.doi.org/10.1364/ao.14.000310" ext-link-type="DOI">10.1364/ao.14.000310</ext-link>, 1975.</mixed-citation></ref>
      <ref id="bib1.bib6"><label>6</label><mixed-citation>Cardenas, L. M., Brassington, D. J., Allan, B. J., Coe, H., Alicke, B.,
Platt, U., Wilson, K. M., Plane, J. M. C., and Penkett, S. A.:
Intercomparison of formaldehyde measurements in clean and polluted
atmospheres, J. Atmos. Chem., 37, 53–80, <ext-link xlink:href="http://dx.doi.org/10.1023/a:1006383520819" ext-link-type="DOI">10.1023/a:1006383520819</ext-link>, 2000.</mixed-citation></ref>
      <ref id="bib1.bib7"><label>7</label><mixed-citation>Chance, K., Palmer, P. I., Spurr, R. J. D., Martin, R. V., Kurosu, T. P.,
and Jacob, D. J.: Satellite observations of formaldehyde over North America
from GOME, Geophys. Res. Lett., 27, 3461–3464, <ext-link xlink:href="http://dx.doi.org/10.1029/2000gl011857" ext-link-type="DOI">10.1029/2000gl011857</ext-link>,
2000.</mixed-citation></ref>
      <ref id="bib1.bib8"><label>8</label><mixed-citation>Chatfield, R. B., Ren, X., Brune, W., and Schwab, J.: Controls on urban
ozone production rate as indicated by formaldehyde oxidation rate an nitric
oxide, Atmos. Environ., 44, 5395–5406, 2010.</mixed-citation></ref>
      <ref id="bib1.bib9"><label>9</label><mixed-citation>Co, D. T., Hanisco, T. F., Anderson, J. G., and Keutsch, F. N.: Rotationally
resolved absorption cross sections of formaldehyde in the
28100–28500 cm(-1) (351–356 nm) spectral region: Implications for in situ
LIF measurements, J. Phys. Chem. A, 109, 10675–10682, <ext-link xlink:href="http://dx.doi.org/10.1021/jp053466i" ext-link-type="DOI">10.1021/jp053466i</ext-link>,
2005.</mixed-citation></ref>
      <ref id="bib1.bib10"><label>10</label><mixed-citation>DiGangi, J. P., Boyle, E. S., Karl, T., Harley, P., Turnipseed, A., Kim, S.,
Cantrell, C., Maudlin III, R. L., Zheng, W., Flocke, F., Hall, S. R.,
Ullmann, K., Nakashima, Y., Paul, J. B., Wolfe, G. M., Desai, A. R., Kajii,
Y., Guenther, A., and Keutsch, F. N.: First direct measurements of
formaldehyde flux via eddy covariance: implications for missing in-canopy
formaldehyde sources, Atmos. Chem. Phys., 11, 10565–10578,
<ext-link xlink:href="http://dx.doi.org/10.5194/acp-11-10565-2011" ext-link-type="DOI">10.5194/acp-11-10565-2011</ext-link>, 2011.</mixed-citation></ref>
      <ref id="bib1.bib11"><label>11</label><mixed-citation>Duncan, B., Yoshida, Y., Olson, J., Sillman, S., Martin, R., Lamsal, L., Hu,
Y., Pickering, K., Retscher, C., Allen, D., and Crawford, J.: Application of
OMI observations to a space-based indicator of NOx and VOC controls on
surface ozone formation, Atmos. Environ., 44, 2213–2223,
<ext-link xlink:href="http://dx.doi.org/10.1016/j.atmosenv.2010.03.010" ext-link-type="DOI">10.1016/j.atmosenv.2010.03.010</ext-link>, 2010.</mixed-citation></ref>
      <ref id="bib1.bib12"><label>12</label><mixed-citation>Edwards, P., Evans, M. J., Commane, R., Ingham, T., Stone, D., Mahajan, A.
S., Oetjen, H., Dorsey, J. R., Hopkins, J. R., Lee, J. D., Moller, S. J.,
Leigh, R., Plane, J. M. C., Carpenter, L. J., and Heard, D. E.: Hydrogen
oxide photochemistry in the northern Canadian spring time boundary layer, J.
Geophys. Res.-Atmos., 116, D22306, <ext-link xlink:href="http://dx.doi.org/10.1029/2011jd016390" ext-link-type="DOI">10.1029/2011jd016390</ext-link>, 2011.</mixed-citation></ref>
      <ref id="bib1.bib13"><label>13</label><mixed-citation>Edwards, P. M., Young, C. J., Aikin, K., deGouw, J., Dubé, W. P., Geiger,
F., Gilman, J., Helmig, D., Holloway, J. S., Kercher, J., Lerner, B., Martin,
R., McLaren, R., Parrish, D. D., Peischl, J., Roberts, J. M., Ryerson, T. B.,
Thornton, J., Warneke, C., Williams, E. J., and Brown, S. S.: Ozone
photochemistry in an oil and natural gas extraction region during winter:
simulations of a snow-free season in the Uintah Basin, Utah, Atmos. Chem.
Phys., 13, 8955–8971, <ext-link xlink:href="http://dx.doi.org/10.5194/acp-13-8955-2013" ext-link-type="DOI">10.5194/acp-13-8955-2013</ext-link>, 2013.</mixed-citation></ref>
      <ref id="bib1.bib14"><label>14</label><mixed-citation>Fortems-Cheiney, A., Chevallier, F., Pison, I., Bousquet, P., Saunois, M.,
Szopa, S., Cressot, C., Kurosu, T. P., Chance, K., and Fried, A.: The
formaldehyde budget as seen by a global-scale multi-constraint and
multi-species inversion system, Atmos. Chem. Phys., 12, 6699–6721,
<ext-link xlink:href="http://dx.doi.org/10.5194/acp-12-6699-2012" ext-link-type="DOI">10.5194/acp-12-6699-2012</ext-link>, 2012.</mixed-citation></ref>
      <ref id="bib1.bib15"><label>15</label><mixed-citation>Fried, A., Olson, J. R., Walega, J. G., Crawford, J. H., Chen, G., Weibring,
P., Richter, D., Roller, C., Tittel, F., Porter, M., Fuelberg, H., Halland,
J., Bertram, T. H., Cohen, R. C., Pickering, K., Heikes, B. G., Snow, J. A.,
Shen, H., O'Sullivan, D. W., Brune, W. H., Ren, X., Blake, D. R., Blake, N.,
Sachse, G., Diskin, G. S., Podolske, J., Vay, S. A., Shetter, R. E., Hall, S.
R., Anderson, B. E., Thornhill, L., Clarke, A. D., McNaughton, C. S., Singh,
H. B., Avery, M. A., Huey, G., Kim, S., and Millet, D. B.: Role of convection
in redistributing formaldehyde to the upper troposphere over North America
and the North Atlantic during the summer 2004 INTEX campaign, J. Geophys.
Res.-Atmos, 113, D17306, <ext-link xlink:href="http://dx.doi.org/10.1029/2007jd009760" ext-link-type="DOI">10.1029/2007jd009760</ext-link>, 2008.</mixed-citation></ref>
      <ref id="bib1.bib16"><label>16</label><mixed-citation>Gilpin, T., Apel, E., Fried, A., Wert, B., Calvert, J., Genfa, Z., Dasgupta,
P. Harder, J. W., Heikes, B., Hopkins, B., Westberg, H.,
Kleindiens, T., Lee, Y.-N., Zhou, X.,
Lonneman, W., and Sewell, S.: Intercomparison of six ambient CH<inline-formula><mml:math display="inline"><mml:msub><mml:mi/><mml:mn mathvariant="normal">2</mml:mn></mml:msub></mml:math></inline-formula>O
measurement techniques, J. Geophys. Res.-Atmos., 102, 21161–21188, 1997.</mixed-citation></ref>
      <ref id="bib1.bib17"><label>17</label><mixed-citation>Hagopian, J.: Carbon nanotubes for stray light suppression, Art. 9417, NASA
Tech. Briefs, 2011.</mixed-citation></ref>
      <ref id="bib1.bib18"><label>18</label><mixed-citation>Hak, C., Pundt, I., Trick, S., Kern, C., Platt, U., Dommen, J.,
Ordóñez, C., Prévôt, A. S. H., Junkermann, W.,
Astorga-Lloréns, C., Larsen, B. R., Mellqvist, J., Strandberg, A., Yu,
Y., Galle, B., Kleffmann, J., Lörzer, J. C., Braathen, G. O., and
Volkamer, R.: Intercomparison of four different in-situ techniques for
ambient formaldehyde measurements in urban air, Atmos. Chem. Phys., 5,
2881–2900, <ext-link xlink:href="http://dx.doi.org/10.5194/acp-5-2881-2005" ext-link-type="DOI">10.5194/acp-5-2881-2005</ext-link>, 2005.</mixed-citation></ref>
      <ref id="bib1.bib19"><label>19</label><mixed-citation>Hottle, J. R., Huisman, A. J., Digangi, J. P., Kammrath, A., Galloway, M.
M., Coens, K. L., and Keutsch, F. N.: A Laser Induced Fluorescence-Based
Instrument for In-Situ Measurements of Atmospheric Formaldehyde, Environ.
Sci. Tech., 43, 790–795, 2009.</mixed-citation></ref>
      <ref id="bib1.bib20"><label>20</label><mixed-citation>Kaiser, J., Li, X., Tillmann, R., Acir, I., Holland, F., Rohrer, F., Wegener,
R., and Keutsch, F. N.: Intercomparison of Hantzsch and
fiber-laser-induced-fluorescence formaldehyde measurements, Atmos. Meas.
Tech., 7, 1571–1580, <ext-link xlink:href="http://dx.doi.org/10.5194/amt-7-1571-2014" ext-link-type="DOI">10.5194/amt-7-1571-2014</ext-link>, 2014.</mixed-citation></ref>
      <ref id="bib1.bib21"><label>21</label><mixed-citation>Luecken, D. J., Hutzell, W. T., Strum, M. L., and Pouliot, G. A.: Regional
sources of atmospheric formaldehyde and acetaldehyde, and implications for
atmospheric modeling, Atmos. Environ., 47, 477–490,
<ext-link xlink:href="http://dx.doi.org/10.1016/j.atmosenv.2011.10.005" ext-link-type="DOI">10.1016/j.atmosenv.2011.10.005</ext-link>, 2012.</mixed-citation></ref>
      <ref id="bib1.bib22"><label>22</label><mixed-citation>Marais, E. A., Jacob, D. J., Kurosu, T. P., Chance, K., Murphy, J. G.,
Reeves, C., Mills, G., Casadio, S., Millet, D. B., Barkley, M. P., Paulot,
F., and Mao, J.: Isoprene emissions in Africa inferred from OMI observations
of formaldehyde columns, Atmos. Chem. Phys., 12, 6219–6235,
<ext-link xlink:href="http://dx.doi.org/10.5194/acp-12-6219-2012" ext-link-type="DOI">10.5194/acp-12-6219-2012</ext-link>, 2012.</mixed-citation></ref>
      <ref id="bib1.bib23"><label>23</label><mixed-citation>Meller, R. and Moortgat, G. K.: Temperature dependence of the absorption
cross sections of formaldehyde between 223 and 323 K in the wavelength range
225–375 nm, J. Geophys. Res.-Atmos., 105, 7089–7101,
<ext-link xlink:href="http://dx.doi.org/10.1029/1999jd901074" ext-link-type="DOI">10.1029/1999jd901074</ext-link>, 2000.</mixed-citation></ref>
      <ref id="bib1.bib24"><label>24</label><mixed-citation>Millet, D. B., Jacob, D. J., Boersma, K. F., Fu, T. M., Kurosu, T. P.,
Chance, K., Heald, C. L., and Guenther, A.: Spatial distribution of isoprene
emissions from North America derived from formaldehyde column measurements by
the OMI satellite sensor, J. Geophys. Res.-Atmos., 113, D02307,
<ext-link xlink:href="http://dx.doi.org/10.1029/2007jd008950" ext-link-type="DOI">10.1029/2007jd008950</ext-link>, 2008.</mixed-citation></ref>
      <ref id="bib1.bib25"><label>25</label><mixed-citation>Möhlmann, G. R.: Formaldehyde detection in air by laser-induced
fluorescence, Appl. Spectrosc., 39, 98–101, <ext-link xlink:href="http://dx.doi.org/10.1366/0003702854249088" ext-link-type="DOI">10.1366/0003702854249088</ext-link>,
1985.</mixed-citation></ref>
      <ref id="bib1.bib26"><label>26</label><mixed-citation>Olaguer, E. P., Rappengluck, B., Lefer, B., Stutz, J., Dibb, J., Griffin,
R., Brune, W. H., Shauck, M., Buhr, M., Jeffries, H., Vizuete, W., and Pinto,
J. P.: Deciphering the Role of Radical Precursors during the Second Texas Air
Quality Study, J. Air Waste Manage., 59, 1258–1277,
<ext-link xlink:href="http://dx.doi.org/10.3155/1047-3289.59.11.1258" ext-link-type="DOI">10.3155/1047-3289.59.11.1258</ext-link>, 2009.</mixed-citation></ref>
      <ref id="bib1.bib27"><label>27</label><mixed-citation>Palmer, P. I., Jacob, D. J., Fiore, A. M., Martin, R. V., Chance, K., and
Kurosu, T. P.: Mapping isoprene emissions over North America using
formaldehyde column observations from space, J. Geophys. Res.-Atmos., 108,
4180, <ext-link xlink:href="http://dx.doi.org/10.1029/2002jd002153" ext-link-type="DOI">10.1029/2002jd002153</ext-link>, 2003.</mixed-citation></ref>
      <ref id="bib1.bib28"><label>28</label><mixed-citation>Palmer, P. I., Abbot, D. S., Fu, T. M., Jacob, D. J., Chance, K., Kurosu, T.
P., Guenther, A., Wiedinmyer, C., Stanton, J. C., Pilling, M. J., Pressley,
S. N., Lamb, B., and Sumner, A. L.: Quantifying the seasonal and interannual
variability of North American isoprene emissions using satellite observations
of the formaldehyde column, J. Geophys. Res.-Atmos., 111, D12315,
<ext-link xlink:href="http://dx.doi.org/10.1029/2005jd006689" ext-link-type="DOI">10.1029/2005jd006689</ext-link>, 2006.</mixed-citation></ref>
      <ref id="bib1.bib29"><label>29</label><mixed-citation>Shibuya, K., Holtermann, D. L., Peacock, J. R., and Lee, E. K. C.:
Fluorescence emission-spectroscopy, radiative lifetimes, and
vibrational-relaxation rates of the 40 and 41 single vibronic levels of D2CO
and HDCO (A1A2), J. Phys. Chem., 83, 940–944, <ext-link xlink:href="http://dx.doi.org/10.1021/j100471a012" ext-link-type="DOI">10.1021/j100471a012</ext-link>, 1979.</mixed-citation></ref>
      <ref id="bib1.bib30"><label>30</label><mixed-citation>Weibring, P., Richter, D., Walega, J. G., and Fried, A.: First demonstration
of a high performance difference frequency spectrometer on airborne
platforms, Opt. Express, 15, 13476–13495, <ext-link xlink:href="http://dx.doi.org/10.1364/oe.15.013476" ext-link-type="DOI">10.1364/oe.15.013476</ext-link>, 2007.</mixed-citation></ref>
      <ref id="bib1.bib31"><label>31</label><mixed-citation>Wert, B. P., Fried, A., Henry, B., and Cartier, S.: Evaluation of inlets
used for the airborne measurement of formaldehyde, J. Geophys. Res.-Atmos.,
107, ACH3.1–ACH3.13, <ext-link xlink:href="http://dx.doi.org/10.1029/2001jd001072" ext-link-type="DOI">10.1029/2001jd001072</ext-link>, 2002.
</mixed-citation></ref><?xmltex \hack{\newpage}?>
      <ref id="bib1.bib32"><label>32</label><mixed-citation>Wisthaler, A., Apel, E. C., Bossmeyer, J., Hansel, A., Junkermann, W.,
Koppmann, R., Meier, R., Müller, K., Solomon, S. J., Steinbrecher, R.,
Tillmann, R., and Brauers, T.: Technical Note: Intercomparison of
formaldehyde measurements at the atmosphere simulation chamber SAPHIR, Atmos.
Chem. Phys., 8, 2189–2200, <ext-link xlink:href="http://dx.doi.org/10.5194/acp-8-2189-2008" ext-link-type="DOI">10.5194/acp-8-2189-2008</ext-link>, 2008.
</mixed-citation></ref><?xmltex \hack{\newpage}?>
      <ref id="bib1.bib33"><label>33</label><mixed-citation>Yokelson, R. J., Burling, I. R., Gilman, J. B., Warneke, C., Stockwell, C.
E., de Gouw, J., Akagi, S. K., Urbanski, S. P., Veres, P., Roberts, J. M.,
Kuster, W. C., Reardon, J., Griffith, D. W. T., Johnson, T. J., Hosseini, S.,
Miller, J. W., Cocker III, D. R., Jung, H., and Weise, D. R.: Coupling field
and laboratory measurements to estimate the emission factors of identified
and unidentified trace gases for prescribed fires, Atmos. Chem. Phys., 13,
89–116, <ext-link xlink:href="http://dx.doi.org/10.5194/acp-13-89-2013" ext-link-type="DOI">10.5194/acp-13-89-2013</ext-link>, 2013.</mixed-citation></ref>

  </ref-list><app-group content-type="float"><app><title/>

    </app></app-group></back>
    </article>