Articles | Volume 10, issue 7
https://doi.org/10.5194/amt-10-2383-2017
https://doi.org/10.5194/amt-10-2383-2017
Research article
 | 
03 Jul 2017
Research article |  | 03 Jul 2017

Evaluation and environmental correction of ambient CO2 measurements from a low-cost NDIR sensor

Cory R. Martin, Ning Zeng, Anna Karion, Russell R. Dickerson, Xinrong Ren, Bari N. Turpie, and Kristy J. Weber

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by Cory Martin on behalf of the Authors (05 Apr 2017)  Author's response    Manuscript
ED: Referee Nomination & Report Request started (10 Apr 2017) by Szymon Malinowski
RR by Anonymous Referee #2 (29 Apr 2017)
ED: Publish subject to minor revisions (Editor review) (08 May 2017) by Szymon Malinowski
AR by Svenja Lange on behalf of the Authors (19 May 2017)  Author's response    Manuscript
ED: Publish subject to technical corrections (01 Jun 2017) by Szymon Malinowski
Download
Short summary
A low-cost sensor for measuring carbon dioxide is evaluated for its performance in detecting concentrations in Earth's atmosphere. After a multivariate regression correcting for environmental variables, the root mean square error between it and a research-grade gas analyzer is less than 0.5 % of the observed average value. This demonstrates the viability for using these sensors in certain real-world atmospheric observing applications.