Effects of variation in background mixing ratios of N2, O2, and Ar on the measurement of δ18O–H2O and δ2H–H2O values by cavity ring-down spectroscopy

Jennifer E. Johnson and Chris W. Rella

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by Jennifer Johnson on behalf of the Authors (30 Jun 2017)  Author's response   Manuscript 
ED: Publish as is (13 Jul 2017) by William Ward
AR by Jennifer Johnson on behalf of the Authors (13 Jul 2017)
Download
Short summary
Cavity ring-down spectrometers (CRDS) can make accurate measurements of trace gases when the inert gas background is constant but are prone to errors when the background is variable. Here, we report that in one CRDS analyzer (L2120-i, Picarro), these errors result from broadening, narrowing, and shifting of the target absorption lines as well as neighboring lines. New spectral acquisition and analysis strategies are needed to make CRDS measurements robust to variable backgrounds.