Articles | Volume 10, issue 12
https://doi.org/10.5194/amt-10-4613-2017
https://doi.org/10.5194/amt-10-4613-2017
Research article
 | 
01 Dec 2017
Research article |  | 01 Dec 2017

Validation of spectroscopic gas analyzer accuracy using gravimetric standard gas mixtures: impact of background gas composition on CO2 quantitation by cavity ring-down spectroscopy

Jeong Sik Lim, Miyeon Park, Jinbok Lee, and Jeongsoon Lee

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by Anna Mirena Feist-Polner on behalf of the Authors (24 Aug 2017)  Author's response
ED: Publish as is (16 Sep 2017) by Christian Brümmer
Download
Short summary
Effect of background gas composition on the spectroscopic measurement of CO2 concentration at ambient levels has been investigated. The wavelength-scanned cavity ring-down spectroscopy was employed to explore the spectral line shape of CO2 at 1603 nm. It is revealed that the instrument response should be corrected with respect to a total pressure broadening coefficient given by a matrix composition. The correction method is validated by high accurate gravimetric standard gas mixture.