Characterization of blackbody inhomogeneity and its effect on the retrieval results of the GLORIA instrument

Anne Kleinert, Isabell Krisch, Jörn Ungermann, Albert Adibekyan, Berndt Gutschwager, and Christian Monte

Viewed

Total article views: 2,726 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
1,699 955 72 2,726 100 94
  • HTML: 1,699
  • PDF: 955
  • XML: 72
  • Total: 2,726
  • BibTeX: 100
  • EndNote: 94
Views and downloads (calculated since 08 Mar 2018)
Cumulative views and downloads (calculated since 08 Mar 2018)

Viewed (geographical distribution)

Total article views: 2,726 (including HTML, PDF, and XML) Thereof 2,566 with geography defined and 160 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 

Cited

Latest update: 23 Nov 2024
Download
Short summary
This study investigates the required accuracy of radiometric calibration sources for remote sensing instruments to properly resolve decadal trends of climate relevant trace species like ozone, water vapor and temperature. The required temperature knowledge of the calibration source is in the order of 100 mK. This is demonstrated by a Monte Carlo simulation. The results are confirmed using real measurements acquired by the GLORIA instrument.