Articles | Volume 11, issue 9
https://doi.org/10.5194/amt-11-5167-2018
https://doi.org/10.5194/amt-11-5167-2018
Research article
 | 
11 Sep 2018
Research article |  | 11 Sep 2018

Performance of the FMI cosine error correction method for the Brewer spectral UV measurements

Kaisa Lakkala, Antti Arola, Julian Gröbner, Sergio Fabian León-Luis, Alberto Redondas, Stelios Kazadzis, Tomi Karppinen, Juha Matti Karhu, Luca Egli, Anu Heikkilä, Tapani Koskela, Antonio Serrano, and José Manuel Vilaplana
Download
Short summary
The performance of the cosine error correction method for correcting spectral UV measurements of the Brewer spectroradiometer was studied. The correction depends on the sky radiation distribution, which can change during one spectral scan. The results showed that the correction varied between 4 and 14 %, and that the relative differences between the reference and the Brewer diminished by 10 %. The method is applicable to other instruments as long as the required input parameters are available.