A calibration procedure which accounts for non-linearity in single-monochromator Brewer ozone spectrophotometer measurements

Zahra Vaziri Zanjani, Omid Moeini, Tom McElroy, David Barton, and Vladimir Savastiouk

Data sets

Brewer 009_25 Oct 2010 Z. Vaziri Zanjani https://doi.org/10.5281/zenodo.2531911

Brewer 009_14 Jul 2010 Z. Vaziri Zanjani https://doi.org/10.5281/zenodo.2531905

Brewer 119_14 Jul 2010_AM, Z. Vaziri Zanjani https://doi.org/10.5281/zenodo.2531955

Brewer 119_ 25 Oct 2010_ AM Z. Vaziri Zanjani https://doi.org/10.5281/zenodo.2531951

Download
Short summary
It is now known that single-monochromator Brewer spectrophotometer ozone and sulfur dioxide measurements suffer from non-linearity at large ozone slant column amounts due to the presence of instrumental stray light caused by scattering within the optics of the instrument. This paper presents a simple and practical method of correcting for the effects of stray light.