Articles | Volume 16, issue 6
https://doi.org/10.5194/amt-16-1583-2023
https://doi.org/10.5194/amt-16-1583-2023
Research article
 | 
27 Mar 2023
Research article |  | 27 Mar 2023

Development of an automated pump-efficiency measuring system for ozonesondes utilizing an airbag-type flowmeter

Tatsumi Nakano and Takashi Morofuji

Download

Interactive discussion

Status: closed

Comment types: AC – author | RC – referee | CC – community | EC – editor | CEC – chief editor | : Report abuse
  • RC1: 'Comment on egusphere-2022-565', Anonymous Referee #1, 12 Sep 2022
    • AC1: 'Reply on RC1', Takashi Morofuji, 14 Sep 2022
  • RC2: 'Comment on egusphere-2022-565', Anonymous Referee #2, 16 Sep 2022
    • AC2: 'Reply on RC2', Takashi Morofuji, 27 Oct 2022
  • EC1: 'Comment on egusphere-2022-565', Roeland Van Malderen, 22 Sep 2022
    • AC3: 'Reply on EC1', Takashi Morofuji, 27 Oct 2022
  • RC3: 'Comment on egusphere-2022-565', Anonymous Referee #3, 29 Sep 2022
    • AC4: 'Reply on RC3', Takashi Morofuji, 29 Nov 2022
    • AC5: 'Reply on RC3', Takashi Morofuji, 29 Nov 2022

Peer review completion

AR: Author's response | RR: Referee report | ED: Editor decision | EF: Editorial file upload
AR by Takashi Morofuji on behalf of the Authors (28 Dec 2022)  Author's response   Author's tracked changes   Manuscript 
ED: Publish subject to minor revisions (review by editor) (12 Jan 2023) by Roeland Van Malderen
AR by Takashi Morofuji on behalf of the Authors (22 Feb 2023)  Author's response   Author's tracked changes   Manuscript 
ED: Publish as is (23 Feb 2023) by Roeland Van Malderen
AR by Takashi Morofuji on behalf of the Authors (02 Mar 2023)  Manuscript 
Download
Short summary
We have developed a system that can automatically measure the pump efficiency of the ECC-type ozonesonde. Operational measurement for 13 years by this system revealed that the efficiency fluctuates in each and slightly increases over time. Those can affect the estimation of total ozone amount by up to 4 %. This result indicates that it is necessary to understand the tendency of the pump correction factor of each ozonesonde in order to detect the actual atmospheric change with high accuracy.