Articles | Volume 17, issue 20
https://doi.org/10.5194/amt-17-6193-2024
https://doi.org/10.5194/amt-17-6193-2024
Research article
 | 
23 Oct 2024
Research article |  | 23 Oct 2024

Water vapor stable isotope memory effects of common tubing materials

Alexandra L. Meyer and Lisa R. Welp
Abstract

Water molecules in vapor can exchange with gaseous water molecules sticking to surfaces of sampling tubing, and exchange rates are unique for each water isotopologue and tubing material. Therefore, water molecules on tubing walls take some time to reach isotopic equilibrium with a new vapor isotopic signal. This creates a memory effect that is observed as attenuation time for signal propagation in continuous stable water vapor isotope measurement systems. Tubing memory effects in δD and δ18O measurements can limit the ability to observe fast changes, and because δD and δ18O memory are not identical, this introduces transient deuterium excess (D-excess, defined as δD-8×δ18O) artifacts in time-varying observations. To our knowledge, a comprehensive performance comparison of commonly used tubing material water exchange properties in laser-based measurement systems has not been published.

We compared how a large isotopic step change propagated through five commonly used tubing materials for water isotopic studies – perfluoroalkoxy (PFA), fluorinated ethylene propylene (FEP), polytetrafluoroethylene (PTFE), high-density polyethylene (HDPE), and copper – at two different temperatures and an airflow rate of 0.635 L min−1 through approximately 100 ft (30.5 m) of 1/4 in. (6.4 mm) outer diameter (o.d.) tubing. All commonly used tubing materials performed similarly to each other in terms of attenuation times, reaching δ18O-location-adjusted δD and δ18O 95 % completion in less than 45 s, with slight variations based on temperature. PFA does appear to perform slightly better than the other materials, although memory metric differences are small. A tubing material commonly used in the early 2000s but reported to have memory effects on δD, Dekabon, was also tested at ambient temperature and changing humidities. The Dekabon isotopic equilibrium was not reached until nearly an hour after source transition, much later than H2O mixing ratios equilibrated. Bev-A-Line XX (used in some soil O2 and CO2 gas studies) was also tested at ambient temperature, but it did not approach isotopic equilibrium until after nearly 6 h of testing. Therefore, we cannot recommend the use of Bev-A-Line XX or Dekabon in water vapor isotope applications. Source transition from heavy to light or from light to heavy affected isotopic transition speed only in experiments where H2O ppmv was changing. While a shorter tubing lengths and smaller inner diameters shorten the delay of signal propagation through the tubing, they did not greatly change the attenuation curves under these conditions for the current commonly used tubing materials tested. However, in Dekabon, attenuation curves were greatly extended with increased tubing length. Our results show that the commonly used plastic tubing materials tested were not inferior to copper in terms of isotopic memory under these conditions, and they are easier to work with and are less expensive than copper.

1 Introduction

In situ laser absorption spectroscopy of water vapor isotopologues has risen in use over the last 2 decades, enabling continuous measurements (Griffith et al., 2006; Kerstel et al., 2006; Lee et al., 2005; Webster and Heymsfield, 2003). All experimental setups inherently attenuate signal variability due to mixing in the analyzer optical cavities and molecular water interactions with surfaces inside the inlet and analyzer system, especially when different H2Ov concentrations lead to wetting and drying of the tubing walls. A wide range of tubing materials, airflow rates, temperatures, and pressures have been used in experimental setups, which may result in different timescales for signal attenuation (Aemisegger et al., 2012; Galewsky et al., 2016; Griffis et al., 2010; Schmidt et al., 2010; Sturm and Knohl, 2010; Tremoy et al., 2011). As condensation in tubing is a concern due to phase change isotopic fractionation, many installations heat the tubing above ambient temperature, use a critical orifice at the tubing inlet to drop pressure in the lines, or do both to keep the tubing air temperature above the dew point temperature (e.g., Griffis et al., 2010; Luo et al., 2019).

Initially Synflex 1300 (also known as Dekabon or Dekoron), commonly used in the carbon dioxide and water eddy-covariance flux community, was used in water vapor isotope observations (Gupta et al., 2009; Lee et al., 2005; Tremoy et al., 2011). Dekabon is an aluminum tape with an ethylene copolymer adhesive film coated on both sides, rolled into a tube, and bonded with a high-density polyethylene jacket (Goodrich Sales, Inc, 2005; New Line Hose and Fittings, personal communication, 29 April 2024). It was eventually found to greatly attenuate the water isotopic signals (Griffis et al., 2010; Schmidt et al., 2010; Sturm and Knohl, 2010; Tremoy et al., 2011) and is no longer commonly used in water vapor isotope studies (Griffis et al., 2010; Schmidt et al., 2010; Steen-Larsen et al., 2014; Sturm and Knohl, 2010; Tremoy et al., 2011). Commonly used tubing material types now include copper (Steen-Larsen et al., 2014) and several types of plastic, such as polytetrafluoroethylene (PTFE, commonly referred to as Teflon) (Griffis et al., 2010; Sturm and Knohl, 2010), perfluoroalkoxy (PFA) (Schmidt et al., 2010; Tremoy et al., 2011), fluorinated ethylene propylene (FEP) (Luo et al., 2019), and high-density polyethylene (HDPE) (Griffis et al., 2010). Some performance testing was conducted, but the details of the experiments and results are sparse (Griffis et al., 2010; Schmidt et al., 2010; Steen-Larsen et al., 2014; Sturm and Knohl, 2010; Tremoy et al., 2011). Fluorinated polymers (FEP, PFA, and PTFE) are commonly used as transfer lines in chemical, pharmaceutical, food processing, and oil and gas industries because of their chemical resistance, weather resistance, and nonstick and dielectric properties (Chemours, n.d.). These materials have found favor in water vapor isotope applications for the same reasons.

Air tubing choices are important because materials may have different affinities, or degrees of attraction, for the isotopologues of water. This affinity causes a delay in the speed at which the isotopologue signals move through the tubing due to exchange rates with water molecules stuck to the walls, called the memory effect. The memory effect is stronger for δD compared with δ18O, presumably due to the stronger hydrogen bonding of the molecules containing deuterium slowing tubing wall exchanges (Griffis et al., 2010; Schmidt et al., 2010; Sturm and Knohl, 2010). This can result in false deuterium-excess (D-excess, defined as δD-8×δ18O) anomalies and is important to minimize when D-excess signals are interpreted in quickly changing atmospheric signals (Galewsky et al., 2016; Managave et al., 2016; Salmon et al., 2019; Sodemann et al., 2017). Some studies have suggested that memory effects may be lessened at higher temperatures and faster airflow rates (Griffis et al., 2010; Pagonis et al., 2017).

It is important to minimize isotopic wall effects in the intake tubing lines and other in-line elements positioned before the analyzer to minimize signal attenuation. Five studies previously reporting memory effects of tubing types tested a maximum of three materials at a time and are summarized in Table 1 (Griffis et al., 2010; Schmidt et al., 2010; Steen-Larsen et al., 2014; Sturm and Knohl, 2010; Tremoy et al., 2011). Most concluded that Dekabon was not suitable for water isotope applications, but they varied with respect to which tubing was preferred across applications. The National Ecological Observatory Network (NEON) selected FEP, which has not been widely used in reported studies, for their monitoring installations (Luo et al., 2019). In this study, we tested five of the commonly used tubing materials under nearly identical conditions at two different temperatures to determine which tubing type and temperature combination resulted in the smallest isotopic signal attenuation. For contrast, we also tested a tubing material known to have memory issues, Dekabon, and Bev-A-Line XX, a tubing not previously used in published water isotope studies but which is being increasingly used in soil O2 and CO2 gas studies (i.e., Brecheisen et al., 2019). Note that Bev-A-Line XX has a patented Hytrel® inner lining and is distinct from Bev-A-Line IV, which has been used in a few published water vapor isotope studies (Havranek et al., 2023; Lee et al., 2005; Simonin et al., 2013). Because Dekabon and Bev-A-Line XX have extremely slow isotope response times, they were only tested at ambient temperature and with changing water concentrations to demonstrate that the source switching in the experimental setup was working properly.

Table 1Literature findings.

* Experimental details and results of source-switching experiments are included in the peer-reviewed, published materials.

Download Print Version | Download XLSX

2 Methods

In this study, we tested PFA, FEP, PTFE, HDPE, and copper at ambient and elevated temperatures using self-regulating heat tape. We switched between two isotopically distinct vapor sources to examine the memory effects of each material. We also tested Bev-A-Line XX and Dekabon at ambient temperature and at two humidities.

2.1 Analyzer

A Los Gatos Research, Inc. (LGR) triple water vapor isotope analyzer (TWVIA) off-axis integrated-cavity-output spectroscopy system (OA-ICOS) was used for testing. An external pump (KNF pump, model N920-2.08) was added to the TWVIA to maximize the turnover rate of air inside the analyzer. The TWVIA itself regulates the outflow to maintain a constant internal pressure, resulting in discontinuous (jumpy) flow rates which averaged 0.635 ± 0.006 L min−1 at standard temperature and pressure with a cell pressure of  40 Torr. This resulted in an  4 s mean residence time of sample air in the analyzer. It is typical to average data over an optimum time interval determined by Allan variance testing to minimize analyzer noise and maximize measurement precision. In this experiment, the objective was to maximize the analyzer response time in order to resolve potential differences in isotopic signal attenuation during travel through inlet tubing. Applying a running mean to the 1 Hz data would have smoothed the response, masking the signal of interest in this study. Measurement uncertainty was estimated using 2 s Allan deviation, which is the lowest time limit of the Allan deviation code output (Guerrier et al., 2020). The Allan deviation at 2 s for δD and δ18O measured over 18 h at approximately 9300 ppm produced by a LGR water vapor isotope standard source (WVISS) was approximately 1.3 ‰ and 0.6 ‰, respectively, propagating to a D-excess precision better than ±3.3 ‰. To demonstrate that this analyzer performance is consistent with other published studies, the full Allan deviation plot of analyzer variance is presented in Fig. S1 in the Supplement.

2.2 Experimental setup

2.2.1 H2O-matched experiments

The memory effect of the tubing material was tested by switching between two sources of moist air with different isotopic values but nearly identical water vapor mixing ratios ( 9200 ppm; Table S1 in the Supplement). A LI-COR model LI-610 portable dew point generator (DPG) was used to create a vapor of approximately 187 ‰ δD, 25.5 ‰ δ18O, and 17.5 ‰ D-excess, measured by the LGR TWVIA without calibration, from water at 5 °C. The second vapor of approximately 32 ‰ δD, 5.8 ‰ δ18O, and 14.0 ‰ D-excess was produced by the WVISS, also measured by the analyzer without calibration. DPG-generated vapor isotopic values for the experiments became isotopically enriched over time as water evaporated from the liquid reservoir. Isotopic δD and δ18O transitions were normalized to a scale from one to zero to compare across experiments and adjust for small source water and analyzer drift over time. For this reason, further calibration of the isotopic measurements was not needed. Five replicate switches were completed for each experiment; the vapor sources switched approximately every 60 min, giving sufficient time to reach a new isotopic equilibrium. We focus on data through the first 20 min, as equilibrium was already established (with the exception of Dekabon and Bev-A-Line XX).

For each experiment, the WVISS programming and internal valve system controlled the switching between the DPG output connected to the WVISS inlet port and the WVISS (Fig. 1) output to the TWVIA. The WVISS was connected to the analyzer by approximately 100 ft ( 30.5 m, lengths listed in Table S1) long sections of 1/4 in. (6.4 mm) outer diameter (o.d.) test tubing for the main experiments. The Swagelok connection to the analyzer included an extra stainless-steel union and 2.5 in. (6.4 cm) thick-walled FEP to protect the analyzer bulkhead union threads from wear during the experiment, but this addition is not expected to affect the results significantly. Sensitivity to tubing length and inner volume were investigated using a short (62 in. or 1.6 m) and a long (99.0 ft or 30.2 m) piece of thick-walled FEP and long piece of thin-walled FEP. Tubing inner diameters (i.d. values, summarized in Table S1) were 3/16 in. (4.8 mm) with the exception of HDPE and thick-walled FEP, which were 1/8 in. (3.2 mm) i.d. Damaged thin-walled FEP tubing was repaired using three stainless-steel Swagelok unions and the Dekabon with one, but this is not expected to affect the results significantly.

https://amt.copernicus.org/articles/17/6193/2024/amt-17-6193-2024-f01

Figure 1Instrument setup for memory effect tests. The WVISS controls switching between WVISS air and dew point generator air (depicted here as an external three-way valve, but it is internal to the WVISS), which is passed through test tubing of up to 100 ft (30.5 m) and either heated or unheated to the analyzer. The flow through the test tubing is controlled by the analyzer and the external pump flow rate.

Download

Tubing and self-regulating heat tape (EasyHeat ADKS-0500, 100 ft or 30.5 m roof and gutter deicing kit) were wrapped in either flexible foam tape (HDPE, PTFE, and thick-walled FEP; AP/Armaflex TAP18230 insulation tape) or rigid foam pipe insulation (copper, thin-walled FEP, and PFA; Tundra brand 1/2 in. or 1.3 cm wall). The thermocouple probe was placed inside the insulation on the side of the tubing opposite the heat tape, about 3 in. (7.6 cm) from the end closest to the analyzer inlet. A data logger recorded the average temperature over the  10 h experiments. During heated-tubing tests, the tubing was allowed to warm up for at least 1 h prior to measurements to let the tubing moisture equilibrate to the elevated temperature and minimize the effects of degassing water molecules adhered to the tubing from previous experiments. Differences in the insulation properties of the two materials used and likely differences in thermocouple placement relative to unavoidable internal gradients in temperature resulted in differences in average temperatures for each experiment, ranging from 48.6 to 75.2 °C (Table S1). We note that this heating design is commonly used under field conditions and represents likely inlet conditions. However, the lack of uniform temperature control leads to potential temperature-induced differences that are hard to quantify. All heated experiments (average 60 ± 8 °C) are significantly warmer than ambient temperature experiments (average 24 ± 1 °C). Dekabon was only tested under ambient conditions and, thus, was not insulated.

Temperature-adjusted tubing residence times were 1.0 ± 0.1 s for short, thick-walled FEP; 19.7 ± 1.6 s for long, thick-walled tubing (FEP and HDPE); and 44.5 ± 3.0 s for long, thin-walled tubing (FEP, PFA, PTFE, and copper). Uncertainties in the tubing residence time (a few seconds) based on length (a couple of inches) and temperature (due to internal gradients and overall temperature fluctuations) were not considered here. Airflow rates through the tested tubing were controlled by the TWVIA itself, making the tubing flow rate as slow as possible and the analyzer flow rate as fast as possible with this set of equipment. The DPG was operated in a continuous fashion, constantly generating humid air. To maintain these constant conditions, a vent was added before the DPG outlet to the WVISS inlet to provide an overflow when the WVISS was pushing its humid airstream to the TWVIA, otherwise the DPG pump would be pushing against a closed valve. A Dwyer rotameter (model VFB-65-SSV) was used to monitor outflow from the vent. The vent airflow rate is not critical to the tubing tests, as it is simply the overflow. An Omega mass flow meter (MFM, 0–30 L min−1 range, model FMA1826A) was used to monitor airflow rates downstream of the TWVIA to verify that analyzer conditions remained unchanged during the experiments. A Mesa Labs Bios Definer 220 primary flow calibrator (Mesa Labs, Lakewood, CO; 50–5000 sccm – standard cubic centimeter per minute; accuracy ± 1 % of reading) was used to validate the airflow rate through the TWVIA and test tubing at the inlet of the TWVIA prior to the experiments but was not included during the actual experiments. When the primary flow calibrator was removed, no change in the TWVIA outlet flow was detected on the Omega MFM. Rotameter flow rates were verified at the beginning of the experiments using the primary flow calibrator. The DPG vent flow rate was  0.9 L min−1 when the DPG was sampled by the TWVIA and  1.5 L min−1 when the WVISS was sampled, consistent with the 0.6 L min−1 flow rate of the analyzer.

2.2.2 H2O-varied experiments

For this set of experiments, the plumbing and flows remained the same. The only difference in this case was that the two different isotopic sources also had different water vapor mixing ratios (Table S1). The DPG was used to create a vapor of approximately 184 ‰ δD, 26.2 ‰ δ18O, 25.4 ‰ D-excess, and  9300 ppmv H2O, measured by the LGR TWVIA without calibration, from water at 5 °C. The second vapor of approximately 20.3 ‰ δD, 8.8 ‰ δ18O, 50.4 ‰ D-excess, and  16 950 ppmv H2O was produced by the WVISS, also measured by the analyzer without calibration. Because data were normalized as above, calibration was not necessary to determine attenuation times. Two to four replicate switches were completed for Dekabon and HDPE tubing depending on the time to reach the new isotopic equilibrium. One replicate of Bev-A-Line XX was run in each direction of the isotopic switch, and the results are presented in Fig. S2. Replicate 5 min switches comparing the performance of Bev-A-Line XX and HDPE can also be found in Fig. S3.

For each of the H2O-varied experiments, source switching was controlled manually, as the TWVIA control of the WVISS unit malfunctioned. The WVISS was connected to the analyzer by approximately 100 ft (30.5 m, lengths listed in Table S1) long sections of 1/4 in. (6.4 mm) outer diameter (o.d.) HDPE, Bev-A-Line XX, or Dekabon tubing. Other tests were done with a short (79 in., or 2 m) section of HDPE or Dekabon to quantify the sensitivity to tubing length and inner volume using high-memory materials. Tubing inner diameters (i.d. values, summarized in Table S1) were 0.2 in. (4.3 mm) with the exception of HDPE, which was 1/8 in. (3.2 mm) i.d. These experiments were conducted under ambient conditions (average 24 ± 1 °C). Temperature-adjusted tubing residence times were 2.8 s for short Dekabon, 42.2 s for long Bev-A-Line XX and Dekabon, and 1.5 and 22.8 s for short and long HDPE, respectively. All other experimental aspects remain the same as detailed in Sect. 2.2.1.

2.3 Data processing

Isotopic values were measured at 1 Hz. No calibration to assign values to the international scale was performed on the isotopic measurements because the transitions were normalized to their starting and ending equilibrium values, resulting in signal transitions from one to zero. Isotopic measurements made with this analyzer are known to vary with water mixing ratio and potentially drift over long periods of time. Normalizing the measurements between sources, as described below, removed any potential influence of instrument or source drift over periods of more than 20 min.

For δD and δ18O, the individual transitions were normalized from one to zero, and replicates were then averaged to characterize the transition memory and uncertainty. Initial δ values (normalized to one) were either the maximum δ value after the source switch indicator in the data file (short, thick FEP and long Bev-A-Line XX) or the average of 5 s on either side of that maximum δ value (all other experiments). Final δ values (normalized to zero) were the average of measurements from 600 to 1200 s after the source switch in H2O-matched experiments. Diverse experimental lengths were used during the H2O-varied experiments, with lengths ranging from 336 to 31 001 s depending on the time to equilibrium (Table S1). The variation in experimental lengths resulted in final δ values (“0”) set as the average of at least the last 50 s depending on the time to equilibrium and length of the experiment (see Table S1 for the exact intervals used to average). D-excess was calculated as δD-8×δ18O. D-excess was not normalized in the same way as δD and δ18O, as the shape of the attenuation curve is different. First, a 10 s running mean was applied. We then calculated the average D-excess value for each replicate over 600–1200 s after the source switch in H2O-matched experiments, whereas we used the time spans that the final δ values were averaged over (the last 50 or 100 s) in the H2O-varied experiments. This average was then subtracted from all data points within a replicate to adjust for small changes in D-excess source waters between replicates, especially in the DPG vapor which undergoes evaporative enrichment and D-excess decrease. These time spans after the source switch (600–1200 s for H2O-matched experiments and the last 50 or 100 s for H2O-varied experiments) visually appear to be conditions of tubing equilibration and were used to calculate the source vapor sample averages given in Table S1 and summarized in Sect. 2.2. Replicates were screened based on successful WVISS-to-DPG and DPG-to-WVISS switching and consistent water vapor mixing ratios, ensuring that vapor source generators were operating properly. Four replicates were discarded from the collected data due to water mixing ratio variability from the WVISS. These discards include one replicate each from heated PFA and 100 ft (30.5 m) Dekabon H2O-matched experiment as well as two from the 100 ft HDPE H2O-varied experiment. Finally, replicates were averaged to reduce noise.

When comparing experiments between different tubing lengths and i.d. values, differences in the internal volume result in different tubing residence times due to advection. The flow in all experiments was estimated to be laminar, with Reynolds numbers calculated between 579 and 870. In Sect. 3.1, we will describe how the experiments are adjusted with respect to the advection delay to compare transitions directly.

Memory analysis included both directions of the isotopic switch. Isotopically enriched-to-depleted (WVISS-to-DPG) figures are presented in the main body of the text, whereas isotopically depleted-to-enriched (DPG-to-WVISS) transitions are available in the Supplement (Figs. S4, S5, S6, and S7).

2.4 Memory quantification

Memory effects can, in some respects, be analogous to a low-pass filter, smoothing high-frequency variability (e.g., Zannoni et al., 2022). Previous studies have approximated the smoothing of a fast-step-change input as an exponential transition and report a threshold time to some percentage of completion, like an e-folding (63 %), 90 %, or 95 % (Aemisegger et al., 2012; Schmidt et al., 2010; Steen-Larsen et al., 2014; Sturm and Knohl, 2010). In some cases, the threshold metrics were obtained from the data directly (Steen-Larsen et al., 2014; Sturm and Knohl, 2010); in others cases, it appears that an exponential function was fit to the data first and the metrics were extracted from the fit (Aemisegger et al., 2012; Schmidt et al., 2010). A second method used in the literature takes the first derivative of the normalized transition (Steen-Larsen et al., 2014) and characterizes an impulse response function using curve fitting (Jones et al., 2017; Kahle et al., 2018). We have quantified memory effect metrics using both methods.

2.4.1 Threshold metrics

We extracted attenuation threshold metrics directly from the normalized and replicate-averaged data (not an exponential fit). An e-folding time corresponds to τ=1/e of the signal transition remaining to reach a new value. In this study, we have chosen to estimate attenuation threshold times at 1τ ( 63 %) and 3τ ( 95 %) completion of the switch to the next δD and δ18O value, denoted as t63 % or t95 %, respectively (Schmidt et al., 2010). These t values are the time that the averaged curve intersects the threshold percent value. We chose not to fit exponential curves to extract an e-folding time, because the measured attenuation curves were not accurately described by an exponential curve (not shown). The 1-standard-deviation envelope was calculated by taking the standard deviation of the two to five replicates at each time step. Errors associated with attenuation threshold times were determined by finding the time that the 1-standard-deviation envelope of the averaged replicates intersects the completion threshold. Because the analyzer measures in discrete 1 s intervals, the raw t63 % and t95 % values output the next second from where the averaged curve intersects the threshold percent value. This leads to slight differences in δ18O-location-adjusted t63 % and t95 % values (discussed in Sect. 2.4.2) compared with the sweepout curves presented in Sect. 3.

D-excess signals of the source transitions are not unidirectional, and memory must be quantified differently. Previous studies have reported that δD signals take longer to equilibrate with the surface of tubing materials compared with δ18O signals due to the isotopic effects of hydrogen binding with the tubing walls (Aemisegger et al., 2012; Griffis et al., 2010; Schmidt et al., 2010; Sturm and Knohl, 2010). The D-substituted hydrogen bonds exchange with the vapor more slowly. This difference between isotope signal speed leads to a D-excess transition that has a transient anomaly until the δD signal propagation catches up to the δ18O signal. The direction of the D-excess transient peak depends on the direction of the isotopic signal switch. In the enriched-to-depleted transition, the enriched δD signal is retained on the tubing walls, creating a transient, positive anomaly in D-excess while approaching equilibrium. However, in a depleted-to-enriched transition, the depleted δD signal is preserved on the tubing walls, creating a negative D-excess anomaly during isotopic equilibration. The absolute value of the maximum transient peak was identified, and the associated errors are given as the standard deviation of the replicate D-excess values at the time of the maximum peak (Table S2). The threshold chosen to measure completion in D-excess transitions is a 3 ‰ threshold within the new equilibrium value (t3 ‰). The 3 ‰ threshold is a conservative estimate of analyzer precision of D-excess measurements if δD precision was 1.0 ‰ and δ18O precision was 0.25 ‰.

To compare the attenuation threshold times across experiments, we adjusted for differences in signal propagation due to the time it takes air to move through the tubing from the WVISS and mix inside the analyzer, controlled by the airflow rate through the instrument, optical cavity size, test tubing volume, and airflow rate (Schmidt et al., 2010), as well as temperature. Smaller tube i.d. values, increased temperature, and shorter tube lengths tested here will all shorten lag times associated with a measurement. The δ18O lag times were calculated via breakpoint analysis to determine the point at which the slope changes. We created a linear model using the first 300 s of data in most cases after the source switched and then utilized the “segmented” function in R software's “segmented” package on the time series to find the breakpoint (Muggeo, 2022). In some cases, different observation intervals were used for short FEP and H2O-varied tests (see Table S1 for exact intervals). The breakpoint lag estimates likely have an error of a few seconds. The exact uncertainty was not quantified. Average measured lag times for 100 ft (30.5 m) thin-walled tubing were 53 s, whereas this value was 1.5 s for the short, thick-walled tubing in the H2O-matched experiments. In the results, the time axis in the plots and quantitative threshold metrics (t63 %, t95 %, and t3 ‰) in the tables were adjusted by the δ18O location time determined by curve fitting (discussed in Sect. 2.4.2) to more easily compare the tubing dimension influence on transition smoothing.

2.4.2 Impulse response method

In the impulse response method, we take advantage of the first derivative of the observed attenuation curves to clearly identify the timing and rates of change. To decrease the noise in the first derivative, it is necessary to reduce noise in the observed attenuation curves. In previous studies, noise reduction has been achieved by fitting a smooth transfer function to the observations. Jones et al. (2017) and Kahle et al. (2018) used a lognormal × lognormal (log–log) function to fit the data, whereas only one lognormal function was used in Steen-Larsen et al. (2014). For our attenuation curves, neither a single- nor double-lognormal fit the observed data well. Our data were most accurately recreated by a transfer function in the form of Eq. (1) (with the exception of the depleted-to-enriched transition for H2O-matched HDPE, H2O-varied depleted-to-enriched 79 in. (2 m) HDPE, and enriched-to-depleted Dekabon in both sets of experiments where an additional normal fit was added):

(1) δ transfer ( t ) = c 1 × 1 + erf log ( t ) - μ 1 σ 1 2 × 1 + erf log ( t ) - μ 2 σ 2 2 × 1 + erf t - μ 3 σ 3 2 + c 2 ,

where t is time since switching, μx is the location parameter of each normal or lognormal distribution, σx is the scale parameter or standard deviation of each distribution, c1 is a scaling factor, and c2 is an offset. The values of μx, σx, and cx are optimized by minimizing the squares of errors using the “DEoptim” global optimization function in the R package of the same name (Ardia et al., 2022). The form of the fitting model here is not that important as long as the observations are faithfully reproduced in the smooth curve fit, as seen in Fig. 2a.

https://amt.copernicus.org/articles/17/6193/2024/amt-17-6193-2024-f02

Figure 2Example of model function fits for the unheated, long, thick-walled FEP experiment. Panel (a) compares normalized and averaged analyzer output (black dots) with the transfer function given in Eq. (1) (red line). Panel (b) compares the impulse function derived from the first derivative of the transfer function fit evaluated every second (black dots) with the fit from Eq. (2) (red line) and the skew-normal impulse function (blue line) used in Jones et al. (2017) and Kahle et al. (2018).

Download

Once a transfer function is fit, the first derivative of the transfer function is calculated to obtain the impulse function. We fit the impulse function using the model in Eq. (2), based on a skew-normal function added to a normal Gaussian function:

(2a)δimpulse(t)=c1×12π×e-x122×121+erfx1×α2+12π×e-x222×c2,(2b)x1=(t-ξ)ω,(2c)x2=(t-μ)σm.

Here, in the skew-normal terms, ξ is the location of the maximum impulse peak, α is shape, and ω is scale; t is time since switching; σm is the standard deviation of the additional PDF and μ is its mean; and c1 and c2 are scaling factors. The parameters are solved for using a two-step method: (1) using the “DEoptim” function (Ardia et al., 2022) to provide an approximate initial guess and (2) utilizing the “nls” nonlinear least-squares function in the “stats” R package of base R (R Core Team, 2023) to provide parameter fine-tuning and uncertainty estimates of each parameter.

While Jones et al. (2017) was able to fit impulse functions of their data solely using a skew-normal PDF fit (a standard normal probability distribution function times a standard normal cumulative distribution function, or PDF × CDF), we most accurately reproduced the first derivative by adding an extra PDF in Eq. (2). Figure 2b shows a comparison of the Jones et al. (2017) impulse function skew-normal fit compared to the impulse function fit that we used in this study. Our impulse function model fits the memory tail in our experiments better than the skew-normal PDF model from Jones et al. (2017).

We extracted two memory metrics from the impulse fit. First, the skew-normal parameters of shape (α, a descriptor of the shape of the curve or other asymmetry of the distribution) and scale (ω, a measure of the spread of the distribution) were used to estimate a mixing time (σs) from Eq. (3). The σs metric has also been called mixing length in Jones et al. (2017) or diffusive length in Kahle et al. (2018), and the analysis time relates to the length along the ice cores in both studies. The σs is a metric of how much mixing occurs due to diffusive flow within the tubing. Error for σs is propagated from the errors associated with shape and scale. Second, we also estimate the standard deviation of the additional PDF (σm) in Eq. (2); this PDF is critical for fitting the memory tail in the observations, as it provides additional information about memory not captured by the skew-normal curve.

(3a)β=α1+α2(3b)σs2=ω2×1-2β2π(3c)σs=σs2
3 Results

3.1 Comparison of residence, lag, and location times in H2O-matched experiments

The residence time of air in the inlets is mathematically predicted using the tubing i.d., length, temperature, pressure within the tubing, and airflow rate through the tubing (Table S2). Residence times are decreased by decreasing the tubing length and inner diameter or increasing the temperature through the tubing and analyzer, as tested here. Average δ18O lag times from breakpoint analysis correlate well with predicted residence times (Fig. S8a). For the long, thick-walled tubing, the calculated residence time is approximately 19.7 ± 1.6 s, with slight variations due to temperature and small length differences, which agrees well with the observed δ18O lag of 23.1 ± 1.2 s. For long, thin-walled tubing, the calculated residence time is approximately 44.5 ± 3.0 s and average δ18O lag time is 53.0 ± 4.0 s (not including Dekabon, due to instrument malfunction). The largest discrepancies between residence and δ18O lag times (< 12.5 s, with the exception of Dekabon) are found in unheated copper and unheated PFA. Tubing roughness was not considered when calculating residence times, as flow was assumed to be laminar and flow rate was measured at the end of the tested tubing closest to the analyzer (and, therefore, should be representative of the actual flow rate in the tubing). For short, thick-walled FEP, the residence time is 1.0 ± 0.1 s and average δ18O lag time is 1.5 ± 1.7 s. Overall, heated-tubing lag and residence times were shorter than their unheated counterparts (Table S2).

Similarly, the location time parameter solved for using the impulse response method is the timing of the maximum peak of the impulse function (or the steepest portion of the attenuation curve; discussed in Sect. 2.4.2). The location time is sensitive to the advection lag and the steepness of the isotopic transition. Our estimated δ18O location time for the long, thick-walled tubing (25.6 ± 1.3 s, Table S2, excluding Dekabon) matches the δ18O lag time above when accounting for the < 5 s between the initial signal change and the maximum slope of the attenuation curve (or peak in the impulse function). Because of this relationship, δ18O location times correlated well with the observed δ18O lag times (Fig. S8b), t63 % estimates from the experiments (Fig. S8c, excluding Dekabon), and residence times (Fig. S8d). The differences in location time between different tubing experiments is not fully explained by differences in residence time predictions. The location time extracted from the δD impulse function is slightly longer than the location time extracted from the δ18O impulse function, but they correlate well (Fig. S8e). The Dekabon δD location time is comparatively much longer ( 30–50 s longer) than the δ18O location time (Table S2). Overall, location is closely related to other methods of timing isotopic transitions, including lag, residence time, and e-folding time, in the tubing materials that we tested, with the exception of Dekabon.

To visualize differences in the curve shape between the tubing materials tested using different internal volumes (due to length and i.d.) and air density (due to temperature), a common transition point was defined. This is similar to adjusting by lag time (e.g., Steen-Larsen et al., 2014) or predicted residence times. Given uncertainties in the breakpoint analysis of lag time and tubing temperature uncertainties which influence residence time, we decided that the δ18O location time was the most self-consistent way to collapse the experiments on top of each other in the figures. Adjusting by δ18O location time also allows comparison to the H2O-varied and Dekabon experiments, as the valve switching time was not precisely recorded by the software.

3.2 Direction of isotopic and water vapor concentration transitions

H2O-varied and H2O-matched experiments for 100 ft (30.5 m) HDPE and Dekabon tubing were used to determine if there was a difference in the enriched-to-depleted and depleted-to-enriched switches due to the isotopologues or net degassing of the tubing walls (Fig. 3). H2O-matched Dekabon experiments did not exhibit clear differences depending on the isotopic switch direction, whereas there were clear differences in δD and D-excess depending on the switch direction for H2O-varied Dekabon (Fig. 3a, c, and e). H2O-varied Dekabon clearly shows longer δ18O-location-adjusted t95 % times in the enriched-to-depleted switch direction compared with the depleted-to-enriched direction. H2O-matched HDPE also does not exhibit clear differences in the switch direction (Fig. 3b, d, and f). H2O-varied HDPE shows a clear difference in δD t95 % between switch direction outside of t95 % error but no clear difference in δ18O or D-excess. Overall, differences in HDPE threshold metric values are much smaller than the differences between Dekabon values. Impulse response metric patterns for both tubing types are mixed.

https://amt.copernicus.org/articles/17/6193/2024/amt-17-6193-2024-f03

Figure 3δ18O-location-adjusted t95 % for δ18O and δD transitions and t3 ‰ times for D-excess in H2O-matched and H2O-varied experiments of isotopic transition directions for long HDPE and Dekabon. Panels (a), (c), and (e) depict Dekabon tests, while panels (b), (d), and (f) depict the HDPE tests. The depleted-to-enriched direction is indicated in black, whereas the enriched-to-depleted direction is indicated in red. Error bars indicate the maximum and minimum threshold times from the standard deviation of the experiment replicates. We saw differences in switch direction only in H2O-varied experiments.

Download

There is a difference in the H2O-varied switch direction but little to no clear and consistent difference in switch direction in H2O-matched experiments. Most results presented in the following sections are H2O matched. We discuss both switch directions in the main body of the text, but present figures of the enriched-to-depleted switch transition only; we provide figures for the depleted-to-enriched transition in the Supplement.

3.3 Tubing material and temperature

3.3.1 Visual inspection of mean attenuation curves

The mean attenuation curves for the enriched-to-depleted transitions for all H2O-matched long, thin-walled tubing experiments and HDPE, which is thick-walled, are compared in Fig. 4, while the depleted-to-enriched results are given in Fig. S4. Attenuation curves for each experiment have been adjusted by the δ18O location time metric to remove the influence of different air lag times caused by different tubing i.d. values and temperature-sensitive air density to more easily compare equilibration times of different tubing dimensions. Therefore, 0 s in these figures indicates the time of most rapid change in the transfer function and the peak of the impulse function for each experiment. The δD signal also underwent δ18O location adjustment to highlight potential differences in equilibration speeds between the two isotopologues. Dekabon stands out as the tubing material with the longest isotopic memory in δD and δ18O and the largest difference between δ18O memory and δD memory (Figs. 4 and S4). There are slight variations with respect to performance within the rest of the tubing material type and temperature combinations. Specifically, thin-walled FEP δD results show the next slowest transitions compared with other tubing experiments and indicate that FEP has the second-largest difference between δ18O memory and δD memory. The δ18O-location-adjusted attenuation curves for δ18O have no consistent difference with respect to where they intercept the t63 % threshold between heated and unheated experiments (Figs. 4 and S4).

https://amt.copernicus.org/articles/17/6193/2024/amt-17-6193-2024-f04

Figure 4Mean attenuation curves for enriched-to-depleted (WVISS-to-DPG) H2O-matched transitions of each tubing type for δ18(a, b), δD (c, d), and D-excess (e, f), plotted as seconds since the δ18O impulse function peak (i.e., δ18O-location-adjusted time). Panels (a), (c), and (e) depict time from 50 s before the peak of the δ18O impulse function for each experiment to 100 s after, while panels (b), (d), and (f) zoom in on time from 5 to 15 s and exclude the Dekabon results. Solid lines indicate unheated experiments, while dashed lines indicate heated experiments. For δD, δ18O, and D-excess, only Dekabon and FEP show clear differences in material type, and only FEP shows clear differences in heated and unheated experiments. The full Dekabon attenuation curve can be found in Fig. S9. The orange curve in panel (b) shows the mean δD of all experiments for comparison with δ18O, and the orange curve in panel (d) shows the mean δ18O for comparison with δD; these means exclude Dekabon. Here, FEP is long and thin-walled as a comparison to the rest of the thin-walled tubing. HDPE is thick-walled and has a smaller i.d. than the other tubing shown here. Horizontal gray lines indicate thresholds of 95 % and 63 % transition completion for δD and δ18O or 3 ‰ for D-excess, while a black line at zero indicates full equilibrium completion. Depleted-to-enriched results are presented in the Supplement.

Download

δD attenuation times were longer compared with δ18O, creating a transient positive D-excess anomaly in the enriched-to-depleted transitions before equilibrating with the new vapor source isotopic values (Fig. 4). In the enriched-to-depleted transition, propagation of the depleted δD signal was delayed relative to the depleted δ18O signal (as shown by the orange lines denoting isotopic means of the non-Dekabon tubing in Fig. 4b and d). The δD signal transition was also delayed relative to δ18O in the depleted-to-enriched transitions (Fig. S4b and d), leading to a transient negative D-excess anomaly (Fig. S4e and f). In Fig. S4, D-excess plots were flipped for easier graphical comparison with the enriched-to-depleted transition. D-excess attenuation times are typically longer than the t95 % times for δD or δ18O (Table S2) while the slower isotopic propagation of the δD signal catches up to δ18O. Different D-excess values between experiments in Fig. 4e and f are caused by D-excess drift of the DPG over the experiments.

3.3.2 Quantitative memory metrics

After δ18O location adjustment, there are few consistent differences between heated and unheated tubing when comparing the same material (Figs. 5, S5, and Table S2). While in δ18O-location-adjusted t95 % most heated-tubing times are similar to or longer than their unheated counterparts, in σm and δ18O-location-adjusted t63 %, heated-tubing times are generally similar to or shorter than their unheated counterparts. In σs, most signal is within error and there are no consistent patterns between switching direction (Table S2). Overall, heated memory metrics are generally either similar to or smaller than those of the unheated memory metrics when comparing the same tubing types without δ18O location adjustment (Table S2), with the exception of HDPE (both directions) and the depleted-to-enriched PTFE t95 % time.

https://amt.copernicus.org/articles/17/6193/2024/amt-17-6193-2024-f05

Figure 5δ18O-location-adjusted t95 % (a, b) and t3 ‰ (c) times comparing heated (red) and unheated (black) experiments for all tubing types. The enriched-to-depleted switch direction is depicted here, while the depleted-to-enriched transition data are located in Fig. S5. We did not see clear differences in tubing temperature influence, and only very small differences were noted between tubing material types. Thick-walled FEP is presented here as a direct comparison to thick-walled HDPE, while the rest of the materials were thin-walled with larger i.d. values. Thin-walled, thick-walled, short, and long FEP experiments are shown in Fig. 6 and discussed in Sect. 3.4.

Download

Each memory metric provides a different ranking of tubing materials based on slight numerical differences in metric values, and all tubing types appear operationally similar with the exception of Dekabon (Table S2). Some common patterns in these rankings do emerge in the H2O-matched experiments. Of the tubing types that we calculated memory metrics for, Dekabon was the worst. The rest of the tubing t95 % and t3 ‰ times are given in Fig 5. We see clusters of tubing material types that have relatively shorter and longer times to equilibration, specifically in the δ18O-location-adjusted δD signal, illustrated in Fig. 5b. Thin-walled FEP, HDPE, and PTFE appear slightly slower to equilibrate than the rest of the tubing materials in the enriched-to-depleted direction, while PFA and copper equilibrate slightly faster. Please note that this figure also includes differences in length and inner diameter. While thick-walled FEP is presented here as a direct comparison to HDPE (which is also thick-walled with a smaller i.d.), the rest of the materials had similar i.d. values to thin-walled FEP. Comparison of different dimensions of FEP experiments are discussed in Sect. 3.4.

There are differences in relative rankings based on temperature, switch direction, and tubing material type, but these relative rankings vary depending on the memory metric used. Based on the t95 %, t63 %, and t3 ‰ times in the enriched-to-depleted direction, PFA and copper appear similar to each other and slightly better than the rest of the tubing material types. However, according to most impulse response metrics (σs and σm), HDPE (thick walled) has the shortest attenuation impulse response time. PFA and copper have the longest δ18O impulse response times in the enriched-to-depleted direction after Dekabon. For D-excess, the best tubing materials in the enriched-to-depleted direction were copper (t3 ‰) and PFA (based on the absolute value of the maximum D-excess peak), while long, thin-walled FEP was the worst for resolving D-excess signal (after Dekabon). We did not calculate impulse response metrics for Bev-A-Line XX due to its inferior performance and greatly extended curve shape, but it is clearly an inferior tubing with long memory times (Fig. S2).

3.4 Tubing inner volume and length

Properties like tubing length and inner diameter affect the mean the transit time through the tubing and the time it takes the signal change to reach the analyzer, but these properties do not appear to greatly influence the shape of the attenuation curve after δ18O location adjustment in the FEP H2O-matched experiments (Figs. 6 and S6). In these H2O-matched experiments, the short, thick-walled and the long, thick-walled tubing δ18O and δD transitions overlap each other (Fig. 6b and d), but the long, thin-walled tubing has a slightly shallower δ18O slope (Fig. 6b) and a bigger delay between the δD and δ18O signal transitions (Fig. 6f). While there is not much separation between curves visually, the quantitative memory metrics varied for δD with length and inner diameter, with longer memory times for longer and larger-volume tubing. Short, thick-walled FEP generally has smaller memory metrics than long, thick-walled FEP, which in turn normally has smaller memory metrics than its thin-walled counterpart. Longer memory metrics were also observed for δ18O for both switching directions of t95 % and in enriched-to-depleted σs, although other metric differences did not consistently show this pattern.

https://amt.copernicus.org/articles/17/6193/2024/amt-17-6193-2024-f06

Figure 6Mean attenuation curves for only FEP tubing for enriched-to-depleted (WVISS-to-DPG) transitions comparing tubing length and inner diameter for δ18(a, b), δD (c, d), and D-excess (e, f), plotted as seconds since the δ18O impulse function peak (i.e., δ18O-location-adjusted time). Panels (a), (c), and (e) depict time from 5 to 100 s, while panels (b), (d), and (f) zoom in on time from 5 to 15 s. Solid lines indicate unheated experiments, while dashed lines indicate heated experiments. Horizontal gray lines indicate thresholds of 95 % and 63 % transition completion for δD and δ18O or 3 ‰ for D-excess, while a black line at zero indicates full equilibration. The δ18O location adjustment for the short tubing is much shorter than that of the long tubing, leading to a line that appears to start abruptly at approximately 3 s.

Download

In the H2O-varied experiments, tubing length influence on the shape of the attenuation curve after δ18O location adjustment depends on the tubing material (Figs. 7 and S7). While the isotopic signals for short and long tubing are similar to each other in both directions in HDPE (panels b and d in Figs. 7 and S7), the isotopic signal transitions are much slower in long Dekabon than in short Dekabon in both switch directions. In long Dekabon, the much shallower δD slope (panels c and d in Figs. 7 and S7) and a bigger delay between the δD and δ18O signal transitions leads to a D-excess anomaly of approximately 120 ‰ (panels e and f in Figs. 7 and S7), the largest D-excess anomaly of all tubing types tested. This D-excess anomaly is much smaller in short Dekabon ( 40 ‰) and demonstrates the more similar signal transitions between δD and δ18O. Long Dekabon also has a much shallower H2O transition slope than the rest of the tubing types tested, including short Dekabon, which reacts more similarly to long HDPE when the water vapor concentration is changed (panels g and h in Figs. 7 and S7). Overall, isotopic transitions lag H2O transitions, as seen when comparing Fig. 7b, d, and h. Short Dekabon consistently has similar or shorter memory metrics than long Dekabon. Short HDPE generally has similar or shorter memory metrics than long HDPE, with the exception of δ18O enriched-to-depleted σm and depleted-to-enriched δD σs and σm. Again, we have effectively normalized for tubing length, volume, and temperature through the δ18O location adjustment; thus, differences in the attenuation curve steepness could be attributed to vapor–wall interactions that are independent of bulk flow.

https://amt.copernicus.org/articles/17/6193/2024/amt-17-6193-2024-f07

Figure 7Mean attenuation curves comparing length for 79 in. (2 m) and 100 ft (30.5 m) HDPE and Dekabon tubing for enriched-to-depleted (WVISS-to-DPG) transitions for δ18(a, b), δD (c, d), D-excess (e, f), and H2(g, h), plotted as seconds since the δ18O impulse function peak (i.e., δ18O-location-adjusted time). Panels (a), (c), (e), and (g) depict time from 50 to 100 s, while panels (b), (d), (f), and (h) zoom in on 5 to 15 s. Solid lines indicate 100 ft (30.5 m) lengths, while dashed lines indicate 79 in. (2 m) lengths. Horizontal gray lines indicate thresholds of 95 % and 63 % transition completion for δD and δ18O or 3 ‰ for D-excess, while the black line at zero indicates full equilibration.

Download

4 Discussion

Previous water vapor isotope studies have tried to identify suitable tubing material to use in sample inlets, and several different materials have been employed. To our knowledge, the results of rigorous testing for wall adsorption/desorption effects leading to memory artifacts have not been published. Theory based on principles of gas chromatography and gas–wall partitioning predicts that the residence time of gases adsorbed on tubing walls is linearly proportional to tubing inner diameter and length and that residence time should decrease at higher temperatures as gas saturation concentration changes (Pagonis et al., 2017). The experiments performed in this study begin to test these predictions for water vapor isotopes.

4.1 Review of material properties

We hypothesized that predictions of tubing material performance could be made based on tubing material properties. Out of material properties commonly reported by manufacturers, we selected two properties that we thought may play a role in fractionating wall effects: water absorption percentage by tubing weight and relative permittivity. Hydrophobic materials that are nonpolar and have a high relative permittivity (also known as the dielectric constant, or a material's ability to prevent electrical fields from forming) should be ideal for water vapor isotope studies, as polar water molecules are affected by and can induce electric fields (Aemisegger et al., 2012). As previously shown, δD signal transitions are slowed compared with δ18O signals, due to isotope-dependent hydrogen-bonding interactions with tubing walls. Limiting these interactions should lead to reduced isotopic attenuation times. Material specifications vary by manufacturer and material purity; however, in general, FEP and PTFE materials are expected to have the least amount of water absorption of the tubing types that we tested (Table 2). The inner liners of Dekabon and Bev-A-Line XX are proprietary information and/or non-disclosable; thus, the information required to complete Table 2 was not available upon request for these materials. However, plastic polymers do have the capability to absorb water at hydrophilic sites and potentially in free volume within the polymer up to  12 % by material weight for ethylene–vinyl alcohol copolymers (Cava et al., 2006), as discussed in Sect. 4.5. Metals have a relative permittivity value of approximately 1 due to their sea of electrons, which, in this case, interact with the polar water molecules. Larger values of relative permittivity are better in this case, as water vapor molecules will be less attracted to the metal. HDPE, FEP, and PTFE have the highest ability to prevent electrical fields. FEP and PTFE may be expected to have the shortest isotopic attenuation times based on combined water absorption percentage and relative permittivity.

Table 2Material properties of tubing-type options and their water absorption percentages and relative permittivity values.

a After being submerged for 24 h, as per the ASTM D570 (Plastic Materials, 2024). This metric is solely for plastic materials (ASTM, 2022). NA represents not applicable.
b Electrical properties of plastic materials (2021).

Download Print Version | Download XLSX

At the airflow rate that we tested, the isotopic memory metrics of FEP and PTFE were not noticeably superior to the other tubing tested. If the material properties listed here correspond to a fractionating effect, their impact may be too small to measure, possibly due to the additional  4 s residence time of the analyzer optical cell and internal plumbing. Alternatively, the material properties listed may impart non-fractionating effects. Slower tubing airflow rates and faster analyzer flow rates may result in a more precise resolution of memory differences between tubing types, but further research would be needed to determine if these or other material properties affect water isotope memory.

4.2 Direction of isotopic and water vapor concentration transition

Quantitative memory metrics were used to determine if there was a difference in the enriched-to-depleted and depleted-to-enriched switches. We focused on the t95 % threshold metric, as the t63 % values and impulse metrics were too small to gain a complete understanding of any differences in either switch direction or between H2O-matched and H2O-varied experiments. Similar to Aemisegger et al. (2012) and their tests with PFA, we found that the enriched-to-depleted switch exhibited longer attenuation times during H2O-varied experiments (Fig. 3 and Table S2). However, we were not able to replicate this finding in the H2O-matched experiments. While Aemisegger et al. (2012) indicated that they found a difference in switch direction regardless of the span of isotopic transition, they do not mention testing the effects of varying the span of water vapor concentration. We posit that the difference in isotopic transition dependent on direction is actually a dependence on the wetting and drying of the tubing and analyzer walls. Related to the claim of Aemisegger et al. (2012) that isotopic adsorption (depleted-to-enriched switch) is faster than the desorption process in the heavy isotopologues, this may be a side effect of the water vapor transition, as it is energetically harder to pull a water vapor molecule off a tubing wall and replace it than it is to simply add more molecules. H2O concentration variation between sources is likely the driving factor of memory metric differences based on transition direction.

4.3 Effects of material and temperature

We found that δD and δ18O attenuation curves between the commonly used tubing materials were slightly different, although operationally similar, at the flow rate, humidity, and temperatures tested (Figs. 4, 5, S4, and S5). Dekabon and Bev-A-Line XX attenuation curves were much longer. Our results are consistent with the assertation of Griffis et al. (2010) that HDPE is similar to PTFE. We were not able to replicate the finding of Steen-Larsen et al. (2014) that copper was better than PTFE in all metrics. In our study, tubing materials performed similarly when comparing all memory metrics: σs, σm, t63 %, t95 %, t3 ‰, and the absolute value of the maximum D-excess peak. However, given differences in D-excess values between sources, we caution overinterpreting the maximum D-excess anomalies between experiments, as evidenced by the different starting points in Fig. 4e. After accounting for differences in tubing i.d. values and length, PFA seems to be one of the better tubing materials by a very small margin.

Warmer temperatures are theoretically predicted to reduce attenuation times (Pagonis et al., 2017) by changing the saturation concentration of gases. Additionally, the lower molar density of the warmer air means that there is a shorter residence time through the tubing, increased molecular movement, and faster wall exchanges; moreover, warmer tubing material means that fewer molecules are stuck to the tubing walls. We found some evidence of reduced attenuation times in heated experiments when comparing δ18O location times and σm from the impulse function method (Table S2). δ18O location times for heated tubes are always shorter than those of their unheated counterparts, and σm values are similar to or shorter for heated tubing in most cases. Calculated residence times and observed lag times were also shorter for heated tubes, although to varying degrees depending on the tubing. By carrying out δ18O location adjustment on the threshold memory metrics, we are effectively removing the effects of temperature on the residence time of air in the tubing. The removal of temperature effects on residence time is why the differences in Figs. 5 and S5 are not consistent between heated and unheated tubing. However, δ18O location adjustment may also remove some wall-effect differences between materials, and the adjustment limits our ability to discuss t63 %, given the similarity with location times discussed in Sect. 3.1. The memory tail, best described by t95 % (Figs. 5 and S5) and σm (Table S2), shows that heated experiments are not consistently slower or faster to equilibrate than their unheated counterparts. Overall, heating the tubing to avoid condensation does not negatively impact the isotopic measurements. Similarly, Aemisegger et al. (2012) found little difference in attenuation times with varying PFA tubing temperatures.

4.4 Effects of tubing inner volume and length

The model in Pagonis et al. (2017) indicates that tubing delays are expected to scale proportionally with tubing length and diameter. The difference in length in the thick-walled FEP long and short experiments was a factor of 19 (99 ft/5.2 ft, or 30.2 m/1.6 m). Both HDPE and Dekabon had a factor of 15 times difference in length between the long and short experiments (100 ft/79 in., or 30.5 m/2 m). While the results show a clear influence of longer memory times in longer tubing compared with short tubing, we were unable to find quantitative evidence of linear dependence on memory metrics like t95 %. Although there are slight differences in the memory metrics that we calculated, this is likely due to the influence of the analyzer. Because the analyzer optical cavity and inner tubing has a residence time of  4 s, we are unable to resolve the residence time (1.0 ± 0.09 s) and memory metrics associated with the short FEP tubing only. Even with the large length difference in FEP, the shape of the isotopic attenuation curves remained similar after location adjustment which removes the length-based residence time differences (Figs. 6, 7, S6, and S7). δ18O location time-adjusted δD t95 % and t63 % times for long, thick-walled FEP tubing were at maximum 6.2 times and 2.6 times greater than the short tubing, respectively. The mixing timescales (σs) and the memory tail metric (σm) both showed less than a doubling between short and long tubing. These modest differences in wall-effect memory metrics may be because the analyzer memory itself makes it impossible to accurately isolate and quantify the short-tubing response.

From the δ18O-location-adjusted comparison of the same material (FEP) with different i.d. values (panels a and c in Figs. 6 and S6), we conclude that a bigger i.d. causes the increased memory. In our experiments, i.d. increased by a factor of 1.5 times between thick- and thin-walled FEP (1/8 in., or 3.2 mm, i.d. compared with 3/16 in., or 4.8 mm, i.d.). There was clear separation in Fig. 6 between thick- and thin-walled long FEP even after isolating the memory tail by adjusting for bulk delay differences with the δ18O location adjustment. The thin-walled FEP had a less steep slope and longer t63 % intercept than the thick-walled tubing. δ18O location time-adjusted memory metrics also show a slight increase in memory with i.d. increase, with a memory metric that was on average 2.1 times larger for δD and 1.98 times larger for δ18O between thin- and thick-walled long FEP tubing (Table S2). The long, thin-walled FEP consistently showed the slowest δD signal transitions of the FEP tubes tested (panels c and d in Figs. 6 and S6), as well as of all the tubing materials tested (panels c and d in Figs. 4 and S4). We also note that PTFE and PFA tubing had the same i.d. value as the thin-walled FEP. In the H2O-matched experiments, 3/16 in. (4.8 mm) PTFE and PFA experiments showed a shorter attenuation threshold time than the thin-walled FEP, whereas they showed longer attenuation times than tubing with smaller i.d. values like HDPE (Figs. 5 and S5). Therefore, the results in Figs. 4 and S4 must be evaluated while considering i.d. differences.

In summary, we found that all tubing dimensions, including i.d. and length, had some effects on the threshold metrics (Figs. 6, 7, S6, and S7), even after removing differences in δ18O location times in signal propagation to the analyzer based on tubing inner volume and the temperature influence on molecular density and the total number of molecules in the tubing. While these overall memory metric differences exist, they are small in the materials and dimensions tested, with the exception of Dekabon. The operational impact among commonly used 1/4 in. (6.4 mm) o.d. tubing inlets is expected to be limited. Tubing length and i.d. play a role in our experiment, consistent with theory that tubing length, i.d., and material properties, such as density and partitioning depth, will affect the attenuation time of chemical compounds on or in a tubing wall (Pagonis et al., 2017). Further tests under faster analyzer and slower tubing airflow rates would be needed to further validate whether these influences are linearly proportional.

4.5 Relative attenuation time differences between δD and δ18O

δD signals have been demonstrated to take longer than δ18O signals to isotopically equilibrate with tubing materials due to isotope-dependent hydrogen-bonding interactions with the tubing walls (Griffis et al., 2010; Schmidt et al., 2010; Sturm and Knohl, 2010). Our results confirm these previous findings. The speed difference between isotopic signal propagation has been reported as a ratio of attenuation times between the slower δD signal and the faster δ18O signal. Published results show an attenuation time that is 1.4–3.5 times greater for δD signals than for δ18O signals, depending on the tubing airflow rate, tubing type, humidity, and memory metric used (Aemisegger et al., 2012; Griffis et al., 2010; Schmidt et al., 2010; Zannoni et al., 2022). As demonstrated in the results, different metrics show different sensitivities to different parts of the attenuation curve. The threshold metrics that we calculated are most similar to the quantification metrics used in earlier studies, and our results (excluding Dekabon) have similar ranges. For t63 %, this ratio ranges from 1.0 to 3.46, whereas the ratio ranges from 1.0 to 4.79 for t95 %. For σm, a metric that we expected to be sensitive to the characteristic long δD memory tail, δD values were 0.9–1.7 times longer than δ18O values. Overall, δD signals are slower than δ18O signals in all tubing materials.

Dekabon attenuation metric δD/δ18O ratios are greater than those of the other tubing materials; signal ratios are up to 14.1 for σm, 61.5 for t63 %, and 71 times longer for δD vs. δ18O t95 % values across all 100 ft (30.5 m) experiments. We speculate that the long equilibration time and large memory metric ratios in Dekabon and Bev-A-Line XX are due to a large number of water molecules bound to the tubing surface and an affinity to bind to water molecules containing deuterium. These tubing materials may have a larger free volume in their molecular structure, an increased number of hydrophilic sites on the surface, or both (Cava et al., 2006), allowing for more water molecules to be held on or in the tubing. To estimate the amount of water Dekabon might hold, we estimated a mass-balance-based tubing reservoir size using the location-adjusted δD t95 % value as a residence time for water molecules in the tubing. This resulted in a reservoir of 0.13 g H2O, or approximately 0.02 % water absorption by tubing weight. This weight percentage is similar to the upper estimate for PTFE, FEP, and PFA (Table 2). While we were unable to confirm the exact composition of the inner liner of Dekabon, it is an ethylene copolymer, potentially similar to the ethylene–vinyl alcohol copolymers (EVOHs) tested in Cava et al. (2006). Applying the Cava et al. (2006) value of 2 % absorption by weight when our experimental humidity ( 30 %) is considered, the amount of water Dekabon could hold is approximately 14.6 g, much greater than the estimated reservoir size based on isotopic residence time. For Bev-A-Line XX, the mass balance reservoir estimate is 1.7 g, or 0.2 % water absorption by tubing weight. This percentage is double the value for HDPE in Table 2 (Plastic Materials, 2024). Because we know that polymers of the same family with different molecular structures can have different affinities and hold varying amounts of water, as in Cava et al. (2006), it stands to reason that the Dekabon and Bev-A-Line XX tubing types have different affinities and hold different amounts of water than the other tubing materials tested.

4.6 Fitting attenuation curves

The overall attenuation curves of the tested tubing material types, lengths, and temperature conditions had effectively the same reverse sigmoidal shape after the δ18O location time adjustment, although this shape is elongated in Dekabon. Previous studies have approximated the attenuation transfer function as an exponential curve (Aemisegger et al., 2012; Schmidt et al., 2010; Sturm and Knohl, 2010), similar to the exponential decay response that would be expected for the residence time distribution function of a continuously stirred reactor (Toson et al., 2019). We found that the exponential function was not a satisfactory fit to our experimental observations. A more appropriate mixing analogy could be the axially dispersed plug flow (ADPF) model (Huang and Seinfeld, 2019), as this better matches the reverse sigmoid curve that we observe. In the ADPF model, there is a bulk flow that has a diffusive “head” that diverges forwards and backwards from the bulk flow, leading to the observed smoothing of the output signal of an input step change. This diffusive head effectively “smears” the observed isotopic signal. While the shape of this transfer function seems appropriate, the Huang and Seinfeld (2019) model does not consider gas–wall exchange effects. The transfer function model that we introduce here fits the observations sufficiently well, but more work is needed to match the formulas with mixing theory.

Likewise, the impulse fitting method that we used is more complicated than previously used techniques (Jones et al., 2017; Kahle et al., 2018). We were able to estimate a mixing time metric (σs) from the skew-normal function and a memory tail metric (σm) from our modified impulse function fitting method. We believe these metrics are signals of diffusion mixing and isotopic wall effects. Mathematically describing the influence of isotopic wall effects using a transfer and impulse function is potentially useful for correcting out memory effects in water vapor isotope measurements, as suggested by Massman and Ibrom (2008) and others (e.g., Aemisegger et al., 2012; Steen-Larsen et al., 2014). Similar corrections have been achieved in the ice core and liquid water isotope analysis communities (e.g., Jones et al., 2017; Kahle et al., 2018; Vallet-Coulomb et al., 2021). We found that more complicated transfer and impulse function models were necessary to fully capture the memory effects in the vapor inlet system compared with the mostly liquid inlet systems described before (e.g., Jones et al., 2017; Kahle et al., 2018; Vallet-Coulomb et al., 2021). This paper should provide a starting point for future work on removing the low-pass filter effects on continuous water vapor measurements.

5 Implications for measurements

Longer attenuation times smooth signal variability and mask high-frequency features, as shown in Sect. 3.4. While lag times are not inherently bad (as long as measurements lag in tandem), signal smoothing caused by memory effects will draw out the memory tail and muddle environmental signals. Therefore, the magnitude and speed of atmospheric signal variability as well as the analyzer and sample intake smoothing are important considerations when planning for ambient water vapor isotopic measurements. We found very small differences among commonly used tubing materials under the experimental conditions tested here. While different analyzer airflow rates are not presented in this study, it is known that the analyzer flow rate strongly influences the sample residence time in the optical cavity of these analyzers and the speed of signal transitions. The Aemisegger et al. (2012) findings that attenuation times were controlled more by analyzer residence times than PFA intake tubing is supported by the results presented in this study.

Prior research clearly identified Dekabon tubing as unsuitable (Griffis et al., 2010; Schmidt et al., 2010; Sturm and Knohl, 2010; Tremoy et al., 2011) – a conclusion which was validated in this experiment. Bev-A-Line XX also performed particularly poorly, and we cannot recommend the use of either tubing in water vapor isotopic studies. We also suggest testing the effect of any in-line elements, like flow meters, mass flow controllers, or filters, on isotopic signal attenuation, especially if they are made from materials not tested in this study. Our experience found a mass flow meter that introduced a large memory effect (not presented here).

5.1 Low atmospheric variability measurements

For stationary measurements with one intake and high airflow rates, tubing selection among commonly used materials is not as much of a concern, as air advecting past the intake typically changes slowly compared with the tubing attenuation timescales that we quantify here. Conroy et al. (2016), for example, observed vapor on Manus Island, Papua New Guinea, that changed by 22.3 ‰ in δ18O and 154.8 ‰ in δD, with the largest change being  25 ‰ δD over a duration of a few hours. The instant isotopic step change in our experiment (19.7 ‰ in δ18O and 155 ‰ in δD for H2O-matched experiments) is extreme compared with typical atmospheric variability at a stationary inlet. For stationary measurements, any of the tested tubing materials besides Dekabon and Bev-A-Line XX should be suitable and would not be expected to produce large transient D-excess artifacts due to memory differences between δD and δ18O.

5.2 High atmospheric variability measurements

For measurements that need a high temporal resolution of small atmospheric isotopic variability, like flux gradient and eddy-covariance setups or airborne observations, extra precautions should be taken. Griffis et al. (2010) used spectral analysis in their eddy-covariance experiments to show that tube memory effects were not a concern for δ18O signals at tubing airflow rates of 12 L min−1 and analyzer airflow rates of 1.5 L min−1. However, one cannot extend that conclusion to slower airflow rates, and analyzer residence times should be compared across analyzer types.

Aircraft campaigns are of special concern because they not only observe at high temporal (and spatial) resolution but also encounter large and rapid isotopic and humidity variability. Especially when conducting vertical profiles, isotopic compositions can vary by hundreds of per mil in δD. Salmon et al. (2019) found δD signal values ranging from 400 ‰ to 175 ‰ δD within an  5 min vertical profile descent between 1200 and 400 m above ground. Similarly, Sodemann et al. (2017) reported flight sections with > 200 ‰ δD variations in under 5 min. While data were collected at 1 Hz, their reported data are a 15 s average, allowing them a 975 m horizonal and 75 m vertical resolution (Sodemann et al., 2017). However, that best-case estimate is based on the data averaging interval and does not consider signal attenuation due to tubing isotopic memory or mixing in the optical cavity (Sodemann et al., 2017). Additionally, averaging over long time periods may not remove D-excess memory bias depending on patterns of increasing or decreasing δ values. The wetting and drying of the measurement system during flights with large changes in altitude, and therefore atmospheric specific humidity, may also increase isotopic attenuation times.

In both eddy-covariance and aircraft measurement situations, one might consider increasing airflow through the analyzer and intake tubing and shortening the length of tubing from an intake pickoff point to the analyzer in slow-analyzer-flow setups, as has been suggested in previous studies (e.g., Griffis et al., 2010). While high airflow rates can easily be achieved in the air intake main lines in both high-frequency measurement situations, the airflow rate through the analyzer is typically limited by the analyzer design and control software. If tubing or in-line elements, like mass flow controllers, affect the speed at which the isotopes are transmitted from the intake to the optical cavity, signals are effectively low-pass filtered (Zannoni et al., 2022). Our experiments show shorter memory effects for shorter tubing compared with longer tubing. Therefore, it is also important to minimize the length of tubing from the intake pickoff point to the analyzer to reduce the residence time of air in the low-flow portion of the system. These considerations should also maximize D-excess data resolution.

5.3 Liquid water measurements

Liquid water isotope analysis is also plagued by memory effects when samples are converted to the vapor phase for laser-based spectral isotopic analysis, especially in applications measuring samples with large isotopic differences in the same batch. Common protocols recommend multiple replicate injections and discarding the first few to remove carryover from the previous sample (Coplen and Wassenaar, 2015; IAEA, 2009; Penna et al., 2012). In both OA-ICOS and cavity ring-down spectroscopy, Penna et al. (2012) found that up to 8 out of 18 injections had to be ignored to limit memory effects when measuring samples with large isotopic differences. When analyzing highly depleted Antarctic samples ranging from 231.7 ‰ to 421.1 ‰ for δD, memory effects of up to 14 ‰ were found in the first injection compared with the “true” value. Liquid water analysis is one example of a case in which airflow rates and temperatures of transfer lines are often fixed by the instrument design. Material properties inside the analyzer are important, but this study finds little difference between commonly used material types. Waiting for equilibrium in the optical cavity may minimize the memory effect, but a time-efficient method to increase sample throughput is to mathematically correct for these repeatable effects, rather than attempting to minimize them (e.g., de Graaf et al., 2020; Vallet-Coulomb et al., 2021; Hachgenei et al., 2022). Another option, as outlined in the case of de Graaf et al. (2020), is to measure small vapor samples on a background of humid air to reduce memory effects. Work is also being done in the ice core community to correct for signal mixing based on transfer function fitting methods (e.g., Jones et al., 2017; Kahle et al., 2018). These memory correction approaches may provide examples of methods to reconstruct input signal variability from smoothed continuous vapor isotope measurements as well.

6 Conclusions

We tested the water isotopic exchange properties of PFA, FEP, PTFE, HDPE, copper, Bev-A-Line XX, and Dekabon. The commonly used materials tested here (not including Bev-A-Line XX and Dekabon) perform similarly. It does not seem necessary to standardize the materials used to measure stable water vapor isotopologues to make accurate and comparable measurements in most situations, when using analyzers with similar residence times. We cannot recommend Bev-A-Line XX or Dekabon for use in water vapor applications due to extremely long attenuation times. At this relative humidity of  33 %, warmer temperatures did shorten the residence time, lag, and location metrics of the impulse function and t63 % threshold times across all long-tubing experiments, but results were not always consistent for t95 %. While heating the tubing makes the isotopic signal move through the setup faster due to an air density decrease, heating did not decrease isotopic signal smoothing in all cases, and there were no consistent differences with temperature after δ18O location time adjustment. However, heating to avoid condensation does not seem to negatively impact the isotopic measurements. While differences may be found among tubing material types at lower or higher humidities, these experiments are beyond the scope of this study. The direction of the isotopic step change in source transitions affected isotopic transition speeds only in experiments where H2O ppmv was changing. Larger tubing i.d. values and lengths were predicted to increase memory metrics proportionally based on gas–wall partitioning theory (Pagonis et al., 2017), and we found that increasing the tubing i.d. and length increased the threshold metrics after removing differences in δ18O location times. The effect of tubing length was most noticeable between 100 ft (30.5 m) and 79 in. (2 m) Dekabon. The other tubing experiments here showed that overall memory metric differences do exist, but they are small in the materials and dimensions tested. Under experimental settings, the operational impact among commonly used 1/4 in. (6.4 mm) o.d. tubing inlets is expected to be limited.

Researchers must understand the limitations of the airflow conditions and wall effects of their instrumental and intake setups to limit signal memory effects, especially if low airflow rates are a constraint or if there are large isotopic variations over short periods of time. Our experience and results from other published studies indicate that maximizing airflow rates through the analyzer is the most effective way to minimize memory effects when accurate high-frequency D-excess measurements are desired. Our results show that these plastic tubing materials are not inferior to copper in terms of isotopic memory under the tested conditions, and they are easier to work with and are less expensive than copper. As with most decisions, environmental conditions, cost, and preference may influence the type of tubing selected.

Code and data availability

All figure data and scripts, as well as an example workup code, are available at https://doi.org/10.4231/Y13T-6775 (Meyer and Welp, 2024).

Supplement

The supplement related to this article is available online at: https://doi.org/10.5194/amt-17-6193-2024-supplement.

Author contributions

ALM and LRW designed the experiments and conducted them. ALM adapted code from LRW, added to the code for this project, and analyzed data. ALM wrote the manuscript draft. ALM and LRW edited the document.

Competing interests

The contact author has declared that neither of the authors has any competing interests.

Disclaimer

Any opinions, findings, and conclusions or recommendations expressed in this material are those of the authors and do not necessarily reflect the views of the National Science Foundation.

Publisher’s note: Copernicus Publications remains neutral with regard to jurisdictional claims made in the text, published maps, institutional affiliations, or any other geographical representation in this paper. While Copernicus Publications makes every effort to include appropriate place names, the final responsibility lies with the authors.

Acknowledgements

The authors are grateful to Meghan Brown for assistance with conducting the experiments. We also wish to acknowledge Matthew Binkley (MS materials engineering) for valuable discussion of material properties. Finally, we thank the editor (Thomas Röckmann), the reviewer (Robbert Moonen), the two anonymous referees, and a community commenter (Jonathan Keinan) for their time, suggestions for improvement, and patience.

Financial support

Alexandra L. Meyer was supported by a Purdue Doctoral Fellowship and Graduate School Summer Research Grant, an Indiana Space Grant, and the National Science Foundation Graduate Research Fellowship Program (grant no. DGE-1333468).

Review statement

This paper was edited by Thomas Röckmann and reviewed by Robbert Moonen and two anonymous referees.

References

Aemisegger, F., Sturm, P., Graf, P., Sodemann, H., Pfahl, S., Knohl, A., and Wernli, H.: Measuring variations of δ18O and δ2H in atmospheric water vapour using two commercial laser-based spectrometers: an instrument characterisation study, Atmos. Meas. Tech., 5, 1491–1511, https://doi.org/10.5194/amt-5-1491-2012, 2012. 

Ardia, D., Mullen, K., Peterson, B., Ulrich, J., and Boudt, K.: DEoptim: Global Optimization by Differential Evolution, Version 2.2-8, CRAN [code], https://cran.r-project.org/package=DEoptim (last access: 16 October 2024), 2022. 

ASTM: ASTM D570-22 Standard Test Method for Water Absorption of Plastics: ASTM International, https://www.astm.org/d0570-22.html (last access: 16 October 2024), 2022. 

Brecheisen, Z. S., Cook, C. W., Heine, P. R., Ryang, J., and Richter, D. d.: Development and deployment of a field-portable soil O2 and CO2 gas analyzer and sampler, PLoS One, 14, e0220176, https://doi.org/10.1371/journal.pone.0220176, 2019. 

Cava, D., Cabedo, L., Gimenez, E., Gavara, R., and Lagaron, J. M.: The effect of ethylene content on the interaction between ethylene–vinyl alcohol copolymers and water: (I) Application of FT-IR spectroscopy to determine transport properties and interactions in food packaging films, Polym. Test., 25, 254–261, https://doi.org/10.1016/j.polymertesting.2005.09.018, 2006. 

Chemours: Fluoropolymer Industrial Safety Fact Sheet, The Chemours Company, FC, LLC, https://www.chemours.com/en/-/media/files/corporate/pfas/fluoropolymers-industrial-safety-fact-sheet.pdf?rev=5c0409a38f274293b1a0c74dfacf023a&hash=98535320CFE5459081A3F0B2F5391CDF (last access: 16 October 2024), n.d. 

Conroy, J. L., Noone, D., Cobb, K. M., Moerman, J. W., and Konecky, B. L.: Paired stable isotopologues in precipitation and vapor: A case study of the amount effect within western tropical Pacific storms, J. Geophys. Res.-Atmos., 121, 3290–3303, https://doi.org/10.1002/2015JD023844, 2016. 

Coplen, T. B. and Wassenaar, L. I.: LIMS for Lasers 2015 for achieving long-term accuracy and precision of δ2H, δ17O, and δ18O of waters using laser absorption spectrometry, Rapid Commun. Mass Sp., 29, 2122–2130, https://doi.org/10.1002/rcm.7372, 2015. 

de Graaf, S., Vonhof, H. B., Weissbach, T., Wassenburg, J. A., Levy, E. J., Kluge, T., and Haug, G. H.: A comparison of isotope ratio mass spectrometry and cavity ring-down spectroscopy techniques for isotope analysis of fluid inclusion water, Rapid Commun. Mass Sp., 34, e8837, https://doi.org/10.1002/rcm.8837, 2020. 

Electrical properties of plastic materials: https://www.professionalplastics.com/professionalplastics/ElectricalPropertiesofPlastics.pdf (last access: 17 December 2021), 2021. 

Galewsky, J., Steen-Larsen, H. C., Field, R. D., Worden, J., Risi, C., and Schneider, M.: Stable isotopes in atmospheric water vapor and applications to the hydrologic cycle, Rev. Geophys., 54, 809–865, https://doi.org/10.1002/2015RG000512, 2016. 

Goodrich Sales, Inc: Synflex, http://www.goodrichsales.com/products/pdfs/1300.pdf (last access: 29 April 2024), 2005. 

Griffis, T. J., Sargent, S. D., Lee, X., Baker, J. M., Greene, J., Erickson, M., Zhang, X., Billmark, K., Schultz, N., Xiao, W., and Hu, N.: Determining the oxygen isotope composition of evapotranspiration using eddy covariance, Bound.-Lay. Meteorol., 137, 307–326, https://doi.org/10.1007/s10546-010-9529-5, 2010. 

Griffith, D. W. T., Jamie, I., Esler, M., Wilson, S. R., Parkes, S. D., Waring, C., and Bryant, G. W.: Real-time field measurements of stable isotopes in water and CO2 by Fourier transform infrared spectrometry, Isot. Environ. Healt. S., 42, 9–20, https://doi.org/10.1080/10256010500503098, 2006. 

Guerrier, S., Balamuta, J., Bakalli, G., Molinari, R., Lee, J., Radi, A., Xu, H., Zhang, Y., and Claussen, N.: avar: Allan Variance. Version 0.1.1, CRAN [code], https://CRAN.R-project.org/package=avar (last access: 22 March 2022), 2020. 

Gupta, P., Noone, D., Galewsky, J., Sweeney, C., and Vaughn, B. H.: Demonstration of high-precision continuous measurements of water vapor isotopologues in laboratory and remote field deployments using wavelength-scanned cavity ring-down spectroscopy (WS-CRDS) technology, Rapid Commun. Mass Sp., 23, 2534–2542, https://doi.org/10.1002/rcm.4100, 2009. 

Hachgenei, N., Vaury, V., Nord, G., Spadini, L., and Duwig, C.: Faster and more precise isotopic water analysis of discrete samples by predicting the repetitions’ asymptote instead of averaging last values, MethodsX, 9, 11, https://doi.org/10.1016/j.mex.2022.101656, 2022. 

Havranek, R. E., Snell, K., Kopf, S., Davidheiser-Kroll, B., Morris, V., and Vaughn, B.: Technical note: Lessons from and best practices for the deployment of the Soil Water Isotope Storage System, Hydrol. Earth Syst. Sci., 27, 2951–2971, https://doi.org/10.5194/hess-27-2951-2023, 2023. 

Huang, Y. and Seinfeld, J. H.: A note on flow behavior in axially-dispersed plug flow reactors with step input of tracer, Atmospheric Environment: X, 1, 100006, https://doi.org/10.1016/j.aeaoa.2019.100006, 2019. 

IAEA: Laser spectroscopic analysis of liquid water samples for stable hydrogen and oxygen isotopes, International Atomic Energy Agency, Vienna, Germany, Training Course Series No. 35, IAEA, Vienna, 2009. 

Jones, T. R., White, J. W. C., Steig, E. J., Vaughn, B. H., Morris, V., Gkinis, V., Markle, B. R., and Schoenemann, S. W.: Improved methodologies for continuous-flow analysis of stable water isotopes in ice cores, Atmos. Meas. Tech., 10, 617–632, https://doi.org/10.5194/amt-10-617-2017, 2017. 

Kahle, E. C., Holme, C., Jones, T. R., Gkinis, V., and Steig, E. J.: A generalized approach to estimating diffusion length of stable water isotopes from ice-core data, J. Geophys. Res.-Earth, 123, 2377–2391, https://doi.org/10.1029/2018JF004764, 2018. 

Kerstel, E. R. T., Iannone, R. Q., Chenevier, M., Kassi, S., Jost, H.-J., and Romanini, D.: A water isotope (2H, 17O, and 18O) spectrometer based on optical feedback cavity-enhanced absorption for in situ airborne applications, Appl. Phys. B, 85, 397–406, https://doi.org/10.1007/s00340-006-2356-1, 2006. 

Lee, X., Sargent, S., Smith, R., and Tanner, B.: In situ measurement of the water vapor 18O/16O isotope ratio for atmospheric and ecological applications, J. Atmos. Ocean. Tech., 22, 555–565, https://doi.org/10.1175/JTECH1719.1, 2005. 

Luo, H., Pingintha-Durden, N., and Smith, D.: NEON sensor command, control and configuration (C3) document: eddy covariance storage exchange (NEON.DOC.000465) Version F, NEON (National Ecological Observatory Network), 2019. 

Managave, S., Jani, R., Narayana Rao, T., Sunilkumar, K., Satheeshkumar, S., and Ramesh, R.: Intra-event isotope and raindrop size data of tropical rain reveal effects concealed by event averaged data, Clim. Dynam., 47, 981–987, https://doi.org/10.1007/s00382-015-2884-7, 2016. 

Massman, W. J. and Ibrom, A.: Attenuation of concentration fluctuations of water vapor and other trace gases in turbulent tube flow, Atmos. Chem. Phys., 8, 6245–6259, https://doi.org/10.5194/acp-8-6245-2008, 2008. 

Meyer, A. and Welp, L. R.: Water vapor stable isotope memory effects of common tubing materials, Purdue University Research Repository, https://doi.org/10.4231/Y13T-6775, 2024. 

Muggeo, V. M. R.: segmented: Regression models with break-points / change-points (with possibly random effects) estimation Version 1.6-0, CRAN [code], https://CRAN.R-project.org/package=segmented (last access: 6 July 2022), 2022. 

Pagonis, D., Krechmer, J. E., de Gouw, J., Jimenez, J. L., and Ziemann, P. J.: Effects of gas–wall partitioning in Teflon tubing and instrumentation on time-resolved measurements of gas-phase organic compounds, Atmos. Meas. Tech., 10, 4687–4696, https://doi.org/10.5194/amt-10-4687-2017, 2017. 

Penna, D., Stenni, B., Šanda, M., Wrede, S., Bogaard, T. A., Michelini, M., Fischer, B. M. C., Gobbi, A., Mantese, N., Zuecco, G., Borga, M., Bonazza, M., Sobotková, M., Čejková, B., and Wassenaar, L. I.: Technical Note: Evaluation of between-sample memory effects in the analysis of δ2H and δ18O of water samples measured by laser spectroscopes, Hydrol. Earth Syst. Sci., 16, 3925–3933, https://doi.org/10.5194/hess-16-3925-2012, 2012. 

Plastic Materials: https://www.curbellplastics.com/materials/plastics/, last access: 26 August 2024. 

R Core Team: R: A Language and Environment for Statistical Computing, http://www.R-project.org/ (last access: 4 October 2023), 2023. 

Salmon, O. E., Welp, L. R., Baldwin, M. E., Hajny, K. D., Stirm, B. H., and Shepson, P. B.: Vertical profile observations of water vapor deuterium excess in the lower troposphere, Atmos. Chem. Phys., 19, 11525–11543, https://doi.org/10.5194/acp-19-11525-2019, 2019. 

Schmidt, M., Maseyk, K., Lett, C., Biron, P., Richard, P., Bariac, T., and Seibt, U.: Concentration effects on laser-based δ18O and δ2H measurements and implications for the calibration of vapour measurements with liquid standards, Rapid Commun. Mass Spectrom., 24, 3553–3561, https://doi.org/10.1002/rcm.4813, 2010. 

Simonin, K. A., Roddy, A. B., Link, P., Apodaca, R., Tu, K. P., Hu, J., Dawson, T. E., and Barbour, M. M.: Isotopic composition of transpiration and rates of change in leaf water isotopologue storage in response to environmental variables, Plant Cell Environ., 36, 2190–2206, https://doi.org/10.1111/pce.12129, 2013. 

Sodemann, H., Aemisegger, F., Pfahl, S., Bitter, M., Corsmeier, U., Feuerle, T., Graf, P., Hankers, R., Hsiao, G., Schulz, H., Wieser, A., and Wernli, H.: The stable isotopic composition of water vapour above Corsica during the HyMeX SOP1 campaign: insight into vertical mixing processes from lower-tropospheric survey flights, Atmos. Chem. Phys., 17, 6125–6151, https://doi.org/10.5194/acp-17-6125-2017, 2017.  

Steen-Larsen, H. C., Sveinbjörnsdottir, A. E., Peters, A. J., Masson-Delmotte, V., Guishard, M. P., Hsiao, G., Jouzel, J., Noone, D., Warren, J. K., and White, J. W. C.: Climatic controls on water vapor deuterium excess in the marine boundary layer of the North Atlantic based on 500 days of in situ, continuous measurements, Atmos. Chem. Phys., 14, 7741–7756, https://doi.org/10.5194/acp-14-7741-2014, 2014. 

Sturm, P. and Knohl, A.: Water vapor δ2H and δ18O measurements using off-axis integrated cavity output spectroscopy, Atmos. Meas. Tech., 3, 67–77, https://doi.org/10.5194/amt-3-67-2010, 2010. 

Toson, P., Doshi, P., and Jajcevic, D.: Explicit residence time distribution of a generalised cascade of continuous stirred tank reactors for a description of short recirculation time (bypassing), Processes, 7, 615, https://doi.org/10.3390/pr7090615, 2019. 

Tremoy, G., Vimeux, F., Cattani, O., Mayaki, S., Souley, I., and Favreau, G.: Measurements of water vapor isotope ratios with wavelength-scanned cavity ring-down spectroscopy technology: new insights and important caveats for deuterium excess measurements in tropical areas in comparison with isotope-ratio mass spectrometry, Rapid Commun. Mass Sp., 25, 3469–3480, https://doi.org/10.1002/rcm.5252, 2011. 

Vallet-Coulomb, C., Couapel, M., and Sonzogni, C.: Improving memory effect correction to achieve high-precision analysis of δ17O, δ18O, δ2H, 17O-excess and d-excess in water using cavity ring-down laser spectroscopy, Rapid Commun. Mass Sp., 35, e9108, https://doi.org/10.1002/rcm.9108, 2021. 

Webster, C. R. and Heymsfield, A. J.: Water isotope ratios D/H, 18O/16O, 17O/16O in and out of clouds map dehydration pathways, Science, 302, 1742–1745, https://doi.org/10.1126/science.1089496, 2003. 

Zannoni, D., Steen-Larsen, H. C., Peters, A. J., Wahl, S., Sodemann, H., and Sveinbjörnsdóttir, A. E.: Non-equilibrium fractionation factors for D/H and 18O/16O during oceanic evaporation in the north-west Atlantic region, J. Geophys. Res.-Atmos., 127, e2022JD037076, https://doi.org/10.1029/2022JD037076, 2022. 

Download
Short summary
Water molecules stick to air intake tubing wall surfaces, smoothing measurements of fast isotopic variability in the atmosphere. We tested this stickiness and saw small differences in isotopic signal speed between materials, tubing inner dimensions, and isotopic switch direction, although no consistent temperature effects. Researchers can confidently compare measurements across observation systems using different commonly used tubing materials and plan for optimal inlet designs of new systems.