Articles | Volume 17, issue 1
https://doi.org/10.5194/amt-17-73-2024
https://doi.org/10.5194/amt-17-73-2024
Research article
 | 
09 Jan 2024
Research article |  | 09 Jan 2024

New insights from the Jülich Ozone Sonde Intercomparison Experiment: calibration functions traceable to one ozone reference instrument

Herman G. J. Smit, Deniz Poyraz, Roeland Van Malderen, Anne M. Thompson, David W. Tarasick, Ryan M. Stauffer, Bryan J. Johnson, and Debra E. Kollonige

Data sets

Julich Ozone Sonde Intercomparison Experiment (JOSIE) data and Balloon Experiment on Standards for Ozonesondes (BESOS) data WOUDC data archive https://doi.org/10.14287/100000014

Download
Short summary
This paper revisits fundamentals of ECC ozonesonde measurements to develop and characterize a methodology to correct for the fast and slow time responses using the JOSIE (Jülich Ozone Sonde Intercomparison Experiment) simulation chamber data. Comparing the new corrected ozonesonde profiles to an accurate ozone UV photometer (OPM) as reference allows us to evaluate the time response correction (TRC) method and to determine calibration functions traceable to one reference with 5 % uncertainty.