the Creative Commons Attribution 4.0 License.
the Creative Commons Attribution 4.0 License.

A relaxed eddy accumulation flask sampling system for 14C-based partitioning of fossil and non-fossil CO2 fluxes
Ann-Kristin Kunz
Lars Borchardt
Andreas Christen
Julian Della Coletta
Markus Eritt
Xochilt Gutiérrez
Josh Hashemi
Rainer Hilland
Armin Jordan
Richard Kneißl
Virgile Legendre
Ingeborg Levin
Susanne Preunkert
Pascal Rubli
Stavros Stagakis
Samuel Hammer
A relaxed eddy accumulation (REA) system was developed and tested, enabling conditional sampling of air for subsequent 14CO2 analysis. This allows a 14C-based estimation of fossil fuel CO2 concentrations in the collected air samples and, thus, an observation-based partitioning of total CO2 fluxes measured in urban environments by eddy covariance into fossil and non-fossil components. This article describes the REA system, evaluates its performance, and assesses uncertainties in the concentration measurements. In the REA system, two separate inlet lines equipped with fast-response valves and loop systems adapted to the technical requirements enable the conditional collection of air in two sets of aluminum cylinders for updraft and downdraft samples, respectively. The switching between updraft sampling, downdraft sampling, and standby mode is thereby determined by the vertical wind measured at 20 Hz by a co-located ultrasonic 3D anemometer. A logger program provides different options for the definition of a deadband, which is used to increase the concentration differences between updraft and downdraft samples. After the sampling interval, the accumulated air is transferred by an automated 24-port flask sampler into 3 L glass flasks, which can be analyzed in the laboratory, and the cylinders are re-evacuated for the next sampling. The REA system was tested in the laboratory, as well as on a tall tower near the city center of Zurich, Switzerland. Between July 2022 and April 2023, 103 REA updraft and downdraft flask pairs for flux measurements and 9 flask pairs for quality control purposes were selected from the tall tower for laboratory analysis based on suitable micro-meteorological conditions. Uncertainties in the CO2 concentration differences between updraft and downdraft flasks were estimated by simulations using 20 Hz in situ measurements of a closed-path gas analyzer and an open-path gas analyzer co-located with the ultrasonic anemometer. The measurements show that there is no significant bias in the concentration differences between updraft and downdraft samples and that uncertainties due to the sampling process are negligible when estimating fossil fuel CO2 signals. In the Zurich measurements, the CO2 concentration differences between the flask pairs agreed with the differences obtained from in situ measurements within −0.005 ± 0.227 ppm. The largest source of uncertainty, as well as the main limitation, in the separation of fossil and non-fossil CO2 signals in Zurich was the small signal-to-noise ratio of the Δ14C differences measured by accelerator mass spectrometry between the updraft and downdraft flasks. The novel REA flask sampling system meets the high technical requirements of the REA method and is a promising technology for observation-based estimation of fossil fuel CO2 fluxes.
- Article
(2705 KB) - Full-text XML
- BibTeX
- EndNote
In view of the overarching aim of reducing anthropogenic greenhouse gas emissions to mitigate climate change, reliable emissions data and timely information on emission reductions are indispensable, especially at the local scale in urban environments where emission reduction efforts are to be assessed. Of central importance in this context is the quantification of fossil fuel CO2 (ffCO2) emissions in cities as cities contribute more than 70 % to global and European ffCO2 emissions (Seto et al., 2014). Officially reported bottom-up emission inventories are usually based on statistical activity data, e.g., fossil fuel consumption, and emission factors for the different emission sectors, such as traffic or industry (Super et al., 2020). Downscaled to urban and local resolutions, they form an important basis for policy decisions, as well as for fundamental research (WMO, 2022). Despite continuous improvements to such inventories, the benefits of bottom-up estimates are currently limited by their coarse spatial and temporal resolutions, large uncertainties in the available methodologies, and the delayed availability of data (Gately and Hutyra, 2017; Lauvaux et al., 2020; Stagakis et al., 2023). To independently validate and refine emission inventories for CO2, atmospheric measurements providing timely, localized, and sector-specific top-down information are therefore indispensable.
The only method that allows direct measurement of vertical atmospheric trace gas fluxes is the eddy covariance (EC) technique, in which vertical wind velocity and trace gas concentrations are both measured at a frequency of 10 to 20 Hz (e.g., Mauder et al., 2021). Although this method assumes a horizontally flat and homogeneous surface area (Foken et al., 2012), studies have shown that EC measurements can also be successfully performed in a complex and heterogeneous urban environment (Grimmond and Christen, 2012; Feigenwinter et al., 2012; Christen, 2014). However, EC-based CO2 flux estimates contain fossil and non-fossil components. Models and measurements have shown that biospheric CO2 fluxes (autotrophic and heterotrophic respiration and photosynthesis) can contribute significantly to the total CO2 flux measured over an urban area, even in winter (e.g., Crawford et al., 2011; Kellett et al., 2013; Hardiman et al., 2017; Wu et al., 2022). In addition, human respiration fluxes typically account for about 5 % of the total annual CO2 flux, depending on the population and emission density (e.g., Kellett et al., 2013; Crawford and Christen, 2015). Considering the increasing use of biofuels as well (e.g., Guo et al., 2015), this results in significant uncertainties in inverse estimates of fossil fuel emissions (Crawford and Christen, 2015; Wu et al., 2018; Stagakis et al., 2023).
In separating fossil and non-fossil CO2 enhancements, radiocarbon has proven to be the ideal tracer since 14C-free fossil fuels dilute the atmospheric ratio compared to a clean background (e.g., Levin et al., 2003; Turnbull et al., 2016). To measure the atmospheric ratio (on the order of 10−12), air is commonly sampled in glass flasks and subsequently analyzed in the laboratory, e.g., by accelerator mass spectrometry of the extracted and catalytically reduced CO2 (Lux, 2018). Based on long-term atmospheric 14CO2 observations, Levin et al. (2003) quantified the ffCO2 enhancements at three German sites and used the radon tracer method to derive ffCO2 fluxes for the respective catchment areas. Maier et al. (2024a) analyzed the ratio of carbon monoxide (CO), which is co-emitted with ffCO2, to get a continuous time series of the excess ffCO2 concentration in Heidelberg, Germany, and used an atmospheric inversion model to obtain ffCO2 fluxes (Maier et al., 2024b). Wu et al. (2022) used the mean ΔCO ΔffCO2 ratio of flask pairs collected at two towers in Indianapolis, USA, to derive ffCO2 fluxes from CO fluxes measured by the flux gradient approach. However, to derive ffCO2 fluxes, all methods rely on the assumption of constant proxy ffCO2 ratios or results from inversion models. Direct 14C-based ffCO2 flux measurements have not been possible so far because the precision and temporal resolution of state-of-the-art lasers for 14C in situ measurements are still far too low for EC measurements (e.g., Lehmuskoski et al., 2021; McCartt and Jiang, 2022). The relaxed eddy accumulation (REA) method (Businger and Oncley, 1990), on the other hand, allows for flux estimation from the concentration differences between two conditionally collected air samples, which can be determined in the laboratory. The ffCO2 fluxes can thus be estimated from the 14C-based ffCO2 concentration differences between updraft and downdraft flask sample pairs.
In the present study, to our knowledge, the first REA system for 14C-based estimation of ffCO2 fluxes was developed. Based on the principles of relaxed eddy accumulation (Sect. 2) and 14C-based fossil fuel CO2 estimation (Sect. 3), the purpose of this article is to describe the novel REA flask sampling system (Sect. 4), as well as to evaluate its performance and assess the uncertainties of the concentration measurements (Sect. 5). As a proof of concept of the 14C-based separation of fossil and non-fossil CO2 components, the ffCO2 concentration differences between updraft and downdraft flasks collected during a field campaign at a tall tower 112 m above ground level near the city center of Zurich, Switzerland, are presented (Sect. 6). This work forms the basis for the derivation and analysis of ffCO2 fluxes in Zurich and for future deployments in other urban environments.
Relaxed eddy accumulation (REA), first described by Businger and Oncley (1990), is a conditional sampling method for measuring turbulent trace gas fluxes using slow-response analyzers. A fast ultrasonic anemometer measures the vertical wind velocity w at a frequency of 10 to 20 Hz. Based on w, the opening and closing of two fast-response sampling valves are controlled in quasi-real time. Any bias in the vertical wind velocity must therefore be removed before activating the valves (Rinne et al., 2021). When there is an updraft eddy and w is above a certain threshold w0, air is collected and accumulated in an updraft reservoir, whereas air is collected in a separate downdraft reservoir when w < −w0.
The range of wind speeds where no air is collected is called the deadband. Under ideal conditions, the mean vertical wind speed over a sampling period of, for example, 30 or 60 min is zero and defines the center of the deadband. The deadband width can be constant or dynamically adjusted in relation to the standard deviation of the vertical wind speed σw so that w0 = (Rinne et al., 2021). The larger the δ, the greater the concentration difference between the updraft and downdraft reservoirs, reducing the relative measurement uncertainty (Rinne et al., 2021). In addition, a larger deadband reduces the switching frequency of the sampling valves, thereby increasing their lifetime (Rinne et al., 2021). However, this also reduces the fraction of time during which air is collected, which reduces the sample volume and the statistical significance. A compromise between a high concentration difference, a sufficient sample volume, and good representativity has to be found (Christen et al., 2006). Since the requirement = 0 can be violated, particularly in urban environments with complex airflow, and since the actual value of is not known before the end of the sampling period, two options to trigger the valves based on vertical wind were implemented in this study (see Sect. 4.1).
The flow rate into the two reservoirs is always zero when the valves are closed and constant when they are open. This relaxes the high technical requirements of the true eddy accumulation method proposed by Desjardins (1972), where the sampling rate is adjusted in quasi-real time in relation to the magnitude of the vertical wind velocity (for a detailed description, see Siebicke and Emad, 2019, and Emad and Siebicke, 2023). However, the constant flow rate prevents a direct assessment of the mean vertical turbulent flux of a trace gas Fc. With REA, Fc is calculated according to Eq. (1) from the mean concentration difference between the updraft and the downdraft sample Δc, the standard deviation of the vertical wind speed σw, the mean molar air density , and an empirical coefficient β.
The proportionality factor β depends on the joint probability distribution of variations in the vertical wind velocity and the trace gas concentration (Milne et al., 1999). Without a deadband and under an ideal joint Gaussian distribution of w and c, β is 0.627 (Wyngaard, 1992; Baker et al., 1992), but the dependence of β on the prevailing atmospheric conditions; the deadband width; and, eventually, the scalar, e.g., the trace gas concentration, itself adds uncertainty into the calculated flux (Siebicke and Emad, 2019). Grönholm et al. (2008) showed that, with a deadband dynamically adjusted based on the standard deviation of the vertical wind velocity, β does not depend on the atmospheric stability, allowing the use of a constant value in flux calculations. Consequently, assuming similarity in the turbulent exchange of two quantities, β can be determined with Eq. (1) from a scalar where the flux is known from EC measurements, e.g., temperature or CO2 (Rinne et al., 2021).
Estimating the ffCO2 difference between updraft and downdraft reservoirs based on 14CO2 analyses of REA sample pairs and assuming scalar similarity between CO2 and 14CO2, Eq. (1) allows us to determine the fossil contribution to a total CO2 flux:
In this paper, only the concentration differences (right-hand side of Eq. 2) and their uncertainties are analyzed. The observation-based ffCO2 fluxes, which can be obtained by multiplying the ratio of the REA samples with from EC measurements, will be presented in a follow-up work.
It should be noted that Eq. (2) is unstable for and/or ΔCO2 close to zero. In this case, a proxy other than CO2 is needed, e.g., CO. Moreover, like any other method based on the eddy covariance principle, REA only provides reasonable estimates of the mean vertical turbulent flux if the micro-meteorological conditions are stationary and if turbulence is well developed during the sampling period (Foken et al., 2012). If a temporal change in concentrations below the measurement height causes a significant storage flux, the measured turbulent fluxes are not representative of the respective surface fluxes at the time (Crawford and Christen, 2014; Bjorkegren et al., 2015). Further, if vertical advection leads to significant flux components due to non-turbulent vertical transport, this cannot be properly captured by eddy covariance and the REA approach (Foken et al., 2012). This implies the consideration of various criteria when selecting suitable samples in any application (see Sect. 4.4).
In determining the contribution of fossil fuel emissions to a measured CO2 signal, 14C has turned out to be an important tracer. The radioactive carbon isotope has a half-life of 5730 years. Consequently, the millions-of-years-old fossil fuels are 14C-free; hence, CO2 emitted by fossil fuel combustion (ffCO2) dilutes the atmospheric ratio. Due to the low 14C abundance, the specific activity of a reservoir or sample is commonly given as the relative deviation (in ‰) from the absolute specific activity of the radiocarbon standard AABS = 0.2261 Bq gC−1 (Eq. 3) (Stuiver and Polach, 1977). This is commonly denoted as Δ14C but is abbreviated to Δ14 in this work to improve the readability of the following equations.
ASN is the 14C sample activity AS normalized to the postulated mean δ13C value of terrestrial wood of −25 ‰ to correct for isotopic fractionation due to biological or physical processes during the sample formation and the measurement routine (Stuiver and Polach, 1977):
Following the general approach of Turnbull et al. (2016) and Maier et al. (2023), any measured atmospheric 14CO2 signal can, analogously to the total CO2 concentration cmeas, be expressed as the sum of a background (bg), a fossil fuel (ff), a biofuel (bf), a nuclear (nuc), a stratospheric (strato), a respiratory (resp), a photosynthetic (photo), and an oceanic (oc) component:
On a local scale, as during REA sampling, it can be assumed that two air parcels differ only in terms of their fossil fuel, biofuel, respiration, and photosynthesis components, while the impact of the other, more distant sinks and sources is the same. Since biofuels and respiration have a similar Δ14 signature, they cannot be distinguished by radiocarbon analysis alone and are therefore summarized in the following as non-fossil (nf) CO2 emissions, with respiration being by far the largest contributor. Thus, the concentration differences between the updraft (↑) and the downdraft (↓) reservoir of an REA measurement can be expressed in the following way:
Combining Eq. (7) and Eq. (8), the difference in cff between updraft and downdraft reservoirs can be estimated via
The corresponding uncertainty is derived according to Gauss' law of error propagation.
Since fossil fuel emissions do not contain any 14C, is, by definition, exactly −1000 ‰. On the contrary, biogenic signatures are heterogeneous and much more uncertain (e.g., Naegler and Levin, 2009b; Maier et al., 2023). To estimate (in the following denoted as ΔffCO2), it is therefore necessary to make assumptions about , and .
Since the Δ notation accounts for mass-dependent isotopic fractionation (Eq. 4), corresponds to the 14C signature of the photosynthesized atmospheric CO2. For our measurements of local CO2 fluxes close to the tall tower, is therefore best approximated by the measured atmospheric signature . As there are always two flasks per REA run, the average of the up and down flasks is chosen: ). However, due to temporal variability, the uncertainty is larger than the measurement uncertainty (∼ 1.8 ‰; see Appendix D) and is therefore set to 10 ‰.
Non-fossil CO2 (from autotrophic and heterotrophic respiration and, to a lesser extent, biofuels) is generally more enriched in 14CO2 because heterotrophically respired CO2 and CO2 from biofuels were taken up by the biosphere several years to decades ago. At that time, the atmospheric Δ14 was higher due to 14CO2 released during nuclear bomb tests in the 1950s–1960s, which constituted the dominant contribution up to the 2000s (Naegler and Levin, 2009a). Since then, the strongest component of the ongoing atmospheric Δ14 decline has been the emission of fossil fuel CO2 (Levin et al., 2010; Turnbull et al., 2016). In total, the Δ14 signature of respiration was found to be larger than atmospheric values by a few tens of permil (e.g., Palonen et al., 2018; Chanca, 2022). Following Maier et al. (2023), an enrichment of the respiration-dominated of 25 ± 12 ‰ is assumed in this study. The atmospheric signature during CO2 uptake of the biospheric is estimated by the mean value in summer, when photosynthesis is most pronounced.
The third unknown is the difference Δcnf = . This can be estimated based on the mean measured total CO2 differences between up and down flasks and the assumption that, in an urban setting, there is always a fossil contribution. Since this is only a very rough and upper estimate, a relative uncertainty of 100 % is reasonable.
The assumptions regarding , , and Δcnf, as well as the specific values used to calculate ΔffCO2 from the REA flasks of the first measurement campaign in Zurich, are summarized in Table 1. Although , , and Δcnf are not well known, the uncertainty in the ΔffCO2 estimates is dominated by the 14CO2 measurement uncertainty, which currently leads to an inherent ΔffCO2 uncertainty of 0.7 ppm at least and 1.2 ppm on average. Details are given in Appendix D.
Table 1Variables used to estimate = ΔffCO2. denotes the Δ14 values of fossil fuels (ff), photosynthetic (photo) and non-fossil (nf) CO2, and flask measurements (meas). = denotes the mean of the updraft and downdraft samples, which is different for each REA run. The atmospheric signature during CO2 uptake of the biosphere is estimated by the mean value in summer (July to September 2022 in the case of the Zurich campaign). Also given are the specific values derived for the first measurement campaign in Zurich.

The relaxed eddy accumulation (REA) flask sampling system consists of an eddy covariance (EC) system with a 3D ultrasonic anemometer and an integrated open-path gas analyzer (IRGASON, Campbell Scientific Inc., Logan, UT, USA); two fast-response valves controlled by a solid-state DC control module (SDM-CD8S, Campbell Scientific Inc., Logan, UT, USA); a data logger (CR6, Campbell Scientific Inc., Logan, UT, USA); and an extension of a regular 24-port ICOS (Integrated Carbon Observation System) automated flask sampler, described, for example, in Levin et al. (2020) (Fig. 1). This allows the collection of updraft and downdraft air samples (Sect. 2) in glass flasks for subsequent laboratory analysis.

Figure 1Schematic setup of the REA flask sampling system. Pa–Pc refers to the pumps that transfer the air from the inlets to the flask sampler and evacuate the buffers after sample transfer. p1–p4 indicates pressure sensors. Blue components at the bottom of the diagram depict the general wiring for data transfer and communication between the data logger and the sampler computer. The green arrows and the filled, unfilled, and hatched circles indicate the air flow and the position of the valves when sampling for buffer set 1. Details of the loop systems are shown in Fig. 2.
As depicted in Fig. 1, there are two inlets, one for updraft conditions and one for downdraft conditions. They are about 20 cm away from the center of the ultrasonic anemometer and are horizontally separated from each other by about 5 cm. The collection of air is controlled by the two solenoid valves located approximately 30 cm behind the inlets. They respond to the data logger's 20 Hz signal, which is based on the vertical wind velocity measurements of the IRGASON (Sect. 4.1). This signal is also sent to the sampler computer, which controls the extended flask sampler. Two pumps (Pa, up and Pa, down in Fig. 1) transfer the sampled air into two separate 50 L cylinders, which are called buffers in the following. The so-called loop systems (Sect. 4.2) avoid flow rate fluctuations during sampling and non-sampling to ensure constant flow rates despite high-frequency switching. Excess air from the updraft and downdraft sides is released through one common outflow. After a successful sampling event, for which both 50 L buffers need to have a pressure between 1.2 and 1.6 bar (Sect. 4.3), the samples are dried and transferred into 3 L glass flasks (pump Pb), and the buffers are evacuated again (pump Pc). Since this takes about 45 min, there is a second pair of buffers that can be filled in the meantime. This allows for a nearly continuous sampling routine. The flask pairs can be sent to the laboratory or re-sampled if, for example, the sampling conditions did not fulfill the requirements (Sect. 4.4). In addition, a third line from the tower top directly to the flask sampler enables a simultaneous sampling of flasks with a regular filling (Levin et al., 2020), which can be used for quality control tests (Sect. 5.3.1). In the following, the individual steps of collecting REA flasks are described in more detail. The following terminology is used: REA run denotes the REA sampling event of usually 1 h; REA ID denotes the unique, consecutively assigned number for each REA run; sampling mode denotes the state of the updraft or downdraft line when air is collected, i.e, when the vertical wind is above or below the deadband; and standby mode denotes the state of the updraft or downdraft line when no air is collected, i.e, when the vertical wind is below or above or within the deadband. A photo of the inlets and the extended flask sampler and specifications of the components of the system are shown in Fig. 3 and Appendix A.
4.1 Conditional collection of air
During an REA run, the data logger controls the opening and closing of the solenoid valves at the inlets according to the in situ measurements of the vertical wind velocity. Depending on the definition of the deadband (Sect. 2), flags are assigned to each 20 Hz wind measurement, denoting the status of the two valves in the current and previous time step. If the two values are different, the corresponding valve is switched. For quality control, the flags, called REA flags in the following, can be used to estimate the sample CO2 concentrations from high-frequency in situ concentration measurements (see Sect. 5.2 and 5.3.2).
Table 2 shows the possible deadband settings implemented in the logger program, which can be selected depending on the scientific question and site-specific requirements.
It should be noted that the digital output from the IRGASON is delayed by several hundred milliseconds depending on the bandwidth of the low-pass filter that is applied to the actual 60 Hz measurements. The larger the bandwidth, the smaller the delay, with a minimum delay of 200 ms (20 Hz bandwidth). In addition, there will also be a short delay between the signals being sent, the valves being physically switched, and the air being sucked in due to the slight underpressure in the line. The impact of these delays on the flask concentration differences is analyzed in Sect. 5.2.1.
4.2 Transfer of collected air into buffers
To pump the collected air at a constant flow rate into the respective buffers while ensuring that the system can switch between sampling and standby mode at any time, loop systems (Fig. 2) were developed. They consist of a membrane pump that runs continuously, a pressure control valve, a mass flow controller (MFC), and two three-way valves. The technical details of the components are given in Appendix A.

Figure 2Schematic setup of the loop systems indicating the air flow in sampling and standby mode. NC: normally closed (closed in standby mode, open in sampling mode); NO: normally open (open in standby mode, closed in sampling mode); CO: constantly open; MFC: mass flow controller; > 1.6 bar: pressure control valve.
When the solenoid valve at the inlet is closed (i.e., in standby mode when no air is sampled), the three-way valves are closed to the outside, causing the air to circulate in two loops. As shown by the red arrows, the air splits behind the pump: one part flows through the MFC and the constantly open (CO) and normally open (NO) positions of the three-way valve connected to the buffer, while the other part flows through the pressure control valve and the second three-way valve until both reach the suction side of the pump again. The air in the intake line between the inlet and the pump does not move. In sampling mode, both three-way valves switch to the normally closed (NC) position. Then, air is pumped through the line and the MFC into the buffer, while excess air leaves the system through the pressure control valve, as indicated by the green arrows. In both modes, the pump is continuously running, and the pressure behind the pump and the flow rate through the loop are constant due to the pressure control valve and the MFC. Thus, in sampling mode, the flow rate into the buffers is approximately constant, and the system is able to switch between sampling and standby mode at any time. The effect of a remaining variability in the flow rates is discussed in Sect. 5.2.3.
Each loop system has an internal volume of Vl = 12.0 ± 1.1 mL, while the volumes of the inlet tubes with length lt and radius rt are Vt = . The pump flow velocity qpump (L min−1) depends on the dimensions of the intake lines (longer and thinner tubes have a higher resistance) and the flow rate set at the MFC. The latter is adjusted based on the length of the REA run and the deadband width δ: the shorter the REA run and the larger the deadband width, the larger the required flow rate into the buffers. Consequently, an air parcel needs the time
from the inlet to the buffer. We refer to tr as the “rinse time” in the following because it is the exact amount of time by which sampling needs to be delayed artificially in order to avoid sampling air from before the event. For this purpose, the valves at the buffers remain closed for the first tr seconds in sampling mode so that the residual air is released through the outflow. On the contrary, the pump stays on, and all valves remain open for an additional tr seconds after the end of the sampling period so that all sample air remaining in the tubes is transferred into the buffers. Due to the site-specific lengths of the intake lines, tr must be individually calculated and adjusted in the sampler software. As shown in Sect. 5.1 and 5.2.2, calculated values agree well with measurements of the travel time of a CO2 pulse, and uncertainties in the flask concentrations resulting from the uncertainty of tr are negligible.
4.3 Transfer of accumulated air into glass flasks
After the sampling period, the accumulated air is transferred from the buffers into 3 L glass flasks in the flask sampler. Thereby, it is dried to a dew point of approximately −40 °C. To (almost) completely replace the initial air content of a flask with the sample air from the buffers, the flask volume is flushed 10 times at atmospheric pressure. Then the flask is filled up to 2 bar. As the performance of the sampler transfer pump does not allow us to use the entire sample volume in the buffers, a minimal sample amount of 60 L is desired, corresponding to a pressure of 1.2 bar in the buffers. Excess air is discarded when the buffers are evacuated. At the same time, the maximum pressure at which the pumps in the loop systems can be operated is 1.6 bar, and so an REA run is automatically stopped when either buffer reaches this threshold. Consequently, an REA sampling event is only successful if the pressures of the updraft and downdraft buffers are between 1.2 and 1.6 bar at the end of sampling.
4.4 Sample selection for laboratory analysis
The REA flasks are analyzed with gas chromatography for CO2, CO, CH4, N2O, SF6, and H2 at the ICOS Flask and Calibration Laboratory (FCL) in Jena, Germany. To measure 14CO2, the flasks are sent to the ICOS Central Radiocarbon Laboratory (CRL) in Heidelberg. There, the CO2 is extracted from the remaining air sample and catalytically reduced to graphite (Lux, 2018). The graphite targets are then analyzed for the ratio with an accelerator mass spectrometer at the Curt-Engelhorn-Centre Archaeometry in Mannheim, Germany. Since the measurement process is complex and because funding is limited, a thorough selection of appropriate samples is necessary. Several criteria are important to consider.
First, it must be ensured that there were no technical problems with the IRGASON or the flask sampler, i.e., that the CO2 signal strength of the IRGASON was above 90 %, that the valves were switching, that the flow rate into the buffers was approximately constant, and that the horizontal wind direction was not from sectors with known flow distortion effects of the ultrasonic anemometer or the tower structure. Second, the assumptions made in the EC and REA methods, namely stationarity and well-developed turbulence (Sect. 2), must have been fulfilled during the sampling period. For this purpose, the 20 Hz measurements of the IRGASON during the sampling period can be analyzed. In this study, the software EddyPro (version 7.0.9, Licor Inc., Lincoln, NE, USA) was used to process the data. Flask samples were discarded if the integral turbulence characteristic test for w (Foken et al., 2004) was greater than 100 % and if the steady-state test for the covariance between the vertical wind w and the CO2 concentration (Foken et al., 2004) was greater than 400 %.
Depending on the aim of the application, further selection criteria may be considered in the selection of REA flasks. If the objective is a 14C-based decomposition of total CO2 surface fluxes, the fossil fuel CO2 differences between updraft and downdraft samples should be greater than the inherent measurement uncertainty of 0.7 ppm that is caused by the current 14CO2 measurement precision (see Appendix D). For times during which photosynthesis is expected to be weak, estimates of total ΔCO2 based on the IRGASON measurements are good indicators.
4.5 Zurich installation
The REA system was installed and operated for the first time in Zurich, Switzerland, from July 2022 to April 2023. The IRGASON and REA inlets were mounted on a radio tower of 16.5 m height on top of a 95.3 m tall high-rise residential building (Fig. 3). The data loggers, the network devices, and the REA flask sampler were placed in a climate-controlled room on the top floor of the building. The building is surrounded by an industrial sector, railway lines, and a busy commuting road to the north; an urban sector (city center) to the southeast; and a green, less densely populated area with a cemetery to the southwest. Sampling periods of 1 h were chosen to match the current resolution of mesoscale inverse models. The deadband was first used with pre-set wind statistics, i.e., adjusted to the mean and standard deviation of the vertical wind speed of the preceding 30 min period (Table 2). In this case, changes in wind statistics during the sampling period often lead to unequal volumes being filled into the updraft and downdraft buffers, and, on average, only every fourth REA sample pair could be successfully transferred into the flask sampler. With a dynamic deadband, implemented and used since the end of October 2022, the success rate increased to about 75 % for the remaining samples. After 12 experimental runs in the beginning of the campaign with varying deadband widths of δ = 0.3, 0.4, or 0.8, the deadband width was always set to δ = 0.7 for the remainder of the campaign. In this case, the solenoid valves switched, on average, 23 ± 11 times per minute, and air was collected around 50 % of the time. To sample sufficient air for filling a flask (Sect. 4.3), the flow rate into the buffers was 4.67 L min−1. With an estimated pump flow velocity of 5 ± 1.5 L min−1 and with 33 m long Synflex tubes with an inner radius of ri = 2.65 mm connecting the inlets with the loop system, the rinse time was set to 8 s (Eq. 10). During this rinse time, the flow rate at the MFCs was 3.5 L min−1. This implies a slight but negligible underrepresentation of the air sampled during the final 8 s of the sampling period; in future applications, the flow rates during sampling and during the rinse time should be equal. In total, 640 REA runs were started; 300 samples were successfully transferred into flasks; and 103 flask pairs were eventually analyzed for greenhouse gases, including 14CO2.

Figure 3Setup of the Zurich campaign. (a) Schematic illustration of the building Hardau II with the mast, where measurements were made. Also given are the corresponding heights in meters above sea level (m a.s.l.) and above ground level (m a.g.l.). zH denotes the mean building height within 1.5 km radius, and zd is the displacement height according to Kanda et al. (2013). (b) Picture of the IRGASON and the inlets for the relaxed eddy accumulation (REA) system, a multi-compound (MGA7) gas analyzer, and an Aerodyne for carbonyl sulfide (COS) measurements. The aforementioned instruments were placed in a room on the top floor of the building and connected by 33 m long Synflex tubings.
The final setup of the REA flask sampling system, including the chosen materials (Table A1) and the flow scheme with the specially designed loop systems (Fig. 1), was the result of many preliminary assessments, simulations, tests, and iterative system improvements required to meet the high technical requirements of the REA method and the sample analysis. However, due to non-idealities and uncertainties in the sampling procedure, the concentration of the air that is collected in the updraft and downdraft flasks may deviate from the “true” sample concentration that would result from a certain temporal variation in the gas concentration and the vertical wind velocity. To evaluate the performance of the system, i.e., to check for biases and quantify the uncertainty of the flask concentrations due to the sampling process, several experiments and simulations were performed. This section describes the following:
The focus was on the uncertainty in the CO2 concentration differences between updraft and downdraft flasks (ΔCO2) rather than on absolute concentrations because only the concentration differences are needed to calculate fluxes (Eq. 1). Moreover, the flask concentration differences were comparable with high-frequency in situ CO2 measurements of the IRGASON and a multi-compound (MGA7) gas analyzer (MIRO Analytical AG, Wallisellen, Switzerland) despite an irregular calibration of the gas analyzers. It must be noted that, when estimating fossil fuel CO2 differences, uncertainties due to the sampling process are negligible compared to the 14C measurement uncertainty (compare Sect. 3 and Appendix D).
5.1 Quality control tests in the laboratory
Several quality control tests were carried out at the ICOS Flask and Calibration Laboratory in Jena to investigate bias or uncertainty due to contamination and memory effects, i.e., the dependence of the measured concentration on the previous sample, for example, due to incomplete evacuation of the buffers. In addition, the rinse time required at the beginning and end of a sampling period was determined by injecting CO2 pulses.
By sampling gas from a cylinder with a known concentration under sampling conditions as close to reality as possible (for details, see Appendix B), it was shown that neither the buffers nor the intake lines or the switching of the valves alter the gas concentrations significantly. As shown in Fig. 4, the CO2 concentrations of test flasks filled with the sampling system using updraft and downdraft intakes, the loop systems, and buffers agree within 1σ with each other and with the cylinder concentration and mostly meet the WMO compatibility goal for CO2 of 0.1 ppm (WMO recommendation for the compatibility of measurements of greenhouse gases and related tracers; Tans and Zellweger, 2014).

Figure 4Deviation of measured flask sample concentrations from the CO2 concentration of the reference cylinder (405.71 ± 0.01 ppm) that was connected to the fast-response solenoid valves. The gray-shaded area highlights the WMO compatibility goal of 0.1 ppm for CO2. Details of the experiments are given in Appendix B.
A memory effect was visible after filling one buffer with pure nitrogen (causing a diluted CO2 concentration of 404.6 ppm measured in the flask compared to 405.8 ppm inserted from the cylinder). During regular operation of the system, however, the differences between two consecutive buffer fillings are much smaller. Based on the measurements during the Zurich campaign, the additional uncertainty of the absolute CO2 concentration is therefore estimated to be about 0.15 ppm, depending on the CO2 difference compared to the previous sample (see Appendix B2). The impact on ΔCO2 is expected to be even smaller (uncertainty of about 0.02 ppm) since the updraft and downdraft samples are usually affected in a similar way.
It was also shown that the time needed to pump an air parcel from the inlet into the buffer as estimated from the flow rates and the volumes of the tubes (Eq. 10) agrees well (within less than 1.5σ) with the measured travel time of an artificial CO2 pulse (see Appendix B3). This confirms the calculation of the rinse time according to Eq. (10).
5.2 ΔCO2 uncertainty simulations
The laboratory tests “only” proved the ability to successfully transfer sample air from the REA inlets into the flasks without altering the gas concentration. Several aspects that would affect the measurements in a real REA run, i.e., with a temporally varying gas concentration, were investigated through simulations using the high-frequency CO2 measurements of the MGA7. For this purpose, the 10 Hz time series of the MGA7 (available since early August 2022, with a few outages, covering 74 REA sampling periods) was upsampled to 20 Hz and synchronized with the IRGASON data by finding the time lag of maximum correlation between the high-frequency CO2 measurements. Mean concentrations of the updraft and the downdraft samples could then be simulated from the high-frequency data using the 20 Hz REA flags (Sect. 4.1). In this section, ΔCO2 uncertainties and potential biases are assessed based on the MGA7 data. To make the in situ estimates comparable to the air samples that were dried during the transfer into the flasks (compare Sect. 5.3), the measured gas densities were converted into dry air molar fractions (see Appendix C1).
5.2.1 ΔCO2 with delayed collection of air
Time lags between a change in the sign of the vertical wind velocity fluctuations w′ = and the actual collection of air in the respective reservoirs are a known source of uncertainty in REA flux measurements (e.g., Baker et al., 1992; Pattey et al., 1993). In the given setup, the IRGASON measures the vertical wind speed at a frequency of 20 Hz so that the maximum delay between the change in wind speed and its detection is 50 ms. As mentioned in Sect. 4.1, the IRGASON output has, at the maximum bandwidth of 20 Hz, a time delay of 200 ms. With a bandwidth of 10 Hz, as used by default in Zurich, there is a 400 ms delay. The response time of the solenoid valves is 50 ms or less (open) or 150 ms or less (closed). Consequently, it is assumed that the collection of air is delayed by up to 500 ms.
The effect of a delayed collection of air was investigated by calculating the expected CO2 difference between updraft and downdraft samples of the Zurich REA runs based on the 20 Hz REA flags considering different time lags (see Appendix C3). As expected, the larger the delay, the smaller the concentration differences as air is “sampled” from within the deadband, while portions of the larger signals are missed. On average, a 500 ms delay reduces ΔCO2 by 0.04 ± 0.06 ppm. If the delay is 100 ms smaller or larger than expected (i.e., 400 or 600 ms), ΔCO2 changes by less than 0.02 ppm. More results are given in Table C2.
5.2.2 ΔCO2 with incorrect rinse time
Compared to systems with a single inlet, the direct separation of updraft and downdraft samples close to the ultrasonic anemometer of the IRGASON prevents biases due to uncertainties in the travel time of air from the inlet to the buffer and a potential mixing of air in front of the valves. The travel time, however, becomes important at the start and end of sampling, where the rinse time determines the opening and closing of the magnet valves at the buffers (Sect. 4.2). Given the uncertainties in the pump speeds and the lengths of the tubing, the calculated rinse time (Eq. 10) in the Zurich setup had an uncertainty of about 2 s. This can lead to the sampling of unwanted air and the loss of wanted air. The impact on the concentration difference between the updraft and downdraft flasks was estimated from MGA7 data by discarding the first 2 s of measurements in sampling mode and adding another 2 s after the sampling end and vice versa (see Appendix C4). There is no systematic bias, and the standard deviation in the change of ΔCO2 is 0.01 ppm.
5.2.3 ΔCO2 with variable sampling rate
Another source of uncertainty is the actual flow rate at which the air is sampled. This flow rate should be constant to ensure a homogeneous weighting of the collected air over time. For this purpose, mass flow controllers were placed right before the buffers. However, the MFCs are slightly affected by humidity and have a response time of 1 to 2 s. This means that, after a change in the pressure difference across the element, it will take up to 2 s for the flow to become constant again. In the REA system, the valves can switch with a frequency of up to 10 Hz, and, with each switching, the pressure behind the MFC changes between atmospheric pressure (standby mode) and the pressure in the buffer (sampling mode). Consequently, the deviation from the desired flow is especially large at the beginning and end of a sampling period, when the pressure in the buffer is 1 mbar or up to 1600 mbar, respectively. In addition, unsynchronized switching of the top valves (20 Hz scan rate) compared to the valves in the loop systems (10 Hz scan rate) can lead to underpressure in the intake line. The greater the underpressure, the higher the flow rate at the inlet when the top valve opens the next time. At the beginning of the Zurich campaign, there were additional biases due to a difference in flow rate during the rinse time compared to during the rest of the sampling period and due to a loss of sample when one of the two buffers reached the maximum pressure of 1.6 bar.
To estimate the effect on the concentration differences between updraft and downdraft flasks, ΔCO2 was simulated 103 times for each REA run during the Zurich campaign, each time with a different weighting of the MGA7 or IRGASON measurements. The weighting was thereby based on actual flow rate measurements of the MFCs (see Appendix C5). While there is no systematic bias, the standard deviation of the ΔCO2 difference between an inhomogeneous and a homogeneous weighting is, on average, 0.03 ± 0.05 ppm. The deviations are largest during sampling periods with high CO2 variability, where the absolute ΔCO2 is also usually larger than average (see Fig. C1). Consequently, the relative uncertainties are less affected. To minimize this uncertainty, the flow rate during the rinse time was adjusted to the flow rate during the sampling period; the maximum buffer pressure, i.e., the pressure at which sampling is stopped, was reduced to 1.55 bar; and an additional pressure regulator was installed in each loop system.
Mean ΔCO2 uncertainties
Table 3 summarizes the estimated ΔCO2 biases and uncertainties due to the different aspects mentioned above. It is important to note that the given means and standard deviations are solely based on data from the 103 REA runs during the Zurich campaign and a small number of quality control tests. The results may be different for other time periods and other sites. However, the estimates show that, except for a delayed collection of air, there is no bias in the concentration differences between updraft and downdraft samples. The standard deviations of the data set, on the other side, are of the order of the ΔCO2 measurement uncertainty at the gas chromatograph (GC). This means that, for individual samples, e.g., those collected during periods of high variability in the CO2 concentration of the ambient air, the flow variability and other non-idealities in the sampling process can be significant sources of ΔCO2 uncertainty. Regular quality control tests and comparisons of flask samples with in situ measurement data are therefore important for independent validation of the performance of the system during a measurement campaign (see Sect. 5.3). Due to dependence on the ambient sampling conditions, the different ΔCO2 uncertainty contributions should, in this case, be considered for each sampling period individually.
Table 3Overview of the estimated biases and uncertainties in the CO2 concentration differences between updraft and downdraft flasks (ΔCO2) due to non-idealities in the sampling procedure. Given are the means (bias) and standard deviations (uncertainty) of the measured or simulated changes in ΔCO2 with respect to the expected values. For completeness, the measurement uncertainty of the gas chromatographic analysis of the flasks is also given. N represents the number of measurements or sampling periods for the underlying data set.

On the contrary, to estimate ffCO2 differences, the measurement uncertainty of 14CO2 is by far the dominant source of uncertainty (see Appendix D). Moreover, when estimating ffCO2 fluxes, β calculated from CO2 for each REA run individually accounts for most of the above-mentioned effects (Pattey et al., 1993). In this case, the uncertainties due to the sampling process are considered to be negligible.
5.3 Quality control tests during the campaign
To ensure that the REA flask sampling system was working as intended, quality control tests were performed regularly throughout the Zurich campaign. In addition, the measured concentration differences of the REA flask pairs were compared with the in situ measurements of the IRGASON and the MGA7.
5.3.1 All-valves-open tests
To check for biases between updraft and downdraft sampling, as well as for leaks or other sources of contamination, “all-valves-open” tests were performed about once a month. This involved continuously filling two buffers with ambient air by turning on the pumps and opening both solenoid valves at the inlets, as well as the valves in the loop systems, until a buffer pressure of about 1.4 bar was reached and the samples could be transferred into the flask sampler. Biases between updraft and downdraft lines would result in concentration differences between the two corresponding flasks. To detect systematic errors that might affect both lines equally, a third flask was sampled simultaneously through a separate tube directly into the flask sampler, bypassing the REA loops and the buffers. In this case, the flow rate through the flask was reduced over sampling time t according to as flask concentrations then best reflect the real mean concentrations (Levin et al., 2020). If the system is working as expected, the concentrations in the three flasks should agree within the WMO compatibility goal. Another quality control would be to compare the absolute concentrations of the flasks with the average concentration of in situ measurements over the sampling period. In Zurich, however, no accurate measurements of ambient air near the inlets were available (neither the IRGASON nor the MGA7 were meant to be calibrated regularly according to WMO standards).
During the Zurich campaign, nine all-valves-open tests were performed. Figure 5 shows the CO2 concentration of each flask compared to the respective mean of the “up” and “down” samples. In tests 2, 3, 7, 8, and 9, an additional flask was sampled directly into the flask sampler. However, it was later found that, at least at the beginning of the sampling periods, the flow rate was lower than intended. This means that the air in the direct flasks could not be exchanged sufficiently, such that the concentration does not represent the actual mean value but is also influenced by the purging period preceding the sampling. This particularly affects the flasks that were collected during a period with a large variability in concentration.

Figure 5Comparison of flask concentrations from all-valves-open tests. Shown are the differences in CO2 between flasks sampled through the updraft and downdraft intakes of the REA system and a flask sampled directly into the flask sampler compared to the mean of the two samples from the REA intake system. Large deviations in tests 5, 6, and 7 (beige-shaded parts with different scales on the y axis) were caused by two leakages in the loop systems. Outside this period, the results mostly agree within the WMO compatibility goals.
It can be seen that, for tests 1, 2, 3, 4, 8, and 9, up and down flasks agree within their measurement uncertainties, indicating that there is no significant bias between the two lines. The difference between the pairs of simultaneously collected flasks is, on average, −0.01 ± 0.03 ppm. The smaller CO2 concentrations of the direct flasks of tests 8 and 9 can be explained by a large CO2 variability with an overall increase over time. The fact that, in tests 2 and 3, up and down flasks also agree with the direct sample within the WMO compatibility goal indicates that there was no bias from the loop system or from the buffers that would have affected the up and down flasks in the same way.
In tests 5, 6, and 7, on the contrary, there are large concentration differences between up, down, and direct flasks (note the different scales on the y axis). This indicates a leak in the system, which also explains the large deviations of the measured CO2 concentrations from the expected CO2 concentrations based on in situ measurements observed at this time (compare with the next section). Unfortunately, due to the long time lag between sampling and the availability of concentration measurements, the leak could only be detected and fixed after several months.
In summary, the results show that, in general, there is no significant bias between updraft and downdraft sampling and that all-valves-open tests help detect leaks.
5.3.2 Flask – in situ comparison
In addition to the all-valves-open tests, the measured concentration differences of the REA flask pairs were compared to in situ measurements from the IRGASON and the co-located MGA7 by averaging the high-frequency data from the periods during which the respective valves were open and air was sampled, as denoted by the 20 Hz REA flags (see Appendix C2). For the final ΔCO2 estimates, uncertainties of ⋅ 0.15 ppm for the IRGASON estimates and ⋅ 0.1 ppm for the MGA7 estimates were assumed based on the precision of the instruments as stated by the manufacturers. For the flask data uncertainty, the GC measurement uncertainties and the individually simulated sampling uncertainties from a memory effect, an uncertainty in the time lag between the vertical wind signal and the conditional collection of air, an uncertainty in the rinse time, and an inhomogeneous weighting of sample air due to variability in the sampling rate were taken into account (compare to Table 3).
The results of 102 REA runs during the Zurich campaign are shown in Fig. 6, which plots the differences between ΔCO2 from flask measurements and in situ estimates from the IRGASON and the MGA7 over the sampling time (for 1 of the 103 REA selected runs, neither IRGASON nor MGA7 data are available). The samples from November 2022 to February 2023, which were contaminated due to a leak (Sect. 5.3), were discarded.

Figure 6Difference between ΔCO2 (up−down) measured in the flasks and estimated from high-frequency measurements of the IRGASON and the MGA7. Error bars represent the flask measurement uncertainty (analysis plus estimated uncertainty due to the sampling process) and an estimated uncertainty of 0.15 ppm for the IRGASON and 0.1 ppm for the MGA7. MGA7 data are only available from the middle of August. The leak period between the end of October and the beginning of February and the measurements by the IRGASON of CO2 signal strength < 90 % and the correlation with the MGA7 < 0.5 were excluded. The boxplot on the right shows the mean, median, and interquartile range of all summer and winter samples.
The fact that the MGA7 estimates agree very well with the flask measurements (mean difference in ΔCO2 < 0.01 ppm, with a standard deviation of 0.23 ppm) provides evidence that there were no major leaks or significant biases due to the sampling process. A potential bias due to a delayed collection of air (Sect. 5.2.1) is therefore considered to be negligible. Four ΔCO2 measurements deviated from the expected value by more than 0.5 ppm. The exact reasons are not known, but, for example, during the REA run on 21 February 2023 (07:00–08:00 LT), the CO2 concentration increased by more than 100 ppm, and the CO2 difference between the updraft and downdraft samples was the highest observed. During the REA run on 11 March 2023 (13:30–14:30 LT), the wind measurements were very spiky, likely due to a rain event earlier in the day, and, according to the REA flags, the valves did not switch as one would expect from the wind data. These sampling periods must therefore be examined more closely before further analysis.
The IRGASON-based ΔCO2 estimates deviate significantly more from the flask measurements, with an average of 0.20 ± 0.33 ppm. This, however, could be linked to the fact that the IRGASON CO2 dry molar fractions were derived from a CO2 density output that does not properly account for high-frequency fluctuations in air temperature in the sensing path because the ambient temperature measured by the EC100 slow-response temperature probe is used in the conversion of absorption measurements into CO2 density (see Appendix C1). As described by Helbig et al. (2016), this causes a systematic bias compared to closed-path gas analyzers due to the high-frequency temperature attenuation. Indeed, the difference between ΔCO2, flasks and ΔCO2, IRGA correlates with the difference between ultrasonic temperature during updraft and downdraft conditions. Helbig et al. (2016) showed that this bias could be reduced significantly by using the ultrasonic anemometer's fast-response temperature. Due to a lack of knowledge, this additional CO2 density output, where the high-frequency temperature is used in the absorption-to-density conversion, was not recorded during the Zurich campaign. Since the differences between IRGASON and flask measurements could be explained by the insufficient correction of spectroscopic effects during high sensible heat fluxes, the good agreement between flask data and MGA7 measurements indicates an overall successful implementation of the REA method.
Figure 7 shows the differences in CO2 and ffCO2 between the updraft and downdraft flasks, measured with the gas chromatograph at the ICOS FCL in Jena and derived from the 14CO2 analyses at the ICOS CRL in Heidelberg (Eq. 9). Of the 103 REA flask pairs selected for laboratory analysis and analyzed for CO2 at the FCL (Sect. 4.5), 3 samples were lost during graphitization at the CRL. Eight sample pairs from the end of the campaign were not analyzed for 14CO2 because their total CO2 difference was small, and the expected significance of the 14CO2 analysis was low compared to the associated costs. This leaves a total of 92 ΔffCO2 estimates. The error bars represent the measurement uncertainties (ΔCO2 uncertainties of about 0.05 ppm in x direction are omitted for clarity). In the analysis of ffCO2, uncertainties due to the sampling process were considered to be negligible (Sect. 5). The colors indicate the month in which the sample was collected.
It can be seen that the largest concentration differences between updraft and downdraft flasks with ΔCO2 of up to 13.6 ppm were collected in February and March, when anthropogenic emissions are usually higher than average due to residential heating. Most of these samples are close to the 1:1 line, which represents the case where the net non-fossil CO2 flux is zero and the measured CO2 fluxes are completely due to fossil fuel emissions. Accordingly, the observed deviations from this line to the right or the left indicate dominant respiratory and other non-fossil sources or photosynthetic signals, respectively. For samples along the x axis, the ffCO2 contribution is approximately zero. In agreement with other studies (e.g., Wu et al., 2022; Miller et al., 2020; Crawford and Christen, 2015; Turnbull et al., 2015), this shows that there are significant non-fossil CO2 signals even in winter.
It should be noted that the applied sample selection criteria do not necessarily exclude sampling periods in which a change in the storage below the sampling height contributes to the measured fluxes. This is the case, for example, when the depth of the atmospheric boundary layer increases rapidly due to convective, turbulent vertical motions generated by radiative heating of the surface in the morning (Stull, 1988). Based on an observed drop in CO2 concentration in the morning and a preliminary investigation of the turbulence structure of the atmosphere, some of the Zurich REA flasks with large CO2 concentration differences between updraft and downdraft samples were probably collected during this time. In this case, the composition of the samples does not necessarily represent the actual surface fluxes during the REA sampling periods but rather represents the integrated nocturnal fluxes. Further interpretation of the results and a subsequent investigation of the corresponding CO2 fluxes therefore require a thorough analysis of the sampling periods.
What also becomes clear in Fig. 7 is that most signals are of the order of 1 ppm or less, which is small compared to the ΔffCO2 uncertainty. The latter was, on average, 1.2 ppm and was primarily due to the mean 14CO2 measurement precision of 1.8 ‰ (Appendix D). Consequently, ffCO2 flux estimates derived from samples with ΔffCO2 < 1.2 ppm will mostly have uncertainties of more than 100 % (compare to Eq. 2). This highlights the importance of further quality control of the existing data set for subsequent calculations and quantitative analyses of ffCO2 fluxes.
A relaxed eddy accumulation (REA) flask sampling system for 14C-based estimation of ffCO2 fluxes for urban eddy covariance (EC) flux measurements was developed and tested in the ICOS Flask and Calibration Laboratory in Jena, Germany, and on a tall tower near the city center of Zurich, Switzerland.
Two fast-response valves are activated by the vertical wind signal from an ultrasonic anemometer of a co-located IRGASON. The conditionally collected air accumulates in two separate cylinders (one for the updraft sample and one for the downdraft sample) and is transferred into 3 L glass flasks after the sampling period using an automated 24-port flask sampler. The samples can thus be analyzed in the laboratory for a variety of gases, including 14CO2. Laboratory tests and quality control tests during the first measurement campaign in Zurich showed that the sampling procedure does not cause a significant bias in the CO2 differences between updraft and downdraft samples. Uncertainties due to the sampling process are negligible when estimating fossil fuel CO2 differences between the updraft and downdraft flasks and, consequently, ffCO2 fluxes as these are dominated by the 14CO2 measurement precision. The novel REA flask sampling system itself fulfills the high technical requirements of relaxed eddy accumulation, providing high-quality data for scientific studies of multiple non-reactive species.
Due to the prerequisites of the EC and REA method, e.g., stationarity and well-developed turbulence, and the costs and efforts associated with the flask analysis, only a limited number of individual sampling periods can be analyzed. In general, operating the system, i.e., scheduling sampling events, analyzing and selecting suitable sampling periods, sending the flasks to the laboratory, etc., requires frequent remote and on-site work.
Given the good agreement between the total CO2 concentration differences measured in the REA flask sample pairs and observed from in situ measurements of an MGA7, the results of the first measurement campaign in Zurich serve as a proof of concept for a 14C-based separation of fossil and non-fossil CO2 signals. As expected, the CO2 differences between updraft and downdraft samples were largest during the heating season in February and March. In this case, fossil fuel emissions were the major contributor. However, even in winter, small photosynthetic and significant non-fossil (respiration and biofuels) signals were observed, highlighting the role of the biosphere in an urban environment.
The main challenge so far was a generally small signal-to-noise ratio of measured 14CO2 differences. Resulting ΔffCO2 uncertainties on the order of 100 % limit the interpretation of individual results. However, a new accelerator mass spectrometer has reduced the Δ14 measurement uncertainty and, thus, the ΔffCO2 uncertainty by about 23 %, which will increase the fraction of usable samples in future applications.
Two improvements are proposed to further increase the concentration differences in future campaigns: first, the two pumps in the loop systems are recommended to be replaced by larger pumps with a higher pump speed, allowing a reduction in the proportion of time that air is collected and, thus, a larger deadband width δ. Second, the so-called hyperbolic relaxed eddy accumulation could be added into the logger program. With this setting, as proposed by Bowling et al. (1999), air would only be collected if the fluctuations in vertical wind speed w′ and concentration c′ are above a certain threshold. For the latter, the 20 Hz CO2 measurements of the IRGASON could be used. Excluding eddies with little flux contribution increases the concentration differences more effectively than a linear deadband with larger deadband width and is recommended for scalars where the detection limit is of concern (Vogl et al., 2021). However, the hyperbolic deadband comes with additional challenges as the threshold for not only vertical wind velocity fluctuations but also CO2 mixing ratio fluctuations would need to be dynamically adjusted. The signal-to-noise ratio could also be increased if multiple graphite targets were to be analyzed or if measurements were to be taken closer to a particular source. However, obtaining and analyzing multiple flask samples would require major changes in the setup of the flask sampler and would multiply the cost and instrument time accordingly. In addition, analyzing turbulent fluxes on a neighborhood scale requires tall-tower measurements within the inertial sublayer. Besides limitations in the availability of the corresponding infrastructure, reducing the measurement height was therefore not an option for our study.
The next step in the process will be to derive the actual ffCO2 fluxes using EC-based total CO2 fluxes and assuming scalar similarity to estimate the β coefficient for each sampling period individually (Eq. 2). This will allow for a fully independent time-resolved evaluation of the Zurich emission inventories, taking into account the changing turbulent flux footprints during each of the sampling intervals. Furthermore, the analysis of CO and other species in the updraft and downdraft samples along with 14CO2 will provide guidance regarding the use of co-emitted species for partitioning total CO2 fluxes into fossil and non-fossil components.

Figure A1Photo of the REA flask sampling system as installed in the control room below the tower at Zurich–Hardau. (1) Screen with flask sampler software. (2) Connection to the CR6 data logger. (3) Intake lines. (4) Loop system. (5) Buffers. (6) Air dryer. (7) Flasks.
B1 Contamination test 1
To test for a potential contamination of the samples during the sampling process, e.g., due to a leak in the line or at the valves or due to surface effects with the surfaces of tubes and buffers, the REA system was set up in the laboratory of the ICOS FCL in Jena. For practical reasons, the IRGASON and the fast-response valves were placed in the same room as the buffers and the flask sampler but were connected through two > 50 m long Synflex tubes. A reference gas cylinder with 407.71 ± 0.01 ppm CO2 was connected to a custom dew point generator. In this way, the humidity of the gas was comparable to that of typical ambient air (dew point of approximately 12 °C), thus avoiding increased or reduced surface effects along the tube and in the buffers, which could result from a change in water availability (Zhao et al., 2020). The humidified gas was then directed through a three-way valve to the solenoid valves and sampled into the respective buffers according to a wind signal from the IRGASON, artificially generated through a fan. From there, the gas was transferred into the flask sampler as during a regular REA run. The test was repeated five times, with slightly different setups as listed in Table B1. The results are given in Fig. 4.
B2 Contamination test 2: memory effect test
Although the first set of contamination tests (Sect. B1) had already shown that the concentration of the sample is not significantly changed by the sampling process, a second experiment was performed to quantify the influence of the previous sampling on the flask concentration. For this purpose, a cylinder with a known gas concentration and pure nitrogen were alternately connected to the loop systems via an approximately 80 m long Synflex tube with an 8 mm outer diameter. As in first contamination tests (Sect. B1), the dry gas from the tanks was humidified through a custom dew point generator to a dew point of approximately 12 °C, comparable to normal sampling conditions of ambient (humid) air. All valves were opened, and one buffer (always buffer 4) was filled six times to approximately 1.2 bar. The cylinder gas was then transferred into the flask sampler, and the buffer was evacuated again (residual buffer pressure of about 0.6 mbar). The CO2 concentrations measured in the respective flasks are given in Table B2.
Table B2CO2 concentrations of the flasks compared to the reference gas cylinder that was connected to the inlets before and after sampling pure nitrogen. The results are given in chronological order of the fillings.

While the CO2 concentrations of flasks 1 and 4 agreed with the cylinder gas within 1.5σ, flasks 3 and 6, collected after flushing the system with nitrogen, i.e., after an intentionally chosen large CO2 change of 405.7 ppm, deviated by more than 1 ppm. Assuming that the bias in the flask concentration cmeas − cref (measured minus reference concentration) depends linearly on the concentration difference between the current and the previous sample cref − , it can therefore be expected that CO − CO = . As an uncertainty, half of the difference between the two measurements was taken. Based on the IRGASON-derived CO2 estimates of the updraft and downdraft samples of all REA sampling periods during the Zurich campaign (702 in total), the mean concentration difference between subsequent buffer fillings was estimated to be 0 ± 52 ppm (assuming that buffer sets 1 and 2 are always used alternately). Consequently, there is no systematic bias in absolute flask concentrations, while the estimated uncertainty is 0.17 ppm.
Since the updraft and downdraft samples are usually affected in a similar way, the effect on concentration differences between them is expected to be even smaller. Multiplying the mean difference between subsequent concentration differences between updraft and downdraft sampling (estimated again from the IRGASON data) of 0 ± 7 ppm with the 0.28 ± 0.05 % derived from the laboratory experiments (Table B2) results in an estimated uncertainty of about 0.02 ppm. Again, there is no systematic bias.
B3 Rinse time measurements
In the field, the rinse time, i.e., the time between the start (end) of the sampling period and the opening (closing) of the valves at the buffers, is calculated from the estimated pump speed and the dimensions of the intake lines according to Eq. (10). To test the validity of this approach and to quantify the associated uncertainty, a cylinder with pure CO2 was connected at the normally closed position of a three-way valve installed in front of the fast-response valves. The loop system was in sampling mode (pump on, three-way valves in the loop systems connecting the pump with the outflow). Then, the three-way valve was opened to the CO2 cylinder for 200 ms, injecting a short CO2 pulse into the line. This pulse was detected at the outflow of the REA sampler by a CO2 sensor (SprintIR-WF-20, Gas Sensing Solutions Ltd, Cumbernauld, UK). The travel time of the CO2 pulse was thus estimated as the time between the opening of the three-way valve and the start of the detected spike. The experiment was repeated several times and for different tubes (see Table B3). The measurements were then compared to the values estimated from the length and inner radius of the tube and the flow rate measured at the outflow of the REA sampler. As shown in Table B3, both values agreed within less than 1σ.
Table B3Length lt and inner diameter rt of the tubes connecting the REA inlets with the loop systems. Together with the pump flow velocity qpump (flow rate measured at the outflow), the travel time from the inlet to the outflow was estimated according to Eq. (10) and compared with , which was measured with a CO2 pulse.

The experiment was also done when the REA valves opened or closed once per second and were therefore only open half the time. As expected, the time until the signal was detected was approximately twice as long as when the valves were permanently open. This suggests that the concept of the rinse time also works well at the start of an REA run when the valves are already switching according to the wind signal.
For uncertainty estimation and quality control, the expected CO2 differences between the updraft and downdraft samples collected in Zurich were calculated from the high-frequency in situ measurements of the IRGASON and the MGA7. This required the synchronization between IRGASON and MGA7 data and the start and end times of the REA runs, despiking of the 20 Hz time series according to Mauder et al. (2013), and conversion of measured gas densities into dry molar fractions. The high-frequency measurements were then averaged over the time periods when the updraft and/or downdraft valve was open and air was collected using the 20 Hz REA flags (Sect. 4.1).
C1 Conversion to dry molar fractions
Flask “concentrations” measured at the gas chromatograph are given in moles of gas per mole of dry air, i.e., molar mixing ratio. To compare them with in situ measurements, the 20 Hz humid air mole fractions χgas recorded by the MGA7 and the mole densities dgas from the IRGASON were converted into dry air molar mixing ratios as shown below (compare to Mauder et al., 2021; EddyPro Software version 7.0 , 2023).
-
MGA7:
-
IRGASON:
va = is the ambient air molar volume. Following Helbig et al. (2016), the ambient air temperature Ta is derived from the fast-response ultrasonic temperature corrected for humidity effects (Schotanus et al., 1983):
For calculating the specific humidity q, the slow-response air temperature T measured by a co-located EC100 thermistor at a resolution of 1 s is used.
The various variables are described in Table C1. Note that, for flux calculations in EddyPro, the raw, high-frequency measurements were not converted from molar densities into mixing ratios, but the Webb–Pearman–Leuning (WPL) density corrections Webb et al. (1980) were applied.
C2 ΔCO2 estimates from IRGASON and MGA7 measurements
To calculate the expected CO2 concentration differences between the updraft and downdraft sample pairs collected in Zurich, the synchronized, despiked, and dry-molar-fraction-converted 20 Hz CO2 measurements were averaged over the periods where, according to the 20 Hz REA flags, air should have been collected in the updraft and/or downdraft reservoir. For the IRGASON data, four sampling periods with low correlation between the CO2 of the IRGASON and the MGA7 (Pearson correlation coefficient of < 0.5) due to the low CO2 signal strength of the IRGASON (< 90 %) were discarded.
Table C1Description of variables used in the conversion of mole fractions and mole densities to molar mixing ratios.

C3 ΔCO2 with delayed collection of air
To estimate the potential bias and uncertainty caused by a certain time lag between the wanted and the actual collection of air, the CO2 differences between REA sample pairs were calculated by shifting the 20 Hz REA flags forward in time and then averaging over the periods where the updraft or downdraft valves were open. These values were then compared to the ΔCO2 estimates calculated without time lag (Sect. C2). Table C2 shows the statistics of the differences in ΔCO2 with and without the time lag for the 74 REA sampling periods in Zurich where MGA7 data are available. The increasingly negative mean ΔCO2 differences show that ΔCO2 is systematically reduced when the collection of air is delayed. This is to be expected since, in this case, air is collected when the vertical wind is within the deadband and CO2 fluctuations are usually smaller than outside the deadband. With a time lag of 500 ms, as is expected in the Zurich setup, ΔCO2 is, on average, 0.04 ± 0.06 ppm smaller than in the ideal case without a delay; 0.06 ppm is therefore considered to be the mean uncertainty due to a 500 ms delay in the collection of air. For 75 % of the REA runs, the change in ΔCO2 is less than 0.06 ppm, whereas the maximum simulated difference is 0.29 ppm. However, there is no systematic bias observed in the comparison between flasks and in situ (Fig. 6).
C4 ΔCO2 with incorrect rinse time
Uncertainty in the time an air parcel needs to travel from the inlet to the buffers (about 2 s in Zurich) can cause sampling of unwanted air and loss of wanted air at the beginning and end of an REA run. If the rinse time were to be longer than the actual travel time by Δtr (s), the first Δtr seconds of sampled air would be lost, while air from Δtr seconds after the intended REA run end would be sampled. Similarly, if the rinse time were to be shorter than the actual travel time by Δtr, unwanted air remaining in the intake lines would be sampled into the buffers, while the sample air from the last Δtr seconds in sampling mode would be lost. This was simulated by discarding and adding 20 Hz CO2 measurements from the corresponding time periods and comparing the estimated concentration differences between updraft and downdraft samples (ΔCO) with the (ideal) estimates where Δtr = 0 s. The statistics from the 74 REA sampling periods in Zurich with MGA7 data are shown in Table C3. The mean difference between ΔCO2 with and without error in the rinse time is 0.001 ppm and so is well below the CO2 measurement uncertainty of the flask samples. For individual REA runs, however, a 2 s overestimation or underestimation of the rinse time can change ΔCO2 by up to 0.5 ppm. The standard deviation of about 0.01 ppm with Δtr = ± 2 s is considered to be the ΔCO2 uncertainty of the Zurich samples due to the 2 s uncertainty in the rinse time. The fact that this standard deviation and the median and maximum differences are about twice as large when Δtr = ± 4 s shows that this uncertainty contribution is site specific.
Table C3ΔCO2 simulated from MGA7 in situ measurements assuming different errors Δtr in the rinse time. Positive or negative Δtr indicate whether the rinse time was overestimated or underestimated. Given are the mean and standard deviation (SD), as well as the 0.25th, 0.5th, 0.75th, and 1st quantiles of the differences in relation to the ideal case without Δtr = 0 for the Zurich REA runs (n = 74).

C5 ΔCO2 with variable sampling rate
Under real sampling conditions, a certain variability of the flow rate and, thus, an inhomogeneous weighting of the concentration over time are to be expected, adding uncertainty into the ΔCO2 flask measurements. To estimate this uncertainty for each REA flask pair, the CO2 concentration differences between updraft and downdraft samples were simulated from the high-frequency CO2 measurements of the MGA7 or, if not available, the IRGASON (compare to Appendix C2) but this time with different weightings for each 20 Hz measurement. A total of 103 weighting functions were therefore calculated by interpolating the 103 flow rate time series (temporal resolution of 5 s) recorded by the two mass flow controllers in the loop systems during the Zurich REA runs in relation to the 20 Hz CO2 in situ measurements of each REA run. This resulted in 103 ΔCO2 estimates for each REA run. The standard deviations of these estimates were then considered to be the ΔCO2 uncertainty due to flow rate variability. The mean uncertainty of the 103 REA runs was 0.03 ± 0.05 ppm. As shown in Fig. C1, the magnitude depends on the CO2 variability of the ambient air. In the extreme case of a standard deviation of the ambient CO2 of more than 50 ppm, the estimated uncertainty can be as high as 0.38 ppm. On the other hand, if CO2 is approximately constant, the weighting does not matter. In the analysis of ΔCO2, e.g., in the comparison of flasks with in situ measurements (Sect. 5.3), it is therefore recommended to consider the uncertainty contributions for each REA sample individually. For the estimation of ΔffCO2, on the other hand, the uncertainty due to flow rate variability is still negligible compared to the 14C measurement uncertainty (Appendix D).
As mentioned in Sect. 3, , , and Δcnf = , which are needed to calculate ΔffCO2, are not known but are estimated as given in Table 1. To illustrate the effects of each variable on the final ΔffCO2 estimate, Fig. D1 shows the differences between ΔffCO2 calculated using the values in Table 1 and ΔffCO2 when one of the variables is changed. For the latter calculation, values over the maximum plausible range (based on and Δcmeas) and uncertainties close to mean measurement uncertainties were chosen (2 ‰ for and and 1 ppm for Δcnf). Therefore, Δ(ΔffCO2) plotted on the y axes represents the absolute error that would be made with the current assumptions if the “true” values were those given on the x axes.
For all samples, higher and Δcnf lead to higher ΔffCO2 estimates. Consequently, with the assumptions of = 9 ± 16 ‰ and Δcnf = 5 ± 5 ppm, ΔffCO2 would be overestimated or underestimated if the actual values were smaller or larger. The magnitude of the effect depends on the difference and is therefore largest for winter samples.
The effect of is sample-specific because the reference values are different for each sample. However, for most samples, ΔffCO2 decreases with increasing .

Figure D1Impact of different assumptions for Δ14 values of CO2 differences between up and down flasks due to respiration and biofuels (non-fossil, nf) and photosynthesis (photo) and the non-fossil signal Δcnf = based on estimates of ΔffCO2 = . For each REA sample, the difference in ΔffCO2 with respect to the reference settings used in this study is shown, where = 10 ± 16 ‰, Δcnf = 5 ± 5 ppm, and = ± 10 ‰. The light-blue ribbons indicate the 1σ uncertainty range due to measurement uncertainties.

Figure D2Contributions to the absolute uncertainty of ΔffCO2 estimates (Eq. 9, with assumptions as given in Table 1) for each REA sample, collected between July 2022 and March 2023. Δ14 measurement uncertainties and an estimated non-fossil CO2 difference between updraft and downdraft samples of 5 ± 5 ppm are the main contributors. The ΔCO2 uncertainties due to the sampling process and the laboratory analysis, as well as the assumptions regarding the Δ14C signatures of photosynthesis, respiration, and biofuels (shown here in black), are negligible.
Varying the variables over the entire reasonable range, the differences can be as large as 2 ppm and thus of the same order of magnitude as ΔffCO2 itself. Nevertheless, due to the large 14CO2 measurement uncertainty (compare to light-blue area in Fig. D1), the deviations from the previous ΔffCO2 estimates are not significant.
The fact that the uncertainty of ΔffCO2 estimates is dominated by the 14CO2 measurement uncertainty can also be seen in Fig. D2. It shows the contributions of each variable to the ΔffCO2 uncertainty in ppm for each REA sample pair. The contributions of the ΔCO2 uncertainty (including all aspects summarized in Table 3) and the assumptions regarding and are summarized as “other assumptions and measurements” (black) and are negligible for all samples. As discussed before, the impact of Δcnf (orange) is especially large for the winter samples but is still secondary compared to the Δ14 measurement uncertainty (blue). The latter has been reduced from approximately 1.3 to 1.0 ppm as the graphite targets were measured with a new accelerator mass spectrometer starting January 2023. This improved the mean Δ14 precision from 2.1 ± 0.3 ‰ to 1.6 ± 0.2 ‰. Overall, the analyses show that the choice of , , and Δcnf has little effect on the ΔffCO2 estimate due to the current Δ14 measurement precision.
The logger program and the data supporting this publication are provided at https://doi.org/10.5281/zenodo.13926680 (Kunz et al., 2024).
IL and SH developed the idea of ffCO2 REA measurements. VL, LB, RK, and ME conducted initial proof-of-principle tests and designed, built, and tested the REA flask sampler together with AK. AC and JH performed numerical proof-of-concept simulations and developed the logger code. JH, RH, AC, and SS operated the IRGASON and MGA7 in Zurich and provided data for sample selection and simulations of flask concentration differences. AK, SH, and IL collected and selected the REA samples during the Zurich campaign. PR exchanged the flasks and maintained the REA sampler in Zurich. XG and JDC were responsible for the flask measurements at the FCL and the CRL, and AJ and SP were responsible for the respective data processing and quality control. AK performed the analysis, created the figures, and prepared the paper with contributions from all of the co-authors.
The contact author has declared that none of the authors has any competing interests.
Publisher’s note: Copernicus Publications remains neutral with regard to jurisdictional claims made in the text, published maps, institutional affiliations, or any other geographical representation in this paper. While Copernicus Publications makes every effort to include appropriate place names, the final responsibility lies with the authors.
The authors have received funding from ICOS Cities, a.k.a. the Pilot Applications in Urban Landscapes – Towards integrated city observatories for greenhouse gases (PAUL) project, from the European Union's Horizon 2020 research and innovation program under grant agreement no. 101037319. Financial support from ICOS Switzerland (ICOS-CH) Phase 3 (Swiss National Science Foundation, grant no. 20FI20_198227) is also acknowledged. Additional support was provided by internal funds and staff at the Universities of Heidelberg and Freiburg and the Max Planck Institute for Biogeochemistry in Jena. We thank Lukas Emmenegger (EMPA, Switzerland) for negotiating and managing the installation of the Zurich Hardau site, Roland Vogt (University of Basel, Switzerland) for installing the eddy covariance (IRGASON) system in Zurich, Sophie Emberger (ETH Zurich, Switzerland) for maintaining the eddy covariance site, and Felix Baab (University of Freiburg, Germany) for constructing the REA inlets and tower valves. In addition, we gratefully thank Steffen Knabe (ICOS FCL, Germany) and the staff of the ICOS FCL and the ICOS CRL for measuring the test and REA flasks. When comparing the REA flask concentrations with the IRGASON CO2 measurements, we appreciated the fruitful discussions with Ivan Bogoev (Campbell Scientific Inc., USA, Logan).
This research has been supported by the EU Horizon 2020 (grant no. 101037319).
This open-access publication was funded by the University of Freiburg.
This paper was edited by Daniela Famulari and reviewed by two anonymous referees.
Baker, J. M., Norman, J. M., and Bland, W. L.: Field-scale application of flux measurement by conditional sampling, Agr. Forest Meteorol., 62, 31–52, https://doi.org/10.1016/0168-1923(92)90004-N, 1992. a, b
Bjorkegren, A. B., Grimmond, C., Kotthaus, S., and Malamud, B. D.: CO2 emission estimation in the urban environment: Measurement of the CO2 storage term, Atmos. Environ., 122, 775–790, https://doi.org/10.1016/j.atmosenv.2015.10.012, 2015. a
Bowling, D. R., Delany, A. C., Turnipseed, A. A., Baldocchi, D. D., and Monson, R. K.: Modification of the relaxed eddy accumulation technique to maximize measured scalar mixing ratio differences in updrafts and downdrafts, J. Geophys. Res., 104, 9121–9133, https://doi.org/10.1029/1999JD900013, 1999. a
Businger, J. A. and Oncley, S. P.: Flux Measurement with Conditional Sampling, J. Atmos. Ocean. Tech., 7, 349–352, https://doi.org/10.1175/1520-0426(1990)007<0349:FMWCS>2.0.CO;2, 1990. a, b
Chanca, I.: Theoretical and experimental approaches using Δ14C for estimating system diagnostic times in the central Amazon rainforest, PhD thesis, Universidade Federal Fluminense, Brazil, https://hdl.handle.net/21.11116/0000-000A-C8A5-A (last access: 5 October 2025), 2022. a
Christen, A.: Atmospheric measurement techniques to quantify greenhouse gas emissions from cities, Urban Climate, 10, 241–260, https://doi.org/10.1016/j.uclim.2014.04.006, 2014. a
Christen, A., Grimmond, C. S. B., Moriwaki, R., Scherer, D., and Vogt, R.: Evaluating conditional sampling strategies for trace-gas flux measurements in urban environments, AMS 6th Symposium on the Urban Environment, Atlanta, GA, 27 January–3 February 2006, https://www.researchgate.net/publication/235297637_Evaluating_conditional_sampling_strategies_for_trace-gas_flux_measurements_in_urban_environments, 2006. a
Crawford, B. and Christen, A.: Spatial variability of carbon dioxide in the urban canopy layer and implications for flux measurements, Atmos. Environ., 98, 308–322, https://doi.org/10.1016/j.atmosenv.2014.08.052, 2014. a
Crawford, B. and Christen, A.: Spatial source attribution of measured urban eddy covariance CO2 fluxes, Theor. Appl. Climatol., 119, 733–755, https://doi.org/10.1007/s00704-014-1124-0, 2015. a, b, c
Crawford, B., Grimmond, C. S. B., and Christen, A.: Five years of carbon dioxide fluxes measurements in a highly vegetated suburban area, Atmos. Environ., 45, 896–905, https://doi.org/10.1016/j.atmosenv.2010.11.017, 2011. a
Desjardins, R. L.: A study of carbon-dioxide and sensible heat fluxes using the eddy correlation technique, PhD thesis, Cornell University, Ithaca, New York, USA, 1972. a
EddyPro Software version 7.0: Instruction Manual, LI-COR, Inc., https://www.licor.com/env/support/EddyPro/manuals.html (last access: 5 October 2025), 2023. a
Emad, A. and Siebicke, L.: True eddy accumulation – Part 2: Theory and experiment of the short-time eddy accumulation method, Atmos. Meas. Tech., 16, 41–55, https://doi.org/10.5194/amt-16-41-2023, 2023. a
Feigenwinter, C., Vogt, R., and Christen, A.: Eddy Covariance Measurements Over Urban Areas, in: Eddy Covariance, edited by: Aubinet, M., Vesala, T., and Papale, D., Springer, Dordrecht, 377–398, https://doi.org/10.1007/978-94-007-2351-1_16, 2012. a
Foken, T., Göckede, M., Mauder, M., Mahrt, L., Amiro, B., and Munger, W.: Post-field data quality control, in: Handbook of Micrometeorology, edited by: Lee, X., Massman, W., and Law, B., Atmospheric and Oceanographic Sciences Library, Springer, Dordrecht, 29, 181–208, https://doi.org/10.1007/1-4020-2265-4_9, ISBN 978-1-4020-2264-6, 2004. a, b
Foken, T., Leuning, R., Oncley, S. R., Mauder, M., and Aubinet, M.: Corrections and Data Quality Control, in: Eddy Covariance, edited by: Aubinet, M., Vesala, T., and Papale, D., Springer, Dordrecht, 85–131, https://doi.org/10.1007/978-94-007-2351-1_4, 2012. a, b, c
Gately, C. K. and Hutyra, L. R.: Large Uncertainties in Urban–Scale Carbon Emissions, J. Geophys. Res., 122, 11242–11260, https://doi.org/10.1002/2017JD027359, 2017. a
Grimmond, S. and Christen, A.: Flux measurements in urban ecosystems, FluxLetter - Newsletter of Fluxnet, 5, 1–8, https://fluxnet.org/wp-content/uploads/FluxLetter_Vol5_No1.pdf (last access: 5 October 2025), 2012. a
Grönholm, T., Haapanala, S., Launiainen, S., Rinne, J., Vesala, T., and Rannik, U.: The dependence of the β coefficient of REA system with dynamic deadband on atmospheric conditions, Environ. Pollut., 152, 597–603, https://doi.org/10.1016/j.envpol.2007.06.071, 2008. a
Guo, M., Song, W., and Buhain, J.: Bioenergy and biofuels: History, status, and perspective, Renew. Sust. Energ. Rev., 42, 712–725, https://doi.org/10.1016/j.rser.2014.10.013, 2015. a
Hardiman, B. S., Wang, J. A., Hutyra, L. R., Gately, C. K., Getson, J. M., and Friedl, M. A.: Accounting for urban biogenic fluxes in regional carbon budgets, Sci. Total Environ., 592, 366–372, https://doi.org/10.1016/j.scitotenv.2017.03.028, 2017. a
Helbig, M., Wischnewski, K., Gosselin, G. H., Biraud, S. C., Bogoev, I., Chan, W. S., Euskirchen, E. S., Glenn, A. J., Marsh, P. M., Quinton, W. L., and Sonnentag, O.: Addressing a systematic bias in carbon dioxide flux measurements with the EC150 and the IRGASON open-path gas analyzers, Agr. Forest Meteorol., 228–229, 349–359, https://doi.org/10.1016/j.agrformet.2016.07.018, 2016. a, b, c
Kanda, M., Inagaki, A., Miyamoto, T., Gryschka, M., and Raasch, S.: A New Aerodynamic Parametrization for Real Urban Surfaces, Bound.-Lay. Meteorol., 148, 357–377, https://doi.org/10.1007/s10546-013-9818-x, 2013. a
Kellett, R., Christen, A., Coops, N. C., van der Laan, M., Crawford, B., Tooke, T. R., and Olchovski, I.: A systems approach to carbon cycling and emissions modeling at an urban neighborhood scale, Landscape Urban Plan., 110, 48–58, https://doi.org/10.1016/j.landurbplan.2012.10.002, 2013. a, b
Kunz, A.-K., Borchardt, L., Christen, A., Della Coletta, J., Eritt, M., Gutiérrez, X., Hashemi, J., Hilland, R., Jordan, A., Kneißl, R., Legendre, V., Levin, I., Preunkert, S., Rubli, P., Stagakis, S., and Hammer, S.: Flask data and logger program supporting the manuscript “A relaxed eddy accumulation flask sampling system for 14C-based partitioning of fossil and non-fossil CO2 fluxes”, Zenodo [code and data set], https://doi.org/10.5281/zenodo.13926680, 2024. a
Lauvaux, T., Gurney, K. R., Miles, N. L., Davis, K. J., Richardson, S. J., Deng, A., Nathan, B. J., Oda, T., Wang, J. A., Hutyra, L., and Turnbull, J.: Policy-Relevant Assessment of Urban CO2 Emissions, Environ. Sci. Technol., 54, 10237–10245, https://doi.org/10.1021/acs.est.0c00343, 2020. a
Lehmuskoski, J., Vasama, H., Hämäläinen, J., Hokkinen, J., Kärkelä, T., Heiskanen, K., Reinikainen, M., Rautio, S., Hirvelä, M., and Genoud, G.: On-Line Monitoring of Radiocarbon Emissions in a Nuclear Facility with Cavity Ring-Down Spectroscopy, Anal. Chem., 93, 16096–16104, https://doi.org/10.1021/acs.analchem.1c03814, 2021. a
Levin, I., Kromer, B., Schmidt, M., and Sartorius, H.: A novel approach for independent budgeting of fossil fuel CO2 over Europe by 14CO2 observations, Geophys. Res. Lett., 30, 2194, https://doi.org/10.1029/2003GL018477, 2003. a, b
Levin, I., Naegler, T., Kromer, B., Diehl, M., Francey, R. J., Gomez-Pelaez, A. J., Steele, L. P., Wagenbach, D., Weller, R., and Worthy, D. E.: Observations and modelling of the global distribution and long-term trend of atmospheric 14CO2, Tellus B, 62, 26–46, https://doi.org/10.1111/j.1600-0889.2009.00446.x, 2010. a
Levin, I., Karstens, U., Eritt, M., Maier, F., Arnold, S., Rzesanke, D., Hammer, S., Ramonet, M., Vítková, G., Conil, S., Heliasz, M., Kubistin, D., and Lindauer, M.: A dedicated flask sampling strategy developed for Integrated Carbon Observation System (ICOS) stations based on CO2 and CO measurements and Stochastic Time-Inverted Lagrangian Transport (STILT) footprint modelling, Atmos. Chem. Phys., 20, 11161–11180, https://doi.org/10.5194/acp-20-11161-2020, 2020. a, b, c
Lux, J. T.: A new target preparation facility for high precision AMS measurements and strategies for efficient 14CO2 sampling, PhD thesis, University of Heidelberg, Germany, https://archiv.ub.uni-heidelberg.de/volltextserver/24767/1/DissertationJohannesLuxFinal.pdf (last access: 5 October 2025), 2018. a, b
Maier, F., Levin, I., Gachkivskyi, M., Rödenbeck, C., and Hammer, S.: Estimating regional fossil fuel CO2 concentrations from 14CO2 observations: challenges and uncertainties, Philos. T. R. Soc. A, 381, 20220203, https://doi.org/10.1098/rsta.2022.0203, 2023. a, b, c
Maier, F., Levin, I., Conil, S., Gachkivskyi, M., Denier van der Gon, H., and Hammer, S.: Uncertainty in continuous ΔCO-based ΔffCO2 estimates derived from 14C flask and bottom-up ΔCO ΔffCO2 ratios, Atmos. Chem. Phys., 24, 8205–8223, https://doi.org/10.5194/acp-24-8205-2024, 2024a. a
Maier, F., Rödenbeck, C., Levin, I., Gerbig, C., Gachkivskyi, M., and Hammer, S.: Potential of 14C-based vs. ΔCO-based ΔffCO2 observations to estimate urban fossil fuel CO2 (ffCO2) emissions, Atmos. Chem. Phys., 24, 8183–8203, https://doi.org/10.5194/acp-24-8183-2024, 2024b. a
Mauder, M., Cuntz, M., Drüe, C., Graf, A., Rebmann, C., Schmid, H. P., Schmidt, M., and Steinbrecher, R.: A strategy for quality and uncertainty assessment of long-term eddy-covariance measurements, Agr. Forest Meteorol., 169, 122–135, https://doi.org/10.1016/j.agrformet.2012.09.006, 2013. a
Mauder, M., Foken, T., Aubinet, M., and Ibrom, A.: Eddy-Covariance Measurements, in: Springer Handbook of Atmospheric Measurements, edited by: Foken, T., Springer Handbooks, Springer, Cham, 1473–1504, https://doi.org/10.1007/978-3-030-52171-4_55, ISBN 978-3-030-52170-7,2021. a, b
McCartt, A. D. and Jiang, J.: Room-Temperature Optical Detection of 14CO2 below the Natural Abundance with Two-Color Cavity Ring-Down Spectroscopy, 7, 3258–3264, https://doi.org/10.1021/acssensors.2c01253, 2022. a
Miller, J. B., Lehman, S. J., Verhulst, K. R., Miller, C. E., Duren, R. M., Yadav, V., Newman, S., and Sloop, C. D.: Large and seasonally varying biospheric CO2 fluxes in the Los Angeles megacity revealed by atmospheric radiocarbon, P. Natl. Acad. Sci. USA, 117, 26681–26687, https://doi.org/10.1073/pnas.2005253117, 2020. a
Milne, R., Beverland, I. J., Hargreaves, K., and Moncrieff, J. B.: Variation of the β coefficient in the relaxed eddy accumulation method, Bound.-Lay. Meteorol., 93, 211–225, https://doi.org/10.1023/A:1002061514948, 1999. a
Naegler, T. and Levin, I.: Observation–based global biospheric excess radiocarbon inventory 1963–2005, J. Geophys. Res., 114, D17302, https://doi.org/10.1029/2008JD011100, 2009a. a
Naegler, T. and Levin, I.: Biosphere–atmosphere gross carbon exchange flux and the δ13CO2 and Δ14CO2 disequilibria constrained by the biospheric excess radiocarbon inventory, J. Geophys. Res., 114, D17303, https://doi.org/10.1029/2008JD011116, 2009b. a
Palonen, V., Pumpanen, J., Kulmala, L., Levin, I., Heinonsalo, J., and Vesala, T.: Seasonal and Diurnal Variations in Atmospheric and Soil Air 14CO2 in a Boreal Scots Pine Forest, Radiocarbon, 60, 283–297, https://doi.org/10.1017/RDC.2017.95, 2018. a
Pattey, E., Desjardins, R. L., and Rochette, P.: Accuracy of the relaxed eddy-accumulation technique, evaluated using CO2 flux measurements, Bound.-Lay. Meteorol., 66, 341–355, https://doi.org/10.1007/BF00712728, 1993. a, b
Rinne, J., Ammann, C., Pattey, E., Paw U, K. T., and Desjardins, R. L.: Alternative Turbulent Trace Gas Flux Measurement Methods, in: Springer Handbook of Atmospheric Measurements, edited by: Foken, T., Springer Handbooks, Springer, Cham, 1505–1530, https://doi.org/10.1007/978-3-030-52171-4_56, ISBN 978-3-030-52170-7, 2021. a, b, c, d, e
Schotanus, P., Nieuwstadt, F. T. M., and de Bruin, H. A. R.: Temperature measurement with a sonic anemometer and its application to heat and moisture fluxes, Bound.-Lay. Meteorol., 26, 81–93, https://doi.org/10.1007/BF00164332, 1983. a
Seto, K. C., Dhakal, S., Bigio, A., Blanco, H., Delgado, G. C., Dewar, D., Huang, L., Inaba, A., Kansal, A., Lwasa, S., McMahon, J. E., Müller, D. B., Murakami, J., Nagendra, H., and Ramaswami, A.: Human Settlements, Infrastructure and Spatial Planning, in: Climate Change 2014: Mitigation of Climate Change: Contribution of Working Group III to the Fifth Assessment Report of the Intergovernmental Panel on Climate Change, edited by: Edenhofer, O., Pichs-Madruga, R., Sokona, Y., Farahani, E., Kadner, S., K, Seyboth, Adler, A., Baum, I., Brunner, S., Eickemeier, P., Kriemann, B., Savolainen, J., Schlömer, S., von Stechow, C., Zwickel, T., and Minx, J. C., Cambridge University Press, Cambridge University Press, Cambridge, United Kingdom and New York, NY, USA, https://doi.org/10.1017/CBO9781107415416.018, 2014. a
Siebicke, L. and Emad, A.: True eddy accumulation trace gas flux measurements: proof of concept, Atmos. Meas. Tech., 12, 4393–4420, https://doi.org/10.5194/amt-12-4393-2019, 2019. a, b
Stagakis, S., Feigenwinter, C., Vogt, R., Brunner, D., and Kalberer, M.: A high-resolution monitoring approach of urban CO2 fluxes. Part 2 – surface flux optimisation using eddy covariance observations, Sci. Total Environ., 903, 166035, https://doi.org/10.1016/j.scitotenv.2023.166035, 2023. a, b
Stuiver, M. and Polach, H. A.: Discussion Reporting of 14C Data, Radiocarbon, 19, 355–363, https://doi.org/10.1017/S0033822200003672, 1977. a, b
Stull, R. B. (Ed.): An Introduction to Boundary Layer Meteorology, in: Atmospheric and Oceanographic Sciences Library, Vol. 13, Springer, Dordrecht, https://doi.org/10.1007/978-94-009-3027-8, 1988. a
Super, I., Dellaert, S. N. C., Visschedijk, A. J. H., and Denier van der Gon, H. A. C.: Uncertainty analysis of a European high-resolution emission inventory of CO2 and CO to support inverse modelling and network design, Atmos. Chem. Phys., 20, 1795–1816, https://doi.org/10.5194/acp-20-1795-2020, 2020. a
Tans, P. and Zellweger, C. (Eds.): GAW Report No. 213, 17th WMO/IAEA Meeting on carbon dioxide, other greenhouse gases and related tracers measurement techniques (GGMT-2013), Beijing, China, 10–13 June 2013, https://www.uncclearn.org/wp-content/uploads/library/gaw_213_en.pdf (last access: 5 October 2025), 2014. a
Turnbull, J. C., Sweeney, C., Karion, A., Newberger, T., Lehman, S. J., Tans, P. P., Davis, K. J., Lauvaux, T., Miles, N. L., Richardson, S. J., Cambaliza, M. O., Shepson, P. B., Gurney, K., Patarasuk, R., and Razlivanov, I.: Toward quantification and source sector identification of fossil fuel CO2 emissions from an urban area: Results from the INFLUX experiment, J. Geophys. Res., 120, 292–312, https://doi.org/10.1002/2014JD022555, 2015. a
Turnbull, J. C., Graven, H., and Krakauer, N. Y.: Radiocarbon in the Atmosphere, in: Radiocarbon and Climate Change, edited by: Schuur, E. A., Druffel, E. R. M., and Trumbore, S. E., Springer, Cham, 83–137, https://doi.org/10.1007/978-3-319-25643-6_4, 2016. a, b, c
Vogl, T., Hrdina, A., and Thomas, C. K.: Choosing an optimal β factor for relaxed eddy accumulation applications across vegetated and non-vegetated surfaces, Biogeosciences, 18, 5097–5115, https://doi.org/10.5194/bg-18-5097-2021, 2021. a
Webb, E. K., Pearman, G. I., and Leuning, R.: Correction of flux measurements for density effects due to heat and water vapour transfer, Q. J. Roy. Meteor. Soc., 106, 85–100, https://doi.org/10.1002/qj.49710644707, 1980. a
WMO: IG3IS Urban Greenhouse Gas Emission Observation and Monitoring Good Research Practice Guidelines, GAW Report No. 275, https://library.wmo.int/idurl/4/58055 (last access: 5 October 2025), 2022. a
Wu, K., Lauvaux, T., Davis, K. J., Deng, A., Coto, I. L., Gurney, K. R., and Patarasuk, R.: Joint inverse estimation of fossil fuel and biogenic CO2 fluxes in an urban environment: An observing system simulation experiment to assess the impact of multiple uncertainties, Elementa: Science of the Anthropocene, 6, 17, https://doi.org/10.1525/elementa.138, 2018. a
Wu, K., Davis, K. J., Miles, N. L., Richardson, S. J., Lauvaux, T., Sarmiento, D. P., Balashov, N. V., Keller, K., Turnbull, J., Gurney, K. R., Liang, J., and Roest, G.: Source decomposition of eddy-covariance CO2 flux measurements for evaluating a high-resolution urban CO2 emissions inventory, Environ. Res. Lett., 17, 074035, https://doi.org/10.1088/1748-9326/ac7c29, 2022. a, b, c
Wyngaard, J. C.: Atmospheric Turbulence, Annu. Rev. Fluid Mech., 24, 205–234, https://doi.org/10.1146/annurev.fl.24.010192.001225, 1992. a
Zhao, J., Deng, S., Zhao, L., Yuan, X., Du, Z., Li, S., Chen, L., and Wu, K.: Understanding the effect of H2O on CO2 adsorption capture: mechanism explanation, quantitative approach and application, Sustainable Energy Fuels, 4, 5970–5986, https://doi.org/10.1039/D0SE01179G, 2020. a
- Abstract
- Introduction
- The relaxed eddy accumulation method
- 14C-based fossil fuel CO2 estimation
- Setup of the REA system
- Quality control of REA flask samples
- ΔCO2 partitioning
- Conclusions
- Appendix A: Setup of the REA flask sampling system
- Appendix B: Laboratory tests
- Appendix C: ΔCO2 simulations
- Appendix D: ΔffCO2 uncertainties
- Code and data availability
- Author contributions
- Competing interests
- Disclaimer
- Acknowledgements
- Financial support
- Review statement
- References
- Abstract
- Introduction
- The relaxed eddy accumulation method
- 14C-based fossil fuel CO2 estimation
- Setup of the REA system
- Quality control of REA flask samples
- ΔCO2 partitioning
- Conclusions
- Appendix A: Setup of the REA flask sampling system
- Appendix B: Laboratory tests
- Appendix C: ΔCO2 simulations
- Appendix D: ΔffCO2 uncertainties
- Code and data availability
- Author contributions
- Competing interests
- Disclaimer
- Acknowledgements
- Financial support
- Review statement
- References