Good performance of low-cost fast-response carbon dioxide sensor based on intercomparison with the standard eddy-covariance system

Üllar Rannik, Ivan Mammarella, Timo Vesala, Pirkko Väkimies, Hilkka Heiskari-Tuohiniemi, and Mika Korkiakoski

Viewed

Total article views: 1,624 (including HTML, PDF, and XML)
HTML PDF XML Total Supplement BibTeX EndNote
954 535 135 1,624 238 115 167
  • HTML: 954
  • PDF: 535
  • XML: 135
  • Total: 1,624
  • Supplement: 238
  • BibTeX: 115
  • EndNote: 167
Views and downloads (calculated since 05 Mar 2026)
Cumulative views and downloads (calculated since 05 Mar 2026)

Viewed (geographical distribution)

Total article views: 1,624 (including HTML, PDF, and XML) Thereof 1,603 with geography defined and 21 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 
Latest update: 28 Sep 2026
Download
Short summary
We present a case study that aimed to develop an inexpensive yet precise fast-response carbon dioxide and water vapour sensor. A working prototype was developed and field-tested against a scientific eddy covariance setup. Attention was paid to internal auto-calibration procedure such that signal drifting was avoided and the instrument was capable of measuring fluxes with high correlation relative to the reference setup.