Articles | Volume 19, issue 18
https://doi.org/10.5194/amt-19-6159-2026
https://doi.org/10.5194/amt-19-6159-2026
Research article
 | 
28 Sep 2026
Research article |  | 28 Sep 2026

Good performance of low-cost fast-response carbon dioxide sensor based on intercomparison with the standard eddy-covariance system

Üllar Rannik, Ivan Mammarella, Timo Vesala, Pirkko Väkimies, Hilkka Heiskari-Tuohiniemi, and Mika Korkiakoski

Viewed

Total article views: 2,202 (including HTML, PDF, and XML)
HTML PDF XML Total Supplement BibTeX EndNote
1,364 649 189 2,202 298 160 218
  • HTML: 1,364
  • PDF: 649
  • XML: 189
  • Total: 2,202
  • Supplement: 298
  • BibTeX: 160
  • EndNote: 218
Views and downloads (calculated since 05 Mar 2026)
Cumulative views and downloads (calculated since 05 Mar 2026)

Viewed (geographical distribution)

Total article views: 2,202 (including HTML, PDF, and XML) Thereof 2,181 with geography defined and 21 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 
Latest update: 18 Oct 2026
Download
Short summary
We present a case study that aimed to develop an inexpensive yet precise fast-response carbon dioxide and water vapour sensor. A working prototype was developed and field-tested against a scientific eddy covariance setup. Attention was paid to internal auto-calibration procedure such that signal drifting was avoided and the instrument was capable of measuring fluxes with high correlation relative to the reference setup.
Share