A newly identified calculation discrepancy of the Sunset semi-continuous carbon analyzer
Abstract. The Sunset semi-continuous carbon analyzer (SCCA) is an instrument widely used for carbonaceous aerosol measurement. Despite previous validation work, in this study we identified a new type of SCCA calculation discrepancy caused by the default multipoint baseline correction method. When exceeding a certain threshold carbon load, multipoint correction could cause significant total carbon (TC) underestimation. This calculation discrepancy was characterized for both sucrose and ambient samples, with two protocols based on IMPROVE (Interagency Monitoring of PROtected Visual Environments) (i.e., IMPshort and IMPlong) and one NIOSH (National Institute for Occupational Safety and Health)-like protocol (rtNIOSH). For ambient samples, the IMPshort, IMPlong and rtNIOSH protocol underestimated 22, 36 and 12% of TC, respectively, with the corresponding threshold being ~ 0, 20 and 25 μgC. For sucrose, however, such discrepancy was observed only with the IMPshort protocol, indicating the need of more refractory SCCA calibration substance. Although the calculation discrepancy could be largely reduced by the single-point baseline correction method, the instrumental blanks of single-point method were higher. The correction method proposed was to use multipoint-corrected data when below the determined threshold, and use single-point results when beyond that threshold. The effectiveness of this correction method was supported by correlation with optical data.