Articles | Volume 8, issue 8
https://doi.org/10.5194/amt-8-3407-2015
https://doi.org/10.5194/amt-8-3407-2015
Research article
 | 
24 Aug 2015
Research article |  | 24 Aug 2015

Comparison of ozone retrievals from the Pandora spectrometer system and Dobson spectrophotometer in Boulder, Colorado

J. Herman, R. Evans, A. Cede, N. Abuhassan, I. Petropavlovskikh, and G. McConville

Viewed

Total article views: 3,657 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
2,123 1,419 115 3,657 134 159
  • HTML: 2,123
  • PDF: 1,419
  • XML: 115
  • Total: 3,657
  • BibTeX: 134
  • EndNote: 159
Views and downloads (calculated since 20 Mar 2015)
Cumulative views and downloads (calculated since 20 Mar 2015)

Cited

Saved (final revised paper)

Saved (final revised paper)

Saved (preprint)

Latest update: 04 Nov 2024
Download
Short summary
Measurements of total column ozone have been obtained by the Dobson #061 spectrophotometer and the Pandora spectrometer system from the roof of the NOAA building in Boulder, Colorado. A comparison of 1 year (17 Dec 2013 to 18 Dec 2014) of ozone data shows that the two instruments are in close agreement after a separate correction was applied to the retrieved ozone data from each instrument. Good agreement was also obtained between Pandora and two satellite data sets, AURA OMI and SUOMI OMPS.