Articles | Volume 8, issue 9
https://doi.org/10.5194/amt-8-3867-2015
https://doi.org/10.5194/amt-8-3867-2015
Research article
 | 
23 Sep 2015
Research article |  | 23 Sep 2015

Comprehensive laboratory and field testing of cavity ring-down spectroscopy analyzers measuring H2O, CO2, CH4 and CO

C. Yver Kwok, O. Laurent, A. Guemri, C. Philippon, B. Wastine, C. W. Rella, C. Vuillemin, F. Truong, M. Delmotte, V. Kazan, M. Darding, B. Lebègue, C. Kaiser, I. Xueref-Rémy, and M. Ramonet

Viewed

Total article views: 4,834 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
2,651 2,051 132 4,834 117 148
  • HTML: 2,651
  • PDF: 2,051
  • XML: 132
  • Total: 4,834
  • BibTeX: 117
  • EndNote: 148
Views and downloads (calculated since 27 Apr 2015)
Cumulative views and downloads (calculated since 27 Apr 2015)

Cited

Saved (final revised paper)

Saved (preprint)

Latest update: 21 Nov 2024
Download
Short summary
We present the results of tests of CRDS instruments in the laboratory (47 instruments) and in the field (15 instruments). We demonstrate that, thanks to rigorous testing, newer models generally perform better than older models, especially in terms of reproducibility between instruments. In the field, we see the importance of individual diagnostics during the installation phase, and we show the value of calibration and target gases that assess the quality of the data.