Articles | Volume 8, issue 10
https://doi.org/10.5194/amt-8-4075-2015
https://doi.org/10.5194/amt-8-4075-2015
Research article
 | 
06 Oct 2015
Research article |  | 06 Oct 2015

H2S interference on CO2 isotopic measurements using a Picarro G1101-i cavity ring-down spectrometer

K. Malowany, J. Stix, A. Van Pelt, and G. Lucic

Viewed

Total article views: 2,888 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
1,694 1,096 98 2,888 98 104
  • HTML: 1,694
  • PDF: 1,096
  • XML: 98
  • Total: 2,888
  • BibTeX: 98
  • EndNote: 104
Views and downloads (calculated since 05 Jun 2015)
Cumulative views and downloads (calculated since 05 Jun 2015)

Cited

Saved (final revised paper)

Saved (preprint)

Latest update: 01 May 2024
Download
Short summary
Portable cavity ring-down spectrometers (e.g. G1101-i) for isotopic CO2 have an interference with elevated (higher than ambient) amounts of hydrogen sulfide (H2S). This results from the overlap of the H2S and CO2 spectral lines in the near-infrared, causing changes in both the 12CO2 and 13CO2 concentrations. This effect can be removed by reacting H2S with a metal scrub before analysis, which will facilitate the application of these instruments in H2S-rich environments (i.e. active volcanoes).