Operation of the Airmodus A11 nano Condensation Nucleus Counter at various inlet pressures and various operation temperatures, and design of a new inlet system

Juha Kangasluoma, Alessandro Franchin, Jonahtan Duplissy, Lauri Ahonen, Frans Korhonen, Michel Attoui, Jyri Mikkilä, Katrianne Lehtipalo, Joonas Vanhanen, Markku Kulmala, and Tuukka Petäjä

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement
Download
Short summary
The paper describes technical aspects of using the Airmodus A11 nCNC at various inlet pressures and how temperature selection affects the performance of the instrument. We also present a sampling box to minimize the inlet losses and make use of the instrument more convenient.