Journal cover Journal topic
Atmospheric Measurement Techniques An interactive open-access journal of the European Geosciences Union
Journal topic

Journal metrics

IF value: 3.668
IF3.668
IF 5-year value: 3.707
IF 5-year
3.707
CiteScore value: 6.3
CiteScore
6.3
SNIP value: 1.383
SNIP1.383
IPP value: 3.75
IPP3.75
SJR value: 1.525
SJR1.525
Scimago H <br class='widget-line-break'>index value: 77
Scimago H
index
77
h5-index value: 49
h5-index49
Volume 9, issue 9
Atmos. Meas. Tech., 9, 4719–4736, 2016
https://doi.org/10.5194/amt-9-4719-2016
© Author(s) 2016. This work is distributed under
the Creative Commons Attribution 3.0 License.
Atmos. Meas. Tech., 9, 4719–4736, 2016
https://doi.org/10.5194/amt-9-4719-2016
© Author(s) 2016. This work is distributed under
the Creative Commons Attribution 3.0 License.

Peer-reviewed comment 22 Sep 2016

Peer-reviewed comment | 22 Sep 2016

Automatic processing of atmospheric CO2 and CH4 mole fractions at the ICOS Atmosphere Thematic Centre

Lynn Hazan, Jérôme Tarniewicz, Michel Ramonet, Olivier Laurent, and Amara Abbaris Lynn Hazan et al.
  • Laboratoire des Sciences du Climat et de l'Environnement (LSCE/IPSL), UMR CEA-CNRS-UVSQ, Gif-sur-Yvette, France

Abstract. The Integrated Carbon Observation System Atmosphere Thematic Centre (ICOS ATC) automatically processes atmospheric greenhouse gases mole fractions of data coming from sites of the ICOS network. Daily transferred raw data files are automatically processed and archived. Data are stored in the ICOS atmospheric database, the backbone of the system, which has been developed with an emphasis on the traceability of the data processing. Many data products, updated daily, explore the data through different angles to support the quality control of the dataset performed by the principal operators in charge of the instruments. The automatic processing includes calibration and water vapor corrections as described in the paper. The mole fractions calculated in near-real time (NRT) are automatically revaluated as soon as a new instrument calibration is processed or when the station supervisors perform quality control. By analyzing data from 11 sites, we determined that the average calibration corrections are equal to 1.7 ± 0.3 µmol mol−1 for CO2 and 2.8 ± 3 nmol mol−1 for CH4. These biases are important to correct to avoid artificial gradients between stations that could lead to error in flux estimates when using atmospheric inversion techniques. We also calculated that the average drift between two successive calibrations separated by 15 days amounts to ±0.05 µmol mol−1 and ±0.7 nmol mol−1 for CO2 and CH4, respectively. Outliers are generally due to errors in the instrument configuration and can be readily detected thanks to the data products provided by the ATC. Several developments are still ongoing to improve the processing, including automated spike detection and calculation of time-varying uncertainties.

Publications Copernicus
Download
Short summary
The ATC automatically processes atmospheric greenhouse gases mole fractions of data sent daily by the ICOS network, this includes calibration and water vapor corrections. Data are stored in a database which has been developed with an emphasis on traceability. Instrument calibration and manual quality control lead to automatic revaluation of the mole fractions calculated in near-real time. Calibration corrections avoid artificial gradients between sites that could lead to error in flux estimates.
The ATC automatically processes atmospheric greenhouse gases mole fractions of data sent daily...
Citation