Automatic processing of atmospheric CO2 and CH4 mole fractions at the ICOS Atmosphere Thematic Centre

Lynn Hazan, Jérôme Tarniewicz, Michel Ramonet, Olivier Laurent, and Amara Abbaris

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement
Download
Short summary
The ATC automatically processes atmospheric greenhouse gases mole fractions of data sent daily by the ICOS network, this includes calibration and water vapor corrections. Data are stored in a database which has been developed with an emphasis on traceability. Instrument calibration and manual quality control lead to automatic revaluation of the mole fractions calculated in near-real time. Calibration corrections avoid artificial gradients between sites that could lead to error in flux estimates.