Articles | Volume 10, issue 10
https://doi.org/10.5194/amt-10-3893-2017
https://doi.org/10.5194/amt-10-3893-2017
Research article
 | 
24 Oct 2017
Research article |  | 24 Oct 2017

Chromatography related performance of the Monitor for AeRosols and GAses in ambient air (MARGA): laboratory and field-based evaluation

Xi Chen, John T. Walker, and Chris Geron

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by John Walker on behalf of the Authors (09 Aug 2017)  Author's response   Manuscript 
ED: Referee Nomination & Report Request started (09 Aug 2017) by Gavin Phillips
RR by Anonymous Referee #1 (30 Aug 2017)
ED: Publish as is (31 Aug 2017) by Gavin Phillips
AR by John Walker on behalf of the Authors (10 Sep 2017)  Manuscript 
Download
Short summary
The MARGA instrument is a powerful research tool for investigating atmospheric chemistry and the exchange of gases and particles between the atmosphere and biosphere. Our study examines the operating characteristics of the instrument with a focus on the software used to calculate air concentrations. We show that more accurate measurements may be obtained by calibrating the instrument with a range of standards in addition to the single internal standard used in online concentration calculations.