Articles | Volume 13, issue 8
https://doi.org/10.5194/amt-13-4111-2020
https://doi.org/10.5194/amt-13-4111-2020
Research article
 | 
03 Aug 2020
Research article |  | 03 Aug 2020

Development of an automatic linear calibration method for high-resolution single-particle mass spectrometry: improved chemical species identification for atmospheric aerosols

Shengqiang Zhu, Lei Li, Shurong Wang, Mei Li, Yaxi Liu, Xiaohui Lu, Hong Chen, Lin Wang, Jianmin Chen, Zhen Zhou, Xin Yang, and Xiaofei Wang

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by Xiaofei Wang on behalf of the Authors (03 Jun 2020)  Author's response   Manuscript 
ED: Publish as is (01 Jul 2020) by Pinhua Xie
AR by Xiaofei Wang on behalf of the Authors (03 Jul 2020)  Manuscript 
Download
Short summary
Single-particle aerosol mass spectrometry (SPAMS) is widely used to detect chemical compositions and sizes of individual aerosol particles. However, it has a major issue: the mass accuracy of high-resolution SPAMS is relatively low. Here we developed an automatic linear calibration method to greatly improve the mass accuracy of SPAMS spectra so that the elemental compositions of organic peaks, such as Cx, CxHy, CxHyOz and CxHyNO peaks, can be directly identified just based on their m / z values.