Improved chloride quantification in quadrupole aerosol chemical speciation monitors (Q-ACSMs)

Anna K. Tobler, Alicja Skiba, Dongyu S. Wang, Philip Croteau, Katarzyna Styszko, Jarosław Nęcki, Urs Baltensperger, Jay G. Slowik, and André S. H. Prévôt

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by Anna Tobler on behalf of the Authors (04 Aug 2020)  Author's response    Manuscript
ED: Publish as is (08 Aug 2020) by Mingjin Tang
Download
Short summary
Some quadrupole aerosol chemical speciation monitors (Q-ACSMs) have had issues with the quantification of particulate chloride, resulting in apparent negative chloride concentrations. We can show that this is due to the different behavior of Cl+ and HCl+, and we present a correction for the more accurate quantification of chloride. The correction can be applied to measurements in environments where the particulate chloride is dominated by NH4Cl.