Increase of the particle hit rate in a laser single-particle mass spectrometer by pulse delayed extraction technology
Ying Chen
Viacheslav Kozlovskiy
Xubing Du
Jinnuo Lv
Sergei Nikiforov
Jiajun Yu
Alexander Kolosov
Wei Gao
Zhen Zhou
Zhengxu Huang
A single-particle mass spectrometer (SPMS) can provide a wealth of valuable information on chemical and physical parameters of individual particles in real time. One of the main performance criteria of the instrument is efficiency of particle detection (hit rate). Most SPMS instruments use constant electrical field (DC) extraction, where stationary high voltage is applied to the extraction electrodes. As the aerosol particles initially carry a certain charge, those with a high amount to charge can be deflected by this electric field and lost, thus decreasing the hit rate. We realized that the delayed extraction technique can eliminate the stochastic dispersion of the particle beam caused by their deflection in the stationary electric field. As the result, the hit rate of the instrument can be significantly improved. Also, as the effect of the deflection in the electric field is mass dependent, it can cause distortion of the measured size distribution of the particles. Hence, the delayed extraction technique can bring the recorded distribution closer to the actual one. We found that the delayed extraction technique provides a mass resolution improvement as well as increases the hit rate. The gain in the hit rate depends on the type of particles. It can be 2 orders of magnitude for model particles and up to 2–4 times for ambient particles. In the present work we report experiments and results showing the effect of the delayed extraction on the beam divergence caused by particle charge, the hit rate improvement, and the effect of the delayed extraction on the measured particle size distribution.
- Article
(3743 KB) - Full-text XML
- Companion paper
-
Supplement
(715 KB) - BibTeX
- EndNote
Aerosol particles have a strong impact on the climate, environment and human health. These effects are closely related to physical and chemical properties of the particles. An in-depth understanding of the physical and chemical properties of individual particles is important for studying various effects of aerosols. A single-particle mass spectrometer (SPMS) is an analytical tool that can provide the particle size and chemical composition of individual particles in real time. There are many publications that have reviewed the principle, structure, and applications of a SPMS in detail (Murphy, 2007; Nash et al., 2006; Noble and Prather, 2000; and Pratt and Prather, 2012). SPMSs usually use an aerodynamic lens to focus and form an aerosol beam and to transfer the beam from the atmosphere into the vacuum system. As a common arrangement, a two-beam sizing laser measures the particle speed and size right before the “sized” particle passes into a high-power pulsed laser beam. When the particles are evaporated and ionized by the laser radiation, positive and negative ions are formed. These ions are accelerated by the strong electric field in the time-of-flight mass spectrometer (TOF MS) and then detected and recorded by a data acquisition system based on a fast analog-to-digital converter card (ADC). Compared to the traditional offline particle methods, the SPMS measurements can be conducted without a preliminary collection and preparation of the particles, providing a rapid real-time analysis. At present, a SPMS is mainly used for the characterization of physical and chemical properties of the particles, atmospheric chemical process analysis, atmospheric aerosol analysis, and other environmental and material science applications (Bi et al., 2011; Kim et al., 2012; Li et al., 2014; Middlebrook et al., 2003; and Roth et al., 2016). A SPMS is fast, and detection time for a single particle is typically on the order of a few milliseconds. However, it is common to accumulate single-particle data over a period of time to obtain statistically significant information about particle diversity. The application of a SPMS to atmospheric aerosol analysis and investigation of chemical processes with atmospheric aerosols requires instruments with a high efficiency of particle detection (hit rate). Hit rate is an important parameter of a SPMS related to its performance. It is generally defined as the ratio of the number of spectra generated by laser ionization to the number of particles detected by sizing lasers. There are many aspects that affect the hit rate of a SPMS, including the particle size, physical and chemical properties (Dall'Osto et al., 2006; Zelenyuk et al., 2009) of the particle, and the instrument design (Gemayel et al., 2016; Zhao et al., 2005).
It is well known (Dodd, 1953; Gunn and Woessner, 1956; and Woessner and Gunn, 1956) that aerosol particles usually carry a certain amount of surface charge. If such particles enter the stationary electrical field in the extraction region of a SPMS, they can be deflected as they move through the ion source region. Once the particle trajectory has shifted from the center of the laser beam, the ionization efficiency can drop down noticeably (Su et al., 2004). Hence the particle deflection in the ion source results in a decrease in the hit rate. In order to reduce the influence of the electrical field on the particle beam trajectory, an electrostatic neutralizer can be installed between the particle introduction and particle ionization regions. The electrostatic neutralizer generally uses a radioactive source based on 210Po, 241Am, or 85Kr, or it uses an X-ray source (Demokritou et al., 2004; Kousaka et al., 1981; Nicosia et al., 2018; Pratt et al., 2009; and Tsai et al., 2005). However, the management of radioactive sources is a difficult task; the equipment is relatively expensive, and it is very inconvenient to use it in field measurements because of strict regulations restricting the handling, transport, and storage of radioactive sources (Nicosia et al., 2018).
Another problem for a SPMS is particle beam divergence. There are many instruments that use Nd:YAG laser with fourth harmonic generation (266 nm wavelength) as the ionization source. Since the Nd:YAG laser with lamp pumping requires a more than 100 µs delay between the lamp ignition and the laser pulse, it is necessary to place the laser ionization source at a considerable distance (about 10 cm) downstream of the second particle-sizing laser. The particle beam is dispersed on the way from the sizing laser to the ionizing laser, and the increase in the beam width leads to a decrease in the hit rate. This problem can be solved using a laser with a shorter delay time, such as excimer or nitrogen pulsed gas lasers (Alaime et al., 1983; Trimborn et al., 2002). It is well known that the delay time for the pulsed gas laser is less than 100 ns. The small delay time makes it possible to reduce the distance between the sizing laser and the ion source, which will be limited by the mass spectrometer design in this case. The effect can be especially bigger for irregular nonspherical particles (Zelenyuk et al., 2006). Shortening the distance between the sizing laser and the ionizing laser can compensate for the influence of the particle shape to some extent (Cziczo et al., 2006; Thomson et al., 1997).
Zelenyuk et al. (2009) also studied the shift of the trigger signals generated by particles with different sizes at the sizing laser, which resulted in the hit rate being affected by the particle size. As a result, the dynamic laser trigger system was developed to implement the trigger compensation for the particles of different sizes, thereby achieving an improvement in the hit rate. However, this method is directed to solve the positional spread of the aerosol particles along the axis; it does not contribute to solving the problem of the lowered hit rate caused by deflection of the aerosol in the direction perpendicular to the particle beam axis. This deflection is caused by the static electrical field and depends on the ion source dimensions and the electrical field strength. Unfortunately, the ion source cannot be built short enough to neglect particle deflection. But it is possible to switch off the electrical field by using a delayed extraction (DE) technique (Li et al., 2018; Chudinov et al., 2019) for ion sampling.
In our previous works we have used the double exponential pulse delayed extraction technique to improve the mass resolution of a SPMS (Chudinov et al., 2019; Li et al., 2018). This solution provides a mass resolution enhancement of about 2–3 times over the broad mass range in comparison to the usual constant electrical field (DC) extraction. It is important that in the delayed extraction, ion source particles pass to the middle of the ion source with no DC electrical field. Therefore, the influence of the particle charge on its trajectory will be neglected in comparison with the usual ion source with a stationary electrical field. In this paper we present our investigation of the effect of stationary and pulse extraction on the particle trajectory and the hit rate of the SPMS instrument.
The SPMS used in this study is a commercial instrument from Hexin Analytical Instrument Co., Ltd. (Guangzhou), which is normally called a single-particle aerosol mass spectrometer (SPAMS). The principle and design of the SPAMS has been described elsewhere in detail (Li et al., 2011). The commercial SPAMS was modified in two ways. First, the delayed extraction technique is used instead of the original DC field extraction technology. The superposition of the rectangular and the exponentially shaped extraction pulses simultaneously improves the resolution over the wide mass range for positive and negative ions. The detailed information on the instrument changes required to implement the delayed extraction is provided elsewhere (Li et al., 2018). After the improvement, a positive ion mass resolution of >1000 full width at half maximum (FWHM) and a negative ion mass resolution of >2000 FWHM were achieved. The second improvement is the use of multichannel superimposed signal acquisition technology. The native signal from the microchannel plate (MCP) detector is equally split in two channels with different amplification ratios and acquired by two equal 8 bit ADC. Thus, the high dynamic range data acquisition system is capable of acquiring signals ranging from 5 to 20 000 mV, and the dynamic range is nearly 40 times higher than that of the original SPAMS. This solution enables one, for instance, to detect high-intensity alkali metal ion peaks together with very weak ion signals from other elements or molecules at the same time (Shen et al., 2018).
The test particles used in the experiment were standard polystyrene latex microspheres (PSL) with particle sizes of 320, 510, 720, 960, and 1400 nm purchased from Duke Scientific. The commercial aerosol generator TSI 9302 was used to atomize the aqueous solution of the PSL and to produce a controlled beam of particles. The airflow with PSL particles was dried by a diffusion drying tube and then passed to the SPAMS for analysis.
For an independent particle size distribution measurement, the glass plate was installed under the ion source as shown in Fig. 1. The distance between the center of the ion source and the plate surface was 31.5 mm. After the glass plate was exposed to the particle flow, the photo of the particles collected on the glass plate surface was acquired by an Olympus CX31RTSF microscope. The density profile of the collected particles was extracted by free ImageJ software. Trajectories of particles shown in Fig. 1 illustrate the deflection in the electrical field. The deflection depends on the charge and mass of the particle. The deflection and trajectory shape is discussed below in the Sect. 3.1
2.1 Key factors affecting the efficiency
Figure 1 shows a schematic representation of the SPAMS ion source. The Nd:YAG laser beam is focused in the center of the ion source where particles are ionized by laser radiation. The 266 nm laser operates in TEM00 mode with a Gaussian distribution. The e−1 width of the focused Gaussian laser beam is about 300 µm. Factors that actually affect the ionization efficiency are the total laser pulse energy and the location of the particle in the focused laser spot. As the particle size is more than 300 times smaller than that of the focused laser beam width, each particle is ionized by a virtually uniform laser power density. But when the beam of the particles is wider than the diameter of the laser beam in the ionization region, the particles in the beam can be exposed to very different laser energy fluencies, resulting in a great difference in ionization efficiency. Therefore, when the particles travel near the edge of the laser spot, it becomes impossible to generate enough ions to detect such particles.
In our experiments, the Ultra compact Q-switched Nd:YAG laser (Quantel Laser) was used with a pulse width of 7.2 ns. The pulse energy of the laser was set to 0.6 mJ, which corresponds to a laser power density in the focal point of about 2×108 W cm−2.
When the delayed extraction technique is not used, the two DC high-voltage potentials are constantly applied to the extraction electrodes of the original bipolar SPAMS ion source. When the charged particles enter the space of the ion source, they start to be deflected under the action of the DC electrical field. The amount of deflection is related to the speed and the mass of the particles, the amount of the surface charge, and the electrical field strength. Thus, during the flight through the ion source, the particles gradually deviate from the laser center point, resulting in the decrease in the ionization efficiency and a significant drop in the hit rate.
When the delayed extraction technique is used, the particles enter the extraction region while the electrical field is kept at zero. Then the laser ionizes the particle, and the high voltage extraction pulse is applied with 100 ns delay after the laser pulse. So there is no electric field between the ion source plates before the particles are ionized by the laser. Thus, the charged aerosol particles are not deflected by the electric field force during their flight, thereby achieving a higher hit rate.
3.1 The experiment with PSL particles
The influence of the electric field on the hit rate was studied by using PSL beads of five different particle sizes. In Fig. 2 the dependencies of the hit rate on the electrical field strength for DC and DE modes are shown. It can be seen that under the same DC electric field strength (Fig. 2a), the hit rate of the particles from 320 to 1400 nm increases, indicating that the effect of the electric field on the small particles is more pronounced. As the DC electric field strength increases, the hit rate of the particles of each size gradually decreases, indicating that the particles are deflected by the electric field force.
Under the DE mode, the hit rate for the particles of all sizes is obviously increased in comparison with the DC case (Fig. 2b). When the pulsed electric field strength exceeds 60 kV m−1, the hit rate of the particles above 520 nm is close to 100 %, indicating that the flight path without an electrical field is essential for the increase in the hit rate of the particles. The hit rate of the 320 nm PSL particles is 40 %, which is much lower than 100 % of other particles. This can be due to the focusing effect of the aerodynamic lens. The divergence of the particle beam itself is higher than that for the particles above 500 nm, which results in a decrease in the hit rate. In addition, when the electric field strength increases from 0 to 60 kV m−1, the hit rate gradually increases from 5 % to 100 %. We speculate that the increase in the hit rate with the electrical field strength in this range may be caused by the increase in the resolving power of the TOF MS (Cotter, 1994). The lower the extraction electrical field strength is, the bigger the turnaround time in the TOF MS is, which in turn decreases the resolving power of the instrument. In the case of a large turnaround time, the peak width increases and the peak height decreases. This, in turn, can result in a loss of measured ion current because of an ADC threshold adjusted to reject the low-intensity peaks accounted as noise. The loss of the recorded ion current can in turn result in a miscount of some particles, thus decreasing the hit rate. Also, for the low extraction electric field, ion divergence can be enhanced, resulting in a loss of ion current during the flight of the ion to the detector.
In Fig. 3 dependencies of the hit rate gain achieved for the DE mode compared to the DC mode case are shown. It can be seen that for different particle sizes, the hit rate gain increases sharply with the increase in the electric field strength, especially for small particles. For 320 nm particles, the hit rate can be increased by more than 100 times at an electrical field strength over 180 kV m−1. This result shows that the DE technique can effectively eliminate the influence of the electric field intensity on the flight path of the particles, avoid the use of additional peripheral devices such as electrostatic neutralizers, and result in an improvement in the hit rate.
It is worth noting that the mass spectra obtained with DE have a higher resolution, but they are approximately the same set of peaks as those obtained with the injection of ions in a constant field (CF). Peaks in the mass spectrum obtained with DE are narrower and, on average, have a larger amplitude. The influence of a ∼100 ns delay is not reflected in the composition of peaks in acquired mass spectra. Mass spectra of PSL particles acquired in DE and DC modes are presented in the Supplement for comparison. The mechanisms of ion formation as part of a complex of processes during laser exposure to a particle are of great interest and are the subject of our further research. In this work we did not focus on these issues, since they are not directly related to the observed increase in the hit rate.
In order to further confirm that the particles are deflected by the action of the electric field, we conducted several further experiments. Taking the 720 nm PSL microsphere as an example, we compared the hit rate of 720 nm PSL beads under the constant DC electrical field only to the hit rate of those under the DC electric field with an electrostatic neutralizer and to the hit rate under the delayed extraction combined with or without a neutralizer. Figure 4 shows the hit rate value for a measurement period of ∼30 min. It can be seen that the particle hit rate after the use of the electrostatic neutralizer has increased from an average level of 20 % to a level exceeding 85 %, indicating that the surface charge of the particles has a great influence on the hit rate. However, the average hit rate of the particles obtained by the DE technique is close to 96 %, which is slightly higher than that of the hit rate measured in the case when the electrostatic neutralizer was used. This difference can be caused by a residual charge remaining on the particles even after the usage of an X-ray neutralizer (TSI 3088). The higher hit rate achieved by delayed extraction is caused by the absence of electric field influence on the particle motion.
A further two experiments were performed to measure particle beam divergence under the influence of the electrical field formed between two electrodes of the extraction region of the SPMS. The particles were collected on the surface of the glass plate, as it is shown in Fig. 1. In the first experiment (Fig. a, b) no voltage was applied to the electrodes, and the resulting spot just represents a 720 nm particle beam profile as a result of the aerodynamic dynamic lens action. The profile was fitted by a Gauss envelope, and the average beam width extracted from the Gaussian curve is a little more than 0.1 mm. The second experiment (Fig. 6a, b) was performed with DC potentials ( V) applied to the extraction region electrodes, in accordance with the usual working conditions of the DC extraction mode. The trace of the same 720 nm PSL beads on the glass plate is shown in Fig. 6a.
The particle divergence is obviously increased in this case. Using the ImageJ program, the trace profile was extracted, and then the Gaussian fitting was uploaded on the data. The space distribution of the particles can be easily transformed into a particle charge distribution as we deal with the calibrated PSL beads. The ion beam displacement at the exit of the extraction region (Fig. 6a, b) was analyzed by Liu et al. (1995), and it can be described by the simple equation:
Here q and mp are particle charge and mass, d and VH are the distance and voltage difference between two flat extraction electrodes, L1 is the electrode height, Up is the particle longitudinal velocity, a is the acceleration of the particle caused by the extraction electrical field, and t is the particle dwell time in the extraction region. Thus, the particle trajectories are parabolic, scaled by their mass and charge. The equation was derived from a simplified assumption that the particle moving between the plates is accelerated by the constant horizontal electrostatic force. Then the particle with the horizontal speed component moves until it hits the glass gathering plate. The resulting particle displacement in the position of the gathering glass plate is as follows:
Here L2 is the distance from the lower edge of the extraction electrode to the gathering glass plate surface. The resulting particle displacement is proportional to the particle charge; so from the trace in Fig. 6a we can extract the particle charge distribution shown in Fig. 6b. The average particle charge (in units of e) extracted from the Gaussian envelope is 60 elementary charges. There are multiple publications in which this electrification charge was measured for different types of particles (Dodd, 1953). The results reported in the literature (40 charges per 700 nm road dust particle; Forsyth et al., 1998; 40 charges per 400 nm PSL particle; Kleefsman et al., 2008) are in reasonable agreement with our estimation made for 720 nm PSL beads.
Using known particle mass, velocity, extraction voltage, and electrodes dimensions, we have extracted the particle displacement per unit charge in the middle point on which the ionizing laser is focused. For the parameters of our instrument this displacement is 12.5 µm z−1, where z is the number of elementary charges on the particle. Hence, a particle with an average of 20 elementary charges will be deflected ∼250 µm from the ionizing laser spot center. Taking into account the Gaussian profile of the focused laser beam of 300 µm, the hit efficiency will be greatly affected by such a displacement.
3.2 Comparison of environmental testing
The comparison of the detection efficiency for the instrument with DC extraction and with DE used for the actual atmospheric particulate matter detection is shown in Fig. 7a and b. Aerosol particles from laboratory room air were analyzed by the SPMS used in this work. In order to ensure that aerosol composition is stable during the experiment, the DC and DE modes of operation were switched alternately after 10 min. Four sets of both DC and DE extraction data were acquired in 80 min. The hit rate was calculated as the ratio of the number of ionized particles to the number of sized particles. The hit rate data were extracted for the groups of particles selected by size; bin value was 20 nm while the whole range was 200–1000 nm. The data of four sets of 10 min each, both for DC and DE extraction modes, were summed. In the DC mode, in 40 min, a total of 12 174 particles was recorded, of which 4569 were ionized, while in the DE mode, 12 030 particles were recorded, and 8228 were ionized for the same time. It is seen from Fig. 7a that the center of the distribution obtained with DC extraction (blue circles) is shifted to the higher particles size. This behavior corresponds to our results obtained for the model PSL particles. Figure 7b shows the hit rate gain from the DE technique to the DC extraction, obtained from the data plotted in Fig. 7a. It should be noted that usage of delayed extraction eliminates the distortion of particle distribution, which results from the dispersion of particle deflection by their size in the case of DC extraction. The integral hit rate for the experiment with the ambient particles is ∼25 % higher in the case of delayed extraction (Fig. 7a), while the gain in hit rate achieved for small particles is ∼4 times bigger than that measured for DC extraction (Fig. 7b). The increased effect for small particles can be explained by Eq. (1). The smaller the m∕q of the particle, the bigger its deflection in the ion source (extraction region). Hence, the hit rate of smaller particles is decreased more, and the gain caused by DE should be higher.
The hit rate of the atmospheric aerosol is lower than that of the standard PSL beads mainly due to the shape of the particles. In the actual atmosphere, the shape of the particles is much more complicated than that of the PSL sphere. For example, the aged carbon particles tend to be spherical, and the hit rate is relatively high, while the fresh black carbon particles exhibit a chain structure, so the divergence during their flight is probably bigger, resulting in lower hit rates (Ghan et al., 2012). In general, using the delayed extraction technique, the hit rate of environmental particles with a size of <500 nm can be noticeably improved, and the obtained particle size distribution becomes closer to the real one.
Aerosols often carry surface charges. When charged particles fly through the SPMS ion source, they can be deflected by the action of constant electric field, resulting in a substantial drop in the hit rate. When delayed extraction is used in a SPMS, the instrument efficiency is improved because the particles are not deflected by the action of the stationary electric field. The absence of the DC electrical field in this method allows one to avoid the use of the auxiliary external devices such as electrostatic neutralizers, maximize the hit rate of the aerosol particles, and decrease the distortion of the measured size distribution of particles. Experiments have shown that for PSL microspheres, the smaller the particle size, the greater the gain in the hit rate. There is no dependence of the hit rate on the measured particle size for particles with a diameter of more than 500 nm when DE is used. For the real application it means that this method minimizes losses in the hit rate, which in turn affects the measured size distribution of the particles.
The experiments with environmental aerosol have also shown that SPAMS with delayed extraction technology can improve the average detection efficiency of the actual atmospheric aerosol by more than 25 % and improve the hit rate for small (<500 nm) particles by more than 4 times. It should be noted that the charge of the particles is connected with their parameters. It means that a preselection of particles based on their charge can be used to extract extra information from the data acquired by a SPMS. It could be done, for example, by using two pairs of deflecting plates before the ion source. By choosing the deflection voltage, only particles with a specific mass∕charge ratio will pass this double deflector and can be ionized in the ion source.
These data can be publicly accessible in free.
The supplement related to this article is available online at: https://doi.org/10.5194/amt-13-941-2020-supplement.
LL designed the study; YC, XD, and LL performed the experiments; LL, VK, SN, XD, and ZXH conducted the data analysis; LL, VK, and SN wrote the paper with the input from all authors.
The authors declare that they have no conflict of interest.
We would like to thank engineer Wang Jingjing of the Guangzhou Hexin Mass Spectrometer Co., Ltd., for his technical support and Alexandre Loboda and Adam Strock for the language proof.
This research has been supported by the Guangdong Provincial Natural Science Foundation-Ph.D. Launched Vertical Collaborative Project (grant no. 2017A030310384), the Guangdong International Cooperation Project (grant no. 2018A05056020), 2013–2020 Program of Basic Research of the Russian State Academies of Sciences (nos. AAAA-A18-118112690060-9, AAAA-A19-119121790091-3), and the National Key Research Project (grant no. 2017YFC0209506).
This paper was edited by Mingjin Tang and reviewed by four anonymous referees.
Alaime, C., Bertrand, J., Bonnet, L., El Masri, Y., Palffy, L., Pellegrin, P., and Prieels, R.: Stabilization of the light output delay of a pulsed nitrogen laser, Nucl. Instrum. Methods, 207, 423–427, https://doi.org/10.1016/0167-5087(83)90653-1, 1983.
Bi, X., Zhang, G., Li, L., Wang, X., Li, M., Sheng, G., Fu, J., and Zhou, Z.: Mixing state of biomass burning particles by single particle aerosol mass spectrometer in the urban area of PRD, China, Atmos. Environ., 45, 3447–3453, https://doi.org/10.1016/j.atmosenv.2011.03.034, 2011.
Chudinov, A., Li, L., Zhou, Z., Huang, Z., Gao, W., Yu, J., Nikiforov, S., Pikhtelev, A., Bukharina, A., and Kozlovskiy, V.: Improvement of peaks identification and dynamic range for bi-polar Single Particle Mass Spectrometer, Int. J. Mass Spectrom., 436, 7–17, https://doi.org/10.1016/j.ijms.2018.11.013, 2019.
Cziczo, D. J., Thomson, D. S., Thompson, T. L., DeMott, P. J., and Murphy, D. M.: Particle analysis by laser mass spectrometry (PALMS) studies of ice nuclei and other low number density particles, Int. J. Mass Spectrom., 258, 21–29, https://doi.org/10.1016/j.ijms.2006.05.013, 2006.
Dall'Osto, M., Harrison, R. M., Beddows, D. C. S., Freney, E. J., Heal, M. R., and Donovan, R. J.: Single-particle detection efficiencies of aerosol time-of-flight mass spectrometry during the North Atlantic marine boundary layer experiment, Environ. Sci. Technol., 40, 5029–5035, https://doi.org/10.1021/es050951i, 2006.
Demokritou, P., Lee, S. J., Ferguson, S. T., and Koutrakis, P.: A compact multistage (cascade) impactor for the characterization of atmospheric aerosols, J. Aerosol Sci., 35, 281–299, https://doi.org/10.1016/j.jaerosci.2003.09.003, 2004.
Dodd, E. E.: The Statistics of Liquid Spray and Dust Electrification by the Hopper and Laby Method, J. Appl. Phys., 24, 73–80, https://doi.org/10.1063/1.1721137, 1953.
Gemayel, R., Hellebust, S., Temime-Roussel, B., Hayeck, N., Van Elteren, J. T., Wortham, H., and Gligorovski, S.: The performance and the characterization of laser ablation aerosol particle time-of-flight mass spectrometry (LAAP-ToF-MS), Atmos. Meas. Tech., 9, 1947–1959, https://doi.org/10.5194/amt-9-1947-2016, 2016.
Ghan, S. J., Liu, X., East er, R. C., Zaveri, R., Rasch, P. J., Yoon, J.-H., and Eaton, B.: Toward a Minimal Representation of Aerosols in Climate Models: Comparative Decomposition of Aerosol Direct, Semidirect, and Indirect Radiative Forcing, J. Climate, 25, 6461–6476, https://doi.org/10.1175/jcli-d-11-00650.1, 2012.
Gunn, R. and Woessner, R. H.: Measurements of the systematic electrification of aerosols, J. Coll. Sci. Imp. U. Tok., 11, 254–259, https://doi.org/10.1016/0095-8522(56)90050-2, 1956.
Kim, M., Cahill, J. F., Su, Y., Prather, K. A., and Cohen, S. M.: Postsynthetic ligand exchange as a route to functionalization of “inert” metal–organic frameworks, Chem. Sci., 3, 126–130, https://doi.org/10.1039/C1SC00394A, 2012.
Kousaka, Y., Okuyama, K., and Endo, Y.: Calibration of differential mobility analyser by visual method, J. Aerosol Sci., 12, 339–348, https://doi.org/10.1016/0021-8502(81)90023-9, 1981.
Li, L., Huang, Z., Dong, J., Li, M., Gao, W., Nian, H., Fu, Z., Zhang, G., Bi, X., Cheng, P., and Zhou, Z.: Real time bipolar time-of-flight mass spectrometer for analyzing single aerosol particles, Int. J. Mass Spectrom., 303, 118–124, https://doi.org/10.1016/j.ijms.2011.01.017, 2011.
Li, L., Li, M., Huang, Z., Gao, W., Nian, H., Fu, Z., Gao, J., Chai, F., and Zhou, Z.: Ambient particle characterization by single particle aerosol mass spectrometry in an urban area of Beijing, Atmos. Environ., 94, 323–331, https://doi.org/10.1016/j.atmosenv.2014.03.048, 2014.
Li, L., Liu, L., Xu, L., Li, M., Li, X., Gao, W., Huang, Z., and Cheng, P.: Improvement in the Mass Resolution of Single Particle Mass Spectrometry Using Delayed Ion Extraction, Int. J. Mass Spectrom., 29, 2105–2109, https://doi.org/10.1007/s13361-018-2037-4, 2018.
Liu, P., Ziemann, P. J., Kittelson, D. B., and McMurry, P. H.: Generating Particle Beams of Controlled Dimensions and Divergence: II. Experimental Evaluation of Particle Motion in Aerodynamic Lenses and Nozzle Expansions, Aerosol Sci. Technol., 22, 314–324, https://doi.org/10.1080/02786829408959749, 1995.
Middlebrook, A. M., Murphy, D. M., Lee, S.-H., Thomson, D. S., Prather, K. A., Wenzel, R. J., Liu, D.-Y., Phares, D. J., Rhoads, K. P., Wexler, A. S., Johnston, M. V., Jimenez, J. L., Jayne, J. T., Worsnop, D. R., Yourshaw, I., Seinfeld, J. H., and Flagan, R. C.: A comparison of particle mass spectrometers during the 1999 Atlanta Supersite Project, J. Geophys. Res.-Atmos., 108, 8424, https://doi.org/10.1029/2001jd000660, 2003.
Murphy, D. M.: The design of single particle laser mass spectrometers, Mass Spectrom. Rev., 26, 150–165, https://doi.org/10.1002/mas.20113, 2007.
Nash, D. G., Baer, T., and Johnston, M. V.: Aerosol mass spectrometry: An introductory review, Int. J. Mass Spectrom., 258, 2–12, https://doi.org/10.1016/j.ijms.2006.09.017, 2006.
Nicosia, A., Manodori, L., Trentini, A., Ricciardelli, I., Bacco, D., Poluzzi, V., Di Matteo, L., and Belosi, F.: Field study of a soft X-ray aerosol neutralizer combined with electrostatic classifiers for nanoparticle size distribution measurements, Particuology, 37, 99–106, https://doi.org/10.1016/j.partic.2017.08.001, 2018.
Noble, C. A. and Prather, K. A.: Real-time single particle mass spectrometry: a historical review of a quarter century of the chemical analysis of aerosols, Mass Spectrom. Rev., 19, 248–274, https://doi.org/10.1002/1098-2787(200007)19:4<248::AID-MAS3>3.0.CO;2-I, 2000.
Pratt, K. A., Mayer, J. E., Holecek, J. C., Moffet, R. C., Sanchez, R. O., Rebotier, T. P., Furutani, H., Gonin, M., Fuhrer, K., Su, Y., Guazzotti, S., and Prather, K. A.: Development and Characterization of an Aircraft Aerosol Time-of-Flight Mass Spectrometer, Anal. Chem., 81, 1792–1800, https://doi.org/10.1021/ac801942r, 2009.
Pratt, K. A. and Prather, K. A.: Mass spectrometry of atmospheric aerosols – Recent developments and applications, Part II: On-line mass spectrometry techniques, 31, 17–48, https://doi.org/10.1002/mas.20330, 2012.
Roth, A., Schneider, J., Klimach, T., Mertes, S., van Pinxteren, D., Herrmann, H., and Borrmann, S.: Aerosol properties, source identification, and cloud processing in orographic clouds measured by single particle mass spectrometry on a central European mountain site during HCCT-2010, Atmos. Chem. Phys., 16, 505–524, https://doi.org/10.5194/acp-16-505-2016, 2016.
Shen, W., Dai, X., Huang, Z. X., Hou, Z. H., Cai, W. G., Du, X. B., Zhou, Z., Li, M., and Li, L.: Improvement of the Dynamic Range of Data Acquisition System in Single Particle Mass Spectrometry, J. Chinese Cinemas., 39, 331–336, https://doi.org/10.7538/zpxb.2017.0119, 2018.
Su, Y., Sipin, M. F., Furutani, H., and Prather, K. A.: Development and Characterization of an Aerosol Time-of-Flight Mass Spectrometer with Increased Detection Efficiency, Anal. Chem., 76, 712–719, https://doi.org/10.1021/ac034797z, 2004.
Thomson, D. S., Middlebrook, A. M., and Murphy, D. M.: Thresholds for Laser-Induced Ion Formation from Aerosols in a Vacuum Using Ultraviolet and Vacuum-Ultraviolet Laser Wavelengths, Aerosol Sci. Technol., 26, 544–559, https://doi.org/10.1080/02786829708965452, 1997.
Trimborn, A., Hinz, K. P., and Spengler, B.: Online analysis of atmospheric particles with a transportable laser mass spectrometer during LACE 98, J. Geophys. Res.-Atmos., 107, LAC 13-11-LAC 13-10, https://doi.org/10.1029/2001JD000590, 2002.
Tsai, C.-J., Lin, J.-S., Deshpande, C., and Liu, L.-C.: Electrostatic Charge Measurement and Charge Neutralization of Fine Aerosol Particles during the Generation Process, Part. Part. Syst. Charact., 22 293–298, https://doi.org/10.1002/ppsc.200500961, 2005.
Woessner, R. H. and Gunn, R.: Measurements related to the fundamental processes of aerosol electrification, J. Coll. Sci. Imp. U. Tok., 11, 69–76, https://doi.org/10.1016/0095-8522(56)90021-6, 1956.
Zelenyuk, A., Cai, Y., and Imre, D.: From Agglomerates of Spheres to Irregularly Shaped Particles: Determination of Dynamic Shape Factors from Measurements of Mobility and Vacuum Aerodynamic Diameters, Aerosol Sci. Technol., 40, 197–217, https://doi.org/10.1080/02786820500529406, 2006.
Zelenyuk, A., Yang, J., Imre, D., and Choi, E.: Achieving Size Independent Hit-Rate in Single Particle Mass Spectrometry, Aerosol Sci. Technol., 43, 305–310, https://doi.org/10.1080/02786820802637915, 2009.
Zhao, Y., Bein, K. J., Wexler, A. S., Misra, C., Fine, P. M., and Sioutas, C.: Field evaluation of the versatile aerosol concentration enrichment system (VACES) particle concentrator coupled to the rapid single-particle mass spectrometer (RSMS-3), J. Geophys. Res.-Atmos., 110, D07S02, https://doi.org/10.1029/2004jd004644, 2005.