the Creative Commons Attribution 4.0 License.
the Creative Commons Attribution 4.0 License.
The NO2 camera based on gas correlation spectroscopy
Leon Kuhn
Jonas Kuhn
Thomas Wagner
Ulrich Platt
Spectroscopic methods have proven to be reliable and of high selectivity by utilizing the characteristic spectral absorption signature of trace gases such as NO2. However, they typically lack the spatiotemporal resolution required for real-time imaging measurements of NO2 emissions. We propose imaging measurements of NO2 in the visible spectral range using a novel instrument, an NO2 camera based on the principle of gas correlation spectroscopy (GCS). For this purpose two gas cells (cuvettes) are placed in front of two camera modules. One gas cell is empty, while the other is filled with a high concentration of the target gas. The filled gas cell operates as a non-dispersive spectral filter to the incoming light, maintaining the two-dimensional imaging capability of the sensor arrays. NO2 images are generated on the basis of the signal ratio between the two images in the spectral window between 430 and 445 nm, where the NO2 absorption cross section is strongly structured. The capabilities and limits of the instrument are investigated in a numerical forward model. The predictions of this model are verified in a proof-of-concept measurement, in which the column densities in specially prepared reference cells were measured with the NO2 camera and a conventional differential optical absorption spectroscopy (DOAS) instrument. Finally, results from measurements at a large power plant, the Großkraftwerk Mannheim (GKM), are presented. NO2 column densities of the plume emitted from a GKM chimney are quantified at a spatiotemporal resolution of frames per second (FPS) and 0.9 m×0.9 m. A detection limit of 2⋅1016 molec. cm−2 was reached. An NO2 mass flux of kg h−1 was estimated on the basis of wind speeds obtained from consecutive images. The instrument prototype is highly portable for building costs of below EUR 2000.
- Article
(12742 KB) - Full-text XML
- BibTeX
- EndNote
Oxides of nitrogen () play an important role in urban air quality. Nitrogen dioxide (NO2) is itself toxic to humans and furthermore contributes to the formation of ozone (O3) and particulate matter. Both NO2 as well as ozone and particulate matter are linked to a variety of diseases, such as asthmatic and cardiovascular diseases (see, e.g., World Health Organization, 2000; Faustini et al., 2014). Recent studies have shown that in many European countries the average annual exposure to NO2 exceeds 10 µg m−3 (European Environment Agency, 2017), which is the exposure limit recommended by the World Health Organization (2021). In other parts of the world exceedances are even higher. Therefore, monitoring NO2 emissions and abundance near the planetary surface is of interest. In many cases NO2 concentration gradients occur on small spatial (sub-meter) and temporal (sub-second) scales, e.g., when measuring the emissions of moving point sources, such as cars, ships, or airplanes. At the same time, examinations of plume geometries, mass fluxes, and chemical reactions that take place in plumes require spatial coverage of the scene. Overall, an imaging method for NO2 with high spatiotemporal resolution could reveal more insight into the quantity and dynamics of NO2 emissions.
In polluted regions NOx emissions are mainly of anthropogenic origin. Combustion processes, which occur, e.g., in car motors or industrial power plants, generate NOx, which, at the time of emission, consists mostly of NO (typically with NO2NOx ratios as low as 5 %–10 %; see, e.g., Kenty et al., 2007, and Carslaw, 2005). Through oxidization processes, such as
or, at very high NO concentrations,
NO is converted to NO2. In addition, other sources of NOx exist, such as geophysical events like lightning strikes, forest fires or soil emissions. Due to photodissociation, i.e.,
an equilibrium between NO2, NO, and O3 (quickly formed by O+O2), called the Leighton relationship, settles in.
There are different remote sensing methods for monitoring of atmospheric trace gases such as NO2. The state-of-the-art method is differential optical absorption spectroscopy (DOAS; Platt and Stutz, 2008), in which the absorption cross sections of the target gases are fitted to the spectrally resolved differential optical depths along a light path. Then the column densities of the target gases are retrieved as fit parameters. DOAS measurements can be based on natural light sources, such as scattered sunlight, or on artificial ones. Modern DOAS spectrographs typically have a spectral resolution of <1 nm and operate in the UV and visible spectral range. The benefits of analyzing spectrally resolved data are high selectivity and low detection limits. However, grating spectrographs are less suited for imaging because spectral mapping leads to a reduced light throughput. Therefore, measurements with sufficient spatial and spectral resolution require rather long acquisition times of many minutes (Bobrowski et al., 2006; Louban et al., 2009). Imaging DOAS (I-DOAS) is typically realized using a push-broom technique, in which one detector dimension is used for spatial resolution and the other for spectral mapping. Consequently, I-DOAS requires scanning a field of view (FOV) column by column or row by row. This strategy was used, for example, by Manago et al. (2018), who report on an imaging DOAS instrument for NO2 based on a hyperspectral camera with a spatial resolution of 640×480 pixels, a FOV, and a frame rate of 0.2 FPS. Although modern hyperspectral cameras can reach adequate spatiotemporal resolution, some problems remain. Methods that rely on a push-broom scheme suffer from time delays between the rows (or columns) of the recorded images. Furthermore, spectrally resolving instruments are usually expensive and bulky.
We propose an imaging instrument for NO2 based on gas correlation spectroscopy (GCS; see, e.g., Ward and Zwick, 1975, Drummond et al., 1995, Wu et al., 2018, and Baker et al., 1986) and demonstrate that an instrument designed to measure only a single trace gas can work by using reduced but specific spectral information in order to maximize spatiotemporal resolution. This is achieved by the use of two 2D photosensors, each equipped with a lens and a glass cell: one filled with air (the “empty” cell) and one filled with a high concentration of NO2.
Figure 1 shows a photograph of an instrument prototype. The NO2 cell functions as a spectral filter to the incoming light, while the empty cell ideally has no effect on the incoming light and serves as a reference. At the same time, the cameras fully resolve the light in two spatial dimensions. This way image data with only two spectral channels (in contrast to about 100 spectral channels used for typical DOAS fitting windows) are obtained. The NO2 column density measured by each pixel of the instrument can then be computed by application of the Lambert–Beer law to the two channels. This principle is explained in more detail in Sect. 2.1. The method is therefore similar to the recently developed filter-correlation-based SO2 camera (Mori and Burton, 2006), the imaging Fabry–Pérot interferometer correlation spectroscopy technique (IFPICS; see, e.g., Kuhn et al., 2019, and Fuchs et al., 2021), or the acousto-optical tunable-filter-based (AOTF) NO2 camera (Dekemper et al., 2016). However, using a gas cell has substantial advantages compared to the listed techniques. While the filter correlation approach through its reduced selectivity only works for large volcanic SO2 emissions, Fabry–Pérot interferometers and AOTFs require collimated light beams within the lens setup, largely reducing the light throughput. In order to further increase selectivity to NO2, both cameras are equipped with an additional bandpass filter with transmission in the region of 425 to 450 nm, where the absorption cross section of NO2 shows strong characteristic features. An instrument of this kind requires NO2 to be stably contained in glass cells. The instrument prototype we present fulfills this requirement. The chemistry of NO2 gas cells is explained in detail by Platt and Kuhn (2019).
The rest of this paper is structured as follows: Sect. 2 deals with the theory of GCS and how it can be utilized for imaging measurements of NO2. We introduce an instrument forward model, which allows for the prediction of instrument responses, detection limits, and cross-sensitivities of a GCS-based NO2 camera under different circumstances. Section 3 presents a prototype of the instrument and lists its detailed technical specifications. Section 4 shows the results of two measurements that have been taken with that instrument prototype. The first is a proof-of-concept measurement with reference cells in an optical laboratory. The purpose of this measurement is to verify the functionality of the instrument and to validate the predictions of the instrument forward model in Sect. 2.2. The second is a measurement of the emissions of the German coal power plant Großkraftwerk Mannheim (GKM). Section 5 concludes.
2.1 Gas correlation spectroscopy
The absorption of light is described by the Lambert–Beer law. It states that for a given incident spectral radiance L0(λ) the spectral radiance L(λ) after traveling along a light path s through absorbing media with absorption cross sections σk(λ) and concentrations ck is given by
Here, in units of molec. cm−2 denotes the column density of the absorbing medium k in the atmosphere and τ is the resulting optical depth. In our application, L0 denotes the radiance spectrum of scattered sunlight. The Lambert–Beer law can be applied to radiances, denoted with L in units of W nm−1 m−2 sr−1, as well as to irradiances, denoted with I in units of W nm−1 m−2. In the following, all absorption cross sections are considered constant; i.e., their slight dependence on pressure and temperature is neglected.
The pixels of a photosensor do not resolve spectrally. Let μp(λ) be the number of photons per wavelength interval and time period in units of ph nm−1 s−1 that a photosensor is exposed to. It will then measure a detector signal J in units of photoelectrons (e−), given by the spectral and temporal integral
where η in units of e denotes the quantum efficiency of the photosensor and texp the exposure time. The wavelength dependence of η and the spectrum of the light source (typically scattered sunlight) usually restrict the integration to the near-ultraviolet (UV), visible, and near-infrared regions of the electromagnetic spectrum. μp(λ) can be expressed as
where L0 denotes the radiance spectrum of the light source, T denotes the transmission of the instrumental setup, e−τ(λ) describes all absorption along the light path according to the Lambert–Beer law, and E denotes the étendue of the instrument in units of mm2 sr. The factor converts radiant flux in units of watts (W) to photon counts per time, i.e., ph s−1, where λ denotes wavelength, and J m denotes the product of Planck's constant and the speed of light.
Figure 2 explains the principle of GCS, assuming (for the sake of simplicity) that the target gas with column density S and absorption cross section σ(λ) is the sole absorber and thus . Two camera modules are placed behind two gas cells, one of which is filled with air (the “empty” cell) and one is filled with a high concentration of the target gas (see Fig. 1).
For a detector pixel with the indices (i,j), the camera with the empty cell will measure
and the camera with the cell containing the target gas will measure
where S(i,j) denotes the column density of the target gas in the FOV of the pixel with indices (i,j) and Sc the column density of the target gas in the gas cell of the instrument. The two measurements J(i,j) and can be interpreted as spectral channels in analogy to the widely used DOAS terminology. In imaging GCS, the instrument response (instrument signal),
is computed for each individual pixel. is the logarithmic signal ratio between the two spectral channels of the instrument and functions as a measure of S(i,j): when S(i,j) is small, incoming light will be only slightly attenuated before it reaches the cells, and thus the signal ratio will be smaller compared to a scenario in which S(i,j) is large, and thus the atmospheric target gas has already attenuated a larger portion of the light that would otherwise have been absorbed by the gas cell. It therefore follows directly that grows monotonically with S(i,j).
When using two camera modules with distinct optical setups, the resulting detector signals are highly sensitive to imperfections in the optical path. For example, small differences in the focal lengths of the camera lenses or dust particles on the lenses or gas cells can induce significant false signals, contributing to . Furthermore, vignetting is immanent to imaging measurements and manifests itself in increasing false signal gradients towards the corners of the image. Even with entirely identical optical setups, the two camera sensors may have slightly different pixel response nonuniformity (PRNU) maps. These effects can be partly corrected by recording reference signals for the channel with the empty cell and for the channel with the filled cell in the zenith direction, where S=0 is assumed. In reality this latter condition does not need to be perfectly fulfilled, although it is important that S is approximately constant throughout the FOV for the reference images. In analogy to Eqs. (7) and (8) the reference signals are given by
and
The measurement signal ratio is then divided by the reference signal ratio, i.e.,
This procedure is also referred to as flat-field correction. In the following section it will be shown that in good approximation holds.
Furthermore, Eq. (12) points towards a crucial benefit of the proposed measurement principle. While other correlation methods for remote sensing typically operate with two channels in different spectral domains (e.g., an on- and off-band channel in filter-spectroscopy-based SO2 cameras), the spectral domain of the two channels is identical in GCS. Additionally, that domain is typically restricted to a few dozen nanometers using a bandpass filter. This makes the instrument insensitive to broadband extinction, i.e., by Rayleigh scattering or due to aerosols, given that their extinction coefficients vary only very slightly throughout the spectral domain the instrument operates in. The instrument response to broadband extinction is examined numerically in Sect. 2.2.
2.2 Instrument model calculation
A numerical forward model was implemented to predict the characteristics of a GCS-based NO2 camera. Specifically, we investigate the shape of the instrument response, the calibration curve, and the signal-to-noise ratio (SNR) as a function of S and Sc, as well as cross-sensitivities to other atmospheric trace gases. This section specifically discusses the application of GCS to measurements of NO2. Other trace gases may, for example, require operating in a different spectral range. Overall, the simulation of realistic conditions of daytime measurements in the atmosphere is the aim. For this, a spectrum of scattered sunlight is used as the light source and atmospheric NO2 column densities are considered in the range from 1016 to 1018 molec. cm−2, as well as integration times on the scale of seconds. The assumed range of NO2 column densities is justified as follows: in order to measure column densities much lower than 1016 molec. cm−2, the exposure time would need to be increased significantly, resulting in poor temporal resolution. At the same time, even strong NO2 pollution in the atmosphere typically does not exceed 1018 molec. cm−2, assuming realistic viewing geometries.
The relevant detector signals are modeled according to Eqs. (7), (8), (10), and (11). In this instrument model, we assume , where Tf denotes the transmission of the bandpass filter and Tl denotes the transmission of the camera lens. Since texp is realistically small enough that I0(λ) is constant throughout exposure and the transmission of the bandpass filter used is effectively a cutoff function outside its transmission band from 430 to 445 nm, the detector signals can be simplified to
The choice of this particular bandpass filter is motivated by the strong, characteristic absorption features that NO2 shows in its transmission range. The absorption cross section of NO2 (Vandaele et al., 2002) is displayed in Fig. 3a with a zoomed-in region close to the transmission band of the bandpass filter.
The model requires a light source radiance spectrum L0. For realistic applications of the instrument the light source will almost exclusively be an atmospheric background spectrum, i.e., a radiance spectrum of scattered sunlight. We use a highly resolved irradiance spectrum in units of W nm−1 m−2 (Chance and Kurucz, 2010) and scale it with a low-resolution radiance spectrum at 400 nm (Pissulla et al., 2009) in units of W nm−1 m−2 sr−1. This way we obtain a radiance spectrum that represents the typical spectral shape of scattered sunlight but maintain the high spectral resolution of the irradiance spectrum. We argue that this is the most realistic general estimation of the background spectrum that we can make. The radiance spectrum used for scaling was recorded at Thessaloniki, Greece, at a sun zenith angle of 21∘. The transmission lines of the bandpass filter Tf(λ) and the camera lenses Tl(λ), as well as the quantum efficiency of the camera sensors η(λ), are provided by the manufacturers. An étendue of sr mm2 was assumed throughout, which was computed on the basis of a fully opened aperture (f number 1.6). Figure 3b shows plots of L0, Tf, Tl, and η. In the following we assume an exposure time of 2 s. The detector signals J, Jc, Jref, and Jc,ref are then calculated by numeric integration, according to the instrument model as described. Figure 4 shows the modeled instrument response (see Eq. 12) as a function of the column density S in the range from 1016 to 1018 molec. cm−2 for different choices of the column density Sc inside the NO2 cell of the instrument.
The instrument response is in good approximation proportional to S. The instrument calibration factor k can be obtained for any fixed value of Sc by sampling the instrument signal for different choices of S and fitting a linear function of the form
to the samples. In order to convert the unitless instrument signal to column densities, the inverse k−1(Sc) in units of molec. cm−2 is used. During measurements, Sc must be determined so that k−1(Sc) can be computed. For this purpose, Sc could be directly measured using a second instrumental setup, such as a DOAS instrument. However, in many measuring scenarios it is more practical to determine Sc on the basis of the acquired images alone. For this purpose, an off-plume region of the imaged scene, where S=0 is assumed, is used, and Sc is approximated by
where cm2 molec.−1 is the absorption cross section of NO2 averaged over the spectral range from 430 to 445 nm. The validity of this approximation was verified numerically, as displayed in Fig. 5. For a cell column density of molec. cm−2 (this value will be reasoned in the following paragraph), the proposed approximation underestimates the true value of Sc by less than 2⋅1017 molec. cm−2.
With this model we can also quantify the signal-to-noise ratio (SNR) in order to estimate the detection limit of the instrument under typical atmospheric conditions. An SNR of 1 is assumed to be the lower limit at which atmospheric column densities of the target gas can be resolved. Photoelectron counting follows Poissonian statistics; i.e., the uncertainty ΔJ of a signal measured by a photosensor is . Thus, the uncertainty of the instrument signal can be expressed in closed form by application of Gaussian uncertainty propagation:
In practice the uncertainties of the reference signals will be comparably small because the exposure time for the recording of Jref and Jc,ref can be chosen to make the contribution of and negligible. Then the uncertainty reduces to
and the SNR can be expressed as
This instrument model only accounts for the photon shot noise and disregards additional possible sources of noise such as dark noise and read-out noise of the photosensors. This is on purpose in order to make the model applicable to different instrumental setups. In practice the shot noise is by far the dominating source of noise due to the large light throughput of the setup, and both dark current and dark noise can be neglected (see Sect. 3 for a more detailed explanation). Figure 6a shows the modeled SNR as a function of the cell column density Sc for different choices of the column density of the target gas S. The highest SNR is reached at approximately molec. cm−2 with a slight dependence on the observed target gas column density S. Figure 6b shows the modeled SNR as a function of the target gas column density S. The red horizontal line marks the resulting detection limit, with SNR=1. With an ideal choice of molec. cm−2, a detection limit of approximately 2⋅1016 molec. cm−2 is reached with an exposure time of 2 s.
The instrument model also allows for the study of the selectivity of the instrument. Equation (13) holds under the assumption that the target gas is the sole absorber. In a realistic measuring scenario many different trace gases other than NO2 could be present in the atmosphere. Cross-sensitivities to other trace gases can be determined on the basis of the instrument model. We define , the false signal of a species X, as the additional contribution to the overall instrument signal that is due to the absorption of X and present the results of a study on the false signals of water vapor (H2O; absorption cross section was taken from Rothman et al., 2013) and the oxygen collision complex (O4; absorption cross section was taken from Thalman and Volkamer, 2013), since both species show possibly relevant absorption features in the spectral range our instrument operates in. Figure 7 shows the absorption cross sections of NO2, H2O, O4, and the transmission line of the bandpass filter used.
The bandpass filter blocks almost all light of wavelengths greater than λ=445 nm. Therefore, most of the O4 absorption is filtered out and is strongly reduced. Water vapor, on the other hand, shows strong absorption features between 440 and 445 nm. Calculating the false signals of the two species requires an assumption of their atmospheric abundance. In reality these column densities can vary strongly with place and time. We therefore use the model to make predictions of the cross-sensitivities assuming large but still realistic column densities of the cross-sensitive species. If the predicted false signals are sufficiently small, the cross-sensitivities can be neglected altogether because the model has then realistically overestimated the induced false signals. For O4 a maximum column density of 1044 molec.2 cm−5 at a light path length of 10 km was assumed. For reference, Peters et al. (2019) report maximal O4 column densities of around 5⋅1043 molec.2 cm−5 during the CINDI-2 measurement campaign. For H2O a maximum column density of 6⋅1023 molec. cm−2 was assumed. This corresponds to a relative humidity of 100 % at a pressure of 1 atm, temperature of 20∘ C, and a light path length of 10 km.
Figure 8 shows the modeled false signals of H2O and O4. The false signal was converted to NO2 column density equivalents using the calibration of the model obtained from Fig. 4, assuming a cell column density of molec. cm−2. Both species induce a negative false signal. When expressed in NO2 signal equivalents, the false signal of O4 is comparably small, reaching around molec. cm−2 assuming the maximal column. The false signal of H2O is an order of magnitude larger, reaching up to molec. cm−2 assuming the maximal column. As discussed, we treat these false signals and the column densities that have generated them as an overestimate of a realistic expectation. In addition, the naturally abundant water vapor of the atmosphere is typically distributed much more homogeneously than strong NO2 concentration gradients from a point source. Under this circumstance any false signal induced by water vapor should be easily separable from the NO2 signal of interest. Water vapor inside the plume of a point source emission, which cannot be separated from the NO2 signal by the argument above, is contained within much shorter light paths (typically on the order of 100–200 m) and is not expected to induce relevant false signals.
In addition to water vapor and O4 the modeled instrument response to broadband extinction was investigated. Rayleigh scattering has a wavelength dependence of λ−4, while extinction due to larger particles shows weaker wavelength dependence. It was verified in a numerical experiment that the instrument response curves displayed in Fig. 4 vary by less than 0.5 % when the assumed irradiance spectra are scaled by λ−4 and λ0=1. This demonstrates that the GCS-based NO2 camera is practically insensitive to broadband extinction. Beside the numeric model presented here, an analytic model was developed as well (see Appendix A).
We have built an instrument prototype based on commercially available hardware. The camera modules use a monochrome progressive-scan CMOS sensor in a in. format with a pixel size of 5.86 µm×5.86 µm and a global shutter. They record images with 1920×1200 (height × width) pixels. A charge signal is digitized by a 16-bit analog–digital converter (ADC). The cameras connect via USB 3 to a controlling computer equipped with corresponding camera software. Image acquisition rates depend on the selected exposure time and the read-out time tread=24.4 ms of the camera sensors. The instrument is therefore limited to a frame rate of 41 FPS at best. However, the read-out time tread can be reduced by using windowing, a feature in which the cameras are advised to only read out a subrange of their sensor arrays. The usability of windowing depends on the imaged scene and whether large parts of the FOV can be neglected. The camera modules have a read-out noise of 7 e−. The thermal dark signal of the camera modules was determined experimentally according to the European Machine Vision Association (EMVA) (see Jähne, 2010). A thermal dark signal of (24±9) e− s−1 at a sensor temperature of 50 ∘C, which is approximately the average operating temperature of the camera modules due to their small form factor, and a doubling temperature of (6.1±0.1) ∘C were found. The camera modules have a full-well depth of 34 000 e−. Given that in bright daylight the exposure times for images within the dynamic range of the camera are typically far below 1 s, the contribution of the dark signal to the total measured camera signal is negligibly small (e.g., below 0.05 % for an exposure time of 30 ms and a sensor saturation of 50 %). Also, the total dark noise (meaning read-out noise + thermal noise) is negligible compared to the photon shot noise of around 130 e− at 50 % saturation. From a technical perspective the retrieval of the camera data follows the typical pattern of digital imaging: inside the camera modules, the incoming photons detach electrons from the semiconductor material of the camera chip (characterized by η). That charge is digitized (characterized by the fixed ADC gain K in units of e− ph−1) and saved as 16-bit grayscale image files. Each camera is equipped with a lens with a focal length of f=25 mm. The full diagonal, vertical, and horizontal opening angles amount to 30, 16, and 25.5∘, respectively. For each camera a bandpass filter with transmission in the range from 430–445 nm was placed between the camera lens and the camera sensor. The gas cells of the instrument are cylindrical with a diameter of 50 nm and a thickness of 10 nm. The NO2 cell was filled from a large reservoir to contain an NO2 column density of 4⋅1018 molec. cm−2 (which is the ideal value according to the results shown in Sect. 2.2, specifically Fig. 6a). The camera behind the NO2 cell is mounted to a tiltable stage, which can be used to adjust its optical axis in vertical and horizontal orientation with milliradian (mrad) precision using two thumb screws. This adjustment is of crucial importance in order to eliminate shifts in the FOVs of the two cameras. All parts are placed inside a closable plastic case. Overall, the instrument is portable and compact while maintaining a reasonable cost of below EUR 2000. Control software with graphical user interface was developed in the Python programming language.
4.1 Proof-of-concept measurement with gas cells
In order to validate the instrument model described in Sect. 2.2 a simple laboratory experiment was performed. Four glass cells were filled with different concentrations of NO2 and measured with both the NO2 camera and a conventional DOAS setup. The light source for the camera measurement was a halogen lamp inside an integrating sphere in front of which the cells were mounted onto a stand with a clamp. An additional series of images was recorded without a cell in the light path, whose average serves as the reference image (Jref, Jc,ref; see Eq. 12). When evaluating the images taken by the NO2 camera, an in-cell pixel set and a background pixel set were defined. The in-cell pixel set contained the pixels inside the cell, while the background pixel set contained pixels of the illuminated entrance of the integrating sphere, not covered by the cell. Due to the varying size of the test cells, the in-cell and background pixel sets were different for each cell. The total acquisition time of the NO2 camera was set to 3 min for each cell, and the exposure time of each camera was chosen such that the camera sensors saturated to approximately 50 %.
First, the column density inside the gas cell of the NO2 camera was estimated as
where and are the camera signals of the camera with the empty cell and the one with the filled cell, respectively, averaged over the background pixels of all images. cm2 molec.−1 is the absorption cross section of NO2 averaged over the spectral range from 430 to 445 nm. A cell column density of molec. cm−2 was obtained. The cell was originally filled with molec. cm−2, but this deviation can be explained by the temperature-dependent NO2 ⇌ N2O4 equilibrium. The lower the temperature, the lower the NO2 concentration inside the gas cell. The calibration of the instrument was obtained from the instrument model as explained in Sect. 2.2. The fit procedure yielded a calibration factor of molec. cm−2. Additionally, the signal offset of the instrument was calculated from the background pixels, which was defined as
Subtraction of from the instrument signal set the average background pixel to zero. The instrument signal of a test cell was determined by averaging over the pixels that were covered by the cell, i.e.,
where J and Jc denote the camera signal with the empty cell and with the filled gas cell, respectively, in the in-cell pixel region.
The uncertainty of these measurements is given by Gaussian error propagation according to Eq. (17). The uncertainties and are obtained by computing the standard deviation of the detector signal in the in-cell region for the two channels, respectively. Figure 9a shows an exemplary image of this measurement. In the center foreground of the image the outline of test cell no. 4 as well as the stand and clamp used to hold it are shown. The offset was subtracted and the flat-field correction was applied using the reference images according to Eq. (12). The camera measured a signal of in the in-cell region of the test cell. Using the calibration factor k−1, a column density of molec. cm−2 was obtained. Within the uncertainty of the measurement this result coincides with that of the DOAS instrument, which measured a column density of molec. cm−2.
Table 1 lists the column densities measured for each cell by the DOAS setup and the NO2 camera. The measurements taken with the NO2 camera show significant uncertainties. For cell no. 1, the relative uncertainty is as large as 230 % and the detection limits, ranging from 2.29⋅1016 molec. cm−2 to 8⋅1016 molec. cm−2, are larger than the prediction of the instrument model, which was 2⋅1016 molec. cm−2 at 2 s of exposure. The reason for this deviation is the use of a different light source: while the instrument model assumed scattered sunlight as the light source, a halogen lamp inside an integrating sphere was used for this experiment. The detection limit is mainly determined by the overall intensity of the light source, which is much lower for such a halogen lamp in the blue spectral range. This increased the statistical uncertainty of the measurement. Additionally, systematic false signals were observed, which were not considered in the instrument model: due to the small diameter of the test cells and the limited interior space of typical optical laboratories there are inevitable perspective shifts between the images of the two cameras when they are oriented so that the test cells are in the center of their FOVs. Small dust particles on the test cell or condensed droplets on its inside can then introduce false signals. In order to smooth out these false signals, the images were convoluted with a rectangle filter of the same size as the average diameter of the observed structures. Table 1 lists the chosen filter size for each cell. The filter sizes were chosen differently for each cell because the larger cells required less smoothing. The cell image shown in Fig. 9 required no smoothing at all (which corresponds to a filter size of 1 pixel). Figure 10 shows a scatter plot of the instrument response of the NO2 camera against the column density measured with the DOAS setup for each test cell. Additionally, the prediction of the instrument model (see Sect. 2.2) with cell column density molec. cm−2 is plotted. The resulting instrument responses to the test cells are in very good agreement with the instrument model, with an average relative deviation of 18 %. The model and measurement coincide for all test cells within the uncertainties of the measurement. Given the overall good agreement between the DOAS instrument, the NO2 camera, and the instrument model, we take these results as proof of concept.
4.2 Measuring the emissions of the coal power plant Großkraftwerk Mannheim
4.2.1 Setup and methodology
We report measurements taken at the Großkraftwerk Mannheim (GKM) with the NO2 camera. The GKM is a power plant located in Mannheim, Germany, which generates electricity based on burning of bituminous coal. It is one of the largest power suppliers of southwestern Germany. The European Pollutant Release and Transfer Register (E-PRTR) lists an emission of 2 890 000 kg of NOx in 2017 (The European Commision, 2017).
The NO2 camera was set up at Backofen-Riedwiesen, 3.6 km south of the GKM (at 49.417745∘ N, 8.505917∘ W; see Fig. 11), on 26 April 2021. The sky was cloud-free on that day. The FOV of the camera at 3.6 km of distance was approximately 1.77 km wide and 1.10 km high. However, it was decided to decrease the read-out time of the camera modules by using windowing (see Sect. 3). Therefore, the true FOV was reduced to 1.22 km width and 0.53 km height. The camera was positioned so that the plume emitted by GKM block 7 was in the center of the FOV. The optical axes of the two cameras were aligned so that no shifts between their images were visible. The measurement started at 08:44 UTC+2. Reference images of the sky at a 45∘ elevation angle were recorded in regular intervals.
During the measurement the camera with the empty cell recorded with an exposure time of texp=2.7 ms and the camera with the NO2 cell recorded with an exposure time of ms. Additionally, the cameras had a read-out time of 10 ms. The exposure times were chosen so that the camera sensors were read out once they were saturated to about 50 %. In order to increase image rate and reduce data volume, 100 consecutive frames were averaged, and these averages were saved. We refer to them as images consisting of 100 frames. This way an image acquisition time of 2 s per 100 frames was achieved. The reference images were recorded in the same manner, although with exposure times ms and ms. This procedure yielded a total of four images J, Jc, Jref, and Jc,ref. The resulting instrument signal image was then computed according to Eq. (12), with all arithmetic operations and the logarithm applied pixel-wise. In order to obtain sensible results, a few corrections had to be applied.
Firstly, the logarithm of the exposure time ratio
was subtracted in order to account for the fact that all four images were acquired with different exposure times.
Secondly, a background image was subtracted, for which the procedure and reasoning are described in the following. The background image was obtained by fitting a 1D polynomial of degree n to each column of a manually selected set of background pixels, obtained by using a freehand selection tool on the images. This was required because the camera signal images showed large signal gradients across the FOV. We suspect that these gradients are a side effect of the flat-field correction, possibly because the sky, against which the reference images were taken, is generally not radiometrically uniform. An exemplary background correction procedure with n=2 is shown in Fig. 12. The original signal image without background correction, as well as our manual choice of the plume and off-plume regions, is displayed in Fig. 12a. Figure 12b shows the background fit on the basis of that choice. Figure 12c shows the resulting instrument signal image, with a clearly visible plume signal. Figure 12d shows that for an exemplary column at y=660, the background fit tailored very closely to the off-plume region and left a residual in the plume region. Subtraction of the fit made the plume signal visible in the residual, which can be seen in Fig. 12e. A weak temporal dependence of the background image was observed, possibly due to changes in the relative position of the sun (see Fig. 13).
Thirdly, a scalar signal offset was subtracted. The column density inside the instrument's cell Sc is expected to vary over the course of the measurement. If the two measurement signals J and Jc undergo flat-field correction by the two reference signals Jref and Jc,ref, unless all signals are recorded with the exact value of Sc, a constant signal offset will add to . For studying time series it is important that this effect is accounted for, i.e., the signal in an off-plume reference region is forced to remain constant, which can be achieved by subtraction of a suitable estimate of . Here, was computed by averaging the signal over a small rectangle in the off-plume region (the patterned rectangle in Fig. 14) for each image individually.
Finally, the resulting signal images were multiplied with the calibration factor k−1, which was obtained from the instrument model (see Sect. 2.2). This required knowledge of Sc. Sc was therefore estimated according to Eq. (19), considering the same background rectangle as in the calculation of , and a value of molec. cm−2 was obtained. With all corrections included, a single camera image was computed via
where each pixel value S(i,j) corresponds to the NO2 slant column density (SCD) measured at pixel (i,j).
4.2.2 Evaluation of an individual camera image
Figure 14 shows the first camera image of the series, calculated according to Eq. (23). To obtain this image, the first six consecutive images of the series were averaged. A background fitting routine with polynomial degree n=2 and the same fit mask as displayed in Fig. 12a were used (the choice of this fit mask is discussed further at the end of this section). A positive NO2 plume signal equalling approximately 5⋅1016 molec. cm−2 was observed to be emitted from the chimney of block 7. At the point of emission, i.e., directly above the chimney (at width 1000 m), the plume was in a fully condensed phase and the instrument signal image shows structures of strong negative and positive signal. This effect can be explained as a consequence of the optical setup inside the instrument: the optical axes of the two cameras inside the instrument were adjusted so that there was no displacement of the imaged objects (i.e., the uncondensed part of the plume) in the center of the FOV. However, displacements towards the corners of the FOV could not be avoided. These displacements manifest themselves as strong false signals when the signal ratio of the two cameras is computed. Given that in this measurement they occurred in an image region of low interest, they were deemed unavoidable and not considered any further. In order to obtain the NO2 SCD and the diameter d of the plume systematically, each column of the NO2 camera signal image was considered as an individual vertical cross section through the plume. It was observed that the shapes of the measured NO2 SCDs along these cross sections coarsely followed that of a Gaussian. Figure 15a shows this observation for an exemplary column at y pixel 660. To each image column j, a Gaussian with amplitude Aj, mean μj, and standard deviation σj was fitted. The NO2 slant column density and the diameter of the plume at column j were then associated with Aj and 2⋅σj, respectively. Columns for which the fit routine did not converge well were ignored. This was considered the case when either the fit failed to converge entirely or the retrieved fit parameters were outside a realistic range ( molec. cm−2 or m), which was the case for approximately 50 % of the columns. The resulting NO2 SCDs and plume diameters are shown in Fig. 15b. The ensemble of all column fits allows for the calculation of an average plume SCD of molec. cm−2 and an average plume diameter of m. These values are represented by the red marker in Fig. 15b.
4.2.3 Uncertainty analysis
It is necessary to discuss the uncertainties of such an evaluation procedure. It was explained in Sect. 2.2 (see specifically Eq. 17) that the measurement has an intrinsic uncertainty of the uncalibrated camera signal due to the Poissonian error of photon counting. This uncertainty propagates directly onto the NO2 SCDs that are obtained upon calibration of the instrument using k−1(Sc) as described in Eq. (23) and was estimated by computing the standard deviation of the measured NO2 SCDs in an off-plume region of a camera image, e.g., the patterned rectangle in Fig. 14. A value of molec. cm−2 was obtained. This is the detection limit of the instrument prototype. In the next step the plume SCDs and diameters were obtained in a Gaussian fit routine for the vertical plume cross sections of all image columns. For a single column j, this introduced additional uncertainties ΔAj, Δμj, and Δσj, which were given by the covariances of the fit parameters of that column. These uncertainties propagate into those of the means over all columns, producing the uncertainties used above ( molec. cm−2 and Δd=17 m). Finally, the uncertainties of the background fitting routine as described in Sect. 4.2.1 were investigated. The camera image shown in Fig. 14 was calculated according to Eq. (23), where was computed using a polynomial of degree n=2 and the same fit mask as displayed in Fig. 12a. Given that this choice of n and the fit mask are subject to our personal assessment, we investigated how much the obtained NO2 SCD and diameter of the plume vary with different choices of n and the fit mask. Figure 16 shows the results of this analysis. Figure 16a–h show the process of the background fitting routine using the freehand fit mask that was described earlier. Figure 16c, e, and g show the resulting background images for , and 4, respectively. Figure 16d, f, and h show the corresponding scatter plots of NO2 SCD and plume diameters as obtained from the Gaussian fit routine. Figure 16i–p show the same procedure with a different fit mask, namely one that makes no assumptions of the plume position and covers the entire FOV. The case n=1 was dismissed, seeing that the background signal is clearly not linear (see Figs. 12 and 13). Intercomparison of Fig. 16d, f, and h as well as Fig. 16l, n, and p shows that for a given fit mask the average NO2 SCD and plume diameter do not vary significantly with the choice of n. Using a full-FOV fit mask yields significantly smaller average values of NO2 SCD and plume diameter. Furthermore, image objects such as the condensed plumes at y pixels 400 and 1200 lead to vertical fragments in the background image (see Fig. 16j). Overall, the background fitting procedure with n=2 and a freehand selection of the plume as displayed in Fig. 16c and d seems to be a sensible choice because the resulting background image does not suffer from vertical fragments and shows fewer signal variations in the off-plume region. In addition, the fit is fastest to compute for n=2. Table 2 contains a quantitative summary of these findings and allows for the estimation of the uncertainty of the background fitting routine. The uncertainty of the NO2 SCDs spans from molec. cm molec. cm−2 to molec. cm molec. cm−2. The mean is 4.04⋅1016 molec. cm−2. Therefore, the overall uncertainty can be estimated as molec. cm−2. In analogy an uncertainty of Δd=34 m for the plume diameter is obtained, which will be used throughout the rest of this paper. With this method an estimate of the overall uncertainty of the evaluation is obtained by including not only the statistical uncertainty of the measurement (noisy data), but also the systematic uncertainty that is immanent to the evaluation method. In the future, more elaborate methods for the separation of the plume and background should be investigated. Generally, this would be achieved by image segmentation, for which a variety of methods exists. However, finding an ideal method that generalizes to other plume shapes and viewing geometries would require a study on its own.
A series of camera images was assembled into a video (see Video supplement), which shows the movement of the plume in the wind direction from 08:53 to 09:05 UTC+2.
4.2.4 Optical flow and mass flux analysis
A mass flux analysis was carried out on the basis of image sequences. Given a camera image as shown in Fig. 14, the mass flux through a vertical cross section of the plume can be computed as
where g mol−1 is the molar weight of NO2, mol−1 the Avogadro number, v the wind speed in the horizontal direction, and S the column density, which is integrated along the vertical (height) axis. v was obtained by running a Farnebäck optical flow retrieval (Farnebäck, 2003) on the in-plume region of consecutive camera images. The optical flow was then divided by the time difference Δt between the images. Figure 17a shows the wind speeds associated with the camera image in Fig. 14. For this image and its successor, a mean horizontal wind velocity of m s−1 was obtained. The average was considered over the plume region only because in the still background the Farnebäck algorithm cannot detect any flow and returns a wind speed of 0. Similar to the column-wise evaluation of the NO2 SCD and plume diameter in Sect. 4.2.2, the NO2 mass flux was computed through each column separately, according to Eq. (24).
Figure 17b shows the NO2 mass flux obtained through the individual columns of the image that was displayed in Fig. 14, plotted against the distance traveled downwind from the point at which the fully condensed part of the plume ended (see Fig. 14 or 17a at width 840 m). This procedure yielded a mean mass flux of kg h−1. The evaluation was extended to obtain average wind speeds and mass fluxes as a function of time. The results are displayed in Fig. 18. Figure 18a shows the mean horizontal wind speed and Fig. 18b shows the mean NO2 mass flux.
Over the observed time frame from 08:55 to 9:25 UTC+2, an overall mean horizontal wind speed of m s−1 and an NO2 mass flux of kg h t yr−1 were obtained.
A combination of several publicly available sources can be used to estimate a reference value for the NOx mass flux of the GKM, which can be compared to the value measured here. Of course, the NO2 camera data only allow computation of the NO2 mass flux, not the NOx mass flux. However, the large FOV of the camera covers a total distance of up to 1000 m downwind from the point of emission. It can therefore be expected that the main chemical conversion processes (see Reactions R1–R3) have reached equilibrium and the Leighton relationship is valid. In that case the mass fluxes of NO2 and NOx should be of comparable magnitude.
The Fraunhofer Institute for Solar Energy Systems (ISE) reports that the GKM was producing 70.6 MW at 09:00 UTC+2 on the day of measurement (see Fraunhofer Institute for Solar Energy Systems, 2021). The European Pollutant Release and Transfer Register lists an NOx emission of the GKM of 2890 t yr−1 in 2017 (The European Commision, 2017). The business report of the GKM of the same year states a mean power production of 1119 MW (Großkraftwerk Mannheim Aktiengesellschaft, 2018). Therefore, the GKM should have been running at approximately 6.3 % of its average power. Assuming that the NOx emission scales linearly with the power produced, an NOx mass flux of Fm=182 t yr−1 is expected. The mean mass flux obtained from the camera data is significantly lower and amounts only to about one-third of this reference value. Given that the reference is an NOx mass flux and the NO2 camera can only detect the NO2 mass flux, such deviations are expected. The NO2NOx ratio of the plume is further investigated in Sect. 4.2.5.
It should be taken into account that this analysis contains two further uncertainties: firstly, although the most recent available data were used, there may be differences in the reference values between 2017 and 2021 (e.g., total mass of yearly emitted NOx or mean power production). The E-PRTR data show a decline in total yearly emitted NOx from 2007 to 2017, and it can be expected that this trend has continued until 2021. It should be taken into account that a comparison between a mean flux observed in a time frame of 30 min and a yearly average reference flux is hardly indicative for the accuracy of our measurement. Secondly, GKM block 7, the emitted plume column densities of which were used for this analysis, was not the only active block at the time of the measurement. During the measurement, emissions from GKM blocks 6 and 8 were observed as well, but the FOV of the NO2 camera was too small to record the plumes emitted from all blocks simultaneously. It is plausible to assume additional emissions of NO2 from GKM block 6 and 8, which could not be examined on the basis of our measurement.
Although the discussed uncertainties do not allow for a definite conclusion on the overall accuracy of the mass flux analysis, we present the results as a demonstration that flux analyses on the basis of image data with high spatiotemporal resolution are feasible.
4.2.5 Estimation of [NO2] [NOx] ratios
The camera images can be used to investigate the conversion of NO to NO2 by the reaction of NO with ambient ozone (see Reaction R1) and direct oxidization by molecular oxygen (see Reaction R2). The NO ratio can be modeled according to Janssen et al. (1988) by the formula
where x is the distance downwind from the point of emission, and [NO2], [NOx], and [O3] denote the concentrations of NO2, NOx, and O3. The model has a parameter , where k is the rate constant for the reaction and v the wind speed, as well as another parameter , where J is the photodissociation frequency of NO2. The rate constant k(T) is temperature-dependent. Lippmann et al. (1980) find the empirical relationship
with temperature T. The photolysis frequency J is often cited as approximately in full sunshine (see, e.g., Platt and Kuhn, 2019) but varies strongly with irradiance (Parrish et al., 1983).
Figure 19 shows an approach to compare the camera measurements with the Janssen model. The parameters of the Janssen model are determined by the wind speed v, the ozone concentration [O3], the photodissociation frequency J, and temperature T. For the wind speed v=0.94 m s−1 was assumed, as obtained from the optical flow procedure in Sect. 4.2.4. The remaining parameters (ozone concentration, photolysis frequency, and temperature) were obtained by fitting the Janssen model to the measured data points. For this, the first 1000 images of the series were averaged (this amounts to a time window from 08:45 to 09:30 UTC+2). Then the vertical integrals of the plume SCD ∫S(h) dh were computed for each individual column, like in the mass flux analysis in Sect. 4.2.4. The concentration ratio associated with each image column was obtained by normalizing this set of integrated SCDs into the interval [0,1]. This is in accordance with the Janssen model, which predicts an initial concentration ratio of 0 with an exponential convergence towards a concentration ratio of ≤1, depending on the model parameters. Figure 19 shows these obtained ratios as black dots plotted against the distance downwind, measured from the point at which the fully condensed part of the plume ended (see Fig. 14 at width 840 m). By running a least-squares fit routine, an ozone mixing ratio of [O3]=17.5 ppb, a temperature of 13.9 ∘C, and a photolysis frequency of s were obtained. As a reference, the closest ground-based air quality measuring station (Mannheim-Nord, DEBW005) measured an ozone mixing ratio of 26.8 ppb at 09:00 UTC+2 (Landesanstalt für Umwelt Baden-Württemberg, 2021). However, it should be taken into consideration that such ground-based measurements may not yield representative values for 200–500 m of altitude. Moreover, temperatures of up to 17.3 ∘C were reported in Mannheim for the day of our measurement (Deutscher Wetterdienst, 2021). Parrish et al. (1983) report similar values of J at solar zenith angles of approximately 60∘, while the solar zenith angle at the beginning of our measurement was 78∘.
Overall, the data points in Fig. 19 coarsely resemble the shape of the Janssen model. However, they oscillate around the prediction of the best fit (red dotted line in Fig. 19). The cause of these oscillations is possibly the alignment of the optical axes of the cameras inside the instrument. It was explained earlier that the camera axes were aligned so that no shifts occur in the center of the FOV due to the displacement of the two cameras. However, shifts towards the corners of the FOV are then inevitable. It was observed that such shifts typically lead to patterns of consecutively increased and decreased false signal in the signal ratio image. The plateau after 400 m of downwind distance agrees with Janssen models assuming ozone mixing ratios of 15–30 ppb. Although such mixing ratios are relatively low for typical polluted urban areas, they are within a realistic order of magnitude. It should be taken into consideration that more recent studies have found initial concentration ratios of 5 %–10 % to be more realistic for the emission from most combustion processes (see, e.g., Kenty et al., 2007, and Carslaw, 2005). This is neglected by the Janssen model, which predicts an initial ratio of zero. Furthermore, as discussed in Sect. 4.2.2, the NO2 camera is incapable of measuring the NO2 SCD of the plume directly after its emission when it is still in a fully condensed phase (see Fig. 14). Figure 19 shows the concentration ratio against the distance downwind, which is measured from the point at which the fully condensed part of the plume ended (at a width of 840 m in Fig. 14). The evaluation shown here neglects the plume chemistry of this early phase. To conclude, a crucial uncertainty is the mapping of the column-wise vertical SCD integrals onto the interval [0,1] on both ends: at the lower end, near the point of emission, the concentration ratio is unmeasurable due to the phase of the plume. At the upper end, far downwind, the mapping could be slightly off due to the oscillations of the measured data points. However, we notice good agreement between the obtained fit parameters and the reasonably picked reference values listed earlier in this section.
We present a prototype of a novel NO2 imaging instrument based on gas correlation spectroscopy: the GCS NO2 camera. It operates by recording images with two cameras, each with a gas cell (cuvette) in front of it; one is filled with air and the other filled with a high concentration of NO2. The instrument acquires images at high spatiotemporal resolutions of up to FPS and 1920×1200 pixels. The instrument response to a wide range of target column densities, ranging up to 1⋅1018 molec. cm−2, has been examined in a numerical instrument model. A linear instrument response has been observed within that range, making the instrument easy to calibrate. An examination of the signal-to-noise ratio has shown that the ideal NO2 column density in the gas cell of the instrument is approximately 4⋅1018 molec. cm−2. Furthermore, under realistic conditions, a detection limit of about 2⋅1016 molec. cm−2 is expected, which was later confirmed using the instrument prototype. In its current form the instrument is easily transportable and highly cost-efficient with a build price of less than EUR 2000.
A study on the cross-sensitivity to trace gases other than NO2 was carried out for water vapor and O4. Under the assumption of realistic column densities of these species the magnitude of the cross-sensitivity of the instrument was predicted to be below an instrument signal equalling molec. cm−2 of NO2. The predictions of the instrument model were verified in a proof-of-concept laboratory measurement, in which four test cells were filled with different concentrations of NO2. Then their column densities were measured with a conventional DOAS setup and the NO2 camera. We noticed agreement between the two instrumental setups within their uncertainties for all test cells and between the camera results and the predictions of the instrument model. The average relative deviation between model prediction and camera result amounted to 18 %.
We present the results of a field measurement at the coal power plant Großkraftwerk Mannheim. The camera measured an average NO2 plume SCD of and an average plume diameter of (78±34) m. In order to increase the SNR of this measurement and smooth the plume signal, sequences of six images were averaged, reducing the effective frame rate to FPS and the resolution to 1350×600 pixels. By examination of an off-plume area the detection limit of this measurement was estimated to be molec. cm−2; however, the uncertainties of the evaluation procedure, mainly the background estimation, increased the overall uncertainty to molec. cm−2. A mass flux analysis was carried out on the basis of image sequences. For this purpose, the optical flow between pairs of consecutive images was estimated with a Farnebäck algorithm, which yielded average horizontal wind speeds of (0.94±0.33) m s−1 and a resulting mean NO2 mass flux of (7.4±4.2) kg h−1 ( t yr−1). The camera measurements showed good agreement with predictions of the Janssen model for plume chemistry when computing the ratio as a function of distance downwind.
In the future, the following improvements to the instrument should be implemented: firstly, the optical setup inside the instrument can be further optimized. By including a beam splitter, the light for both sensor arrays could be collected from a mutual lens, thus eliminating the need to correct for differences in the two lenses as a potential error source, especially the cumbersome background fitting routine described in Sect. 4.2.1. Additionally, there are camera modules with much lower read-out time than the ones used in our prototype, increasing the overall photon budget available for measurements. Secondly, the instrument would benefit from thermal stabilization in order to maintain a more stable NO2 column inside its gas cell. This way, the evaluation procedure would rely less on successfully determining Sc (see Sect. 2.2) and (see Sect. 4.2.1) from an off-plume region of the camera images. Thirdly, when measuring NO2 emissions from a strong source as in Sect. 4.2, the evaluation routine could be made significantly less ambiguous by implementing an automated image segmentation algorithm to separate the plume and off-plume regions of the individual images.
The instrument model presented in Sect. 2.2 allows forward modeling of the measuring process with highly resolved radiance spectra and absorption cross sections. However, the integral terms that occur in the instrument response do not allow for a closed-form expression of . Starting from Eq. (12), we simplify the expression for the instrument response by assuming a constant radiance spectrum, and quantum efficiency η(λ)=const. We restrict the model to some spectral range and define . The final assumption is that the cross section of the target gas consists of only two representative absorption strengths, σstrong and σweak. To determine both, we compute the median of σNO2 and define σweak and σstrong as the mean absorption strength below and above the median, respectively. The absorption cross section can then be expressed as
where 1I is the indicator function on an interval I. The instrument response then only depends on the integrals of transmission terms of the following form.
Equation (12) then takes the following form.
This equation can be applied to arbitrary absorption cross sections; however, σweak and σstrong must be estimated anew for each absorption cross section. The analytical term in Eq. (A6) could be further simplified if a gas without a broadband contribution to its absorption cross section were considered. In that case, σweak≈0 and the column in the gas cell Sc could be chosen so that . The approximation of the instrument signal would then simplify to
The true instrument signal , as obtained in Sect. 2.1, and the analytical approximation in Eq. (A6) are plotted in Fig. A1. The spectral range of choice was 430–445 nm. The analytical approximation underestimates the true instrument response by around 25 % but is equally linear in S. The deviation can be corrected by tweaking the choice of σweak and σstrong, although good candidates cannot be known a priori. The derived analytical expression allows for quick approximation of the sensitivity of a GCS measurement.
All data are available from the authors upon request.
LK, JK, TW, and UP developed the question of research. LK and JK conducted the laboratory and field measurements. LK and JK developed the instrument model and the instrument prototype. LK characterized the instrument, evaluated the data, and wrote the paper, with all authors contributing by revising it within several iterations.
At least one of the (co-)authors is a member of the editorial board of Atmospheric Measurement Techniques. The peer-review process was guided by an independent editor, and the authors also have no other competing interests to declare.
Publisher's note: Copernicus Publications remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This research has been supported by the German Research Foundation (DFG) (grant no. PL 193/23-1).
The article processing charges for this open-access publication were covered by the Max Planck Society.
This paper was edited by Thomas von Clarmann and reviewed by two anonymous referees.
Baker, R. L., Mauldin III, L. E., and Russell III, J. M.: Design and Performance of the Halogen Occultation Experiment (HALOE) Remote Sensor, in: SPIE Proceedings, edited by: Mollicone, R. A. and Spiro, I. J., SPIE, 30th Annual Technical Symposium, San Diego, United States https://doi.org/10.1117/12.936511, 1986.
Bobrowski, N., Hönninger, G., Lohberger, F., and Platt, U.: IDOAS: A new monitoring technique to study the 2D distribution of volcanic gas emissions, J. Volcanol. Geoth. Res., 150, 329–338, https://doi.org/10.1016/j.jvolgeores.2005.05.004, 2006.
Carslaw, D. C.: Evidence of an increasing NO2NOX emissions ratio from road traffic emissions, Atmos. Environ., 39, 4793–4802, https://doi.org/10.1016/j.atmosenv.2005.06.023, 2005.
Chance, K. and Kurucz, R.: An improved high-resolution solar reference spectrum for earth's atmosphere measurements in the ultraviolet, visible, and near infrared, J. Quant. Spectrosc. Ra., 111, 1289–1295, https://doi.org/10.1016/j.jqsrt.2010.01.036, 2010.
Dekemper, E., Vanhamel, J., Van Opstal, B., and Fussen, D.: The AOTF-based NO2 camera, Atmos. Meas. Tech., 9, 6025–6034, https://doi.org/10.5194/amt-9-6025-2016, 2016.
Deutscher Wetterdienst: https://www.dwd.de/, last access: 6 December 2021.
Drummond, J., Bailak, G., and Mand, G.: The Measurements of Pollution in the Troposphere (MOPITT) Instrument, in: Applications of Photonic Technology, edited by: Lampropoulos, G. A., Chrostowski, J., and Measures, R. M., Springer US, 179–200, https://doi.org/10.1007/978-1-4757-9247-8_38, 1995.
European Environment Agency: Air quality in Europe – 2017 report, https://www.eea.europa.eu/ds_resolveuid/86Y5RVUQT2 (last access: 10 October 2021), 2017.
Farnebäck, G.: Two-Frame Motion Estimation Based on Polynomial Expansion, Scandinavian Conference on Image Analysis, Junge 2003, Norrköping, Sweden, 2749, 363–370, https://doi.org/10.1007/3-540-45103-X_50, 2003.
Faustini, A., Rapp, R., and Forastiere, F.: Nitrogen dioxide and mortality: review and meta-analysis of long-term studies, Eur. Respir. J., 44, 744–753, https://doi.org/10.1183/09031936.00114713, 2014.
Fraunhofer Institute for Solar Energy Systems: https://www.energy-charts.info/, last access: 6 September 2021.
Fuchs, C., Kuhn, J., Bobrowski, N., and Platt, U.: Quantitative imaging of volcanic SO2 plumes using Fabry–Pérot interferometer correlation spectroscopy, Atmos. Meas. Tech., 14, 295–307, https://doi.org/10.5194/amt-14-295-2021, 2021.
Großkraftwerk Mannheim Aktiengesellschaft: Geschäftsbericht 2017, https://www.gkm.de/media/?file=373_gkm_gb_2017.pdf&download (last access: 7 July 2021), 2018.
Janssen, L., Van Wakeren, J., Van Duuren, H., and Elshout, A.: A classification of no oxidation rates in power plant plumes based on atmospheric conditions, Atmos. Environ., 22, 43–53, https://doi.org/10.1016/0004-6981(88)90298-3, 1988.
Jähne, B.: EMVA 1288 Standard for Machine Vision, Optik & Photonik, 5, 53–54, https://doi.org/10.1002/opph.201190082, 2010.
Kenty, K. L., Poor, N. D., Kronmiller, K. G., McClenny, W., King, C., Atkeson, T., and Campbell, S. W.: Application of CALINE4 to roadside NONO2 transformations, Atmos. Environ., 41, 4270–4280, https://doi.org/10.1016/j.atmosenv.2006.06.066, 2007.
Kuhn, L.: A GCS NO2 Camera image series of the emissions of the Großkraftwerk Mannheim (coloured), TIB AV-Portal [video],https://doi.org/10.5446/55859, 2022.
Kuhn, J., Platt, U., Bobrowski, N., and Wagner, T.: Towards imaging of atmospheric trace gases using Fabry–Pérot interferometer correlation spectroscopy in the UV and visible spectral range, Atmos. Meas. Tech., 12, 735–747, https://doi.org/10.5194/amt-12-735-2019, 2019.
Landesanstalt für Umwelt Baden-Württemberg: https://www.lubw.baden-wuerttemberg.de/luft/messwerte-immissionswerte?comp=LUQX&id=DEBW005#karte, last access: 6 September 2021.
Lippmann, H. H., Jesser, B., and Schurath, U.: The rate constant of NO + O3 → NO2 + O2 in the temperature range of 283–443 K, Int. J. Chem. Kinet., 12, 547–554, https://doi.org/10.1002/kin.550120805, 1980.
Louban, I., Bobrowski, N., Rouwet, D., Inguaggiato, S., and Platt, U.: Imaging DOAS for volcanological applications, B. Volcanol., 71, 753–765, https://doi.org/10.1007/s00445-008-0262-6, 2009.
Manago, N., Takara, Y., Ando, F., Noro, N., Suzuki, M., Irie, H., and Kuze, H.: Visualizing spatial distribution of atmospheric nitrogen dioxide by means of hyperspectral imaging, Appl. Optics, 57, 5970, https://doi.org/10.1364/AO.57.005970, 2018.
Mori, T. and Burton, M.: The SO2 camera: A simple, fast and cheap method for ground-based imaging of SO2 in volcanic plumes, Geophys. Res. Lett., 33, 24, https://doi.org/10.1029/2006gl027916, 2006.
Parrish, D., Murphy, P., Albritton, D., and Fehsenfeld, F.: The measurement of the photodissociation rate of NO2 in the atmosphere, Atmos. Environ., 17, 1365–1379, https://doi.org/10.1016/0004-6981(83)90411-0, 1983.
Peters, E., Ostendorf, M., Bösch, T., Seyler, A., Schönhardt, A., Schreier, S. F., Henzing, J. S., Wittrock, F., Richter, A., Vrekoussis, M., and Burrows, J. P.: Full-azimuthal imaging-DOAS observations of NO2 and O4 during CINDI-2, Atmos. Meas. Tech., 12, 4171–4190, https://doi.org/10.5194/amt-12-4171-2019, 2019.
Pissulla, D., Seckmeyer, G., Cordero, R., Blumthaler, M., Klotz, B., Webb, A., Kift, R., Smedley, A., Bais, A., Kouremeti, N., Cede, A., Herman, J., and Kowalewski, M.: Comparison of atmospheric spectral radiance measurements from five independently calibrated systems, Photoch. Photobio. Sci., 8, 516–27, https://doi.org/10.1039/b817018e, 2009.
Platt, U. and Kuhn, J.: Caution with spectroscopic NO2 reference cells (cuvettes), Atmos. Meas. Tech., 12, 6259–6272, https://doi.org/10.5194/amt-12-6259-2019, 2019.
Platt, U. and Stutz, J. (Eds.): Differential Optical Absorption Spectroscopy, 1st edn., Springer, Berlin, Heidelberg, https://doi.org/10.1007/978-3-540-75776-4, 2008.
Rothman, L., Gordon, I., Babikov, Y., Barbe, A., Chris Benner, D., Bernath, P., Birk, M., Bizzocchi, L., Boudon, V., Brown, L., Campargue, A., Chance, K., Cohen, E., Coudert, L., Devi, V., Drouin, B., Fayt, A., Flaud, J.-M., Gamache, R., Harrison, J., Hartmann, J.-M., Hill, C., Hodges, J., Jacquemart, D., Jolly, A., Lamouroux, J., Le Roy, R., Li, G., Long, D., Lyulin, O., Mackie, C., Massie, S., Mikhailenko, S., Müller, H., Naumenko, O., Nikitin, A., Orphal, J., Perevalov, V., Perrin, A., Polovtseva, E., Richard, C., Smith, M., Starikova, E., Sung, K., Tashkun, S., Tennyson, J., Toon, G., Tyuterev, V., and Wagner, G.: The HITRAN2012 molecular spectroscopic database, J. Quant. Spectrosc. Ra., 130, 4–50, https://doi.org/10.1016/j.jqsrt.2013.07.002, 2013.
Thalman, R. and Volkamer, R.: Temperature dependent absorption cross-sections of O2–O2 collision pairs between 340 and 630 nm and at atmospherically relevant pressure, Phys. Chem. Chem. Phys., 15, 15371–15381, https://doi.org/10.1039/C3CP50968K, 2013.
The European Commision: European Pollutant Release and Transfer Register, https://prtr.eea.europa.eu/ (last access: 6 September 2021), 2017.
Vandaele, A., Hermans, C., Fally, S., Carleer, M., Colin, R., Marienne, M., Jenouvrier, A., and Coquart, B.: High-resolution Fourier transform measurement of the NO2 visible and near-infrared absorption cross sections: Temperature and pressure effects, J. Geophys. Res., 107, 4348, https://doi.org/10.1029/2001JD000971, 2002.
Ward, T. V. and Zwick, H. H.: Gas cell correlation spectrometer: GASPEC, Appl. Optics, 14, 2896–2904, https://doi.org/10.1364/AO.14.002896, 1975.
World Health Organization: WHO air quality guidelines for Europe, http://www.euro.who.int/__data/assets/pdf_file/0017/123083/AQG2ndEd_7_1nitrogendioxide.pdf?ua=1 (last access: 10 October 2021), 2000.
World Health Organization: WHO global air quality guidelines: particulate matter (PM2.5 and PM10), ozone, nitrogen dioxide, sulfur dioxide and carbon monoxide, https://apps.who.int/iris/bitstream/handle/10665/345329/9789240034228-eng.pdf?sequence=1&isAllowed=y (last access: 10 October 2021), 2021.
Wu, K., Feng, Y., Yu, G., Liu, L., Li, J., Xiong, Y., and Li, F.: Development of an imaging gas correlation spectrometry based mid-infrared camera for two-dimensional mapping of CO in vehicle exhausts, Opt. Express, 26, 8239, https://doi.org/10.1364/OE.26.008239, 2018.