Full characterization and calibration of a transfer standard monitor for atmospheric radon measurements

Roger Curcoll, Claudia Grossi, Stefan Röttger, and Arturo Vargas

Download

Interactive discussion

Status: closed

Comment types: AC – author | RC – referee | CC – community | EC – editor | CEC – chief editor | : Report abuse
  • CC1: 'Comment on egusphere-2023-2680', Scott Chambers, 28 Nov 2023
    • AC1: 'Reply on CC1', Roger Curcoll Masanes, 20 Dec 2023
      • CC2: 'Reply on AC1', Scott Chambers, 22 Dec 2023
        • AC2: 'Reply on CC2', Roger Curcoll Masanes, 12 Jan 2024
          • CC3: 'Reply on AC2', Scott Chambers, 18 Jan 2024
  • RC1: 'Comment on egusphere-2023-2680', Anonymous Referee #1, 25 Jan 2024
    • AC3: 'Reply on RC1', Roger Curcoll Masanes, 11 Mar 2024
  • RC2: 'Comment on egusphere-2023-2680', Anonymous Referee #2, 12 Feb 2024
    • AC4: 'Reply on RC2', Roger Curcoll Masanes, 11 Mar 2024

Peer review completion

AR: Author's response | RR: Referee report | ED: Editor decision | EF: Editorial file upload
AR by Roger Curcoll Masanes on behalf of the Authors (11 Mar 2024)  Author's response   Author's tracked changes   Manuscript 
ED: Publish as is (26 Mar 2024) by Ute Karstens
AR by Roger Curcoll Masanes on behalf of the Authors (27 Mar 2024)  Manuscript 
Download
Short summary
This paper presents a new user-friendly version of the Atmospheric Radon MONitor (ARMON). The efficiency of the instrument is of 0.0057 s-1, obtained using different techniques at Spanish and German chambers. The total calculated uncertainty of the ARMON for hourly radon concentrations above 5 Bq m-3 is lower than 10 % (k = 1). Results confirm that the ARMON is suitable to measure low-level radon activity concentrations and to be used as a transfer standard to calibrate in situ radon monitors.