A new, low-cost instrument for measuring total column ozone was tested against a MK III Brewer spectrophotometer for six months. Using the "Global Irradiance" method from the Norwegian GUV network, it showed ozone values matching Direct Sun Brewer measurements with a 1.8 % standard deviation. Despite temperature and cloud dependencies, the instrument is robust and easy to use. Fully coverage of the UV-A/B spectrum allows for further UV-radiation analysis. Total cost is less than EUR 3000.