Characterization of a high detection-sensitivity atmospheric pressure interface time-of-flight mass spectrometer

Fabian Schmidt-Ott, Somnath Bhowmick, Alexandros Lekkas, Dimitris Papanastasiou, Anne Maisser, and George Biskos

Data sets

Characterization of a High Detection-Sensitivity Atmospheric Pressure Interface Time-of-Flight Mass Spectrometer Fabian Schmidt-Ott https://doi.org/10.5281/zenodo.15854115

Download
Short summary
We characterized a novel Atmospheric-Pressure-Interface Time-of-Flight (TOF) Mass Spectrometer. By accumulating ions in a trap prior the TOF mass analyzer, we achieve a limit of detection in the 10-3 to 10-6 ppq range with temporal resolutions in the order of 1 s to 10 min, respectively. This makes it highly relevant for atmospheric measurements, as it enables probing short-lived, low-concentration species of high importance in atmospheric chemistry and new particle formation.