Intercomparison of seven collocated ground-based infrared spectrometer radiance observations and retrieved thermodynamic profiles

David D. Turner, Bianca Adler, Laura Bianco, James M. Wilczak, Vincent Michaud-Belleau, and Luc Rochette

Data sets

ASSIST infrared spectrometer side-by-side comparison dataset (Boulder, CO in Sep-Oct 2023) David Turner et al. https://doi.org/10.5281/zenodo.17228423

Download
Short summary
It is critical that a network of ground-based instruments that measure temperature and humidity profiles be well calibrated, so that differences between any two profiles can be attributed to atmospheric differences and not instrument calibration issues. This study evaluated the relative accuracy of 7 ground-based infrared spectrometers and their ability to measure these profiles, and found that the profile bias was much smaller than the uncertainty in the retrieved profiles themselves.