Articles | Volume 19, issue 6
https://doi.org/10.5194/amt-19-2265-2026
https://doi.org/10.5194/amt-19-2265-2026
Research article
 | 
01 Apr 2026
Research article |  | 01 Apr 2026

Fast and reproducible δ13C-CO2 analysis from 1 mL of ambient atmospheric air using continuous-flow IRMS: from sampling to storage to analysis

Joana Sauze, Marie-Laure Tiouchichine, Alexandru Milcu, and Clément Piel

Viewed

Total article views: 2,106 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
1,853 207 46 2,106 56 53
  • HTML: 1,853
  • PDF: 207
  • XML: 46
  • Total: 2,106
  • BibTeX: 56
  • EndNote: 53
Views and downloads (calculated since 20 Aug 2025)
Cumulative views and downloads (calculated since 20 Aug 2025)

Viewed (geographical distribution)

Total article views: 2,106 (including HTML, PDF, and XML) Thereof 2,087 with geography defined and 19 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 
Latest update: 01 Apr 2026
Download
Short summary
We present a simple workflow for high-precision δ¹³C-CO₂ analysis from 1 mL air samples using cryogenic pre-concentration and continuous-flow isotope-ratio mass spectrometry (IRMS). The method combines vial conditioning, dual sealing, and −80 °C storage to minimise gas loss. It achieves ±0.1 ‰ precision and stable isotopic values for up to one week of storage. This low-cost approach enables high-frequency δ¹³C-CO₂ measurements in volume-limited systems such as microcosm or chamber experiments.
Share