Development and validation of an integrated Ambient Air Test Facility (AATF) for multi-instrument aerosol characterization
Download
- Final revised paper (published on 21 Jul 2026)
- Supplement to the final revised paper
- Preprint (discussion started on 11 Jan 2026)
- Supplement to the preprint
Interactive discussion
Status: closed
Comment types: AC – author | RC – referee | CC – community | EC – editor | CEC – chief editor
| : Report abuse
-
RC1: 'Comment on egusphere-2025-6043', Anonymous Referee #1, 30 Jan 2026
- AC1: 'Reply on RC1', Paul Johns, 09 Feb 2026
-
RC2: 'Comment on egusphere-2025-6043', Anonymous Referee #2, 06 Jun 2026
- AC2: 'Reply on RC2', Paul Johns, 10 Jun 2026
Peer review completion
AR – Author's response | RR – Referee report | ED – Editor decision | EF – Editorial file upload
AR by Vasanthi Sivaprakasam on behalf of the Authors (12 Jun 2026)
Author's response
Author's tracked changes
Manuscript
ED: Referee Nomination & Report Request started (15 Jun 2026) by Johannes Schneider
RR by Anonymous Referee #1 (21 Jun 2026)
ED: Publish subject to technical corrections (22 Jun 2026) by Johannes Schneider
AR by Vasanthi Sivaprakasam on behalf of the Authors (25 Jun 2026)
Manuscript
This manuscript describes the development and validation of the so-called Ambient Air Test Facility (AATF), a large-scale setup which enables intercomparison of aerosol instruments under controlled laboratory conditions. The setup comprises several aerosol generators, a 14-meter-long flow tube for aerosol homogenisation and isokinetic sampling ports. The scale of the facility is impressive, allowing parallel measurements with various devices-under-test.
The manuscript is well-structured and easy to read, and the figures are informative. I am puzzled, however, by the lack of reference instruments that would enable a more meaningful intercomparison of aerosol instruments. I am also sceptical about the lack of validation for particles larger than 5 micrometers.
Questions:
To summarise, I appreciate the effort that went into developing the setup, but I would recommend a more thorough validation for large particle sizes. In addition, to make measurements more quantitative, I would recommend the use of calibrated instruments to provide reference data.