Articles | Volume 19, issue 21
https://doi.org/10.5194/amt-19-6921-2026
https://doi.org/10.5194/amt-19-6921-2026
Research article
 | 
02 Nov 2026
Research article |  | 02 Nov 2026

Assessing greenhouse gas measurement consistency and traceability in the UK and Ireland atmospheric monitoring network using whole and synthetic air reference materials

Emmal Safi, Joe Pitt, Grant Forster, Christopher Rennick, Simon O'Doherty, Dickon Young, Kanokrat Charoenpornpukdee, Nick Garrard, Dafina Kikaj, Andrew C. Manning, Gerard Spain, Damien Martin, Morgan Lopez, Michel Ramonet, and Kieran Stanley

Download

Interactive discussion

Status: closed

Comment types: AC – author | RC – referee | CC – community | EC – editor | CEC – chief editor | : Report abuse
  • RC1: 'Comment on egusphere-2026-1802', Anonymous Referee #1, 14 Jul 2026
  • RC2: 'Comment on egusphere-2026-1802', Anonymous Referee #2, 14 Sep 2026

Peer review completion

AR – Author's response | RR – Referee report | ED – Editor decision | EF – Editorial file upload
AR by Emmal Safi on behalf of the Authors (12 Oct 2026)  Author's response   Author's tracked changes   Manuscript 
ED: Publish as is (16 Oct 2026) by Andreas Zahn
AR by Emmal Safi on behalf of the Authors (20 Oct 2026)  Manuscript 
Download
Short summary
We compared whole and synthetic-air reference materials (RMs) for investigating offsets in greenhouse gas measurements so data from different sites/networks can be trusted when combined. By sending whole and synthetic-air (RMs) to six sites and measuring under routine field conditions, we found whole-air gives the most precise results, especially for carbon dioxide and nitrous oxide. Synthetic-air worked fairly well for methane but synthetic-air RMs needs lower uncertainty before wide use.
Share
Altmetrics
Final-revised paper
Preprint