Sensitive detection of n-alkanes using a mixed ionization mode proton-transfer-reaction mass spectrometer

Omar Amador-Muñoz, Pawel K. Misztal, Robin Weber, David R. Worton, Haofei Zhang, Greg Drozd, and Allen H. Goldstein

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement
Download
Short summary
Proton transfer reaction mass spectrometry (PTR-MS) was used to detect n-alkanes that generally have a lower proton affinity than water and therefore proton transfer (PT) by reaction with H3O+ is not an effective mechanism for their detection. In this study, we developed a method using a conventional PTR-MS to detect n-alkanes by optimizing ion source and drift tube conditions to vary the relative amounts of different primary ions (H3O+, O2+, NO+) in the reaction chamber (drift tube).