Accuracy, precision, and temperature dependence of Pandora total ozone measurements estimated from a comparison with the Brewer triad in Toronto

Xiaoyi Zhao, Vitali Fioletov, Alexander Cede, Jonathan Davies, and Kimberly Strong

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement
Download
Short summary
This study evaluates the performance of the recently developed Pandora spectrometer by comparing it with the Brewer reference triad. The instrument random uncertainty, total column ozone temperature dependence, and ozone air mass dependence have been determined using two Pandora and six Brewer instruments. In general, Pandora and Brewer instruments both have very low random uncertainty and air mass dependence. However, the Brewer has smaller ozone temperature dependence than Pandora.