The “Golden Points” and nonequilibrium correction of high-accuracy frost point hygrometers

Yann Poltera, Beiping Luo, Frank G. Wienhold, and Thomas Peter

Data sets

The “Golden Points” and nonequilibrium correction of high-accuracy frost point hygrometers - Dataset Yann Poltera et al. https://doi.org/10.3929/ethz-b-000732964

Download
Short summary
Chilled mirror hygrometers are the most accurate instruments for measuring water vapor in the upper troposphere and lower stratosphere. The “Golden Points” and nonequilibrium correction are a new post-processing technique for these instruments, which can correct frost point data using mirror reflectance information to achieve unprecedented accuracy of better than 4 % in the H2O mixing ratio, even under rapidly changing humidity conditions, from the ground to the middle stratosphere.