An analytical system for stable isotope analysis on carbon monoxide using continuous-flow isotope-ratio mass spectrometry

S. L. Pathirana, C. van der Veen, M. E. Popa, and T. Röckmann

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement
Download
Short summary
CO is established as an important indirect greenhouse gas, as it is the major sink for the OH∙. We have developed a fully automated system for the determination of δ13C and δ18O in atmospheric CO. The blank signal of the Schütze reagent is 1-3 % of the typical sample size. The repeatability is 0.1 ‰ for δ13C and 0.2 ‰ for δ18O. The analytical repeatability for the mole fraction is ~0.7 nmol mol-1 for 100 mL of ambient air (185.4 nmol mol-1 of CO). A single measurement is performed in 18 min.