Articles | Volume 9, issue 6
https://doi.org/10.5194/amt-9-2669-2016
https://doi.org/10.5194/amt-9-2669-2016
Research article
 | 
24 Jun 2016
Research article |  | 24 Jun 2016

Water vapor inhibits hydrogen sulfide detection in pulsed fluorescence sulfur monitors

Anders B. Bluhme, Jonas L. Ingemar, Carl Meusinger, and Matthew S. Johnson

Viewed

Total article views: 3,397 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
1,772 1,424 201 3,397 187 235
  • HTML: 1,772
  • PDF: 1,424
  • XML: 201
  • Total: 3,397
  • BibTeX: 187
  • EndNote: 235
Views and downloads (calculated since 19 Jan 2016)
Cumulative views and downloads (calculated since 19 Jan 2016)

Cited

Saved (final revised paper)

Latest update: 10 Oct 2026
Download
Short summary
Hydrogen sulfide (H2S) is a malodorous, very poisonous, and flammable gas. It can be detected as SO2 using fluorescence after conversion using a hot catalyst. This technique is well established and as such also recommended by authorities such as the EPA. Our paper describes how at a relative humidity as low as 5 %, significant amounts of H2S pass the instrument undetected. At ambient levels of relative humidity, up to 1/3 of all H2S passes the instrument unnoticed.