Caution with spectroscopic NO2 reference cells (cuvettes)

Ulrich Platt and Jonas Kuhn

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by Ulrich Platt on behalf of the Authors (05 Sep 2019)  Author's response   Manuscript 
ED: Publish as is (10 Sep 2019) by Hendrik Fuchs
AR by Ulrich Platt on behalf of the Authors (17 Sep 2019)

Post-review adjustments

AA: Author's adjustment | EA: Editor approval
AA by Ulrich Platt on behalf of the Authors (22 Nov 2019)   Author's adjustment   Manuscript
EA: Adjustments approved (25 Nov 2019) by Hendrik Fuchs
Download
Short summary
Measurements of atmospheric trace gases by absorption spectroscopy are frequently supported by recording the amount of trace gas in absorption cells. These are typically small glass (or quartz) cylinders containing the gas to be studied. Here we show in the example of NO2-absorption cells that the effective amount of gas seen by the instrument can deviate greatly from expected values (by orders of magnitude in severe cases). Some suggestions for improving the situation are discussed.