Measurement of NOx and NOy with a thermal dissociation cavity ring-down spectrometer (TD-CRDS): instrument characterisation and first deployment

Nils Friedrich, Ivan Tadic, Jan Schuladen, James Brooks, Eoghan Darbyshire, Frank Drewnick, Horst Fischer, Jos Lelieveld, and John N. Crowley

Data sets

Measurement of NOx and NOy with a thermal dissociation cavity ring-down spectrometer (TD-CRDS): Instrument characterisation and first deployment N. Friedrich https://doi.org/10.5281/zenodo.4134659

Download
Short summary
We present a new instrument for the measurement of NOx and NOy based on a combination of the thermal dissociation of NOy to NOx and cavity ring-down spectroscopic detection of NO2. It features a denuder to separate the contributions of gas-phase and particulate nitrates to NOy. We provide a detailed characterization of the instrument and briefly outline results from first deployments.