Dead time effect on the Brewer measurements: correction and estimated uncertainties

Ilias Fountoulakis et al.

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement
Download
Short summary
The dead time (DT) is characteristic for each Brewer spectrophotometer and non-proper correction of the raw data for its effect may lead to important errors in UV, AOD and TOC measurements. Quantitative estimation of the DT-related uncertainties shown that a 2 ns error in the DT may lead to errors greater than 1 % in TOC. The operational algorithm for the DT calculation and correction is validated and the development of new methods for the estimation of DT is described.