Articles | Volume 11, issue 4
https://doi.org/10.5194/amt-11-2225-2018
https://doi.org/10.5194/amt-11-2225-2018
Research article
 | 
18 Apr 2018
Research article |  | 18 Apr 2018

Aggregated particles caused by instrument artifact

Ashley M. Pierce, S. Marcela Loría-Salazar, W. Patrick Arnott, Grant C. Edwards, Matthieu B. Miller, and Mae S. Gustin

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by Ashley Pierce on behalf of the Authors (14 Mar 2018)  Author's response   Manuscript 
ED: Referee Nomination & Report Request started (15 Mar 2018) by Pierre Herckes
RR by Anonymous Referee #2 (15 Mar 2018)
ED: Publish as is (16 Mar 2018) by Pierre Herckes
AR by Ashley Pierce on behalf of the Authors (22 Mar 2018)  Author's response   Manuscript 
Download
Short summary
This paper investigates the possible sources of anomalous particulate matter collected at a high-elevation site during June to November 2014. Particles were collected on a sample filter that were > 2.5 µm in aerodynamic diameter, on a system that theoretically should not collect particulate matter that large. These samples indicated that either the observed particles had unique dimensions and behavior or that there was an issue with the particulate monitor inlet setup.