Articles | Volume 15, issue 10
https://doi.org/10.5194/amt-15-3189-2022
https://doi.org/10.5194/amt-15-3189-2022
Peer-reviewed comment
 | 
25 May 2022
Peer-reviewed comment |  | 25 May 2022

Comment on “Comparison of ozone measurement methods in biomass burning smoke: an evaluation under field and laboratory conditions” by Long et al. (2021)

Noah Bernays, Daniel A. Jaffe, Irina Petropavlovskikh, and Peter Effertz
Abstract

Long et al. (2021) conducted a detailed study of possible interferences in measurements of surface O3 by UV spectroscopy, which measures the UV transmission in ambient and O3-scrubbed air. While we appreciate the careful work done in this analysis, there were several omissions, and in one case, the type of scrubber used was misidentified as manganese dioxide (MnO2) when in fact it was manganese chloride (MnCl2). This misidentification led to the erroneous conclusion that all UV-based O3 instruments employing solid-phase catalytic scrubbers exhibit significant positive artifacts, whereas previous research found this not to be the case when employing MnO2 scrubber types. While the Long et al. (2021) study, and our results, confirm the substantial bias in instruments employing an MnCl2 scrubber, a replication of the earlier work with an MnO2 scrubber type and no humidity correction is needed.

Dates
1 Introduction

Ozone (O3) is a key hazardous atmospheric pollutant. In the USA, more than 100 million people live in regions that do not meet the National Ambient Air Quality Standards. Wildfires exacerbate O3 pollution (Crutzen et al., 1979; Crutzen and Andreae, 1990; Jaffe et al., 2013, 2020; Brey and Fischer, 2016; Gong et al., 2017). Given that smoke contains literally hundreds of different compounds, it is important to address possible interferences in measurements of O3. Long et al. (2021) conducted a detailed study of possible interferences in UV measurements of O3, which is the method most commonly used. In the UV method, O3 is measured at 254 nm in a sample airstream and in an airstream where O3 has been removed, usually by a solid-state catalytic scrubber. Long et al. (2021) provide an excellent discussion of this method, which we will not repeat here. However, one of the most important aspects in this measurement is the nature of the scrubber that is used to remove O3. For the scrubber, various companies have used manganese dioxide (MnO2), Hopcalite (a mixture of manganese and copper oxides), and manganese chloride (MnCl2). Long et al. (2021) compared multiple UV instruments with an NO chemiluminescence instrument, a method which is presumably free from interferences. Long et al. (2021) found a significant bias of 16–24 ppb O3 ppm−1 of CO in one type of UV O3 analyzer (Thermo Fisher 49i) that was tested without humidity correction, as compared to the NO chemiluminescence method. The bias was correlated with smoke tracers, such as CO and total hydrocarbons. Other instruments were tested with a humidity correction and found to have a much smaller bias which Long et al. (2021) attributed to the humidity correction. According to Long et al. (2021), the scrubber types on these instruments were similar, but in fact they were not, as discussed below, and this leads to significant uncertainty in their conclusions.

Long et al. (2021) did not cite our earlier study (Gao and Jaffe, 2017). In this work, we conducted a comparison between two UV-based O3 analyzers (Dasibi 1008-RS and Ecotech Serinus 10) and an NO chemiluminescent analyzer in wildfire plumes at the Mt. Bachelor Observatory (MBO) during the 2015 wildfire season. Gao and Jaffe (2017) found no significant bias in the UV analyzers relative to the NO chemiluminescent analyzer in moderate smoke plumes, up to approximately 1 ppm of carbon monoxide (CO). Both of these UV analyzers used an MnO2 scrubber. The precision and bias of instrumentation used in Gao and Jaffe (2017)'s study along with the quality assurance methods are outlined in the paper and are sufficient to meet Long et al. (2021)'s data quality objectives. A key question is the following: why were Long et al. (2021)'s results different from Gao and Jaffe (2017)'s results? We address this question below.

2 Scrubber type misidentified

Long et al. (2021) cite the Thermo Fisher Scientific model 49i series instrument's scrubber type as MnO2 (as do others: Kleindienst et al., 1993; Spicer et al., 2010; Turnipseed et al., 2017). However, according to David Sherwin, a Technical Application Specialist III who has been working at Thermo for 18 years (David Sherwin, personal communication, 2021), and Nathan Bernardini, a Technical Application Specialist II who has been working at Thermo for close to 5 years (Nathan Bernardini, personal communication, 2022), the scrubbers in the 49c and 49i series have always used MnCl2 not MnO2. While we have not done chemical tests on the scrubber, we feel that the manufacturer is in the best place to know what is inside their instrument. The names and email addresses of the Thermo Fisher scientists with whom we communicated, as well as screenshots of our email correspondence, can be found in the author's final comment in the discussion phase of manuscript submission. Please note that there is no info about the scrubber type in the manual. Due to this scrubber type misidentification, Long et al. (2021) did not test any O3 analyzer with a true MnO2 scrubber and without humidity correction, the most common way these instruments are deployed.

3 Recent data from the Mt. Bachelor Observatory confirm bias with MnCl2 scrubber type

The Mt. Bachelor Observatory is a high-elevation research station in the US Pacific Northwest that has been used for many years to study O3 and other pollutants (e.g., Jaffe et al., 2018). Starting in 2018, we have deployed two O3 instruments at MBO, the Ecotech Serinus 10, previously used in the Gao and Jaffe (2017) study, and a Thermo Fisher 49c, a similar instrument to the one used in Long et al. (2021)'s study which uses the same scrubber and no humidity correction. Generally, the Ecotech and Thermo Fisher instruments agree well, but in a particularly strong period of wildfire smoke, we saw a substantial difference in the two measurements. Figure 1 shows data from a 3-week period in September–October 2020, when we experienced heavy smoke at MBO. The slope (0.0112 ppb O3 ppb−1 CO) is smaller but of the same order of magnitude as that reported by Long et al. (2021) for comparisons of the Thermo Fisher to the NO chemiluminescent instrument (0.016–0.024 ppb O3 ppb−1 CO).

https://amt.copernicus.org/articles/15/3189/2022/amt-15-3189-2022-f01

Figure 1Difference in O3 readings between the Thermo Fisher and Ecotech UV instruments vs. CO for a 3-week period starting 14 September 2020. During this period, the Thermo Fisher instrument gave readings that were up to 45 ppb higher than the Ecotech instrument. Values are hourly averages.

Download

In the absence of smoke, we see good agreement between the two measurements. Figure 2 shows the agreement between the Thermo and Ecotech instruments during non-smoke periods (defined as CO <200 ppb), with a root mean squared difference of less than 1 ppb. Given our earlier comparison establishing that the Ecotech instrument did not show significant bias (Gao and Jaffe, 2017), we contend that these findings corroborate Long et al. (2021)'s conclusion that the Thermo Fisher instrument exhibits a significant positive bias at high CO levels. We believe the MnCl2 scrubber in the 49i is the primary cause for the discrepancy between the findings of Long et al. (2021) and Gao and Jaffe (2017).

https://amt.copernicus.org/articles/15/3189/2022/amt-15-3189-2022-f02

Figure 2O3 measured by the Thermo Fisher 49c instrument vs. Ecotech instrument at MBO during non-smoke periods (defined as CO <200 ppb). Data are hourly averages of all valid data for both instruments in 2020. The inset shows a difference plot (Thermo minus Ecotech) vs. CO for the same data. The root mean squared error (difference) of the Thermo vs. Ecotech plot is 0.9 ppb, and the linear regression line has a slope of 1.055, a y intercept of −2.4 ppb, and an R2 of 0.98.

Download

4 Nafion dryer vs. scrubber impacts on O3 measurements: need for further research

When Long et al. (2021) put a Nafion dryer on their Thermo Fisher instrument midway through the study, the bias was reduced by an order of magnitude. We agree with Long et al. (2021) that the Nafion dryer reduced not only water vapor but probably also scrubbed many of the VOCs that were causing the bias. While Nafion is known to transfer O3 and lower-molecular-weight alkanes efficiently, it will remove more complex VOCs that are likely responsible for the bias in UV instruments (Perma Pure LLC, 2022). Similar tests with and without a Nafion drier were not done for the other instruments. The Nafion-dried 2B-205 instrument (hereafter 2B) in Long et al. (2021)'s study showed O3 artifacts an order of magnitude lower than the non-dried UV analyzers, but this can be explained by the 2B's MnO2-containing Hopcalite scrubber acting similarly to a pure MnO2 scrubber. We note that current EPA recommendations are to include Nafion dryers for UV O3 instruments (Halliday et al., 2020), and we see no downside to this recommendation. But given that this remains a recommendation, as well as to interpret past data, we suggest that future experiments on O3 bias include instruments with a true MnO2 scrubber with, and without, humidity correction, as the most common field setup does not include a drying system.

Data availability

Data from the Mt. Bachelor Observatory are archived at the University of Washington's Research Works Archive (http://hdl.handle.net/1773/48597, last access: 18 May 2022; Jaffe, 2020).

Author contributions

NB completed most of the data analysis and instrument calibrations and wrote the first draft of the article. DAJ is the principal investigator for the Mt. Bachelor Observatory (MBO) and contributed to the overall analysis and article preparation. IP and PE are responsible for the duplicate (Thermo) O3 measurements at MBO and contributed to data interpretation and final preparation of the article.

Competing interests

The contact author has declared that neither they nor their co-authors have any competing interests.

Disclaimer

Publisher’s note: Copernicus Publications remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Financial support

This research has been supported by the National Science Foundation (grant no. AGS-1447832) and the National Oceanic and Atmospheric Administration (grant no. RA-133R-16-SE-0758).

Review statement

This paper was edited by Glenn Wolfe and reviewed by two anonymous referees.

References

Brey, S. J. and Fischer, E. V.: Smoke in the city: how often and where does smoke impact summertime ozone in the United States?, Environ. Sci. Technol., 50, 1288–1294, https://doi.org/10.1021/acs.est.5b05218, 2016. 

Crutzen, P. J. and Andreae, M. O.: Biomass burning in the tropics: Impact on atmospheric chemistry and biogeochemical cycles, Science, 250, 1669–1678, 1990. 

Crutzen, P. J., Heidt, L. E., Krasnec, J. P., Pollock, W. H., and Seiler, W.: Biomass burning as a source of atmospheric gases CO, H2, N2O, NO, CH3Cl, and COS, Nature, 282, 253–256, 1979. 

Gao, H. and Jaffe, D. A.: Comparison of ultraviolet absorbance and NO-chemiluminescence for ozone measurement in wildfire plumes at the Mount Bachelor Observatory, Atmos. Environ., 166, 224–233, https://doi.org/10.1016/j.atmosenv.2017.07.007, 2017. 

Gong, X., Kaulfus, A., Nair, U., and Jaffe, D. A: Quantifying O3 impacts in urban areas due to wildfires using a generalized additive model, Environ. Sci. Technol., 51, 13216–13223, https://doi.org/10.1021/acs.est.7b03130, 2017. 

Halliday, H., Johnson, C., Long, R., Vanderpool, R., and Whitehill, A.: Recommendations for Nationwide Approval of Nafion™ Dryers Upstream of UV-Absorption Ozone Analyzers, U.S. Environmental Protection Agency, Washington, DC, EPA/600/R-20/390, https://cfpub.epa.gov/si/si_public_record_Report.cfm?dirEntryId=350170&Lab=CEMM (last access: 18 May 2022), 2020. 

Jaffe, D.: Mt. Bachelor Observatory final atmospheric data (v1) for Year 2020, University of Washington's Research Works Archive [data set], http://hdl.handle.net/1773/48597 (last access: 18 May 2022), 2020. 

Jaffe, D. A., Wigder, N., Downey, N., Pfister, G., Boynard, A., and Reid, S. B.: Impact of wildfires on ozone exceptional events in the western U.S., Environ. Sci. Technol., 47, 11065–11072, https://doi.org/10.1021/es402164f, 2013. 

Jaffe, D. A., Cooper, O. R., Fiore, A. M., Henderson, B. H., Tonnesen, G. S., Russell, A. G., Henze, D. K., Langford, A. O., Lin, M., and Moore, T.: Scientific assessment of background ozone over the U.S.: Implications for air quality management, Elementa: Science of the Anthropocene, 6, 56, https://doi.org/10.1525/elementa.309, 2018.  

Jaffe, D. A., O'Neill, S. M., Larkin, N. K., Holder, A. L, Peterson, D. L., Halofsky, J. E., and Rappold, A. G.: Wildfire and prescribed burning impacts on air quality in the United States, J. Air Waste Manage., 70, 583–615, https://doi.org/10.1080/10962247.2020.1749731, 2020. 

Kleindienst, T. E., Hudgens, E. E., Smith, D. F., McElroy, F. F., and Bufalini, J. J.: Comparison of chemiluminescence and ultraviolet ozone monitor responses in the presence of humidity and photochemical pollutants, J. Air Waste Manage., 43, 213–222, https://doi.org/10.1080/1073161X.1993.10467128, 1993. 

Long, R. W., Whitehill, A., Habel, A., Urbanski, S., Halliday, H., Colón, M., Kaushik, S., and Landis, M. S.: Comparison of ozone measurement methods in biomass burning smoke: an evaluation under field and laboratory conditions, Atmos. Meas. Tech., 14, 1783–1800, https://doi.org/10.5194/amt-14-1783-2021, 2021. 

Perma Pure LLC: Compounds Removed by Nafion™ Tubing Dryers https://www.permapure.com/environmental-scientific/resources/all-about-nafion-and-faq/, last access: 8 January 2022. 

Spicer, C. W., Joseph, D. W., and Ollison, W. M.: A re-examination of ambient air ozone monitor interferences, J. Air Waste Manage., 60, 1353–1364, https://doi.org/10.3155/1047-3289.60.11.1353, 2010. 

Turnipseed, A. A., Andersen, P. C., Williford, C. J., Ennis, C. A., and Birks, J. W.: Use of a heated graphite scrubber as a means of reducing interferences in UV-absorbance measurements of atmospheric ozone, Atmos. Meas. Tech., 10, 2253–2269, https://doi.org/10.5194/amt-10-2253-2017, 2017. 

Short summary
Ozone is an important pollutant that impacts millions of people worldwide. It is therefore important to ensure accurate measurements. A recent surge in wildfire activity in the USA has resulted in significant enhancements in ozone concentration. However given the nature of wildfire smoke, there are questions about our ability to accurately measure ozone. In this comment, we discuss possible biases in the UV measurements of ozone in the presence of smoke.