A measurement system for CO2 and CH4 emissions quantification of industrial sites using a new in situ concentration sensor operated on board uncrewed aircraft vehicles
Jean-Louis Bonne
Ludovic Donnat
Grégory Albora
Jérémie Burgalat
Nicolas Chauvin
Delphine Combaz
Julien Cousin
Thomas Decarpenterie
Olivier Duclaux
Nicolas Dumelié
Nicolas Galas
Catherine Juery
Florian Parent
Florent Pineau
Abel Maunoury
Olivier Ventre
Marie-France Bénassy
Lilian Joly
Download
- Final revised paper (published on 26 Jul 2024)
- Supplement to the final revised paper
- Preprint (discussion started on 02 Jan 2023)
- Supplement to the preprint
Interactive discussion
Status: closed
-
RC1: 'Comment on amt-2022-334', Anonymous Referee #1, 03 Jun 2023
Bonne et al., presented a new light-weight open path laser absorption spectrometer for in situ CO2 and CH4 concentration measurements, and applied the analyzer in the fields to quantify the emission rates of a series of controlled releases of CH4 and CO2 at the TotalEnergies TADI test platform at Lacq, France and of two offshore oil and gas platforms in the North Sea. Gas LiDAR was used to obtain vertical profiles of wind speeds and directions on site, which were used in the mass balance approach to calculate the emission rates. The paper is well structured, and both the laboratory and field experiments results could be useful to further develop the application of quantification of greenhouse gas emissions using UAVs. On the other hand, the paper would have benefited from more stringent discussion of the uncertainties of the measurements and the quantification using the mass balance approach. The paper is suitable for publication after addressing my following concerns.
General comments:
- It would be very helpful for the readers to see some real measurements from the instrument, e.g., time series of measurements, and concentrations along the measurement plane. Note that the precision and linearity were derived from laboratory experiments under stable temperatures, which may not represent the real conditions in the field. Do the authors have an idea about the stability or the accuracy of the concentration measurements in the field? Is there any field comparison data?
- What does the analyzer measure? Note that greenhouse gas concentrations are reported as dry mole fractions. How were the water vapor effects accounted for? I understand that the analyzer also measures H2O, however, which was not shown or discussed.
- For the mass balance approach, even though the measurement frequency is rather high, interpolation of measurements along the vertical direction was needed. How was the interpolation done? I guess it was done in a similar way as Kriging. Again, it would be very helpful if some figures can be shown.
- The addition of a LiDAR during the measurements is certainly appreciated. However, the uncertainty of LiDAR measurements was not sufficiently discussed, especially when most flights were performed at low altitudes. Are LiDAR wind measurements at low altitudes reliable? What are the associated uncertainties?
Some detailed comments:
L11-37: the abstract can be written more concisely.
L145: still, some more brief description of the system will be appreciated, e.g., how temperature and pressure variations are taken into account. How is water vapor used? How is the spectra data used to derive the mole fractions?
L246: Similar flight protocols have been presented earlier, e.g., in Andersen et al., 2021, and Morles et al., 2022. Also, a minimum wind speed was used to select appropriate flights for analyses. How was the wind speed considered in this paper?
L249: under the wind should be downwind
L270: what is the unit of Cp? Note that the water vapor effect needs to be considered, as is in eq. (4) in Andersen et al., 2021.
L337: plan should be plane
Reference:
Andersen, T., Vinkovic, K., Vries, M.d., Kers,B., Necki, J., Swolkien, J., Roiger, A., Peters, W., Chen, H.:Quantifying methane emissions from coal mining ventilation shafts using an unmanned aerial vehicle (UAV)-based active AirCore system, Atmos. Environ.: X, Volume 12, 100135, ISSN 2590-1621, https://doi.org/10.1016/j.aeaoa.2021.100135, 2021.
Citation: https://doi.org/10.5194/amt-2022-334-RC1 -
AC1: 'Reply on RC1', Jean-Louis Bonne, 15 Sep 2023
Our answers to the referee comments are presented in bold characters.
- RC1: 'Comment on amt-2022-334', Anonymous Referee #1, 03 Jun 2023
Bonne et al., presented a new light-weight open path laser absorption spectrometer for in situ CO2 and CH4 concentration measurements, and applied the analyzer in the fields to quantify the emission rates of a series of controlled releases of CH4 and CO2 at the TotalEnergies TADI test platform at Lacq, France and of two offshore oil and gas platforms in the North Sea. Gas LiDAR was used to obtain vertical profiles of wind speeds and directions on site, which were used in the mass balance approach to calculate the emission rates. The paper is well structured, and both the laboratory and field experiments results could be useful to further develop the application of quantification of greenhouse gas emissions using UAVs. On the other hand, the paper would have benefited from more stringent discussion of the uncertainties of the measurements and the quantification using the mass balance approach. The paper is suitable for publication after addressing my following concerns.
General comments:
- It would be very helpful for the readers to see some real measurements from the instrument, e.g., time series of measurements, and concentrations along the measurement plane. Note that the precision and linearity were derived from laboratory experiments under stable temperatures, which may not represent the real conditions in the field. Do the authors have an idea about the stability or the accuracy of the concentration measurements in the field? Is there any field comparison data?
A figure has been added (Figure 6) showing an example of CH4 concentration measurements during a flight of the TADI campaign.
Concerning the stability of the concentration measurements in the field, it indeed differs from the laboratory measurements, as the perturbations linked with the UAV vibrations and air circulation increase the noise level compared to the tests in-lab. More details have been provided about the in-flight precision of the instrument: to estimate the noise level of each flight, we consider the median value of the rolling 1 second standard deviation.
Concerning the in-flight accuracy of the instrument, no experiment has been performed allowing an intercomparison with a reference instrument. This could be the subject of future work.
- What does the analyzer measure? Note that greenhouse gas concentrations are reported as dry mole fractions. How were the water vapor effects accounted for? I understand that the analyzer also measures H2O, however, which was not shown or discussed.
Our analyzer does not perform any spectral measurement of H2O. However, an InterMet iMet-4 radiosonde has been integrated to the analyzer with an acquisition frequency of 1Hz, which does record temperature (repeatability of 0.2°C, response time at 2s for still air and <1 s at 5 m.s-1 and 1000hPa) and relative humidity with response time of 0.6s at 25°C (5.2s at 5°C) and repeatability of 5%. The humidity is considered for the calculation of the concentration based on the absorption spectra. The dilution effect of the water vapor is considered for the mass balance calculations.
- For the mass balance approach, even though the measurement frequency is rather high, interpolation of measurements along the vertical direction was needed. How was the interpolation done? I guess it was done in a similar way as Kriging. Again, it would be very helpful if some figures can be shown.
We don’t interpolate the greenhouse gases concentrations vertically: we first calculate the integral along the y axis of (concentration above background x wind) for each horizontal transect. This provides a linear flux for each level (in g.s-1.m-1). We linearly interpolate these values between the different vertical levels to calculate the vertical integration which provides the complete flux. For the upper and lower levels, if the values do not reach zero, we fit the distribution with a gaussian curve to complete the missing values.
Therefore, there is no Kriging of the concentration values in our calculation.
The description has been modified to describe these different steps more precisely. An example figure has been added (Figure 7) to illustrate the vertical interpolation of all horizontally integration values along each transect.- The addition of a LiDAR during the measurements is certainly appreciated. However, the uncertainty of LiDAR measurements was not sufficiently discussed, especially when most flights were performed at low altitudes. Are LiDAR wind measurements at low altitudes reliable? What are the associated uncertainties?
The LiDAR measurements are certified by the manufacturer and reliable (controlled against tall tower equipped with anemometers, precision of 0.1 m.s-1 and 0.5°) from 10 to 200 m above the LiDAR window (typically 1 m above the ground). Complementary measurements at 1.5 m above the ground are performed by a sonic anemometer.
The largest source of uncertainty is therefore the interpolation between the monitoring levels of the LiDAR. For the values between 1.5 and 11 m, interpolation was performed with a very simple logarithmic function matching the values at 1.5 m and at 11 m. This might be a source of errors for our method. Associated uncertainties are however hard to estimate. This has been added in the revised manuscript.
Uncertainties linked with the LiDAR measurements are also linked with the low frequency of the measurements (15 to 20s), which does not allow to represent the turbulences of the low atmosphere.
This was already discussed in the article on the TADI results, but this discussion has been extended.
Some detailed comments:
L11-37: the abstract can be written more concisely.
The abstract has been reworked and is now more concise.
L145: still, some more brief description of the system will be appreciated, e.g., how temperature and pressure variations are taken into account. How is water vapor used? How is the spectra data used to derive the mole fractions?
More details have been added about the instrument: there is no spectroscopic measurement of the water vapor but the humidity is measured by the iMet-4 sensor. The effect of water vapor is considered in the theoretical spectral absorption model used to derive concentrations from the absorption spectra. The humidity is also used to calculate dry basis concentrations, considering the dilution effect of water vapor.
We believe the technical details on the spectroscopy are not the subject of the publication and have been already detailed in the papers presenting the original AMULSE instrument (Joly et al. 2016, 2020) from which the AUSEA instruments were adapted, keeping the same spectroscopic principles.
L246: Similar flight protocols have been presented earlier, e.g., in Andersen et al., 2021, and Morles et al., 2022. Also, a minimum wind speed was used to select appropriate flights for analyses. How was the wind speed considered in this paper?
More details about the wind conditions of all flight have been added in this revised manuscript. Concerning the validation experiments during the TADI 2019 and 2021 campaigns, no direct link has been found between the mean wind speed, the wind speed variability or the wind direction variability, and the relative errors of the quantifications. During the TADI 2019 and 2021 campaigns, many flights were performed under low or unstable wind conditions (under the classification of Morales et al. 2022). However, no significant effect on the error quantification has been found. It is possible that other variable parameters between the different flights also affect the uncertainty of the method and partially hide the link between wind speed and quantification error.
L249: under the wind should be downwind
This has been corrected.
L270: what is the unit of Cp? Note that the water vapor effect needs to be considered, as is in eq. (4) in Andersen et al., 2021.
The concentrations are expressed in g.m-3. We consider the effect of dilution on water vapour and use dry basis concentrations of water.
L337: plan should be plane
This has been corrected.
Citation: https://doi.org/10.5194/amt-2022-334-AC1
-
RC2: 'Comment on amt-2022-334', David Noone, 19 Jun 2023
Major points
The paper outlines an extension of an earlier instrument development, to now simultaneously obtain methane and CO2 concentrations. The reported precision and stability over ambient measurement ranges are impressive, and able to provide critical information for a variety of emission and monitoring tasks. The performance of the instrument is impressive, but accomplishing this with the weight, power and size limits marks the limits of current best practices. The paper findings focus on comparison with other instrumentation during a field experiment.The description of the new instrumentation is quite brief, and could be strengthened with some more detail to satisfy the reader of the additional performance relative to earlier work. The lack of detail doesn’t fully document the technical aspects of the instrument to a degree that paper could stand on its own as a work describing instrument development. On the other hand, the presentation of a series of field based results is a sound demonstration of the strategy as a whole and is a useful pairing to the instrumentation overview.
The single largest limitation in the paper, however, is the lack of an error budget in the flux estimates. There is a description of the instrument precision, but there is no attempt to account for errors in the flux estimate. Issues on homogeneity of wind, plume structure, plume sampling, temporal variability, variation in the background and limitations in the assumptions leading to a mass balance approach (“model error”) are all missing. In fact, the direct propagation of measurement precision on the final flux estimate was not given. This needs to be remedied. The revised paper should described an approach (and ideally a full account) of errors in the estimation method. One advantage the authors have is that they have occasion to have knowledge of the actual emission during controlled release, which allows a check not only on the quoted means (as is done), but some assessment of those means relative to the expected error.
I have a hunch that once done, the precision of the instrument will not be the leading source of uncertainty. As such, the authors can claim that their instrument meets the measurement need for these types of problems. However, unless the error budget is done, one can not place any confidence in the fact the instrument or methods are robust enough.
The paper presets results of source flux estimates but does not describe if these are useful. Are the estimates sufficiently robust (small enough error) to be of practical use? A discussion section (and possibly a motivating section in the introduction) that lays out the measurement need in terms of total error (perhaps a percentage or perhaps a total value) is needed.
During revision, the authors should look for clarity in what they wish to use as a title and what to emphasize in the abstract. At the outset, the paper has focus on instrumentation, but in the end claims the focus is on flux estimation technique, while the bulk of the results focus on case study applications. I caution that the lack of specific development of the estimation technique make the claim that there is method development difficult to argue. The bulk flux estimate from a mass balance method was not developed, and the error budget not described. While I do suggest the error budget be added during revision, it is also wise to ensure the text focuses toward the comparative natured of the results.
To this end care is needed as, for instance in the conclusion (L497-501) appears to suggest the technique itself was devised here. This is overstating the matter. Re-phrase sections like this to highlight that it is the combination of the sensor, the UAV platform, and the application of a mass balance method when used together that represent a measurement system. The essential new work done here, to my mind, is in all three areas: using a newly developed sensor designed for UAV work, using data from the sensor to drive a mass balance calculation (albeit, without error characterization), and then testing it in pragmatic cases that illustrate possible wide application.
Minor pointsL24. TADI should be spelled out in the abstract.
L75. “atmospheric concentration measurements can be operated´ should be instruments that are operated. (measurements can be obtained by remotely operated instruments, …etc)
L84 aircrafts - > aircraft (no “s”). Also, perhaps surface platforms rather than slow. Cars and ships can move quickly, and more quickly that some aircraft. Clarify your point, does speed matter, is the type of platform the key here.
L88: Check phrasing “solutions …. To types of solutions” (you men “challenges?”). Also you might say UAV plat forms hold potential for these characteristics. It is not clear that they are currently cost saving, for instance.
L89: You make the case for boats, but I suspect you excluded autonomous boats, which would have some similar benefits as autonomous aircraft.
L105. Check phrasing. The ”L” in DIAL is LIDAR: differential absorption LIDAR.
L109: What does “high quality” mean in this context?
L128: It would be useful to also not some of the limitations of open path measurement here. Close path DAS approaches have advantages as well.
L 155: embarks is the wrong word.
L170/171: exploited is the wrong word
L180: In this section, it would be useful to provide some comparative values from other published work. Perhaps quote precision from commercially available instruments for comparison? (i.e., your Picarro G2401 )
L200: Not essential, but I’m curious of you looked at the linear fit residuals. Was there an obvious curvature?
L244: Missing work. Under the wind conditions…? (Check this sentence)
L270: What procedure do you use to interpolate/extrapolate your measurements to span x an z.
L275: Section 3.2 would be a good place to include a description of the methodology to estimate flux errors. How do you assign uncertainties in your estimates based on the limitations of assumptions in this formulation (constant wind, homogeneous background, imperfectly known plume structure, etc)? What is the full error budget?
L325: Clarify what it is that the error is related to. Is this the estimated total emitted mass related to that recovered from measurements?
L325: Similarly, there is a need to quote uncertainty here. You have an estimate of “q” but the confidence levels for that estimate are not known. (I would expect that only a fraction on the total uncertainty is arising from the precision limitations of the instruments. Is this true?)
L345: You mean modify, rather than modulate.
L346: plan, not plane. Missing word: “allows *the sampling strategy* to adapt…”
L 355: Ultrasonic (no “s”)
L365: Mention of higher uncertainty. I agree this should be the case. However, a budget calculation performed here for your set up would be useful to show/prove this.
L370: “Our flux estimates…” (not performances)
L370. You mean state of the art. However, you might say “commonly used methods”
L403: This seems a major limitation. It would be useful to estimate the magnitude of the flux error associated wit this. What would you expect if you had vertical wind information? Can you illustrate this using the earlier data with the doppler lidar wind data?
L428: Here you certainly mean the simple mass balance method has a known set of causes of error. They can be large, due to the aspects of intermittency you imply.
L465: 7 flights were completed. (not exploited)
L482: These two sentences capture some discussion about possible sources of error, and is welcome. It needs to be expanded to give confidence in the characterization of error from the method as a whole, but particularly the impact of the new instrument measurement precision on the final uncertainty. I would guess it is not the leading source of error.
L515: It is claimed that the estimates are precise. This has not been shown. The sensor itself has a precision which is admirable, and needs to be compared to lab-based (etc) commercial instruments to show this. Even then, there is no assessment of how precise the flux estimates are following from the use of the UAV concentration measures within a simplified, and underconstrained mass balance calculation.
Citation: https://doi.org/10.5194/amt-2022-334-RC2 -
AC4: 'Reply on RC2', Jean-Louis Bonne, 15 Sep 2023
Our answers to the referee comments are presented in bold characters.
- RC2: 'Comment on amt-2022-334', David Noone, 19 Jun 2023
Major points
The paper outlines an extension of an earlier instrument development, to now simultaneously obtain methane and CO2 concentrations. The reported precision and stability over ambient measurement ranges are impressive, and able to provide critical information for a variety of emission and monitoring tasks. The performance of the instrument is impressive, but accomplishing this with the weight, power and size limits marks the limits of current best practices. The paper findings focus on comparison with other instrumentation during a field experiment.
The description of the new instrumentation is quite brief, and could be strengthened with some more detail to satisfy the reader of the additional performance relative to earlier work. The lack of detail doesn’t fully document the technical aspects of the instrument to a degree that paper could stand on its own as a work describing instrument development. On the other hand, the presentation of a series of field based results is a sound demonstration of the strategy as a whole and is a useful pairing to the instrumentation overview.
The technical details of the instrument development have been provided in earlier publications presenting the original AMULSE instrument (Joly et al. 2016, 2020), which is the core of the instrument and has not been modified. More details are given about the changes which have been made to adapt it to the UAV carrier and the telemetry. Providing more details on the spectroscopy is not our aim for this paper.
The single largest limitation in the paper, however, is the lack of an error budget in the flux estimates. There is a description of the instrument precision, but there is no attempt to account for errors in the flux estimate. Issues on homogeneity of wind, plume structure, plume sampling, temporal variability, variation in the background and limitations in the assumptions leading to a mass balance approach (“model error”) are all missing. In fact, the direct propagation of measurement precision on the final flux estimate was not given. This needs to be remedied. The revised paper should described an approach (and ideally a full account) of errors in the estimation method. One advantage the authors have is that they have occasion to have knowledge of the actual emission during controlled release, which allows a check not only on the quoted means (as is done), but some assessment of those means relative to the expected error.
I have a hunch that once done, the precision of the instrument will not be the leading source of uncertainty. As such, the authors can claim that their instrument meets the measurement need for these types of problems. However, unless the error budget is done, one can not place any confidence in the fact the instrument or methods are robust enough.
More details have been provided concerning the evaluation of different sources of uncertainties: wind and concentration measurement uncertainties, signal to noise ratio of the concentrations, wind variability. Evaluating the impact of these uncertainties on the errors of quantification did not reveal much link between the variability of these parameters and the errors of quantification, apart from a limited effect of the signal to noise ratio. It seems that most of the uncertainty of the final method is either linked with a combination of these effects, and/or to the effect of hidden parameters which are more difficult to quantify (undersampling or oversampling of the plume with variable gaps between transects, different durations to perform the full scan of the plume depending of the size of the plume, distance of the source and monitoring flight plane, extrapolation of top/bottom of the plumes if not monitored, angle between transects and wind directions, uncertainty associated with vertical interpolation, etc.).
A direct propagation of the uncertainties is however still missing in our algorithm, and will be the subject of future development. But based on the evaluation of the different sources of uncertainties which are possible to quantify, it is possible that the direct propagation of uncertainties finally does not match the errors of quantification.
Therefore, we chose to evaluate the uncertainties empirically, based on the results of the controlled release experiments of the TADI campaigns, which is representative of a global uncertainty on the field, knowing that it is a worst case scenario (low and unstable wind conditions, low elevated plumes and UAV flight elevation below the LiDAR monitoring levels) and the most of the real world field measurements are performed under better conditions (plumes at higher elevation with stronger wind and LiDAR monitoring levels matching the UAV elevation).
The paper presets results of source flux estimates but does not describe if these are useful. Are the estimates sufficiently robust (small enough error) to be of practical use? A discussion section (and possibly a motivating section in the introduction) that lays out the measurement need in terms of total error (perhaps a percentage or perhaps a total value) is needed.
During revision, the authors should look for clarity in what they wish to use as a title and what to emphasize in the abstract. At the outset, the paper has focus on instrumentation, but in the end claims the focus is on flux estimation technique, while the bulk of the results focus on case study applications. I caution that the lack of specific development of the estimation technique make the claim that there is method development difficult to argue. The bulk flux estimate from a mass balance method was not developed, and the error budget not described. While I do suggest the error budget be added during revision, it is also wise to ensure the text focuses toward the comparative natured of the results.
To this end care is needed as, for instance in the conclusion (L497-501) appears to suggest the technique itself was devised here. This is overstating the matter. Re-phrase sections like this to highlight that it is the combination of the sensor, the UAV platform, and the application of a mass balance method when used together that represent a measurement system. The essential new work done here, to my mind, is in all three areas: using a newly developed sensor designed for UAV work, using data from the sensor to drive a mass balance calculation (albeit, without error characterization), and then testing it in pragmatic cases that illustrate possible wide application.
These suggestions have been taken into account for the revision of the manuscript.
Minor pointsL24. TADI should be spelled out in the abstract.
This has been modified.
L75. “atmospheric concentration measurements can be operated´ should be instruments that are operated. (measurements can be obtained by remotely operated instruments, …etc)
Thank you for this language correction. Done.
L84 aircrafts - > aircraft (no “s”). Also, perhaps surface platforms rather than slow. Cars and ships can move quickly, and more quickly that some aircraft. Clarify your point, does speed matter, is the type of platform the key here.
Ok, corrected.
L88: Check phrasing “solutions …. To types of solutions” (you men “challenges?”). Also you might say UAV plat forms hold potential for these characteristics. It is not clear that they are currently cost saving, for instance.
Ok, corrected.
L89: You make the case for boats, but I suspect you excluded autonomous boats, which would have some similar benefits as autonomous aircraft.
Yes, in terms of speed and distance to the source for example, but not it terms of vertical elevation.
L105. Check phrasing. The ”L” in DIAL is LIDAR: differential absorption LIDAR.
Indeed. This has been corrected.
L109: What does “high quality” mean in this context?
Good remark, this actually depends from the expected signal to noise ratio and thus from the context. In the context of oil and gas industries, high range of concentrations are measured, so there is a need for high sensitivity range but the precision can be lower than for low and diffuse sources.
L128: It would be useful to also not some of the limitations of open path measurement here. Close path DAS approaches have advantages as well.
Indeed, close path DAS are less sensitive to external perturbations such as temperature or pressure changes, or external light sources. This has been added to the manuscript.
L 155: embarks is the wrong word.
“embarks” was replaced by “is equipped with”
L170/171: exploited is the wrong word
“Exploited” was replaced by “analysed”
L180: In this section, it would be useful to provide some comparative values from other published work. Perhaps quote precision from commercially available instruments for comparison? (i.e., your Picarro G2401 )
A comparison with different instrument has been added, although the technics and constraints of applications of these instruments can be very different (particularly for a Picarro analyzer).
L200: Not essential, but I’m curious of you looked at the linear fit residuals. Was there an obvious curvature?
No trend is being observed within the certified Picarro measurement range (below 20 ppm CH4).
L244: Missing work. Under the wind conditions…? (Check this sentence)
I don’t understand this comment.
L270: What procedure do you use to interpolate/extrapolate your measurements to span x an z.
More details have been provided to describe these steps. Integration along the y axis is performed with a simple linear interpolation, as the horizontal resolution of the measurement is very high. The interpolation of the measurements along the z axis between the horizontal transects. Measurements can also be extrapolated vertically if the lowest of highest horizontal transects do not reach the top or bottom of plume.
L275: Section 3.2 would be a good place to include a description of the methodology to estimate flux errors. How do you assign uncertainties in your estimates based on the limitations of assumptions in this formulation (constant wind, homogeneous background, imperfectly known plume structure, etc)? What is the full error budget?
More details have been given on the sources of uncertainties, although a complete uncertainty calculation is unfortunately still missing but will be the subject of future work.
L325: Clarify what it is that the error is related to. Is this the estimated total emitted mass related to that recovered from measurements?
The values presented were corresponding to the average of all relative errors of all quantifications, but this was not a good indicator and has been this has been removed.
L325: Similarly, there is a need to quote uncertainty here. You have an estimate of “q” but the confidence levels for that estimate are not known. (I would expect that only a fraction on the total uncertainty is arising from the precision limitations of the instruments. Is this true?)
More details on the influence of different sources of uncertainties (among which the instrumental uncertainties) are provided in the revised manuscript.
L345: You mean modify, rather than modulate.
Indeed. This has been corrected.
L346: plan, not plane. Missing word: “allows *the sampling strategy* to adapt…”
Ok. The sentence has been simplified.
L 355: Ultrasonic (no “s”)
Ok, this has been corrected.
L365: Mention of higher uncertainty. I agree this should be the case. However, a budget calculation performed here for your set up would be useful to show/prove this.
Same answer as L275.
L370: “Our flux estimates…” (not performances)
Ok, this has been corrected.
L370. You mean state of the art. However, you might say “commonly used methods”
Indeed. This has been corrected.
L403: This seems a major limitation. It would be useful to estimate the magnitude of the flux error associated wit this. What would you expect if you had vertical wind information? Can you illustrate this using the earlier data with the doppler lidar wind data?
This would be a major limitation for field applications with sources at high elevation. But in the case our validation experiments during the TADI campaign, the use of a LiDAR wasn’t the most appropriate wind monitoring method as well, since most of our measurement were performed at low elevation, below the LiDAR elevation range and thus the wind values had to be interpolated. This problematic is already discussed in the manuscript: multiple sonic anemometers distributed along a vertical mast would have been more appropriate in our case.
L428: Here you certainly mean the simple mass balance method has a known set of causes of error. They can be large, due to the aspects of intermittency you imply.
The reference emissions from the platform are indeed daily averaged values, which do not reflect the intermittency of the platform processes.
L465: 7 flights were completed. (not exploited)
Ok, this has been corrected.
L482: These two sentences capture some discussion about possible sources of error, and is welcome. It needs to be expanded to give confidence in the characterization of error from the method as a whole, but particularly the impact of the new instrument measurement precision on the final uncertainty. I would guess it is not the leading source of error.
More details have been given in terms of comparing the different sources of errors. The instrumental uncertainty is very small compared to the signal levels of our experiments. It is difficult to isolate the influence of all different sources of uncertainties based on our experiment: no direct link can be found between errors of the quantification and uncertainties associated with concentrations measurements or wind variability. It is possible that all these parameters play a combined role on the final quantification errors, and that they may be influence by other parameters more difficult to quantify. Therefore, we estimate that the results of these validation experiments in a whole are representative of the final uncertainty of our emissions quantification system.
L515: It is claimed that the estimates are precise. This has not been shown. The sensor itself has a precision which is admirable, and needs to be compared to lab-based (etc) commercial instruments to show this. Even then, there is no assessment of how precise the flux estimates are following from the use of the UAV concentration measures within a simplified, and underconstrained mass balance calculation.
This sentence has been modified. We agree that the method is lacking a complete estimation of uncertainties. This will be the subject of future improvements of the method.
Citation: https://doi.org/10.5194/amt-2022-334-AC4
-
AC4: 'Reply on RC2', Jean-Louis Bonne, 15 Sep 2023
-
EC1: 'Comment on amt-2022-334', Darin Toohey, 22 Jun 2023
Please address the comments of both reviewers in your revised paper, with particular emphasis on showing example(s) of observations and analysis of uncertainties in the flux estimates.
Citation: https://doi.org/10.5194/amt-2022-334-EC1 -
AC2: 'Reply on EC1', Jean-Louis Bonne, 15 Sep 2023
A revised version of the manuscript will be provided, with more details on the analysis of different sources of uncertainties. An example of the observations during a flight of the TADI campaigns has also been included.
Citation: https://doi.org/10.5194/amt-2022-334-AC2 -
AC3: 'Reply on EC1', Jean-Louis Bonne, 15 Sep 2023
Publisher’s note: the content of this comment was removed on 15 September 2023 since the comment was posted by mistake.
Citation: https://doi.org/10.5194/amt-2022-334-AC3
-
AC2: 'Reply on EC1', Jean-Louis Bonne, 15 Sep 2023