Articles | Volume 12, issue 9
https://doi.org/10.5194/amt-12-4965-2019
https://doi.org/10.5194/amt-12-4965-2019
Research article
 | 
13 Sep 2019
Research article |  | 13 Sep 2019

Addition of fast gas chromatography to selected ion flow tube mass spectrometry for analysis of individual monoterpenes in mixtures

Michal Lacko, Nijing Wang, Kristýna Sovová, Pavel Pásztor, and Patrik Španěl

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by Patrik Španěl on behalf of the Authors (18 Jun 2019)  Author's response   Manuscript 
ED: Referee Nomination & Report Request started (27 Jun 2019) by Yoshiteru Iinuma
RR by Anonymous Referee #3 (15 Jul 2019)
RR by Anonymous Referee #1 (23 Jul 2019)
ED: Publish subject to minor revisions (review by editor) (23 Jul 2019) by Yoshiteru Iinuma
AR by Patrik Španěl on behalf of the Authors (30 Jul 2019)  Author's response   Manuscript 
ED: Publish subject to technical corrections (13 Aug 2019) by Yoshiteru Iinuma
AR by Patrik Španěl on behalf of the Authors (13 Aug 2019)  Author's response   Manuscript 
Download
Short summary
The soft chemical ionization analytical technique of selected ion flow tube mass spectrometry, SIFT-MS, was enhanced by a fast GC pre-separation unit to identify individual isomers. Experiments were carried out with two GC columns, MXT-1 and MXT-Volatiles, using two reagent ions, H3O+ and NO+, on monoterpene samples (an artificial mixture and coniferous needles). Analyses of product ion ratios allowed for quantification of multiple monoterpenes in partially separated chromatograms.