Determination of the multiple-scattering correction factor and its cross-sensitivity to scattering and wavelength dependence for different AE33 Aethalometer filter tapes: a multi-instrumental approach

Jesús Yus-Díez, Vera Bernardoni, Griša Močnik, Andrés Alastuey, Davide Ciniglia, Matic Ivančič, Xavier Querol, Noemí Perez, Cristina Reche, Martin Rigler, Roberta Vecchi, Sara Valentini, and Marco Pandolfi

Download

Interactive discussion

Status: closed

Comment types: AC – author | RC – referee | CC – community | EC – editor | CEC – chief editor | : Report abuse
  • RC1: 'Comment on amt-2021-46', Anonymous Referee #1, 23 Apr 2021
  • RC2: 'Comment on amt-2021-46', Anonymous Referee #2, 23 Apr 2021
  • RC3: 'Comment on amt-2021-46', Anonymous Referee #1, 27 Apr 2021

Peer review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by Jesús Yus on behalf of the Authors (19 Jul 2021)  Author's response    Author's tracked changes    Manuscript
ED: Referee Nomination & Report Request started (20 Jul 2021) by Pierre Herckes
RR by Anonymous Referee #2 (17 Aug 2021)
ED: Publish subject to technical corrections (25 Aug 2021) by Pierre Herckes
Download
Short summary
Here we characterize the multiple-scattering factor, C, of the dual-spot Aethalometer AE33 and its cross-sensitivity to scattering and wavelength dependence for three background stations: urban, regional and mountaintop. C was obtained for two sets of filter tapes: M8020 and M8060. The cross-sensitivity to scattering and wavelength dependence of C were determined by inter-comparing with other absorption and scattering measurements including multi-angle off-line absorption measurements.