Articles | Volume 14, issue 1
https://doi.org/10.5194/amt-14-355-2021
https://doi.org/10.5194/amt-14-355-2021
Research article
 | 
15 Jan 2021
Research article |  | 15 Jan 2021

On the calibration of FIGAERO-ToF-CIMS: importance and impact of calibrant delivery for the particle-phase calibration

Arttu Ylisirniö, Luis M. F. Barreira, Iida Pullinen, Angela Buchholz, John Jayne, Jordan E. Krechmer, Douglas R. Worsnop, Annele Virtanen, and Siegfried Schobesberger

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by Arttu Ylisirniö on behalf of the Authors (10 Nov 2020)  Author's response   Manuscript 
ED: Referee Nomination & Report Request started (16 Nov 2020) by Bin Yuan
RR by Anonymous Referee #2 (24 Nov 2020)
ED: Publish as is (24 Nov 2020) by Bin Yuan
AR by Arttu Ylisirniö on behalf of the Authors (02 Dec 2020)  Manuscript 
Download
Short summary
FIGAERO-ToF-CIMS enables online volatility measurements of chemical compounds in ambient aerosols. Previously published volatility calibration results however differ from each other significantly. In this study we investigate the reason for this discrepancy. We found a major source of error in the widely used syringe deposition method and propose a new method for volatility calibration by using atomized calibration compounds.